CEDIA CYCLOSPORINE PLUS HIGH RANGE CONTROLS 4 AND 5
K030616 · Microgenics Corp. · LAS · Mar 31, 2003 · Clinical Toxicology
Device Facts
Record ID
K030616
Device Name
CEDIA CYCLOSPORINE PLUS HIGH RANGE CONTROLS 4 AND 5
Applicant
Microgenics Corp.
Product Code
LAS · Clinical Toxicology
Decision Date
Mar 31, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3280
Device Class
Class 1
Indications for Use
The CEDIA® Cyclosporine Plus High Range Controls 4 and 5 are in-vitro diagnostic medical devices intended for use as an assayed quality control material to monitor the precision of laboratory testing procedures for cyclosporine.
Device Story
CEDIA® Cyclosporine Plus High Range Controls 4 and 5 are in-vitro diagnostic quality control materials. Used by laboratory personnel to monitor precision of cyclosporine testing procedures. Controls are assayed, providing known values to verify assay performance and ensure accuracy of patient test results. Used in clinical laboratory environments as part of standard quality assurance protocols.
Clinical Evidence
No clinical data provided; device is an in-vitro diagnostic quality control material.
Technological Characteristics
In-vitro diagnostic quality control material; assayed for cyclosporine; liquid form; intended for use with clinical laboratory testing procedures.
Indications for Use
Indicated for use as an assayed quality control material to monitor the precision of laboratory testing procedures for cyclosporine in clinical laboratory settings.
Regulatory Classification
Identification
A clinical toxicology control material is a device intended to provide an estimation of the precision of a device test system and to detect and monitor systematic deviations from accuracy resulting from reagent or instrument defects. This generic type of device includes various single, and multi-analyte control materials.
Related Devices
K032842 — MICROGENICS CYCLOSPORINE CONTROL KIT · Microgenics Corp. · Oct 17, 2003
K051074 — MAS IMMUNOSUPPRESSANT CONTROLS · Microgenics Corp. · Jun 6, 2005
K092993 — ADVIA CENTAUR CYCLOSPORINE MASTER CURVE MATERIAL · Siemens Healthcare Diagnostics · Nov 20, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
MAR 3 1 2003
Rueyming Loor, Ph.D. Senior Director, Research and Development Microgenics Corporation 46360 Fremont Blvd Fremont, CA 94538
k030616 Re: Trade/Device Name: CEDIA® Cyclosporine Plus High Range Controls 4 and 5 Regulation Number: 21 CFR 862.3280 Regulation Name: Clinical toxicology control material Regulatory Class: Class I Product Code: LAS Dated: February 25, 2003 Received: February 26, 2003
Dear Dr. Loor:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820).
{1}------------------------------------------------
Page 2 –
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## ATTACHMENT E
## INDICATIONS FOR USE STATEMENT
KO30616
510(K) Number (if known): ___________________________________________________________________________________________________________________________________________________
Device Name: CEDIA® Cyclosporine Plus High Range Controls 4 and 5
Indications For Use:
The CEDIA® Cyclosporine Plus High Range Controls 4 and 5 are in-vitro diagnostic medical devices intended for use as an assayed quality control material to monitor the precision of laboratory testing procedures for cyclosporine.
T. Can Cooper
(Diyision Sign-Off) Division of Clinical Laboratory De 510(k) Number --
Prescription Use X
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.