OMEGA LABORATORIES HAIR DRUG SCREENING ASSAY FOR METHAMPHETAMINE AND 4-METHYLENEDIOXYMETHAMPHETAMINE (MDMA)
Device Facts
| Record ID | K101973 |
|---|---|
| Device Name | OMEGA LABORATORIES HAIR DRUG SCREENING ASSAY FOR METHAMPHETAMINE AND 4-METHYLENEDIOXYMETHAMPHETAMINE (MDMA) |
| Applicant | Omega Laboratories, Inc. |
| Product Code | LAF · Clinical Toxicology |
| Decision Date | Aug 16, 2011 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.3610 |
| Device Class | Class 2 |
Indications for Use
The Omega Laboratories Hair Drug Screening Assay for Methamphetamine and MDMA is a laboratory developed test that is intended to be used for the determination of the presence of Methamphetamine and 3,4-Methylenedioxymethamphetamine (MDMA) in human hair from the crown of the head. The Omega Laboratories Hair Drug Screening Assay (AMP) utilizes International Diagnostics Systems Corp One-Step Methamphetamine ELISA for Hair Testing Kit for the qualitative detection of Methamphetamine and MDMA at or above 500 pg/mg of hair for the purpose of identifying the use of Methamphetamine and MDMA. D-Methamphetamine is used as the standard for the assay. To confirm a screen positive result a more specific alternate chemical method, such as Gas Chromatography/Mass Spectrometry (GC/MS) operating in the selected ion monitoring (SIM) mode is the preferred method with deuterated internal standards. Professional judgment should be applied to any drug of abuse test result, particularly when presumptive positive results are obtained.
Device Story
Laboratory-based screening assay for Methamphetamine and MDMA in human hair; utilizes ELISA reagents and micro-plate reader; qualitative detection at 500 pg/mg cutoff; intended for professional laboratory use; results require confirmation via GC/MS; aids in identifying drug use; clinical decision-making relies on professional judgment of presumptive positive results.
Clinical Evidence
Bench testing only. Performance studies evaluated precision, agreement, cosmetic treatment effects, cross-reactivity, and environmental contamination. Results demonstrated substantial agreement with the predicate device.
Technological Characteristics
Enzyme-linked immunosorbent assay (ELISA) reagents; micro-plate reader; qualitative detection; cutoff 500 pg/mg; utilizes International Diagnostics Systems Corp One-Step Methamphetamine ELISA kit.
Indications for Use
Indicated for qualitative detection of Methamphetamine and MDMA in human head hair at or above 500 pg/mg to identify drug use.
Regulatory Classification
Identification
A methamphetamine test system is a device intended to measure methamphetamine, a central nervous system stimulating drug, in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of methamphetamine use or overdose.
Special Controls
*Classification.* Class II (special controls). A methamphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Predicate Devices
- Quest Diagnostics HairCheck-DT (Amphetamines) (k051161)
Related Devices
- K142855 — Omega Laboratories Hair Drug Screening Assay Methamphetamine (Meth) and -3, 4-methylenedioxymethamphetamine (MDMA) · Omega Laboratories, Inc. · Jul 28, 2015
- K051161 — QUEST DIAGNOSTICS HAIRCHECK-DT (AMPHETAMINES) · Quest Diagnostics, Inc. · Nov 22, 2005