← Product Code [LDJ](/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/LDJ) · K990818

# RAPIDONE - MARIJUANA TEST (K990818)

_American Bio Medica Corp. · LDJ · Jul 7, 1999 · Clinical Toxicology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/LDJ/K990818

## Device Facts

- **Applicant:** American Bio Medica Corp.
- **Product Code:** [LDJ](/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/LDJ.md)
- **Decision Date:** Jul 7, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.3870
- **Device Class:** Class 2
- **Review Panel:** Clinical Toxicology

## Indications for Use

'RapidOne' - Marijuana Test is a one-step, lateral flow immunoassay for the detection of marijuana in urine. 'RapidOne' - Marijuana Test is intended for use in the qualitative detection of marijuana metabolite, 11-nor-9-carboxy-delta-9-THC in human urine at 50 ng/ml. 'RapidOne' - Marijuana Test is intended for professional use. It is not intended for over the counter sale to non-professionals. The assay is easy to perform, but should not be used without proper supervision. This immunoassay is a simplified qualitative screening method that provides only a preliminary result for use in determining the need for additional or confirmatory testing, i.e., gas-chromatography/mass spectrometry (GC/MS) or HPLC. 'RapidOne' - Marijuana Test provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a more confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

## Device Story

RapidOne Marijuana Test is a lateral flow, one-step immunoassay for qualitative detection of 11-nor-9-carboxy-delta-9-THC in human urine. Device consists of a membrane strip with immobilized drug conjugate and dried colloidal gold-antibody complex. Principle: competitive immunoassay; drug in urine competes with immobilized drug conjugate for limited antibody binding sites. Negative result: visible line in test area (no drug present). Positive result: absence of color band in test area (drug present). Control line confirms test validity. Intended for professional use in clinical settings; requires supervision. Provides preliminary results; clinical judgment and confirmatory testing (GC/MS) required for definitive diagnosis. Benefits include rapid, simplified screening for drug abuse.

## Clinical Evidence

Bench testing only. Comparison study of 90 urine samples (50 drug-free, 40 positive) against 'Instacheck' and Syva EMIT-II. All 40 positive specimens (39-2200 ng/ml) confirmed by GC/MS. Both immunoassays correctly identified all negative and positive samples. Reproducibility evaluated over 5 days; 100% agreement at 50 ng/ml cutoff.

## Technological Characteristics

Lateral flow immunoassay; membrane strip with immobilized drug conjugate and colloidal gold-antibody complex. Qualitative visual readout (colored bands). No external energy source required. Standalone test device.

## Regulatory Identification

A cannabinoid test system is a device intended to measure any of the cannabinoids, hallucinogenic compounds endogenous to marihuana, in serum, plasma, saliva, and urine. Cannabinoid compounds include delta-9-tetrahydrocannabinol, cannabidiol, cannabinol, and cannabichromene. Measurements obtained by this device are used in the diagnosis and treatment of cannabinoid use or abuse and in monitoring levels of cannabinoids during clinical investigational use.

## Special Controls

*Classification.* Class II (special controls). A cannabinoid test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

## Predicate Devices

- Forefront Diagnostics, Inc. 'Instacheck' Drug Screen -Marijuana Test ([K971959](/device/K971959.md))

## Reference Devices

- Syva EMIT-II

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

JUL - 7 1999

K990818

# 510(k) Summary

### Submitter's Name/Address:

American Bio Medica Corporation 300 Fairview Avenue Hudson, N.Y. 12534

# Contact Person:

Henry Wells Vice President of Product Development Phone: 518-822-8882 Fax: 518-822-0391

## Date of Preparation of this Summary:

Device Trade of Proprietary Name:

Device Common/Usual Name or Classification Name:

June 3, 1999

'RapidOne'- Marijuana Test

Cannabinoid test system

Classification Number/Class

21 C.F.R. § 862.3870 (Class II)

This 510(k) Summary is being submitted in accordance with the requirements of 21 CFR 807.92.

The assigned 510(k) is: K990818

Predicate Device: Forefront Diagnostics, Inc. 'Instacheck' Drug Screen -Marijuana Test (510(k) No. K971959)

## Test Description:

The assay employed in the 'RapidOne'- Marijuana Test is based on the same principle of highly specific reaction between antigens and antibodies.

Each assay is a one-step, immunoassay in which a specially labeled drug (drug conjugate) competes with drug which may be present in the sample for the limited number of binding sites on the antibody. The test device consists of a membrane strip onto which a drug conjugate has been immobilized. A colloidal gold-antibody complex is dried at one end of the membrane. In the absence of any drug in the urine sample, the colloidal goldantibody complex moves with the urine by capillary action to contact the immobilized drug conjugate. An antibody-antigen reaction occurs forming a visible line in the 'test' area. The formation of a visible line in the test area occurs when the test is negative.

When drug is present in the urine sample, the drug or metabolite will compete with the immobilized drug conjugate in the test area for the limited antibody sites on the colloidal gold-antibody complex. If sufficient amount of drug is present, it will fill all of the available binding sites, thus preventing attachment of the labeled antibody to the drug conjugate. An absence of a color band (line) in the test area is indicative of a positive result.

{1}------------------------------------------------

A control band (line), comprised of a different antibody/antigen reaction, is present on the membrane strip. The control line is not influenced by the presence of absence of drug in the urine, and therefore, should be present in all reactions.

A negative urine will produce two colored bands, and a positive sample will produce only one band.

#### IntendedUse:

'RapidOne' - Marijuana Test is used for the qualitative detection of 11-nor-9-carboxydelta-9-THC in human urine. This immunoassay is a simplified qualitative screening method that provides only a preliminary result for use in determining the need for additional or confirmatory testing, i.e., gas chromatography/mass spectrometry (GC/Ms).

### Performance Characteristics:

'RapidOne'-Marijuana Test will detect 50 ng/ml of 11-nor-9-carboxy-delta-9-THC in urine.

'RapidOne'-Marijuana Test was compared to'Instacheck' Drug Screen-Marijuana Test. Ninety (90) samples were selected for evaluation, fifty (50) of which were found to be drug-free and thirty (40) tested as positive by Syva EMIT-II. The forty positive specimens were confirmed and quantified by GC/MS. Both immunoassays correctly identified all of the specimens which contained no drug as negative and determined the 40 drug-containing specimens, ranging in concentration of 39 ng/ml to 2200 ng/ml, to be positive.

Reproducibility was evaluated using control urines containing concentrations above and below the stated cut-off. Negative urines were also used. Each sample, at each concentration of analyte, was tested 4 times, twice daily, for 5 days.

| Conc. (ng/ml) | #  | Result    | Precision |
|---------------|----|-----------|-----------|
| 0             | 40 | 40/40 neg | >99%      |
| 37.5          | 40 | 32/40 pos | 20%       |
| 50.0          | 40 | 40/40 pos | >99%      |
| 75.0          | 40 | 40/40 pos | >99%      |

### Conclusion:

'RapidOne'-Marijuana Test is substantially equivalent to 'Instacheck' - Drug Screen -MarijuanaTest

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name in a circular arrangement around a symbol. The symbol consists of a stylized caduceus, which is a traditional symbol of medicine, with three intertwined snakes and a staff.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

JUL - 7 - 7 - 00

American Bio Medica Corporation c/o Mr. John B. Dubeck Keller And Heckman LLP 1001 G Street, N.W. Suite 500 West Washington, D.C. 20001

Re: K990818 Trade Name: 'RapidOne' - Marijuana Test Regulatory Class: II Product Code: LDJ Dated: June 16, 1999 Received: June 16, 1999

Dear Mr. Dubeck:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

### Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Autman

Steven I. Gutman, M.D. M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

l 1 Page of

K990818 510(k) Number (if known):

'RapidOne' - Marijuana Test Device Name:

Indications For Use:

'RapidOne' - Marijuana Test is a one-step, lateral flow immunoassay for the detection of marijuana in urine. 'RapidOne' - Marijuana Test is intended for use in the qualitative detection of marijuana metabolite, 11-nor-9-carboxy-delta-9-THC in human urine at 50 ng/ml.

'RapidOne' - Marijuana Test is intended for professional use. It is not intended for over the counter sale to non-professionals. The assay is easy to perform, but should not be used without proper supervision. This immunoassay is a simplified qualitative screening method that provides only a preliminary result for use in determining the need for additional or confirmatory testing, i.e., gas-chromatography/mass spectrometry (GC/MS) or HPLC.

'RapidOne' - Marijuana Test provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a more confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

Sean Cooper

(Division Sign-Off)
Division of Clinical Laboratory vices
510(k) Number K990818

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR

1 1

Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/LDJ/K990818](https://fda.innolitics.com/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/LDJ/K990818)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
