N-ACETYLPROCAINAMIDE ASSAY FOR THE BAYER IMMUNO 1 SYSTEM
K971713 · Bayer Corp. · LAN · Jun 11, 1997 · Clinical Toxicology
Device Facts
Record ID
K971713
Device Name
N-ACETYLPROCAINAMIDE ASSAY FOR THE BAYER IMMUNO 1 SYSTEM
Applicant
Bayer Corp.
Product Code
LAN · Clinical Toxicology
Decision Date
Jun 11, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3320
Device Class
Class 2
Indications for Use
This in vitro diagnostic procedure is intended to quantitatively measure N-acetylprocainamide, an antiarrhythmic drug, in human serum or plasma (lithium heparin) using EMIT* (Enzyme Multiplied Immunoassay Technique) technology on a Bayer Immuno I™ system. Measurements of N-acetylprocainamide are used in the diagnosis and treatment of procainamide overdose and in monitoring serum or plasma levels of N-acetylprocainamide to ensure appropriate therapy. This diagnostic method is not intended for use on any other system.
Device Story
In vitro diagnostic assay for quantitative measurement of N-acetylprocainamide (NAPA) in human serum or plasma; utilizes Enzyme Multiplied Immunoassay Technique (EMIT) on Bayer Immuno 1 system; intended for clinical laboratory use to monitor therapeutic levels and manage procainamide overdose; provides concentration values to clinicians for therapeutic decision-making.
Clinical Evidence
Bench testing only. Performance compared to predicate using 99 samples. Correlation: y = 0.99x - 0.03, r = 0.99, Syx = 0.29 µg/mL. Precision (total) reported at 5.5% @ 1.7 µg/mL, 4.3% @ 4.2 µg/mL, and 6.5% @ 8.7 µg/mL. Minimum detectable concentration: 0.11 µg/mL.
Technological Characteristics
In vitro diagnostic immunoassay using EMIT technology. Designed for use on Bayer Immuno 1 system. Analyte: N-acetylprocainamide. Sample matrix: human serum or plasma (lithium heparin).
Indications for Use
Indicated for quantitative measurement of N-acetylprocainamide in human serum or plasma to assist in diagnosis/treatment of procainamide overdose and therapeutic drug monitoring.
Regulatory Classification
Identification
A digoxin test system is a device intended to measure digoxin, a cardiovascular drug, in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of digoxin overdose and in monitoring levels of digoxin to ensure appropriate therapy.
Predicate Devices
Syva EMIT® N-Acetylprocainamide Assay
Related Devices
K971716 — PROCAINAMIDE ASSAY FOR THE BAYER IMMUNO 1 SYSTEM · Bayer Corp. · Jun 27, 1997
K983812 — NAPA EIA ASSAY · Diagnostic Reagents, Inc. · Nov 16, 1998
K032564 — DIMENSION N-ACETYLPROCAINAMIDE (NAPA) FLEX REAGENT CARTRIDGE METHOD, MODEL DF111 · Dade Behring, Inc. · Oct 31, 2003
K011620 — EMIT 2000 N-ACETYLPROCAINAMIDE ASSAY, MODEL OSR4N229 · Syva Co., Dade Behring, Inc. · Jun 18, 2001
Submission Summary (Full Text)
{0}------------------------------------------------
## SUMMARY OF SAFETY AND EFFECTIVENESS
## N-Acetylprocainamide Assay for Bayer Immuno 1° System
Listed below is a comparison of the performance between the Immuno 1 N-Acetylprocainamide (NAPA) assay and a Elsstor of a confed is a confed clearance of substantial equivalence (Syva EMIT® N-Acetylprocainamide Assay, Behring Diagnostics Inc.). The information used in the Summary of Safety and Effectiveness was extracted from the Immuno I N-Acetylorocainamide Method Sheet and the Syva EMIT® N-Acetylprocainamide Assay Insert Sheet.
## INTENDED USED
This in vitro method is intended to quantitatively measure n-acetylprocainamide, the pharmacologically active meabolite for procainamide, an antiarrhythmic drug, in human serum or plasma (lithium heparin) using Syva EMIT® N-Acetylprocainamide Assay on a Bayer Immuno-1 system. Measurements of n-acetylprocainamide are used in the diagnosis and treatment of procainamide overdose and in monitoring levels of n-acetylprocainanide to ensure appropriate therapy.
| METHOD | Immuno 1 NAPA Assay | Syva EMIT® NAPA Assay<br>(predicate Device) |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
| Part No. | T01-3990-51 | 4N024UL |
| Minimum Detectable Conc. | 0.11 µg/mL | 0.25 µg/mL |
| Precision | (Total)<br>5.5% @ 1.7 µg/mL<br>4.3% @ 4.2 µg/mL<br>6.5% @ 8.7 µg/mL | (Between-Run)<br>3.9% @ 1.7 µg/mL<br>4.5% @ 4.5 µg/mL<br>4.5% @ 10.8 µg/mL |
| Correlation | y = 0.99x - 0.03<br>where<br>y = Immuno 1 N-Acetylprocainamide Assay<br>x = Syva EMIT® N-Acetylprocainamide Assay*<br>n = 99<br>r = 0.99<br>Syx = 0.29 µg/mL | |
*This assay was performed on COBAS FARA II" Instrument using parameters and protocol specified in Behring Application Sheet.
Gabriel J. Murray, Jr.
Gabriel J. Muraca, Jr. Manager Regulatory Affairs Bayer Corporation 511 Benedict Avenue Tarrytown, New York 10591-5097
ﺳﮯ
5/7/97
---
Date
{1}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human figures, represented by flowing lines, suggesting a sense of movement and connection.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
JUN 111 1997
Gabriel J. Muraca, Jr. Manager Regulatory Affairs Bayer Corporation 511 Benedict Avenue Tarrytown, New York 10591-5097
K971713 Re :
N-Acetylprocainamide Assay for the Bayer Immuno 1™ System 发 Regulatory Class: II Product Code: LAN Dated: May 7, 1997 Received: May 9, 1997
Dear Mr. Muraca:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
{2}------------------------------------------------
Paqe 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
" If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven Sitman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Page 1 of 1
510(k) Number (if known):
N-Acetylprocainamide (NAPA) Device Name:_
Indications For Use:
This in vitro diagnostic procedure is intended to quantitatively measure N-acetylprocainamide, an antiarrhythmic drug, in human serum or plasma (lithium heparin) using EMIT* (Enzyme Multiplied Immunoassay Technique) technology on a Bayer Immuno I™ system. Measurements of N-acetylprocainamide are used in the diagnosis and treatment of procainamide overdose and in monitoring serum or plasma levels of N-acetylprocainamide to ensure appropriate therapy. This diagnostic method is not intended for use on any other system.
## T WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF
| | Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|------------------------------------------|--------------------------------------------------------|--------------------------|
| | (Division Sign-Off) | |
| | Division of Clinical Laboratory Devices | |
| 510(k) Number | K171713 | |
| Prescription Use<br>(Per 21 CFR 801.109) | OR | Over-The-Counter Use |
| | | (Optional Format 1-2-96) |
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.