K040198 · Labone, Inc. · DIO · Mar 5, 2004 · Clinical Toxicology
Device Facts
Record ID
K040198
Device Name
LABONE COCAINE METABOLITE MICRO-PLATE EIA
Applicant
Labone, Inc.
Product Code
DIO · Clinical Toxicology
Decision Date
Mar 5, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3250
Device Class
Class 2
Attributes
3rd-Party Reviewed
Intended Use
The LabOne MicroPlate Cocaine Enzyme Immunoassay (EIA) is intended for the qualitative determination of cocaine and cocaine metabolites in oral fluid collected with the OraSure® Oral Fluid Specimen Collection Device. It is a screen test with a cutoff set at 10 ng/ml of benzylecgonine per ml of oral fluid. (See cross-reactivity section for information on cross-reactivity with cocaine). The test is intended for laboratory use only. For in vitro diagnostic use. The LabOne Cocaine Metabolite Microplate EIA Calibrators are intended for medical purposes for use with the LabOne Cocaine Metabolite Microplate EIA to establish points of reference that are used in determination of values in the measurement of cocaine in saliva. The LabOne Cocaine Metabolite Microplate EIA Controls are intended for use as an assayed quality control matrix to monitor the precision and accuracy of the laboratory testing procedures for cocaine metabolite.
Device Story
Solid phase competitive enzyme immunoassay (EIA) for qualitative detection of cocaine and benzoylecgonine in oral fluid. Input: oral fluid collected via OraSure® device. Principle: competitive binding assay; produces colorimetric signal proportional to metabolite concentration. Used in laboratory settings by trained personnel. Output: preliminary qualitative result. Clinical decision-making: preliminary positive results require confirmation via GC/MS/MS. Benefits: rapid screening for cocaine metabolites in clinical or forensic contexts.
Clinical Evidence
Bench testing only. Study compared 126 specimens (108 clinical, 18 diluted) against GC/MS/MS. Positive agreement 84.6% (95% CI: 65%-96%); negative agreement 97.3% (95% CI: 92%-99%). Precision evaluated over 20 days with 120 replicates per level; intra-assay CVs 2.50%-3.45%, inter-assay CVs 2.67%-6.76%. Interference testing performed for common drugs, endogenous compounds, and consumables; false positives noted with ascorbic acid (>5000 mg/L) and orange juice.
Technological Characteristics
Solid phase competitive enzyme immunoassay (EIA). Micro-plate format. Includes assay-specific calibrators and controls for quality monitoring. In vitro diagnostic use.
Indications for Use
Indicated for qualitative screening of cocaine and cocaine metabolites in oral fluid collected via OraSure device. Intended for laboratory use only; prescription use only. Provides preliminary analytical results; requires confirmation by alternative chemical method (e.g., GC/MS).
Regulatory Classification
Identification
A cocaine and cocaine metabolite test system is a device intended to measure cocaine and a cocaine metabolite (benzoylecgonine) in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of cocaine use or overdose.
Special Controls
*Classification.* Class II (special controls). A cocaine and cocaine metabolite test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Predicate Devices
STC Cocaine Metabolite Microplate (k973651)
Related Devices
K142129 — RAPIDFRET ORAL FLUID ASSAY FOR COCAINE, RAPIDFRET ORAL FLUID COCAINE CALIBRATOR SET, RAPIDFRET ORAL FLUID COCAINE CONTROL SET · Biophor Diagnostics, Inc. · Jul 24, 2015
K062929 — QUEST DIAGNOSTICS URINE COCAINE METABOLITE EIA · Quest Diagnostics, Inc. · Dec 18, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
## MAR - 5 2004
## 510(k) Summary------LabOne Cocaine Metabolite Micro-Plate ElA
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92
The assigned 510(k) number is 15040198
#### Date of Summary:
Dec. 17, 2003
#### Correspondent:
| Name: | Liuming Yu |
|---------------|---------------------------|
| Address: | 10101 Renner Boulevard |
| | Lenexa, Kansas 66210-9752 |
| Phone Number: | 913-895-2308 |
| Fax Number: | 913-888-4186 |
#### Product Name:
Common Name: Cocaine Metabolite Micro-Plate EIA Trade Name: LabOne Cocaine Metabolite Micro-Plate EIA Classification Number: 862.3250
#### Substantially Equivalent Device:
Product: STC Cocaine Metabolite Micro-Plate EIA Manufactured by: OraSure® Technologies, Inc. 510(k) Number: K973651
#### Product Description:
LabOne Cocaine Metabolite Micro-Plate ELA is a solid phase competitive enzyme immunoassay for the determination of cocaine metabolites in oral fluid collected with the OraSure® Oral Specimen Collection Device.
#### Intended Use:
LabOne Cocaine Metabolite Micro-Plate EIA is a competitive micro-immunoassay for the qualitative detection of cocaine and its metabolites in oral fluid collected with the OraSure® Oral Fluid Specimen Collection Device.
#### Comparison:
LabOne Cocaine Metabolite Micro-Plate EIA, when used to determine cocaine and cocaine metabolites in oral fluid collected with the OraSure® Oral Fluid Specimen Collection Device, is substantially equivalent to the STC Cocaine Metabolite Micro-Plate ELA.
Page 3 SK
TX
II
{1}------------------------------------------------
## Comparison Performance Data:
Performance characteristic studies on precision, analytical sensitivity, interference and antibody I cross-reactivity showed that the LabOne Cocaine Metabolite Micro-Plate EIA is substantially equivalent to the STC Cocaine Metabolite Micro-Plate EIA.
equivalent to the BTC Octains recimens and diluted samples with both the LabOne Cocaine Kesults Screence from patent speciale Metabolite Micro-Plate also showed that the MS with this from this two test systems are substantially equivalent when using GC/MS/MS results as reference.
### Conclusion:
LabOne Cocaine Metabolite Micro-Plate EIA can be used to screen cocaine and cocaine Labone Cocains Metaoone Interes inthe OraSure® Oral Fluid Specimen Collection Device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. In the center of the circle is a stylized image of an eagle.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
MAR - 5 2004
LabOne, Inc. c/o:Ms. Laura L. Danielson TUV America, Inc. 1775 Old Highway 8 New Brighton, MN 55112
Re: k040198
Trade/Device Name: LabOne Cocaine Metabolite Micro-Plate EIA Regulation Number: 21 CFR 862.3250 Regulation Name: Cocaine and cocaine metabolite test system Regulatory Class: Class II Product Code: DIO Dated: February 19, 2004 Received: February 20, 2004
Dear Ms. Danielson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination docs not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practicc requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{3}------------------------------------------------
#### Page 2
This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter with and h your hinding of substantial equivalence of your device to a legally premits to licated device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, If you destre specific and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Tou indy obtain of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Jean M. Cooper, MS, DVM.
Jean M. Cooper, MS, DVM.
Yean M. Cooper, MS, D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indication for Use Statement
# 510(k) Number (if known) K O40 | 98
## Device Name: LabOne Cocaine Metabolite Micro-Plate EIA
### Indication For Use:
The LabOne Inc Cocaine Metabolite Micro-Plate Enzyme ImmunoAssay (EIA) is intended for the qualitative detection of cocaine and its metabolite, benzylecgonine, in oral fluid with the Orasure® Oral Specimen Device. It is a screen test with a cutoff set at 10 ng/ml of benzoylecgonine per ml of oral fluid. (See Cross Reactivity section for information on cross-reactivity with cocaine). This test is intended for laboratory use only. For in vitro diagnostic use.
The LabOne Inc Cocaine Metabolite Micro-Plate Enzyme ImmunoAssay (ELA) provides only a preliminary result. Clinical consideration and professional judgment must be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. To obtain confirmed analytical results a more specific alternate chemical method is needed. Gas chromatography/mass spectroscopy/mass spectroscopy (GC/MS/MS) is the recommended confirmatory method.
The LabOne Cocaine Metabolite Micro-Plate EIA Calibrators are intended for medical purposes for use with the LabOne Cocaine Metabolite Micro-Plate EIA to establish points of reference that are used in determination of values in the measurement of cocaine metabolite in saliva.
The LabOne Cocaine Metabolite Micro-Plate EIA Controls are intended for use as an assayed quality control matrix to monitor the precision and accuracy of the laboratory testing procedures for cocaine metabolite.
Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Carol C. Benson
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K040198
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches