Acro Rapid MDMA Urine Test is a lateral flow, rapid immunoassay for the qualitative detection of MDMA in human urine at a cutoff of 500 ng/mL. The test is used to obtain a visual qualitative result and is intended for laboratory use only. This assay provides only a preliminary result. Clinical consideration and professional judgment must be applied to a drug test result, particularly in evaluating a preliminary positive result. In order to obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas Chromatography/ Mass Spectroscopy (GC/MS) analysis is preferred.
Device Story
Single-use immunochromatographic cassette test; detects MDMA in human urine. User adds urine to test well; sample migrates via capillary action. Principle: competitive immunochemical reaction between immobilized drug-protein conjugate and drug in sample for limited binding sites on labeled antibody-colloidal gold conjugate. Negative result: line forms at test region (drug concentration below cutoff). Positive result: no line forms (drug competes for binding sites). Procedural control line confirms sample addition. Used in laboratory settings by professional staff. Visual output provides preliminary screening; requires GC/MS confirmation for clinical decision-making.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Lateral flow, rapid immunoassay. Qualitative visual readout. 500 ng/mL cutoff. In vitro diagnostic device for laboratory use.
Indications for Use
Indicated for the qualitative detection of MDMA in human urine at a 500 ng/mL cutoff. Intended for professional and laboratory use only. Provides preliminary results requiring confirmation by GC/MS.
Regulatory Classification
Identification
A methamphetamine test system is a device intended to measure methamphetamine, a central nervous system stimulating drug, in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of methamphetamine use or overdose.
Special Controls
*Classification.* Class II (special controls). A methamphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Related Devices
K040203 — MONITECT MDMA DRUG SCREEN CASSETTE TEST, MODEL MC22; FASTECT II MDMA DRUG SCREEN DIPSTICK TEST, MODEL MF22 · Branan Medical Corp. · Mar 25, 2004
K013180 — RAPIDONE-ECSTASY TEST · American Bio Medica Corp. · Nov 26, 2001
K022589 — ACON MDMA ONE STEP ECSTASY TEST STRIP & ACON MDMA ONE STEP ECSTASY TEST DEVICE · ACON Laboratories, Inc. · Oct 28, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circle around the eagle. The logo is black and white and appears to be a scanned image.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Gregory Lee, Ph.D. Consultant Acro Biotech LLC 9500 7th Street Unit M Rancho Cucamonga, CA 91730
AUG - 7 2006
Re: k060355
> Trade/Device Name: Acro Biotech Rapid MDMA Urine Test Regulation Number: 21 CFR§862.3610 Regulation Name: Methamphetamine test system Regulatory Class: Class II Product Code: DJC Dated: June 30, 2006 Received: July 3, 2006
Dear Dr. Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{1}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Alberto G
Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## FDA INDICTAIONS FOR USE FORM
## 510(K) NUMBER (IF KNOWN)_ ٢٠٥٠ 03 5
Device Name: Acro Rapid MDMA Urine Test
## Indications for Use
Acro Rapid MDMA Urine Test is a lateral flow, rapid immunoassay for the qualitative detection of MDMA in human urine at a cutoff of 500 ng/mL. The test is used to obtain a visual qualitative result and is intended for laboratory use only.
This assay provides only a preliminary result. Clinical consideration and professional judgment must be applied to a drug test result, particularly in evaluating a preliminary positive result. In order to obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas Chromatography/ Mass Spectroscopy (GC/MS) analysis is preferred.
> Prescription Use >< (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
> Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Jivision Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.