← Product Code [NFT](/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/NFT) · K244043

# AllTest Multi-Drug Rapid Test Cup ; AllTest Multi-Drug Test Cup (K244043)

_Hangzhou AllTest Biotech Co., Ltd. · NFT · Feb 27, 2025 · Clinical Toxicology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/NFT/K244043

## Device Facts

- **Applicant:** Hangzhou AllTest Biotech Co., Ltd.
- **Product Code:** [NFT](/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/NFT.md)
- **Decision Date:** Feb 27, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.3100
- **Device Class:** Class 2
- **Review Panel:** Clinical Toxicology

## Indications for Use

AllTest Multi-Drug Rapid Test Cup tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Secobarbital, Benzodiazepines, Cocaine, 2ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencycline, Nortriptyline, Marijuana, Tramadol, Propoxyphene, Fentanyl and 6monoacetylmorphine in human urine at the cutoff concentrations of: [Table of drugs and cutoffs]. AllTest Multi-Drug Rapid Test Cup can be a single drug test cup or used for any combination of the above listed analytes. It is for in vitro diagnostic use only. It is intended for OTC use. The tests may yield positive results for the prescription drugs when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these crugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. GC/MS or LC/MS is the recommended confirmatory method.

## Device Story

Lateral flow immunochromatographic assay cup for qualitative detection of drugs of abuse in human urine. Device uses competitive binding principle; drug in urine competes with drug-protein conjugate for limited antibody binding sites. Absence of colored test line indicates positive result (drug concentration at or above cutoff); presence of colored line indicates negative result. Control line confirms proper procedure. Used in OTC settings; operated by lay users. Provides preliminary screening results; requires professional clinical judgment and confirmatory testing (GC/MS or LC/MS) for definitive diagnosis. Benefits patient by providing rapid, point-of-care drug screening.

## Clinical Evidence

No clinical trials performed. Evidence consists of bench testing (precision, specificity, interference) and a lay-user study. Lay-user study included 280 participants (diverse backgrounds, ages 20->50) testing spiked urine samples at various concentrations relative to cutoffs. Results showed high agreement with LC-MS/MS confirmed concentrations. Analytical performance confirmed via LC-MS/MS comparison.

## Technological Characteristics

Lateral flow immunochromatographic assay. Materials include pre-coated membrane, drug monoclonal antibody (mouse) conjugate, and drug-protein conjugate. Form factor is a test cup. Connectivity: None. Energy source: None (capillary action). Software: None. Sterilization: Not specified.

## Regulatory Identification

An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.

## Special Controls

*Classification.* Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

## Predicate Devices

- AllTest Multi-Drug Rapid Test Cup ([K233019](/device/K233019.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

February 27, 2025

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

Hangzhou Alltest Biotech Co., Ltd % Jenny Xia Director LSI International Inc. 504E Diamond Ave., Suite H Gaithersburg, Maryland 20877

Re: K244043

Trade/Device Name: AllTest Multi-Drug Rapid Test Cup ; AllTest Multi-Drug Test Cup Regulation Number: 21 CFR 862.3100 Regulation Name: Amphetamine Test System Regulatory Class: Class II Product Code: NFT, PTH, NGL, NFV, NFY, PTG, NGG, QBF, NGM, QAW, NFW, NGI Dated: December 28, 2024 Received: December 30, 2024

Dear Jenny Xia:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Joseph A. Joseph A. Kotarek -S Date: 2025.02.27 Kotarek -S 10:58:43 -05'00'

Joseph Kotarek, Ph.D. Branch Chief Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

EF

510(k) Number (if known) K244043

#### Device Name

AllTest Multi-Drug Rapid Test Cup : AllTest Multi-Drug Test Cup

#### Indications for Use (Describe)

AllTest Multi-Drug Rapid Test Cup tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Secobarbital, Benzodiazepines, Cocaine, 2ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencycline, Nortriptyline, Marijuana, Tramadol, Propoxyphene, Fentanyl and 6monoacetylmorphine in human urine at the cutoff concentrations of:

| Drug (Identifier)                                                                           | Cut-off level     |
|---------------------------------------------------------------------------------------------|-------------------|
| Amphetamine (AMP)                                                                           | 500 or 1000 ng/mL |
| Buprenorphine (BUP)                                                                         | 10 ng/mL          |
| Secobarbital (BAR)                                                                          | 300 ng/mL         |
| Benzodiazepines (BZO)                                                                       | 300 ng/mL         |
| Cocaine (COC)                                                                               | 150 or 300 ng/mL  |
| 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)                                    | 300 ng/mL         |
| Methamphetamine (MET)                                                                       | 500 or 1000 ng/mL |
| Methylenedioxymethamphetamine (MDMA)                                                        | 500 ng/mL         |
| Morphine (MOP/OPI)                                                                          | 300 or 2000 ng/mL |
| Methadone (MTD)                                                                             | 300 ng/mL         |
| Oxycodone (OXY)                                                                             | 100 ng/mL         |
| Phencyclidine (PCP)                                                                         | 25 ng/mL          |
| Nortriptyline (TCA)                                                                         | 1000 ng/mL        |
| Marijuana (THC)                                                                             | 50 ng/mL          |
| Tramadol (TRA)                                                                              | 100 ng/mL         |
| Propoxyphene (PPX)                                                                          | 300 ng/mL         |
| Fentanyl(FYL)                                                                               | 1 ng/mL           |
| 6-monoacetylmorphine (6-MAM)                                                                | 10 ng/mL          |
| AllTest Multi-Drug Rapid Test Cup can be a single drug test cup or used for any combination |                   |

est Multi-Drug Rapid Test Cup can be a single drug test cup or used for any combination of the above listed analytes. It is for in vitro diagnostic use only. It is intended for OTC use.

The tests may yield positive results for the prescription drugs when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these crugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. GC/MS or LC/MS is the recommended confirmatory method.

AllTest Multi-Drug Test Cup tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Buprenorphine, Secodiazepines, Cocaine, 2- ethylidene-1.5dimethyl-3,3- diphenylpyrrolidine, Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Nortriptyline, Marijuana, Tramadol, Propoxyphene Fentanyl and 6-monoacetylmorphine in human urine at the cutoff concentrations of:

| Drug (Identifier)   | Calibrator    | Cut-off (ng/mL) |
|---------------------|---------------|-----------------|
| Amphetamine (AMP)   | d-Amphetamine | 500 or 1000     |
| Buprenorphine (BUP) | Buprenorphine | 10              |

{4}------------------------------------------------

| Secobarbital (BAR)                                           | Secobarbital                                          | 300         |
|--------------------------------------------------------------|-------------------------------------------------------|-------------|
| Benzodiazepines (BZO)                                        | Oxazepam                                              | 300         |
| Cocaine (COC)                                                | Benzoylecgonine                                       | 150 or 300  |
| 2-ethylidene-1,5-dimethyl-<br>3,3-diphenylpyrrolidine (EDDP) | 2-ethylidene-1,5-dimethyl-<br>3,3-diphenylpyrrolidine | 300         |
| Methamphetamine (MET)                                        | d-Methamphetamine                                     | 500 or 1000 |
| Methylenedioxymethamphetamine (MDMA)                         | d,l-Methylenedioxymethamphetamine                     | 500         |
| Morphine (MOP/OPI)                                           | Morphine                                              | 300 or 2000 |
| Methadone (MTD)                                              | Methadone                                             | 300         |
| Oxycodone (OXY)                                              | Oxycodone                                             | 100         |
| Phencyclidine (PCP)                                          | Phencyclidine                                         | 25          |
| Nortriptyline (TCA)                                          | Nortriptyline                                         | 1000        |
| Marijuana (THC)                                              | 1-nor-Δ9-THC-9 COOH                                   | 50          |
| Tramadol (TRA)                                               | Tramadol                                              | 100         |
| Propoxyphene (PPX)                                           | Propoxyphene                                          | 300         |
| Fentanyl(FYL)                                                | Fentanyl                                              | 1           |
| 6-monoacetylmorphine (6-MAM)                                 | 6-monoacetylmorphine                                  | 10          |

AllTest Multi-Drug Test Cup can be a single drug test cup or used for any combination of the above listed analytes. It is for in vitro diagnostic use only.

The tests may yield positive results for the prescription drugs when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these crugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.

The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. GC/MS or LC/MS is the recommended confirmatory method.

Type of Use (Select one or both, as applicable)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

# 510(k) SUMMARY K244043

- December 28, 2024 1. Date: Hangzhou AllTest Biotech Co., Ltd. Submitter: 2. Plant Bldg. 3, 4, 5, No. 550 Yinhai Street, Baiyang Street, Hangzhou ETDZ, Jianggan District 3. Contact person: Jenny Xia LSI International Inc. 504 East Diamond Ave., Suite H Gaithersburg, MD 20877 Telephone: 301-525-6856 Email: jxia@lsi-consulting.org
- Device Name: AllTest Multi-Drug Rapid Test Cup 4. AllTest Multi-Drug Test Cup Classification: Class II ਹੈ.

| Product Code                                                                                                                           | Regulation Section                                      | Panel      |
|----------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|------------|
| Target Drug                                                                                                                            |                                                         |            |
| NFT<br>Amphetamine (AMP)                                                                                                               | 862.3100, Amphetamine Test System                       | Toxicology |
| PTH<br>Secobarbital (BAR)                                                                                                              | 862.3150, Barbiturate Test System                       | Toxicology |
| NGL<br>Buprenorphine (BUP)<br>Fentanyl (FYL)<br>Morphine (MOP/OPI)<br>Oxycodone (OXY)<br>6-Monoacetylmorphine(6-MAM)<br>Tramadol (TML) | 862.3650, Opiate Test System                            | Toxicology |
| NGI<br>Morphine (MOP/OPI)                                                                                                              | 862.3640, Morphine Test System                          | Toxicology |
| NFV<br>Oxazepam (BZO)                                                                                                                  | 862.3170, Benzodiazepine Test System                    | Toxicology |
| NFY<br>Cocaine (COC)                                                                                                                   | 862.3250, Cocaine and cocaine<br>metabolite test system | Toxicology |
| PTG<br>2-ethylidene-1,5-dimethyl-3,3-<br>diphenylpyrrolidine (EDDP)<br>Methadone (MTD)                                                 | 862.3620, Methadone Test System                         | Toxicology |
| NGG<br>Methylenedioxymethamphetamine<br>(MDMA)                                                                                         | 862.3610,<br>Methamphetamine Test System                | Toxicology |

{6}------------------------------------------------

| Methamphetamine (MET) |                                                        |            |
|-----------------------|--------------------------------------------------------|------------|
| QBF                   | 862.3700 Propoxyphene test system.                     | Toxicology |
| Propoxyphene(PPX)     |                                                        |            |
| NGM                   | Unclassified                                           | Toxicology |
| Phencyclidine (PCP)   |                                                        |            |
| QAW                   | 862.3910 Tricyclic antidepressant drugs<br>test system | Toxicology |
| Nortriptyline (TCA)   |                                                        |            |
| NFW                   | 862.3870, Cannabinoids Test System                     | Toxicology |
| Cannabinoids (THC)    |                                                        |            |

#### 6. Predicate Devices:

AllTest Multi-Drug Rapid Test Cup (K233019)

#### Intended Use 7.

AllTest Multi-Drug Rapid Test Cup tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Buprenorphine, Secobarbital, Benzodiazepines, Cocaine, 2- ethylidene-1,5-dimethyl-3,3diphenylpyrrolidine, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Nortriptyline, Marijuana, Tramadol , Propoxyphene , Fentanyl and 6-monoacetylmorphine in human urine at the cutoff concentrations of:

| Drug (Identifier)                                        | Cut-off (ng/mL) |
|----------------------------------------------------------|-----------------|
| Amphetamine (AMP)                                        | 500 or 1000     |
| Buprenorphine (BUP)                                      | 10              |
| Secobarbital (BAR)                                       | 300             |
| Benzodiazepines (BZO)                                    | 300             |
| Cocaine (COC)                                            | 150 or 300      |
| 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP) | 300             |
| Methamphetamine (MET)                                    | 500 or 1000     |
| Methylenedioxymethamphetamine (MDMA)                     | 500             |
| Morphine (MOP/OPI)                                       | 300 or 2000     |
| Methadone (MTD)                                          | 300             |
| Oxycodone (OXY)                                          | 100             |
| Phencyclidine (PCP)                                      | 25              |
| Nortriptyline (TCA)                                      | 1000            |
| Marijuana (THC)                                          | 50              |
| Tramadol (TRA)                                           | 100             |
| Propoxyphene (PPX)                                       | 300             |
| Fentanyl (FYL)                                           | 1               |
| 6-monoacetylmorphine (6-MAM)                             | 10              |

AllTest Multi-Drug Rapid Test Cup can be a single drug test cup or used for any combinations of the above listed analytes. It is for in vitro diagnostic use only. It is intended for OTC use.

{7}------------------------------------------------

The tests may yield positive results for the prescription drugs when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.

The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. GC/MS is the recommended confirmatory method.

AllTest Multi-Drug Test Cup tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Buprenorphine, Secobarbital, Benzodiazepines, Cocaine, 2- ethylidene-1,5-dimethyl-3,3diphenylpyrrolidine, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Nortriptyline, Marijuana, Tramadol, Propoxyphene Fentanyl and 6-monoacetylmorphine in human urine at the cutoff concentrations of:

| Drug (Identifier)                                            | Calibrator                                            | Cut-off (ng/mL) |
|--------------------------------------------------------------|-------------------------------------------------------|-----------------|
| Amphetamine (AMP)                                            | d-Amphetamine                                         | 500 or 1000     |
| Buprenorphine (BUP)                                          | Buprenorphine                                         | 10              |
| Secobarbital (BAR)                                           | Secobarbital                                          | 300             |
| Benzodiazepines (BZO)                                        | Oxazepam                                              | 300             |
| 2-ethylidene-1,5-dimethyl-3,3-<br>diphenylpyrrolidine (EDDP) | 2-ethylidene-1,5-dimethyl-3,3-<br>diphenylpyrrolidine | 300             |
| Methamphetamine (MET)                                        | d-Methamphetamine                                     | 500 or 1000     |
| Methylenedioxymethamphetamine (MDMA)                         | d,l-Methylenedioxymethamphetamine                     | 500             |
| Morphine (MOP/OPI)                                           | Morphine                                              | 300 or 2000     |
| Methadone (MTD)                                              | Methadone                                             | 300             |
| Oxycodone (OXY)                                              | Oxycodone                                             | 100             |
| Phencyclidine (PCP)                                          | Phencyclidine                                         | 25              |
| Nortriptyline (TCA)                                          | Nortriptyline                                         | 1000            |
| Marijuana (THC)                                              | 11-nor-Δ9-THC-9 COOH                                  | 50              |
| Tramadol (TRA)                                               | Tramadol                                              | 100             |
| Propoxyphene (PPX)                                           | Propoxyphene                                          | 300             |
| Fentanyl (FYL)                                               | Fentanyl                                              | 1               |
| 6-monoacetylmorphine (6-MAM)                                 | 6-monoacetylmorphine                                  | 10              |

AllTest Multi-Drug Test Cup can be a single drug test cup or used for any combination of the above listed analytes. It is for in vitro diagnostic use only.

The tests may yield positive results for the prescription drugs when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.

The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. GC/MS is the recommended confirmatory method.

{8}------------------------------------------------

#### 8. Device Description

AllTest Multi-Drug Rapid Test Cup and AllTest Multi-Drug Test Cup are immunochromatographic assays that use a lateral flow system for the qualitative detection of single or multiple drugs in human urine.

The devices are a cup format. Each test device is sealed with sachets of desiccant in an aluminum pouch. The device is in a ready-to-use format and no longer requires assembly before use.

#### 9. Substantial Equivalence Information

| Similarities                   |                                                                                                  |                     |                          |
|--------------------------------|--------------------------------------------------------------------------------------------------|---------------------|--------------------------|
| Item                           | Device                                                                                           | Predicate (K233019) |                          |
| Intended use                   | Qualitative detection of drugs of abuse in urine. For prescription use or over-the-counter use   | Same.               |                          |
| Methodology                    | Competitive binding, lateral flow immunochromatographic assay based on antigen-antibody reaction | Same                |                          |
| Type of Test                   | Qualitative                                                                                      | Same                |                          |
| Specimen Type                  | Human urine                                                                                      | Same                |                          |
| Target Drug and Cut Off Values | Target Drugs                                                                                     | Cutoff (ng/mL)      | Same except that no      |
|                                | Amphetamine(AMP)                                                                                 | 1000 or 500         | FYL                      |
|                                | Secobarbital (BAR)                                                                               | 300                 | TRA                      |
|                                | Buprenorphine (BUP)                                                                              | 10                  | 6-MAAM                   |
|                                | Oxazepam (BZO)                                                                                   | 300                 | PPX                      |
|                                | Cocaine (COC)                                                                                    | 150                 |                          |
|                                | 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)                                         | 300                 |                          |
|                                | Methylenedioxymethamphetamine (MDMA)                                                             | 500                 |                          |
|                                | Methamphetamine (MET)                                                                            | 1000 or 500         |                          |
|                                | Morphine (MOP300/OPI2000)                                                                        | 2000 or 300         |                          |
|                                | Methadone (MTD)                                                                                  | 300                 |                          |
|                                | Oxycodone (OXY)                                                                                  | 100                 |                          |
|                                | Phencyclidine (PCP)                                                                              | 25                  |                          |
|                                | Propoxyphene(PPX)                                                                                | 300                 |                          |
|                                | Nortriptyline (TCA)                                                                              | 1000                |                          |
|                                | Cannabinoids (THC)                                                                               | 50                  |                          |
|                                | 6-Monoacetylmorphine(6-MAM)                                                                      | 10                  |                          |
|                                | Fentanyl (FYL)                                                                                   | 1                   |                          |
|                                | Tramadol (TRA)                                                                                   | 100                 |                          |
| Differences                    |                                                                                                  |                     |                          |
| Configurations                 | Test cup                                                                                         |                     | Test cup & Test Dip Card |

{9}------------------------------------------------

# 10. Standard/Guidance Document Reference (if applicable)

None referenced.

# 11. Test Principle

AllTest Multi-Drug Rapid Test Cup or AllTest Multi-Drug Test Cup is a competitive immunoassay that is used to screen for the presence of various drugs and drug metabolites in urine. It is chromatographic absorbent device in which, drugs within a urine sample, competitively combined to a limited number of drug monoclonal antibody (mouse) conjugate binding sites.

When the test is activated, the urine is absorbed into each test strip by capillary action, mixes with the respective drug monoclonal antibody conjugate, and flows across a pre-coated membrane. When drug within the urine sample is below the detection level of the test, respective drug monoclonal antibody conjugate binds to the respective drug-protein conjugate immobilized in the Test Region (T) of the test strip. This produces a colored Test line in the Test Region (T) of the strip, which, regardless of its intensity, indicates a negative test result.

When sample drug levels are at or above the detection level of the test, the free drug in the sample binds to the respective drug monoclonal antibody conjugate, preventing the respective drug monoclonal antibody conjugate from binding to the respective drug-protein conjugate immobilized in the Test Region (T) of the device. This prevents the development of a distinct colored band in the test region, indicating a preliminary positive result.

To serve as a procedure control, a colored line will appear at the Control Region (C) of each strip, if the test has been performed properly.

## 12. Performance Characteristics

## A. Analytical performance

# a. Precision/Reproducibility:

Precision studies were carried out for samples with concentrations of +100% cutoff, +50% cutoff, +25% cutoff, cutoff, -25% cutoff, -50% cutoff, -75% cut off and -100% cutoff. Other samples were prepared by spiked target drug in drug-free urine samples. Each drug concentration was confirmed by LC-MS/MS. For each concentration, tests were performed two runs per day for 25 days using three lots of test cups. The results obtained are summarized in the following table for Tramadol (TRA), Propoxyphene (PPX), Fentanyl(FYL) and 6-monoacetylmorphine (6-MAM). The rest data were reported in the cleared AllTest submission of k233019.

## TRA

| Concentration<br>(ng/mL) | +100%<br>cutoff | +75%<br>cutoff | +50%<br>cutoff | +25%<br>cutoff | Cutoff | -25%<br>cutoff | -50%<br>cutoff | -75%<br>cutoff | -100%<br>cut-off |
|--------------------------|-----------------|----------------|----------------|----------------|--------|----------------|----------------|----------------|------------------|
| Lot Number               | 193             | 175.0          | 152.5          | 129.1          | 101.5  | 74.3           | 49.2           | 25.6           | 0                |

{10}------------------------------------------------

| Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 22-/28+ | 50-/0+ | 50-/0+ | 50-/0+ |        |
|-------|--------|--------|--------|--------|---------|--------|--------|--------|--------|
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 25-/25+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 24-/26+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ |

### PPX

| Concentration | +100%  | +75%   | +50%   | +25%   | Cutoff  | -25%   | -50%   | -75%   | -100%   |
|---------------|--------|--------|--------|--------|---------|--------|--------|--------|---------|
| (ng/mL)       | cutoff | cutoff | cutoff | cutoff |         | cutoff | cutoff | cutoff | cut-off |
| Lot Number    | 613.3  | 528.7  | 435.4  | 371.3  | 304.6   | 229.7  | 146.9  | 79.9   | 0       |
| Lot 1         | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 22-/28+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+  |
| Lot 2         | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 23-/27+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+  |
| Lot 3         | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 20-/30+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+  |

#### FYL

| Concentration<br>(ng/mL) | +100%<br>cutoff | +75%<br>cutoff | +50%<br>cutoff | +25%<br>cutoff | Cutoff  | -25%<br>cutoff | -50%<br>cutoff | -75%<br>cutoff | -100%<br>cut-off |
|--------------------------|-----------------|----------------|----------------|----------------|---------|----------------|----------------|----------------|------------------|
| Lot Number               | 2.05            | 1.79           | 1.53           | 1.21           | 0.96    | 0.71           | 0.48           | 0.26           | 0                |
| Lot 1                    | 0-/50+          | 0-/50+         | 0-/50+         | 0-/50+         | 25-/25+ | 50-/0+         | 50-/0+         | 50-/0+         | 50-/0+           |
| Lot 2                    | 0-/50+          | 0-/50+         | 0-/50+         | 0-/50+         | 23-/27+ | 48-/2+         | 50-/0+         | 50-/0+         | 50-/0+           |
| Lot 3                    | 0-/50+          | 0-/50+         | 0-/50+         | 0-/50+         | 24-/26+ | 49-/1+         | 50-/0+         | 50-/0+         | 50-/0+           |

### 6-MAM

| Concentration | +100%  | +75%   | +50%   | +25%   | Cutoff  | -25%   | -50%   | -75%   | -100%   |
|---------------|--------|--------|--------|--------|---------|--------|--------|--------|---------|
| (ng/mL)       | cutoff | cutoff | cutoff | cutoff |         | cutoff | cutoff | cutoff | cut-off |
| Lot Number    | 20.6   | 17.0   | 15.3   | 12.2   | 9.8     | 7.0    | 5.3    | 2.6    | 0       |
| Lot 1         | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 24-/26+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+  |
| Lot 2         | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 25-/25+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+  |
| Lot 3         | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 23-/27+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+  |

## b. Linearity/assay reportable range:

Not applicable. This device is intended for qualitative use only.

### c. Stability:

The device is stable at 2-30°C for 24 months based on real time stability study.

### d. Analytical specificity/Interference:

To test the specificity, drug metabolites and other components that are likely to cross-react in urine samples were spiked into drug-free urine samples were tested using three lots of the device. The results obtained are summarized in the following table for Tramadol (TRA),

{11}------------------------------------------------

Propoxyphene (PPX), Fentanyl(FYL) and 6-monoacetylmorphine (6-MAM). The rest data were reported in the cleared AllTest submission of k233019.

Percent cross-reactivity, provided in the below table, was calculated as the cutoff concentration divided by the concentration of analyte tested that yielded a positive result, multiplied by 100. If no cross-reactivity was observed, the highest concentration tested is shown.

| Drug/Cutoff                    | Compound                     | Minimum<br>concentration<br>required to obtain<br>a positive result<br>(ng/mL) | % Cross-<br>Reactivity |
|--------------------------------|------------------------------|--------------------------------------------------------------------------------|------------------------|
| TRA 100                        | Tramadol                     | 100                                                                            | 100%                   |
| TRA 100                        | N-Desmethyl-cis-tramadol     | 200                                                                            | 50%                    |
| TRA 100                        | O-Desmethyl-cis-tramadol     | 1000                                                                           | 10%                    |
| TRA 100                        | Venlafaxine                  | 100000                                                                         | 0.1%                   |
| TRA 100                        | (±)-O-Desmethylvenlafaxine   | 100000                                                                         | 0.1%                   |
| PPX 300                        | (+)-Propoxyphene             | 300                                                                            | 100%                   |
| PPX 300                        | (+)-Norpropoxyphene          | 500                                                                            | 60%                    |
| 6-MAM 10                       | 6-Monoacetylmorphine         | 10                                                                             | 100%                   |
| 6-MAM 10                       | Hydrocodone                  | >100000                                                                        | <0.01%                 |
| 6-MAM 10                       | Hydromorphone                | >100000                                                                        | <0.01%                 |
| 6-MAM 10                       | Morphine                     | >100000                                                                        | <0.01%                 |
| 6-MAM 10                       | Oxymorphone                  | >100000                                                                        | <0.01%                 |
| 6-MAM 10                       | Procaine                     | >100000                                                                        | <0.01%                 |
| 6-MAM 10                       | Thebaine                     | >100000                                                                        | <0.01%                 |
| 6-MAM 10                       | Diacetylmorphine (heroin)    | 300                                                                            | 3.3%                   |
| 6-MAM 10                       | Acetylcodeine                | >100000                                                                        | <0.01%                 |
| 6-MAM 10                       | Buprenorphine                | >100000                                                                        | <0.01%                 |
| 6-MAM 10                       | Dihydrocodeine               | >100000                                                                        | <0.01%                 |
| 6-MAM 10                       | Nalorphine                   | >100000                                                                        | <0.01%                 |
| 6-MAM 10                       | Mitragynine (kratom)         | >100000                                                                        | <0.01%                 |
| 6-MAM 10                       | Norbuprenorphine             | >100000                                                                        | <0.01%                 |
| 6-MAM 10                       | s-Monoacetylmorphine         | 10                                                                             | 100%                   |
| 6-MAM 10                       | Codeine                      | >100000                                                                        | <0.01%                 |
| 6-MAM 10                       | Ethylmorphine                | >100000                                                                        | <0.01%                 |
| 6-MAM 10                       | Levorphanol tartrate         | >100000                                                                        | <0.01%                 |
|                                | Morphine-3-β-D-glucuronide   | >100000                                                                        | <0.01%                 |
|                                | Norcodeine                   | >100000                                                                        | <0.01%                 |
|                                | Normorphine                  | >100000                                                                        | <0.01%                 |
|                                | Oxycodone                    | >100000                                                                        | <0.01%                 |
|                                | Dextromethorphan             | >100000                                                                        | <0.01%                 |
|                                | Imipramine hydrochloride     | >100000                                                                        | <0.01%                 |
|                                | Levacetylmethadol (LAAM)     | >100000                                                                        | <0.01%                 |
|                                | Meperidine                   | >100000                                                                        | <0.01%                 |
|                                | (±)-Methadone                | >100000                                                                        | <0.01%                 |
|                                | Naloxone hydrochloride       | >100000                                                                        | <0.01%                 |
|                                | Naltrexone hydrochloride     | >100000                                                                        | <0.01%                 |
|                                | Naproxen                     | >100000                                                                        | <0.01%                 |
|                                | Noroxycodone HCL             | >100000                                                                        | <0.01%                 |
|                                | Noroxymorphone HCL           | >100000                                                                        | <0.01%                 |
|                                | (+)-Norpropoxyphene maleate  | >100000                                                                        | <0.01%                 |
|                                | Oxymorphone-3β-D-glucuronide | >100000                                                                        | <0.01%                 |
|                                | Tapentadol HCl               | >100000                                                                        | <0.01%                 |
|                                | Tramadol hydrochloride       | >100000                                                                        | <0.01%                 |
|                                | Chlordiazepoxide             | >100000                                                                        | <0.01%                 |
|                                | Clobazam                     | >100000                                                                        | <0.01%                 |
|                                | D-Amphetamine                | >100000                                                                        | <0.01%                 |
|                                | (±)-Amphetamine              | >100000                                                                        | <0.01%                 |
| FYL 1                          | Fentanyl                     | 1                                                                              | 100%                   |
|                                | Acetyl fentanyl              | 1                                                                              | 100%                   |
|                                | Acrylfentanyl                | 1                                                                              | 100%                   |
|                                | ω-1-Hydroxyfentanyl          | 20000                                                                          | 0.005%                 |
|                                | Isobutyryl fentanyl          | 1                                                                              | 100%                   |
|                                | Ocfentanil                   | 2.5                                                                            | 40%                    |
|                                | Butyryl fentanyl             | 2.5                                                                            | 40%                    |
|                                | Furanyl fentanyl             | 5                                                                              | 20%                    |
|                                | Valeryl fentanyl             | 10                                                                             | 10%                    |
|                                | (±) β-hydroxythiofentanyl    | 2                                                                              | 50%                    |
|                                | 4-Fluoro-isobutyrylfentanyl  | 50                                                                             | 2%                     |
|                                | Para-fluorobutyryl fentanyl  | 4                                                                              | 25%                    |
| Para-fluoro fentanyl           | 3                            | 33.3%                                                                          |                        |
| (±)-3-cis-methyl fentanyl      | 50                           | 2%                                                                             |                        |
| Carfentanil                    | 2                            | 50%                                                                            |                        |
| Sufentanil                     | 10                           | 10%                                                                            |                        |
| Alfentanil                     | 5000                         | 0.02%                                                                          |                        |
| Cyclopropylfentanyl            | 3                            | 33.3%                                                                          |                        |
| Despropionyl fentanyl (4-ANPP) | 2500                         | 0.04%                                                                          |                        |
| Isotonitaze                    | >100000                      | <0.001%                                                                        |                        |
| AH-7921 HCL                    | >100000                      | <0.001%                                                                        |                        |
| Remifentanil                   | >100000                      | <0.001%                                                                        |                        |
| Norfentanyl                    | >100000                      | <0.001%                                                                        |                        |
| Acetyl norfentanyl             | >100000                      | <0.001%                                                                        |                        |
| Norcarfentanil                 | >100000                      | <0.001%                                                                        |                        |
| 6-Acetyl morphine              | >100000                      | <0.001%                                                                        |                        |
| Amphetamine                    | >100000                      | <0.001%                                                                        |                        |
| Buprenorphine                  | >100000                      | <0.001%                                                                        |                        |
| Buprenorphineglucuronide       | >100000                      | <0.001%                                                                        |                        |
| Codeine                        | >100000                      | <0.001%                                                                        |                        |
| Dextromethorphan               | >100000                      | <0.001%                                                                        |                        |
| Dihydrocodeine                 | >100000                      | <0.001%                                                                        |                        |
| EDDP                           | >100000                      | <0.001%                                                                        |                        |
| EMDP                           | >100000                      | <0.001%                                                                        |                        |
| Fluoxetine                     | >100000                      | <0.001%                                                                        |                        |
| Heroin                         | >100000                      | <0.001%                                                                        |                        |
| Hydrocodone                    | >100000                      | <0.001%                                                                        |                        |
| Hydromorphone                  | >100000                      | <0.001%                                                                        |                        |
| Ketamine                       | >100000                      | <0.001%                                                                        |                        |
| Levorphanol                    | >100000                      | <0.001%                                                                        |                        |
| Meperidine                     | >100000                      | <0.001%                                                                        |                        |
| Methadone                      | >100000                      | <0.001%                                                                        |                        |
| Morphine                       | >100000                      | <0.001%                                                                        |                        |
| Morphine-3-glucuronide         | >100000                      | <0.001%                                                                        |                        |
| Naloxone                       | >100000                      | <0.001%                                                                        |                        |
| Naltrexone                     | >100000                      | <0.001%                                                                        |                        |

{12}------------------------------------------------

{13}------------------------------------------------

{14}------------------------------------------------

| Norbuprenorphine     | >100000 <0.001% |
|----------------------|-----------------|
| Norcodeine           | >100000 <0.001% |
| Norketamine          | >100000 <0.001% |
| Normeperidine        | >100000 <0.001% |
| Normorphine          | >100000 <0.001% |
| Noroxycodone         | >100000 <0.001% |
| Oxycodone            | >100000 <0.001% |
| Oxymorphone          | >100000 <0.001% |
| Pentazocine (Talwin) | >100000 <0.001% |
| Pipamperone          | >100000 <0.001% |
| Risperidone          | >100000 <0.001% |
| Tapentadol           | >100000 <0.001% |
| Thioridazine         | >100000 <0.001% |
| Tilidine             | >100000 <0.001% |
| Tramadol             | >100000 <0.001% |
| Tramadol-O-Desmethyl | >100000 <0.001% |
| Tramadol-N-Desmethyl | >100000 <0.001% |

To evaluate potential interference, non-structurally related compounds were added to drug-free urine and to urine samples containing the target drugs at 50% below and 50% above each corresponding cutoff.

Compounds that show no interference at a concentration of 100µg/mL or specified concentrations are summarized in the following table.

| Acetaminophen                      | D-Pseudoephedrine                           | Norfentanyl                            |
|------------------------------------|---------------------------------------------|----------------------------------------|
| Acetone (1000mg/dL)                | Duloxetine                                  | Noscapine                              |
| Acetophenetidin                    | 4-Dimethyl-aminoantipyrine                  | (+)-Naproxen                           |
| Acetylsalicylic Acid               | 5, 5-Diphenylhydantoin                      | 19-Norethindrone                       |
| Acyclovir                          | Ecgonine methyl ester                       | Octopamine                             |
| Albumin (100mg/dL)                 | EMDP                                        | O-Hydroxyhippuric acid                 |
| Albuterol sulfate (Proair HFA)     | Ephedrine                                   | Olanzapine                             |
| Alpha Methadol                     | Erythromycin                                | Omeprazole                             |
| Aminophylline                      | Esomeprazole Magnesium (except<br>MTD test) | Oxalic acid (100 mg/dL)                |
| Aminopyrine                        | Estradiol                                   | Oxolinic acid                          |
| Amoxicillin                        | Estrone                                     | Oxymetazoline                          |
| Ampicillin                         | Ethanol (1%)                                | Paliperidone                           |
| Aripiprazole                       | Fenofibrate                                 | Papaverine                             |
| Aspartame                          | Fenoprofen                                  | Penicillin-G                           |
| Aspirin                            | Fentanyl(except FYL test)                   | PenicillinV Potassium                  |
| Atomoxetine                        | Fluoxetine Hydrochloride                    | Perphenazine                           |
| Atorvastatin Calcium               | Fluphenazine                                | Phenacetin                             |
| Atropine                           | Fotemustine                                 | Phenelzine                             |
| Azithromycin                       | Furosemide                                  | Phenethylamine                         |
| Baclofen                           | Gabapentin                                  | Phenylethylamine                       |
| Benzilic acid                      | Galactose (10mg/dL)                         | Phenylpropanolamine                    |
| Benzocaine                         | Gamma Globulin(500mg/dL)                    | Prednisone                             |
| Benzoic Acid                       | Gatifloxacin                                | Pregablin                              |
| Benzphetamine                      | Gemfibrozil                                 | Procaine                               |
| Bilirubin                          | Gentisic acid                               | Promazine                              |
| Boric Acid (1%)                    | Glucose (3000mg/dL)                         | Promethazine                           |
| Bupropion                          | Guaiacol glyceryl ether                     | Propoxyphene (except PPX test)         |
| Caffeine                           | Hemoglobin                                  | Propranolol                            |
| Cannabidiol                        | Hydralazine                                 | Pseudoephedrine                        |
| Captopril                          | Hydrochlorothiazide                         | Pyridoxine                             |
| Carbamazepine                      | Hydrocortisone                              | Pyrilamine                             |
| Carfentanil (except FYL test)      | 3-Hydroxytyramine                           | Pyrogallol                             |
| Carisoprodol                       | Ibuprofen                                   | Quetiapine                             |
| Cefradine                          | Isoxsuprine                                 | Quinine                                |
| Cephalexin                         | (+/-)-Isoproterenol                         | Quinolinic Acid                        |
| Chloralhydrate                     | Ketamine                                    | R-(-)-Apomorphine                      |
| Chloramphenicol                    | Ketoprofen                                  | Ranitidine                             |
| Chlordiazepoxide (except BZO test) | LAAM HCl                                    | Riboflavin (10mg/dL)                   |
| Chloroquine (except MET test)      | Labetalol                                   | Rifampicin                             |
| Chlorothiazide                     | L-Ascorbic Acid                             | Risperidone                            |
| Chlorpromazine                     | L-Ephedrine                                 | Salicylic Acid                         |
| Cholesterol                        | L-Epinephrine                               | Serotonin(5- Hydroxytyramine)          |
| Ciprofloxacin Hydrochloride        | Levofloxacin Hydrochloride                  | Sertraline                             |
| Citalopram                         | Levonorgestrel                              | Sildenafil Citrate                     |
| Clarithromycin                     | Levothyroxine Sodium                        | Simvastatin                            |
| Clonidine                          | Lidocaine Hydrochloride                     | Sulfamethazine                         |
| Clozapine                          | Lisinopril                                  | Sulindac                               |
| Conjugated Estrogens               | Loperamide                                  | Telmisartan                            |
| Cortisone                          | Loratadine                                  | Tetracycline                           |
| Creatine Hydrate                   | L-phenylephrine                             | Tetrahydrocortisone 3-(β-Dglucuronide) |
| Creatinine                         | Magnesium                                   | Tetrahydrocortisone, 3-acetate         |
| Cyclobenzaprine                    | Maprotiline                                 | Tetrahydrozoline                       |
| Cyclodextrin                       | Meperidine                                  | Theophylline                           |
| (-)-Cotinine                       | Meprobamate                                 | Thiamine                               |
| (+)-Chlorpheniramine               | Methapyrilene                               | Thioridazine                           |
| D,L-Epinephrine                    | Methaqualone                                | Tramadol (except TRA test)             |
| D,L-Octopamine                     | Methoxyphenamine (except AMP/MET test)      | Triamterene                            |
| d,l-Propranolol                    | Methylphenidate                             | Trifluoperazine                        |
| D,L-Tryptophan                     | Metoprolol Tartrate                         | Trimethobenzamide                      |
| D,L-Tyrosine                       | Metronidazole (300ug/ml)                    | Trimethoprim                           |
| Delorazepam (except BZO test)      | Mifepristone                                | Tryptamine                             |
| Deoxycorticosterone                | N-Acetylprocainamide                        | Tyramine                               |
| Desloratadine                      | NaCl(4000mg/dL)                             | Urea (2000mg/dL)                       |
| Dextromethorphan                   | Nalidixic Acid                              | Uric Acid                              |
| Diclofenac                         | Naloxone hydrochloride (except OXY test)    | Valproic acid (250ug/ml)               |

{15}------------------------------------------------

{16}------------------------------------------------

| Diflunisal                     | Naltrexone     | Venlafaxine HCl (except TRA test) |
|--------------------------------|----------------|-----------------------------------|
| Digoxin                        | Niacinamide    | Verapamil                         |
| Diphenhydramine HCl            | Nicotine       | Vitamin B2                        |
| Disopyramide (except MTD test) | Nicotinic Acid | Vitamin C                         |
| Dopamine HCl                   | Nifedipine     | Zaleplon                          |
| Doxepin                        | Nitroglycerin  | Zomepirac sodium salt             |
| Doxylamine                     | Nordoxepin     |                                   |

Interference by pH and specific gravity were also evaluated using pooled urine specimens with concentrations of 0 (drug-free), at 50% below and 50% above each corresponding cutoff. The results demonstrated that pH levels of 4 to 9 and specific gravity levels of 1.000 to 1.035 do not affect the results of the assays.

# B. Method comparison study

The method comparison studies for the device were performed in-house with three operators. Operators ran 80 (40 negative and 40 positive) unaltered urine clinical samples for each drug. The samples were blind labeled and compared to LC-MS/MS results. The results are presented in the table below for Tramadol (TRA), Propoxyphene (PPX), Fentanyl(FYL) and 6-monoacetylmorphine (6-MAM). The rest data were reported in the cleared AllTest submission of k233019.

| Drug<br>test | Test Cup<br>Result |   | Drug-<br>Free | Low<br>Negative by<br>LC-MS/MS<br>(less than -<br>50%) | Near Cutoff<br>Negative by<br>LC-MS/MS<br>(Between -<br>50% and the<br>Cutoff) | Near Cutoff<br>Positive by<br>LC-MS/MS<br>(Between<br>the cutoff<br>and +50%) | High Positive<br>by LC-<br>MS/MS<br>(greater than<br>+50%) |
|--------------|--------------------|---|---------------|--------------------------------------------------------|--------------------------------------------------------------------------------|-------------------------------------------------------------------------------|------------------------------------------------------------|
| TRA          | Operator           | + | 0             | 0                                                      | 1                                                                              | 11                                                                            | 29                                                         |
| TRA          | A                  | - | 12            | 14                                                     | 13                                                                             | 0                                                                             | 0                                                          |
| TRA          | Operator           | + | 0             | 0                                                      | 0                                                                              | 10                                                                            | 29                                                         |
| TRA          | B                  | - | 12            | 14                                                     | 14                                                                             | 1                                                                             | 0                                                          |
| TRA          | Operator           | + | 0             | 0                                                      | 1                                                                              | 10                                                                            | 29                                                         |
| TRA          | C                  | - | 12            | 14                                                     | 13                                                                             | 1                                                                             | 0                                                          |
| PPX          | Operator           | + | 0             | 0                                                      | 1                                                                              | 10                                                                            | 30                                                         |
| PPX          | A                  | - | 12            | 13                                                     | 14                                                                             | 0                                                                             | 0                                                          |
| PPX          | Operator           | + | 0             | 0                                                      | 1                                                                              | 10                                                                            | 30                                                         |
| PPX          | B                  | - | 12            | 13                                                     | 14                                                                             | 0                                                                             | 0                                                          |
| PPX          | Operator           | + | 0             | 0                                                      | 0                                                                              | 9                                                                             | 30                                                         |
| PPX          | C                  | - | 12            | 13                                                     | 15                                                                             | 1                                                                             | 0                                                          |
| 6-MAM        | Operator           | + | 0             | 0                                                      | 1                                                                              | 14                                                                            | 26                                                         |
| 6-MAM        | A                  | - | 12            | 11                                                     | 16                                                                             | 0                                                                             | 0                                                          |
| 6-MAM        | Operator           | + | 0             | 0                                                      | 1                                                                              | 13                                                                            | 26                                                         |
| 6-MAM        | B                  | - | 12            | 11                                                     | 16                                                                             | 1                                                                             | 0                                                          |

{17}------------------------------------------------

|     | Operator | + | 0  | 0  | 2  | 13 | 26 |
|-----|----------|---|----|----|----|----|----|
|     | C        | - | 12 | 11 | 15 | 1  | 0  |
| FYL | Operator | + | 0  | 0  | 2  | 22 | 16 |
|     | A        | - | 12 | 13 | 13 | 2  | 0  |
|     | Operator | + | 0  | 0  | 3  | 22 | 16 |
|     | B        | - | 12 | 13 | 12 | 2  | 0  |
|     | Operator | + | 0  | 0  | 2  | 22 | 16 |
|     | C        | - | 12 | 13 | 13 | 2  | 0  |

Discordant Results are summarized below.

| Drug  | Operator | Sample Number | LC/MS/MS<br>Result (ng/mL) | Accurate Result |
|-------|----------|---------------|----------------------------|-----------------|
| TRA   | A, C     | S2148         | 91.6                       | +               |
|       | B, C     | S2157         | 103.0                      | -               |
| PPX   | A        | S2052         | 286.3                      | +               |
|       | B        | S2017         | 280.6                      | +               |
|       | C        | S2018         | 319.8                      | -               |
| 6-MAM | A, C     | S2249         | 9.5                        | +               |
|       | B, C     | S2236         | 9.6                        | +               |
|       | B, C     | S2228         | 10.2                       | -               |
| FYL   | A, C     | S1425         | 1.02                       | -               |
|       | A        | S1420         | 0.87                       | +               |
|       | A, B     | S1447         | 1.01                       | -               |
|       | A, B     | S1442         | 0.91                       | +               |
|       | B, C     | S1402         | 0.92                       | +               |
|       | B        | S1422         | 1.08                       | -               |
|       | B        | S1465         | 0.92                       | +               |
|       | C        | S1453         | 0.82                       | +               |
|       | C        | S1439         | 1.02                       | -               |

# C. Lay person study

A lay user study was performed at three intended user sites with 280 lay persons. 121 male and 159 female tested AllTest Multi-Drug Rapid Test Cup. They had diverse educational and professional backgrounds and their age range from 20 to > 50. Urine samples were prepared at the following concentrations; -100%, +/-75%, +/-50%, +/-25% of the cutoff by spiking drug(s) into drug freepooled urine specimens. The concentrations of the samples were confirmed by LC-MS/MS. Each sample was aliquoted into individual containers and blind-labeled. Each participant was provided with the package insert, 1 blind labeled sample and a device. The results are summarized below.

Result of AllTest Multi-Drug Rapid Test Cup: Configuration 1

| Drug | Cutoff | Results | Concentration |
|------|--------|---------|---------------|
|------|--------|---------|---------------|

{18}------------------------------------------------

|       | (ng/mL)       |                                                              | -100% cutoff | -75% cutoff | -50% cutoff | -25% cutoff | +25% cutoff | +50% cutoff | +75% cutoff |
|-------|---------------|--------------------------------------------------------------|--------------|-------------|-------------|-------------|-------------|-------------|-------------|
| AMP   | 1000          | Negative                                                     | 20           | 20          | 20          | 19          | 1           | 0           | 0           |
|       | 1000          | Positive                                                     | 0            | 0           | 0           | 1           | 19          | 20          | 20          |
|       | 1000          | Total                                                        | 20           | 20          | 20          | 20          | 20          | 20          | 20          |
|       | 1000          | Agreement (%)                                                | 100%         | 100%        | 100%        | 95.0%       | 95.0%       | 100%        | 100%        |
| BAR   | 300           | Negative                                                     | 20           | 20          | 20          | 19          | 1           | 0           | 0           |
|       | 300           | Positive                                                     | 0            | 0           | 0           | 1           | 19          | 20          | 20          |
|       | 300           | Total                                                        | 20           | 20          | 20          | 20          | 20          | 20          | 20          |
|       | 300           | Agreement (%)                                                | 100%         | 100%        | 100%        | 95.0%       | 95.0%       | 100%        | 100%        |
| BZO   | 300           | Negative                                                     | 20           | 20          | 20          | 19          | 2           | 0           | 0           |
|       | 300           | Positive                                                     | 0            | 0           | 0           | 1           | 18          | 20          | 20          |
|       | 300           | Total                                                        | 20           | 20          | 20          | 20          | 20          | 20          | 20          |
|       | 300           | Agreement (%)                                                | 100%         | 100%        | 100%        | 95.0%       | 90.0%       | 100%        | 100%        |
| BUP   | 10            | Negative                                                     | 20           | 20          | 20          | 19          | 2           | 0           | 0           |
|       | 10            | Positive                                                     | 0            | 0           | 0           | 1           | 18          | 20          | 20          |
|       | 10            | Total                                                        | 20           | 20          | 20          | 20          | 20          | 20          | 20          |
|       | 10            | Agreement (%)                                                | 100%         | 100%        | 100%        | 95.0%       | 90.0%       | 100%        | 100%        |
| COC   | 300           | Negative                                                     | 20           | 20          | 20          | 19          | 1           | 0           | 0           |
|       | 300           | Positive                                                     | 0            | 0           | 0           | 1           | 19          | 20          | 20          |
|       | 300           | Total                                                        | 20           | 20          | 20          | 20          | 20          | 20          | 20          |
|       | 300           | Agreement (%)                                                | 100%         | 100%        | 100%        | 95.0%       | 95.0%       | 100%        | 100%        |
| EDDP  | 300           | Negative                                                     | 20           | 20          | 20          | 19          | 1           | 0           | 0           |
|       | 300           | Positive                                                     | 0            | 0…

---

**Source:** [https://fda.innolitics.com/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/NFT/K244043](https://fda.innolitics.com/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/NFT/K244043)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
