Healgen Secobarbital Test is an immunochromatographic assay for the qualitative determination of Secobarbital in human urine at a Cut-Off concentration of 300 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format. The test may yield preliminary positive results even when the prescription drug Secobarbital is ingescribed doses; it is not intended to distinguish between prescription use or abuse of this drug. There is no uniformly recognized cutoff concentration level for Secobarbital in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use. Healgen Buprenorphine Test is an immunochromatographic assay for the qualitative determination of Buprenorplaine in human urine at a Cut-Off concentration of 10 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format. The test may yield preliminary positive results even when the prescription drug Burrenorphine is ingested, at prescribed doses; it is not intended to distinguish between prescription use or abuse of this drug. There is no uniformly recognized cutoff concentration level for Buprenorphine in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use. Healgen Methadone Test is an immunochromatographic assay for the qualitative determination of Methadone in human urine at a Cut-Off concentration of 300 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use.
Device Story
Lateral flow immunochromatographic assays for qualitative detection of Secobarbital, Buprenorphine, and Methadone in human urine. Device formats include strip, cassette, dip card, and cup. Principle: competitive binding; target drug in sample competes with drug-conjugate for limited antibody binding sites. If drug concentration is below cutoff, antibody-coated particles bind to immobilized drug-conjugate, forming a visible line. If drug concentration exceeds cutoff, binding sites are saturated, preventing line formation. Control line confirms proper test performance. Used in point-of-care or home settings by lay users or professionals. Results are visual; interpreted by the user to determine presence of drug above specified cutoff. Preliminary positive results require confirmation via GC/MS. Benefits include rapid, accessible screening for drugs of abuse.
Clinical Evidence
Bench testing only. Precision, interference, specificity, and method comparison studies performed. Method comparison against GC/MS using 80 clinical samples per format showed high concordance. Lay-user study (n=560 per analyte, total 1680) demonstrated correct results ranging from 90-100% across various concentrations relative to cutoff. No clinical studies required.
Technological Characteristics
Lateral flow chromatographic immunoassay. Competitive binding principle using monoclonal mouse antibodies. Formats: strip, cassette, dip card, cup. Specimen: human urine. Stability: 4-30°C for 24 months. No electronic components or software algorithms.
Indications for Use
Indicated for qualitative detection of Secobarbital (300 ng/mL), Buprenorphine (10 ng/mL), and Methadone (300 ng/mL) in human urine. Intended for prescription and OTC use to provide preliminary screening results. Not for distinguishing between prescription use and abuse. Requires confirmatory testing (GC/MS).
Regulatory Classification
Identification
A barbiturate test system is a device intended to measure barbiturates, a class of hypnotic and sedative drugs, in serum, urine, and gastric contents. Measurements obtained by this device are used in the diagnosis and treatment of barbiturate use or overdose and in monitoring levels of barbiturate to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). A barbiturate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
K143535 — CR3 Keyless Split Sample Cup Secobarbital- Methadone · Guangzhou Wondfo Biotech Co., Ltd. · Jan 13, 2015
K152551 — First Sign® Drug of Abuse Buprenorphine Cup Test, First Sign® Drug of Abuse Buprenorphine Dip Card Test, First Sign® Drug of Abuse Butalbital Cup Test, First Sign® Drug of Abuse Butalbital Dip Card Test, First Sign® Drug of Abuse Morphine Cup Test, First Sign® Drug of Abuse Morphine Dip Card Test. · W.H.P.M., Inc. · Nov 9, 2015
K131232 — WONDFO METHADONE URINE TEST (MTD 200), WONDOFO MORPHINE URINE TEST (MOP 100) · Guangzhou Wondfo Biotech Co., Ltd. · May 31, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 24, 2015
HEALGEN SCIENTIFIC LLC C/O JOE SHIA REGULATORY CONSULTANT 504 EAST DIAMOND AVE. SUITE F GAITHERSBURG MD 20877
Re: K150791
Trade/Device Name: Healgen Secobarbital Test (Strip, Cassette, Cup, Dip Card), Healgen Buprenorphine Test (Strip, Cassette, Cup, Dip Card), Healgen Methadone Test (Strip, Cassette, Cup, Dip Card) Regulation Number: 21 CFR 862.3150 Regulation Name: Barbiturate test system Regulatory Class: II Product Code: DIS, DJR, DJG Dated: March 15, 2015 Received: March 25, 2015
Dear Mr. Joe Shia:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the
{1}------------------------------------------------
electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Katherine Serrano -S
For : Courtney H. Lias, Ph.D.
Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
#### Indications for Use
#### 510(k) Number (if known) k150791
Device Name Healgen Secobarbital Test (Strip, Cassette, Cup, Dip Card) Healgen Buprenorphine Test (Strip, Cassette, Cup, Dip Card) Healgen Methadone Test (Strip, Cassette, Cup, Dip Card)
Indications for Use (Describe)
Healgen Secobarbital Test is an immunochromatographic assay for the qualitative determination of Secobarbital in human urine at a Cut-Off concentration of 300 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.
The test may yield preliminary positive results even when the prescription drug Secobarbital is ingescribed doses; it is not intended to distinguish between prescription use or abuse of this drug. There is no uniformly recognized cutoff concentration level for Secobarbital in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use.
Healgen Buprenorphine Test is an immunochromatographic assay for the qualitative determination of Buprenorplaine in human urine at a Cut-Off concentration of 10 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.
The test may yield preliminary positive results even when the prescription drug Burrenorphine is ingested, at prescribed doses; it is not intended to distinguish between prescription use or abuse of this drug. There is no uniformly recognized cutoff concentration level for Buprenorphine in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use.
Healgen Methadone Test is an immunochromatographic assay for the qualitative determination of Methadone in human urine at a Cut-Off concentration of 300 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.
The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> ☒ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☒ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.FORM FDA 3881 (8/14)
PSC Publishing Services (301) 443-6740
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
- 1. Date: April 21, 2015
- 2. Submitter: HEALGEN SCIENTIFIC LLC 5213 Maple St Bellaire, TX 77401
- 3. Contact person: Jiangiu Fang HEALGEN SCIENTIFIC LLC 5213 Maple St Bellaire, TX 77401 Telephone: 713-733-8088 Fax: 713-733-8088 Email: bryan@healgen.com
- 4. Device Name: Healgen Secobarbital Test (Strip, Cassette, Cup, Dip Card) Healgen Buprenorphine Test (Strip, Cassette, Cup, Dip Card) Healgen Methadone Test (Strip, Cassette, Cup, Dip Card)
#### Classification:
| Product Code | CFR # | Panel |
|--------------|---------------------------------------------|------------|
| DIS | 21 CFR, 862.3150<br>Barbiturate Test System | Toxicology |
| DJR | 21 CFR, 862.3620<br>Methadone Test System | Toxicology |
| DJG | 21 CFR, 862.3650<br>Opiate Test System | Toxicology |
- 5. Predicate Devices: k132812 UCP Multi-Drug Test Key Cups
- 6. Intended Use / Indications for Use
Healgen Secobarbital Test is an immunochromatographic assay for the qualitative determination of Secobarbital in human urine at a Cut-Off concentration of 300 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.
The test may yield preliminary positive results even when the prescription drug Secobarbital is ingested, at prescribed doses; it is not intended to distinguish between prescription use or abuse of this drug. There is no uniformly recognized cutoff concentration level for Secobarbital in urine. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and
{5}------------------------------------------------
professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use.
Healgen Buprenorphine Test is an immunochromatographic assay for the qualitative determination of Buprenorphine in human urine at a Cut-Off concentration of 10 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.
The test may yield preliminary positive results even when the prescription drug Buprenorphine is ingested, at prescribed doses; it is not intended to distinguish between prescription use or abuse of this drug. There is no uniformly recognized cutoff concentration level for Buprenorphine in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use.
Healgen Methadone Test is an immunochromatographic assay for the qualitative determination of Methadone in human urine at a Cut-Off concentration of 300 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.
The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use.
- 7. Device Description
Healgen Secobarbital Test. Healgen Buprenorphine Test and Healgen Methadone Test are immunochromatographic assays for Secobarbital, Buprenorphine and Methadone. Each assay test is a lateral flow system for the qualitative detection of Secobarbital, Buprenorphine and Methadone (target analyte) in human urine. The products are in vitro diagnostic devices, which come in the form of: Strips, Cassettes, DipCards, or Cups. Each product contains a Test Device (in one of the four formats), and a package insert. Each test device is sealed with a desiccant in an aluminum pouch.
#### 8. Substantial Equivalence Information
A summary comparison of features of the Healgen Secobarbital Test, Healgen Buprenorphine Test and Healgen Methadone Test and the predicate device is provided in Table 1, Table 2 and Table 3.
#### Table 1: Features Comparison of Healgen Secobarbital Test and the Predicate Device
{6}------------------------------------------------
| Item | Device | Predicate – K132812 |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|
| Intended Use | For the qualitative determination of drugs<br>of abuse in human urine. | Same |
| Drug Analyte | Secobarbital | the class of<br>barbiturates<br>including<br>secobarbital |
| Methodology | Competitive binding, lateral flow<br>immunochromatographic assays based on<br>the principle of antigen antibody<br>immunochemistry. | Same |
| Specimen Type | Human Urine | Same |
| Cut-Off Values | 300 ng/mL | Same |
| Intended<br>Population | For over-the-counter and prescription<br>uses. | Same |
| Configurations | Strip, Cassette, Cup, Dip Card | Cup |
### Table 2: Features Comparison of Healgen Buprenorphine Test and the Predicate Device
| Item | Device | Predicate -<br>K132812 | | |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------|------------------------|--|--|
| Intended Use | For the qualitative determination of<br>drugs of abuse in human urine. | Same | | |
| Drug Analyte | Buprenorphine | Same | | |
| Methodology | Competitive binding, lateral flow<br>immunochromatographic assays based<br>on the principle of antigen antibody<br>immunochemistry. | Same | | |
| Specimen Type | Human Urine | Same | | |
| Cut-Off Values | 10 ng/mL<br>Same | | | |
| Intended<br>Population | For over-the-counter and prescription<br>uses. | Same | | |
| Configurations | Strip, Cassette, Cup, Dip Card<br>Cup | | | |
Table 3: Features Comparison of Healgen Methadone Test and the Predicate Device
{7}------------------------------------------------
| Item | Device | Predicate – K132812 |
|---------------------|----------------------------------------------------------------------------------------------------------------------------|---------------------|
| Intended Use | For the qualitative determination of drugs of abuse in human urine. | Same |
| Drug Analyte | Methadone | Same |
| Methodology | Competitive binding, lateral flow immunochromatographic assays based on the principle of antigen antibody immunochemistry. | Same |
| Specimen Type | Human Urine | Same |
| Cut-Off Values | 300 ng/mL | Same |
| Intended Population | For over-the-counter and prescription uses. | Same |
| Configurations | Strip, Cassette, Cup, Dip Card | Cup |
### 9. Test Principle
Healgen Secobarbital Test, Healgen Buprenorphine Test and Healgen Methadone Test are rapid tests for the qualitative detection of Secobarbital, Buprenorphine and Methadone in urine samples. Each assay test is a lateral flow chromatographic immunoassay. During testing, a urine specimen migrates upward by capillary action. If target drugs are present in the urine specimen below its cut-off concentration, it will not saturate the binding sites of its specific antibody (monoclonal mouse antibody) coated on the particles. The antibody-coated particles will then be captured by immobilized drug-conjugate and a visible colored line will show up in the test line region. The colored line will not form in the test line region if the target drug level exceeds its cut-off concentration because it will saturate all the binding sites of the antibody coated on the particles. A band should form in the control region of the devices regardless of the presence of drug or metabolite in the sample to indicate that the test has been performed properly.
#### 10. Performance Characteristics
- 1. Analytical Performance
- a. Precision
Precision studies were carried out for samples with concentrations of -100% cut-off, -50% cut-off, -25% cut-off, at the cut-off, +25% cut-off, +75% cut-off, +75% cut-off and +100% cut-off. These samples were prepared by spiking drug in negative samples. Each drug concentration was confirmed by GC/MS. All sample aliquots were blind labeled and randomized by the person who prepared samples and did not take part in the sample testing. For each concentration, tests were performed two runs per day for 25 days by three different operators for each format of devices. Different set of operators tested each format. The results obtained are summarized in the following tables:
{8}------------------------------------------------
# Secobarbital
Strip Format
| | Result | | | | | | | | |
|-----------------|------------------|-----------------|-----------------|-----------------|---------|-----------------|-----------------|-----------------|------------------|
| Drug | -100%<br>Cut-off | -75%<br>Cut-off | -50%<br>Cut-off | -25%<br>Cut-off | Cut-off | +25%<br>Cut-off | +50%<br>Cut-off | +75%<br>Cut-off | +100%<br>Cut-off |
| Lot: BAR1201001 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 22-/28+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot: BAR1201002 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 25-/25+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot: BAR1201003 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 23-/27+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
Cassette Format
| Drug | Result | -100% Cut-off | -75% Cut-off | -50% Cut-off | -25% Cut-off | Cut-off | +25% Cut-off | +50% Cut-off | +75% Cut-off | +100% Cut-off |
|-----------------|--------|---------------|--------------|--------------|--------------|---------|--------------|--------------|--------------|---------------|
| Lot: BAR1201004 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 22-/28+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- | |
| Lot: BAR1201005 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 25-/25+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- | |
| Lot: BAR1201006 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 17-/33+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- | |
### Dip Card Format
| | Result -100% -75% | -50% - | -25% | | +25% | | +50% +75% +100% |
|-----------------|-----------------------|---------------------------------------|------|--------------------------------------------------------------------------------|------|--|---------------------------------------|
| Drug | | Cut-off Cut-off Cut-off Cut-off | | Cut-off | | | Cut-off Cut-off Cut-off Cut-off |
| Lot: BAR1201007 | | | | 50-/0+ 50-/0+ 50-/0+ 50-/0+ 23-/27+ 50+/0- 50+/0- 50+/0- | | | |
| Lot: BAR1201008 | | | | 50-/0+ 50-/0+ 50-/0+ 50-/0+ 21-/29+ 50+/0- 50+/0- 50+/0- 50+/0- | | | |
| Lot: BAR1201009 | | | | 50-/0+ 50-/0+ 50-/0+ 50-/0+ 23-/27+ 50+/0- 50+/0- 50+/0- 50+/0- | | | |
CUP Format
| | Result | | | | | | | | |
|-----------------|------------------|-----------------|-----------------|-----------------|---------|-----------------|-----------------|-----------------|------------------|
| Drug | -100%<br>Cut-off | -75%<br>Cut-off | -50%<br>Cut-off | -25%<br>Cut-off | Cut-off | +25%<br>Cut-off | +50%<br>Cut-off | +75%<br>Cut-off | +100%<br>Cut-off |
| Lot: BAR1201010 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 27-/23+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot: BAR1201011 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 25-/25+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot: BAR1201012 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 22-/28+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
# Buprenorphine
Strip Format
| Drug | Result | -100%<br>Cut-off | -75%<br>Cut-off | -50%<br>Cut-off | -25%<br>Cut-off | +25%<br>Cut-off | +50%<br>Cut-off | +75%<br>Cut-off | +100%<br>Cut-off |
|-----------------|--------|------------------|-----------------|-----------------|-----------------|-----------------|-----------------|-----------------|------------------|
| Lot: BUP1204001 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 25-/25+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot: BUP1204002 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 24-/26+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot: BUP1204003 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 27-/23+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
Cassette Format
{9}------------------------------------------------
| | Result | -100%<br>Cut-off | -75%<br>Cut-off | -50%<br>Cut-off | -25%<br>Cut-off | Cut-off | +25%<br>Cut-off | +50%<br>Cut-off | +75%<br>Cut-off | +100%<br>Cut-off |
|-----------------|--------|------------------|-----------------|-----------------|-----------------|---------|-----------------|-----------------|-----------------|------------------|
| Drug | | | | | | | | | | |
| Lot: BUP1204004 | | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 26-/24+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot: BUP1204005 | | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 24-/26+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot: BUP1204006 | | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 26-/24+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
### Dip Card Format
| | Result | -100%<br>Cut-off | -75%<br>Cut-off | -50%<br>Cut-off | -25%<br>Cut-off | Cut-off | +25%<br>Cut-off | +50%<br>Cut-off | +75%<br>Cut-off | +100%<br>Cut-off |
|-----------------|--------|------------------|-----------------|-----------------|-----------------|---------|-----------------|-----------------|-----------------|------------------|
| Drug | | | | | | | | | | |
| Lot: BUP1204007 | | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 28-/22+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot: BUP1204008 | | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 22-/28+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot: BUP1204009 | | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 28-/22+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
### CUP Format
| | Result | -100%<br>Cut-off | -75%<br>Cut-off | -50%<br>Cut-off | -25%<br>Cut-off | Cut-off | +25%<br>Cut-off | +50%<br>Cut-off | +75%<br>Cut-off | +100%<br>Cut-off |
|------|-----------------|------------------|-----------------|-----------------|-----------------|---------|-----------------|-----------------|-----------------|------------------|
| Drug | Lot: BUP1204010 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 26-/24+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| | Lot: BUP1204011 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 24-/26+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| | Lot: BUP1204012 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 25-/25+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
### Methadone
## Strip Format
| | Result | -100%<br>Cut-off | -75%<br>Cut-off | -50%<br>Cut-off | -25%<br>Cut-off | +25%<br>Cut-off | +50%<br>Cut-off | +75%<br>Cut-off | +100%<br>Cut-off |
|-----------------|--------|------------------|-----------------|-----------------|-----------------|-----------------|-----------------|-----------------|------------------|
| Drug | | | | | | | | | |
| Lot: MTD1202001 | | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 20-/30+ | 50+/0- | 50+/0- | 50+/0- |
| Lot: MTD1202002 | | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 27-/23+ | 50+/0- | 50+/0- | 50+/0- |
| Lot: MTD1202003 | | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 26-/24+ | 50+/0- | 50+/0- | 50+/0- |
### Cassette Format
| Drug | Result | -100%<br>Cut-off | -75%<br>Cut-off | -50%<br>Cut-off | -25%<br>Cut-off | Cut-off | +25%<br>Cut-off | +50%<br>Cut-off | +75%<br>Cut-off | +100%<br>Cut-off |
|-----------------|--------|------------------|-----------------|-----------------|-----------------|---------|-----------------|-----------------|-----------------|------------------|
| Lot: MTD1202004 | | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 19-/31+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot: MTD1202005 | | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 25-/25+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot: MTD1202006 | | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 24-/26+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
### Dip Card Format
| Drug | Result | -100%<br>Cut-off | -75%<br>Cut-off | -50%<br>Cut-off | -25%<br>Cut-off | Cut-off | +25%<br>Cut-off | +50%<br>Cut-off | +75%<br>Cut-off | +100%<br>Cut-off |
|-----------------|--------|------------------|-----------------|-----------------|-----------------|---------|-----------------|-----------------|-----------------|------------------|
| Lot: MTD1202007 | | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 28-/22+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot: MTD1202008 | | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 24-/26+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
{10}------------------------------------------------
| Drug | Result | -100%<br>Cut-off | -75%<br>Cut-off | -50%<br>Cut-off | -25%<br>Cut-off | +25%<br>Cut-off | +50%<br>Cut-off | +75%<br>Cut-off | +100%<br>Cut-off |
|-----------------|--------|------------------|-----------------|-----------------|-----------------|-----------------|-----------------|-----------------|------------------|
| Lot: MTD1202009 | | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 27-/23+ | 50+/0- | 50+/0- | 50+/0- |
#### CUP Format
| | Result | -100%<br>Cut-off | -75%<br>Cut-off | -50%<br>Cut-off | -25%<br>Cut-off | Cut-off | +25%<br>Cut-off | +50%<br>Cut-off | +75%<br>Cut-off | +100%<br>Cut-off |
|-----------------|--------|------------------|-----------------|-----------------|-----------------|---------|-----------------|-----------------|-----------------|------------------|
| Drug | | | | | | | | | | |
| Lot: MTD1202010 | | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 22-/28+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot: MTD1202011 | | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 25-/25+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot: MTD1202012 | | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 22-/28+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
### b. Linearity
Not applicable.
- c. Stability
The devices are stable at 4-30°C for 24 months based on the accelerated stability study at 45°C and real time stability determination at both 4 ℃ and 30℃.
Control materials are not provided with the device. The labeling provides information on how to obtain control materials.
- d. Cut-off
A total of 150 samples equally distributed at concentrations of -50% cut-off; -25% cut-off; cut-off; +25% cut-off; +50% cut-off were tested using three different lots of each device by three different operators. Results were all positive at and above +25% cut-off and all negative at and below -25% cut-off for Secobarbital, Buprenorphine and Methadone. The following cut-off values for the test devices have been verified.
| Test | Calibrator | Cut-off<br>(ng/mL) |
|--------------------|---------------|--------------------|
| Secobarbital Test | Secobarbital | 300 |
| Buprenorphine Test | Buprenorphine | 10 |
| Methadone Test | Methadone | 300 |
### e. Interference
Potential interfering substances found in human urine of physiological or pathological conditions were added to drug-free urine and target drugs urine with concentration at 25% above cut-off levels. These urine samples were tested using three batches of each device for all formats.
{11}------------------------------------------------
Compounds that showed no interference at a concentration of 100µg/mL are summarized in the following tables. There were no differences observed for different formats.
| Acetophenetidin | Ethyl-p-aminobenzoate | Phencyclidine |
|-----------------------------------|------------------------------------------|------------------------------------|
| N-Acetylprocainamide | Fenoprofen | Phenelzine |
| Acetylsalicylic Acid<br>(Aspirin) | Furosemide | Phentermine |
| Aminopyrine | Gentisic acid | Phenylephrine-L |
| Amitriptyline | Hemoglobin | Phenylethylamine |
| Amoxicillin | Hydralazine | Phenylpropanolamine |
| D-Amphetamine | (+/-)-4-Hydroxyamphetamine<br>HCL | Prednisolone Acetate |
| L-Amphetamine | Hydrochlorothiazide | Prednisone |
| Amphetamine Sulfate | Hydrocodone | Procaine(Novocaine) |
| Ampicinine(Ampicillin) | Hydrocortisone | Promazine |
| Apomorphine | a -Hydroxyhippuric acid | Promethazine |
| L-Ascorbic Acid | p-Hydroxymethamphetamine | Propoxyphene,d- |
| Aspartame | Ibuprofen | Propranolol |
| Atropine | Imipramine | Pseudoephedrine HCL |
| Benzilic acid | Isoxsuprine | Quinidine |
| Benzphetamine | Isoproterenol-(+/-) | Quinine |
| Bezoic Acid | Ketamine | Ranitidine(Zantac) |
| Bilirubin | Labetalol | Salicylic Acid |
| Caffeine | Levorphanol | Secobarbital |
| Chloramphenicol | Loperamide | Serotonin |
| Chlordiazepoxide<br>HCL | Maprotiline | Sulfamethazine |
| Chloroquine | Meprobamate | Sulindac |
| Chlorothiazide | Methadone | Temazepam |
| Chlorpheniramine | Methoxyphenamine | 11-Nor-△9-Tetrahydrocannab<br>inol |
| Chlorpromazine | (+/-)-Methylenedioxyampheta<br>mine(MDA) | Tetracycline |
| Cholesterol | Methylphenidate | Tetrahydrozoline |
| Clomipramine | Nalbuphine | Thiamine |
| Clonidine hydrochloride | Nalidixic acid | L-Thyroxine |
| Cortisone | Naloxone hydrochloride | ThioridazineHydrochloride |
| Cotinine(-) | Naltrexone hydrochloride | Triamterene |
| Creatinine | Naproxen | Triflupromazine<br>Hydrochloride |
# Secobarbital
{12}------------------------------------------------
| Deoxyepinephrine | Niacinamide | Trimethoprim |
|----------------------|-----------------|---------------|
| Dextromethorphan | Nifedipine | Trimipramine |
| Diazepam | Norethindrone | Tryptamine |
| Diflunisal | Norpropoxyphene | DL-Tryptophan |
| Digoxin | Noscapine | Tyramine |
| Doxylamine | Oxazepam | D/L-Tyrosine |
| Ecgonine methylester | Oxymetazoline | Uric Acid |
| R(-)-Epinephrine | Papaverine | Verapamil |
| Erythromycin | Penicillin | Zomepirac |
| Estrone-3-sulfate | Perphenazine | |
## Buprenorphine
| Acetophenetidin | Ethyl-p-aminobenzoate | Phencyclidine |
|-----------------------------------|---------------------------------------|--------------------------------|
| N-Acetylprocainamide | Fenoprofen | Phenelzine |
| Acetylsalicylic Acid<br>(Aspirin) | Furosemide | Phenobarbital |
| Aminopyrine | Gentisic acid | Phentermine |
| Amitriptyline | Hemoglobin | Phenylephrine-L |
| Amoxicillin | Hydralazine | Phenylethylamine |
| Amobarbital | (+/-)-4-Hydroxyamphetamine<br>HCL | Phenylpropanolamine |
| D-Amphetamine | Hydrochlorothiazide | Prednisolone Acetate |
| L-Amphetamine | Hydrocodone | Prednisone |
| Amphetamine Sulfate | Hydrocortisone | Procaine(Novocaine) |
| Ampicinine(Ampicillin) | a -Hydroxyhippuric acid | Promazine |
| Apomorphine | p-Hydroxymethamphetamine | Promethazine |
| L-Ascorbic Acid | Ibuprofen | Propoxyphene,d- |
| Aspartame | Imipramine | Propranolol |
| Atropine | Isoxsuprine | Pseudoephedrine HCL |
| Benzilic acid | Isoproterenol-(+/-) | Quinidine |
| Benzphetamine | Ketamine | Quinine |
| Bezoic Acid | Labetalol | Ranitidine(Zantac) |
| Bilirubin | Levorphanol | Salicylic Acid |
| Caffeine | Loperamide | Secobarbital |
| Chloramphenicol | Maprotiline | Serotonin |
| Chlordiazepoxide HCL | Meprobamate | Sulfamethazine |
| Chloroquine | Methadone | Sulindac |
| Chlorothiazide | Methoxyphenamine | Temazepam |
| Chlorpheniramine | (+/-)-Methylenedioxyamphetamine (MDA) | 11-Nor-△9-Tetrahydrocannabinol |
{13}------------------------------------------------
| Chlorpromazine | Methylphenidate | Tetracycline |
|-------------------------|--------------------------|---------------------------|
| Cholesterol | Nalbuphine | Tetrahydrozoline |
| Clomipramine | Nalidixic acid | Thiamine |
| Clonidine hydrochloride | Naloxone hydrochloride | L-Thyroxine |
| Cortisone | Naltrexone hydrochloride | ThioridazineHydrochloride |
| Cotinine(-) | Naproxen | Triamterene |
| Creatinine | Niacinamide | Triflupromazine |
| | | Hydrochloride |
| Deoxyepinephrine | Nifedipine | Trimethoprim |
| Dextromethorphan | Norethindrone | Trimipramine |
| Diazepam | Norpropoxyphene | Tryptamine |
| Diflunisal | Noscapine | DL-Tryptophan |
| Digoxin | Oxazepam | Tyramine |
| Doxylamine | Oxymetazoline | D/L-Tyrosine |
| Ecgonine methylester | Papaverine | Uric Acid |
| R(-)-Epinephrine | Penicillin | Verapamil |
| Erythromycin | Pentobarbital | Zomepirac |
| Estrone-3-sulfate | Perphenazine | |
#### Methadone
| Methadone | | |
|-----------------------------------|-----------------------------------|----------------------|
| Acetophenetidin | Ethyl-p-aminobenzoate | Phenelzine |
| N-Acetylprocainamide | Fenoprofen | Phenobarbital |
| Acetylsalicylic Acid<br>(Aspirin) | Furosemide | Phentermine |
| Aminopyrine | Gentisic acid | Phenylephrine-L |
| Amitriptyline | Hemoglobin | Phenylethylamine |
| Amoxicillin | Hydralazine | Phenylpropanolamine |
| Amobarbital | (+/-)-4-Hydroxyamphetamine<br>HCL | Prednisolone Acetate |
| D-Amphetamine | Hydrochlorothiazide | Prednisone |
| L-Amphetamine | Hydrocodone | Procaine(Novocaine) |
| Amphetamine Sulfate | Hydrocortisone | Promazine |
| Ampicinine(Ampicillin) | a -Hydroxyhippuric acid | Promethazine |
| Apomorphine | p-Hydroxymethamphetamine | Propoxyphene,d- |
| L-Ascorbic Acid | Ibuprofen | Propranolol |
| Aspartame | Imipramine | Pseudoephedrine HCL |
| Atropine | Isoxsuprine | Quinidine |
| Benzilic acid | Isoproterenol-(+/-) | Quinine |
| Benzphetamine | Ketamine | Ranitidine(Zantac) |
{14}------------------------------------------------
| Bezoic Acid | Labetalol | Salicylic Acid |
|-------------------------|--------------------------------------|----------------------------------|
| Bilirubin | Levorphanol | Secobarbital |
| Caffeine | Loperamide | Serotonin |
| Chloramphenicol | Maprotiline | Sulfamethazine |
| Chlordiazepoxide HCL | Meprobamate | Sulindac |
| Chloroquine | Methoxyphenamine | Temazepam |
| Chlorothiazide | (+/-)-Methylenedioxyamphetamine(MDA) | 11-Nor-△9-Tetrahydrocannabinol |
| Chlorpheniramine | Methylphenidate | Tetracycline |
| Chlorpromazine | Nalbuphine | Tetrahydrozoline |
| Cholesterol | Nalidixic acid | Thiamine |
| Clomipramine | Naloxone hydrochloride | L-Thyroxine |
| Clonidine hydrochloride | Naltrexone hydrochloride | ThioridazineHydrochloride |
| Cortisone | Naproxen | Triamterene |
| Cotinine(-) | Niacinamide | Triflupromazine<br>Hydrochloride |
| Creatinine | Nifedipine | Trimethoprim |
| Deoxyepinephrine | Norethindrone | Trimipramine |
| Dextromethorphan | Norpropoxyphene | Tryptamine |
| Diazepam | Noscapine | DL-Tryptophan |
| Diflunisal | Oxazepam | Tyramine |
| Digoxin | Oxymetazoline | D/L-Tyrosine |
| Doxylamine | Papaverine | Uric Acid |
| Ecgonine methylester | Penicillin | Verapamil |
| R(-)-Epinephrine | Pentobarbital | Zomepirac |
| Erythromycin | Perphenazine | |
| Estrone-3-sulfate | Phencyclidine | |
- f. Specificity
To test the specificity, drug metabolites and other components that are likely to interfere in urine samples were tested using three batches of each device for all formats. The obtained lowest detectable concentration was used to calculate the cross-reactivity. There were no differences observed for different formats.
| Secobarbital<br>(Cut-off=300 ng/mL) | Result | % Cross-Reactivity |
|-------------------------------------|-----------------------|--------------------|
| Secobarbital | Positive at 300 ng/mL | 100% |
| Amobarbital | Positive at 300 ng/mL | 100% |
{15}------------------------------------------------
| Alphenol | Positive at 750 ng/mL | 40% |
|--------------------|-------------------------|------|
| Aprobarbital | Positive at 250 ng/mL | 120% |
| Butabarbital | Positive at 2500 ng/mL | 12% |
| Butathal | Positive at 2500 ng/mL | 12% |
| Butalbital | Positive at 2500 ng/mL | 12% |
| Cyclopentobarbital | Positive at 500 ng/mL | 60% |
| Pentobarbital | Positive at 2500 ng/mL | 12% |
| Phenobarbital | Positive at 25000 ng/mL | 1% |
| Buprenorphine<br>(Cut-off=10 ng/mL) | Result | % Cross-Reactivity |
|--------------------------------------|----------------------|--------------------|
| Buprenorphine | Positive at 10 ng/mL | 100% |
| Buprenorphine -3-D-Glucuronide | Positive at 10 ng/mL | 100% |
| Norbuprenorphine | Positive at 20 ng/mL | 50% |
| Norbuprenorphine<br>-3-D-Glucuronide | Positive at 20 ng/mL | 50% |
| Morphine | Negative at 100000 | Not detected |
| Oxymorphone | Negative at 100000 | Not detected |
| Hydromorphone | Negative at 100000 | Not detected |
| Methadone<br>(Cut-off=300 ng/mL) | Result | % Cross-Reactivity |
|----------------------------------|------------------------|--------------------|
| Methadone | Positive at 300 ng/mL | 100% |
| Doxylamine | Positive at 5000 ng/mL | 6% |
| EDDP | Negative at 100000 | Not detected |
| EMDP | Negative at 100000 | Not detected |
| LAAM HCI | Negative at 100000 | Not detected |
| Alpha Methadol | Negative at 100000 | Not detected |
- g. Effect of Urine Specific Gravity and Urine pH
To investigate the effect of urine specific gravity and urine pH, urine samples with of 1.000 to 1.035 specific gravity or urine samples with pH 4 to 9 were spiked with target drugs at 25% below and 25% above cut-off levels. These samples were tested using three batches of each device for all formats. Results were all positive for samples at and above +25% cut-off and all negative for samples at and below -25% Cut-Off. There were no differences observed for different formats.
#### 2. Comparison Studies
The method comparison studies for the Secobarbital Test, the Buprenorphine Test and the Methadone Test were performed in-house with three different laboratory assistants for each format
{16}------------------------------------------------
of the device. Operators ran 80 (40 negative and 40 positive) unaltered clinical samples. The samples were blind labeled and compared to GC/MS results. The results are presented in the tables below:
| Secobarbital | | | | | | |
|-----------------|----------|----------|----------------------------------------------------|-------------------------------------------------------------------------|--------------------------------------------------------------------------------|--------------------------------------------------------|
| Strip<br>format | | Negative | Low<br>Negative<br>by GC/MS<br>(less than<br>-50%) | Near Cutoff<br>Negative by<br>GC/MS<br>(Between<br>-50% and<br>cut-off) | Near<br>Cutoff<br>Positive by<br>GC/MS<br>(Between<br>the cut-off<br>and +50%) | High<br>Positive by<br>GC/MS<br>(greater<br>than +50%) |
| Viewer A | Positive | 0 | 0 | 0 | 14 | 24 |
| | Negative | 10 | 15 | 15 | 2 | 0 |
| Viewer B | Positive | 0 | 0 | 0 | 14 | 24 |
| | Negative | 10 | 15 | 15 | 2 | 0 |
| Viewer C | Positive | 0 | 0 | 0 | 15 | 24 |
| | Negative | 10 | 15 | 15 | 1 | 0 |
### Discordant Results of Secobarbital Strip
| Viewer | Sample Number | GC/MS Result | Strip Format<br>Viewer Results |
|----------|---------------|--------------|--------------------------------|
| Viewer A | BAR06 | 303 | Negative |
| Viewer A | BAR72 | 302 | Negative |
| Viewer B | BAR06 | 303 | Negative |
| Viewer B | BAR72 | 302 | Negative |
| Viewer C | BAR06 | 303 | Negative |
| Cassette<br>format | | Negative | Low<br>Negative<br>by GC/MS<br>(less than<br>-50%) | Near<br>Cutoff<br>Negative<br>by GC/MS<br>(Between<br>-50% and<br>cut-off) | Near Cutoff<br>Positive by<br>GC/MS<br>(Between<br>the cut-off<br>and +50%) | High<br>Positive by<br>GC/MS<br>(greater<br>than<br>+50%) |
|--------------------|----------|----------|----------------------------------------------------|----------------------------------------------------------------------------|-----------------------------------------------------------------------------|-----------------------------------------------------------|
| Viewer A | Positive | 0 | 0 | 0 | 13 | 24 |
| | Negative | 10 | 15 | 15 | 3 | 0 |
| Viewer B | Positive | 0 | 0 | 0 | 14 | 24 |
| | Negative | 10 | 15 | 15 | 2 | 0 |
| Viewer C | Positive | 0 | 0 | 0 | 13 | 24 |
| | Negative | 10 | 15 | 15 | 3 | 0 |
Discordant Results of Secobarbital Cassette
{17}------------------------------------------------
| Viewer | Sample Number | GC/MS Result | Cassette Format<br>Viewer Results |
|----------|---------------|--------------|-----------------------------------|
| Viewer A | BAR06 | 303 | Negative |
| Viewer A | BAR16 | 308 | Negative |
| Viewer A | BAR72 | 302 | Negative |
| Viewer B | BAR06 | 303 | Negative |
| Viewer B | BAR72 | 302 | Negative |
| Viewer C | BAR72 | 302 | Negative |
| Viewer C | BAR06 | 303 | Negative |
| Viewer C | BAR16 | 308 | Negative |
| Cup format | | Negative | Low Negative by GC/MS (less than -50%) | Near Cutoff Negative by GC/MS (Between -50% and cut-off) | Near Cutoff Positive by GC/MS (Between the cut-off and +50%) | High Positive by GC/MS (greater than +50%) |
|------------|----------|----------|----------------------------------------|----------------------------------------------------------|--------------------------------------------------------------|--------------------------------------------|
| Viewer A | Positive | 0 | 0 | 0 | 13 | 24 |
| | Negative | 10 | 15 | 15 | 3 | 0 |
| Viewer B | Positive | 0 | 0 | 0 | 13 | 24 |
| | Negative | 10 | 15 | 15 | 3 | 0 |
| Viewer C | Positive | 0 | 0 | 0 | 14 | 24 |
| | Negative | 10 | 15 | 15 | 2 | 0 |
## Discordant Results of Secobarbital Cup
| Viewer | Sample Number | GC/MS Result | Cup Format Viewer Results |
|----------|---------------|--------------|---------------------------|
| Viewer A | BAR06 | 303 | Negative |
| Viewer A | BAR16 | 308 | Negative |
| Viewer A | BAR72 | 302 | Negative |
| Viewer B | BAR06 | 303 | Negative |
| Viewer B | BAR16 | 308 | Negative |
| Viewer B | BAR72 | 302 | Negative |
| Viewer C | BAR06 | 303 | Negative |
| Viewer C | BAR72 | 302 | Negative |
{18}------------------------------------------------
| Dip Card<br>format | | Negative | Low<br>Negative<br>by GC/MS<br>(less than<br>-50%) | Near Cutoff<br>Negative by<br>GC/MS<br>(Between<br>-50% and<br>cut-off) | Near Cutoff<br>Positive by<br>GC/MS<br>(Between<br>the cut-off<br>and +50%) | High<br>Positive by<br>GC/MS<br>(greater<br>than<br>+50%) |
|--------------------|----------|----------|----------------------------------------------------|-------------------------------------------------------------------------|-----------------------------------------------------------------------------|-----------------------------------------------------------|
| Viewer A | Positive | 0 | 0 | 0 | 13 | 24 |
| | Negative | 10 | 15 | 15 | 3 | 0 |
| Viewer B | Positive | 0 | 0 | 0 | 13 | 24 |
| | Negative | 10 | 15 | 15 | 3 | 0 |
| Viewer C | Positive | 0 | 0 | 0 | 13 | 24 |
| | Negative | 10 | 15 | 15 | 3 | 0 |
### Discordant Results of Secobarbital Dip Card
| Viewer | Sample Number | GC/MS Result | Dip Card Format<br>Viewer Results |
|----------|---------------|--------------|-----------------------------------|
| Viewer A | BAR06 | 303 | Negative |
| Viewer A | BAR16 | 308 | Negative |
| Viewer A | BAR72 | 302 | Negative |
| Viewer B | BAR06 | 303 | Negative |
| Viewer B | BAR16 | 308 | Negative |
| Viewer B | BAR72 | 302 | Negative |
| Viewer C | BAR06 | 303 | Negative |
| Viewer C | BAR16 | 308 | Negative |
| Viewer C | BAR72 | 302 | Negative |
### Buprenorphine
| Strip<br>format | | Negative | Low<br>Negative<br>by GC/MS<br>(less than<br>-50%) | Near Cutoff<br>Negative by<br>GC/MS<br>(Between<br>-50% and<br>cut-off) | Near Cutoff<br>Positive by<br>GC/MS<br>(Between<br>the cut-off<br>and +50%) | High<br>Positive by<br>GC/MS<br>(greater<br>than<br>+50%) |
|-----------------|----------|----------|----------------------------------------------------|-------------------------------------------------------------------------|-----------------------------------------------------------------------------|-----------------------------------------------------------|
| Viewer A | Positive | 0 | 0 | 0 | 13 | 24 |
| | Negative | 10 | 15 | 15 | 3 | 0 |
| Viewer B | Positive | 0 | 0 | 0 | 13 | 24 |
| | Negative | 10 | 15 | 15 | 3 | 0 |
| Viewer C | Positive | 0 | 0 | 0 | 14 | 24 |
| | Negative | 10 | 15 | 15 | 2 | 0 |
## Discordant Results of Buprenorphine Strip
{19}------------------------------------------------
| Viewer | Sample Number | GC/MS Result | Strip Format<br>Viewer Results | |
|----------|---------------|--------------|--------------------------------|--|
| Viewer A | BUP23 | 10.2 | Negative | |
| Viewer A | BUP67 | 10.6 | Negative | |
| Viewer A | BUP12 | 10.4 | Negative | |
| Viewer B | BUP23 | 10.2 | Negative | |
| Viewer B | BUP67 | 10.6 | Negative | |
| Viewer B | BUP12 | 10.4 | Negative | |
| Viewer C | BUP23 | 10.2 | Negative | |
| Viewer C | BUP12 | 10.4 | Negative | |
| Cassette<br>format | | Negative | Low<br>Negative<br>by GC/MS<br>(less than<br>-50%) | Near Cutoff<br>Negative by<br>GC/MS<br>(Between<br>-50% and<br>cut-off) | Near Cutoff<br>Positive by<br>GC/MS<br>(Between<br>the cut-off<br>and +50%) | High<br>Positive by<br>GC/MS<br>(greater<br>than<br>+50%) |
|--------------------|----------|----------|----------------------------------------------------|-------------------------------------------------------------------------|-----------------------------------------------------------------------------|-----------------------------------------------------------|
| Viewer A | Positive | 0 | 0 | 0 | 14 | 24 |
| | Negative | 10 | 15 | 15 | 2 | 0 |
| Viewer B | Positive | 0 | 0 | 0 | 14 | 24 |
| | Negative | 10 | 15 | 15 | 2 | 0 |
| Viewer C | Positive | 0 | 0 | 0 | 14 | 24 |
| | Negative | 10 | 15 | 15 | 2 | 0 |
## Discordant Results of Buprenorphine Cassette
| Viewer | Sample Number | GC/MS Result | Cassette Format Viewer Results |
|----------|---------------|--------------|--------------------------------|
| Viewer A | BUP23 | 10.2 | Negative |
| Viewer A | BUP12 | 10.4 | Negative |
| Viewer B | BUP23 | 10.2 | Negative |
| Viewer B | BUP12 | 10.4 | Negative |
| Viewer C | BUP23 | 10.2 | Negative |
| Viewer C | BUP12 | 10.4 | Negative |
| Dip Card<br>format | | Negative | Low<br>Negative<br>by GC/MS<br>(less than<br>-50%) | Near Cutoff<br>Negative by<br>GC/MS<br>(Between<br>-50% and<br>cut-off) | Near Cutoff<br>Positive by<br>GC/MS<br>(Between<br>the cut-off<br>and +50%) | High<br>Positive by<br>GC/MS<br>(greater<br>than<br>+50%) |
|--------------------|----------|----------|----------------------------------------------------|-------------------------------------------------------------------------|-----------------------------------------------------------------------------|-----------------------------------------------------------|
| | | | 0 | 0 | 13 | 24 |
| Viewer A | Positive | 0 | | | | |
{20}------------------------------------------------
| | Negative | 10 | 15 | 15 | 3 | 0 |
|----------|----------|----|----|----|----|----|
| Viewer B | Positive | 0 | 0 | 0 | 14 | 24 |
| | Negative | 10 | 15 | 15 | 2 | 0 |
| Viewer C | Positive | 0 | 0 | 0 | 13 | 24 |
| | Negative | 10 | 15 | 15 | 3 | 0 |
| Discordant Results of Buprenorphine Dip Card | | | |
|----------------------------------------------|--|--|--|
| | | | |
| Viewer | Sample Number | GC/MS Result | Dip Card Format<br>Viewer Results |
|----------|---------------|--------------|-----------------------------------|
| Viewer A | BUP23 | 10.2 | Negative |
| Viewer A | BUP67 | 10.6 | Negative |
| Viewer A | BUP12 | 10.4 | Negative |
| Viewer B | BUP23 | 10.2 | Negative |
| Viewer B | BUP12 | 10.4 | Negative |
| Viewer C | BUP23 | 10.2 | Negative |
| Viewer C | BUP67 | 10.6 | Negative |
| Viewer C | BUP12 | 10.4 | Negative |
| Cup<br>format | | Negative | Low<br>Negative<br>by GC/MS<br>(less than<br>-50%) | Near Cutoff<br>Negative by<br>GC/MS<br>(Between<br>-50% and<br>cut-off) | Near Cutoff<br>Positive by<br>GC/MS<br>(Between<br>the cut-off<br>and +50%) | High<br>Positive by<br>GC/MS<br>(greater<br>than<br>+50%) |
|---------------|----------|----------|----------------------------------------------------|-------------------------------------------------------------------------|-----------------------------------------------------------------------------|-----------------------------------------------------------|
| Viewer A | Positive | 0 | 0 | 0 | 13 | 24 |
| | Negative | 10 | 15 | 15 | 3 | 0 |
| Viewer B | Positive | 0 | 0 | 0 | 14 | 24 |
| | Negative | 10 | 15 | 15 | 2 | 0 |
| Viewer C | Positive | 0 | 0 | 0 | 14 | 24 |
| | Negative | 10 | 15 | 15 | 2 | 0 |
Discordant Results of Buprenorphine Cup
| Viewer | Sample Number | GC/MS Result | Cup Format<br>Viewer Results |
|----------|---------------|--------------|------------------------------|
| Viewer A | BUP23 | 10.2 | Negative |
| Viewer A | BUP67 | 10.6 | Negative |
| Viewer A | BUP12 | 10.4 | Negative |
| Viewer B | BUP23 | 10.2 | Negative |
| Viewer B | BUP12 | 10.4 | Negative |
| Viewer C | BUP23 | 10.2 | Negative |
| Viewer C | BUP12 | 10.4 | Negative |
{21}------------------------------------------------
### Methadone
| Strip<br>format | | Negative | Low<br>Negative<br>by GC/MS<br>(less than<br>-50%) | Near Cutoff<br>Negative by<br>GC/MS<br>(Between<br>-50% and<br>cut-off) | Near Cutoff<br>Positive by<br>GC/MS<br>(Between<br>the cut-off<br>and +50%) | High<br>Positive by<br>GC/MS<br>(greater<br>than<br>+50%) |
|-----------------|----------|----------|----------------------------------------------------|-------------------------------------------------------------------------|-----------------------------------------------------------------------------|-----------------------------------------------------------|
| Viewer A | Positive | 0 | 0 | 0 | 14 | 24 |
| | Negative | 10 | 15 | 15 | 2 | 0 |
| Viewer B | Positive |…
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.