EMIT 2000 VALPROIC ACID ASSAY, MODEL OSR4G229UL

K011947 · Syva Co. · LEG · Jul 2, 2001 · Clinical Toxicology

Device Facts

Record IDK011947
Device NameEMIT 2000 VALPROIC ACID ASSAY, MODEL OSR4G229UL
ApplicantSyva Co.
Product CodeLEG · Clinical Toxicology
Decision DateJul 2, 2001
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.3645
Device ClassClass 2

Indications for Use

The Emit® 2000 Valproic Acid Assay is a homogeneous enzyme immunoassay intended for use in the quantitative analysis of valproic acid in human serum or plasma. Measurements obtained from this device are used in the diagnosis and treatment of valproic acid overdose or in monitoring levels of valproic acid to ensure appropriate therapy. These reagents are packaged specifically for use on a variety of OLYMPUS® analyzers.

Device Story

The Emit® 2000 Valproic Acid Assay is a homogeneous enzyme immunoassay for quantitative measurement of valproic acid in human serum or plasma. The device consists of reagents packaged in HDPE wedge-shaped bottles, designed for use on OLYMPUS® AU400/600™, AU800/1000™, and AU2700™ series analyzers. The system utilizes barcode labels for assay-specific information. It is intended for clinical laboratory use to assist healthcare providers in monitoring drug levels for therapeutic management or diagnosing overdose. The assay provides quantitative results that inform clinical decisions regarding valproic acid dosing.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on identical operating principles, design, and performance characteristics to the predicate device.

Technological Characteristics

Homogeneous enzyme immunoassay. Reagents contained in HDPE wedge-shaped bottles. Compatible with OLYMPUS® AU400/600™, AU800/1000™, and AU2700™ series analyzers. Includes barcode labeling for automated assay identification.

Indications for Use

Indicated for quantitative analysis of valproic acid in human serum or plasma for diagnosis and treatment of overdose or therapeutic drug monitoring.

Regulatory Classification

Identification

A neuroleptic drugs radioceptor assay test system is a device intended to measure in serum or plasma the dopamine receptor blocking activity of neuroleptic drugs and their active metabolites. A neuroleptic drug has anti-psychotic action affecting principally psychomotor activity, is generally without hypnotic effects, and is a tranquilizer. Measurements obtained by this device are used to aid in determining whether a patient is taking the prescribed dosage level of such drugs.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K011947 # JUL - 2 2001 ### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS For Emit® 2000 Valproic Acid Assay ## 1. Manufacturer and Contact Information: | Manufacturer: | Syva Company - Dade Behring Inc.20400 Mariani Ave.Cupertino, CA 95014 | |---------------|-----------------------------------------------------------------------| |---------------|-----------------------------------------------------------------------| Susan L. Collins Contact Information: Regulatory Affairs Syva Company - Dade Behring, Inc. 20400 Mariani Avenue Tel: 408 - 366-3840 Fax: 408 - 366-3725 - Date Summary Prepared: 2. June 20, 2001 ### Device Trade Name: 3. Emit® 2000 Valproic Acid Assay ### Common Name: 4. Enzyme Immunoassay, Valproic Acid #### Device Classification Name: 5. The Clinical Chemistry and Clinical Toxicology Devices Panel have classified "Valproic Acid test system" as Class II, 21 CFR Part 862.3645, 91 LEG. #### 6. Intended Use: The Emit® 2000 Valproic Acid Assay is a homogeneous enzyme immunoassay intended for use in the quantitative analysis of Valproic Acid in human serum or plasma. These reagents are packaged specifically for use on a variety of OLYMPUS® analyzers. #### Device Description: 7. The modified assay is similar to the predicate device with minor differences in the packaging of the product. The modified assay has a smaller fill volume of the reagents into different shaped (wedge) reagent bottles. Both the predicate and modified device reagent bottles are made of the same material (HDPE). The modified reagent bottles incorporate a barcode label with assay specific information and are compatible with the OLYMPUS® AU400/600™, AU800/1000™ and AU2700™ Series Analyzers. #### Substantial Equivalence 8. The modified device has the same operating principles, design, manufacturing materials, method of manufacture, assay performance characteristics and intended use as the predicate device. In conclusion the modified Emit® 2000 Valproic Acid Assay is substantially equivalent to the predicate Emit® 2000 Valproic Acid Assay. {1}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three lines forming its body and wings. Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 JUL - 2 2001 Ms. Susan L. Collins Regulatory Affairs Syva Company - Dade Behring Inc. 20400 Mariani Avenue Cupertino, CA 95014 510(k) Number: K011947 Re: Trade/Device Name: Emit® 2000 Valproic Acid Assay Regulation Number: 862.3645 Regulatory Class: II Product Code: LEG Dated: June 20, 2001 Received: June 21, 2001 Dear Ms. Collins: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device mendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations (1 remarket 1 pp . J . be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such periours. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect register. Frease now might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {2}------------------------------------------------ ## Page 2 This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed notification. The I Dri imating of cassion for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and If you desire specific daily of your avec devices), please contact the Office of Compliance at additionally 607.10 for in the claighestions on the promotion and advertising of your device, (301) 594-4508. Additionally, for questions of (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small information on your responsibilities and mober (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Steven Butman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use Statement ### Device Name: Emit® 2000 Valproic Acid Assay 510(k) Number (if known): Kolliqy7 ### Indications for Use: The Emit® 2000 Valproic Acid Assay is a homogeneous enzyme immunoassay intended for use in the quantitative analysis of valproic acid in human serum or plasma. Measurements obtained from this device are used in the diagnosis and treatment of valproic acid overdose or in monitoring levels of valproic acid to ensure appropriate therapy. These reagents are packaged specifically for use on a variety of OLYMPUS® analyzers. Fred Lacy ion of Clinical Laboratory Deviees Ko11947 t) Number (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) 1 Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use_ (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...