AMEDICA DRUG SCREEN METHAPHETAMINE TEST

K023101 · Amedica Biotech, Inc. · LAF · Dec 16, 2002 · Clinical Toxicology

Device Facts

Record IDK023101
Device NameAMEDICA DRUG SCREEN METHAPHETAMINE TEST
ApplicantAmedica Biotech, Inc.
Product CodeLAF · Clinical Toxicology
Decision DateDec 16, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3610
Device ClassClass 2

Indications for Use

The Amedica Drug Screen Methamphetamine Test is an in vitro diagnostic test for the rapid detection of methamphetamine in human urine at a cutoff of 1000 ng/ml. This test kit is used to obtain a visual, qualitative result and is intended for professional use. It is not intended for over the counter sale. This assay provides only a preliminary result. A more specific alternative chemical method is needed to obtain a confirmed result.

Device Story

Lateral flow, competitive immunoassay for methamphetamine detection in human urine; utilizes colloidal gold particles. Urine specimen migrates via capillary action; methamphetamine competes with immobilized conjugates for antibody binding sites. Negative result: colored line appears in test region (insufficient drug to saturate antibodies). Positive result: no colored line in test region (drug concentration >1000 ng/ml). Built-in control line confirms test validity. Intended for professional use in laboratory settings; provides preliminary results requiring confirmation by alternative chemical methods (e.g., GC/MS).

Clinical Evidence

Clinical performance evaluated via correlation study using blind-labeled clinical specimens compared against GC/MS. Study demonstrated >94% agreement with GC/MS results. Additional clinical site study performed at two certified laboratories confirmed suitability for professional use.

Technological Characteristics

Membrane-based lateral flow competitive immunoassay; colloidal gold label; 1000 ng/ml cutoff; visual qualitative readout; manual operation; no electronic components or software.

Indications for Use

Indicated for professional use for the rapid, qualitative detection of methamphetamine in human urine at a 1000 ng/ml cutoff. Not for point-of-care or over-the-counter use.

Regulatory Classification

Identification

A methamphetamine test system is a device intended to measure methamphetamine, a central nervous system stimulating drug, in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of methamphetamine use or overdose.

Special Controls

*Classification.* Class II (special controls). A methamphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K023/01 ## 3. Summary of Safety and Effectiveness ( As required by 21CFR 807.92(c)) DEC 1 6 2002 Device name: Amedica Drug Screen Methamphetamine Test Design and Materials: Membrane based one-step, lateral flow, competitive immunoassay use colloidal gold for visual detection. The test cutoff is1000 ng/ml. Intended Use: The Amedica Drug Screen Methamphetamine Test is a immunochromatographic assay for the rapid detection of methamphetamine in human urine at a cutoff concentration of 1000 ng/ml. This assay has not been evaluated at point-of-care locations and is intended for use by healthcare professionals. This assay provides only a preliminary result. A more specific alternative chemical method is needed to obtain a confirmed result. Test Principle and Description: The Amedica Drug Screen Methamphetamine Test is based on the principle of highly specific competitive immunochemical reactions between antigens and antibodies for the analysis of specific substances in urine. During testing, a urine specimen moves along membrane on the strip by capillary action. When methamphetamine concentration in the urine is below 1000 ng/ml, it is not enough to saturate all of the binding sites of the antibody-coated colored particles in the test strip. The unsaturated antibody-coated particles will then be captured by methamphetamine conjugates immobilized on the strip and a colored line will appear in the test line region. The test result is negative. If the methamphetamine level is above 1000 ng/ml, it is sufficient to occupy all of the binding sites on the antibody-coated particles. The saturated antibody-coated particles will not be captured by methamphetamine conjugate coated on the strip. The colored line will not form in the test region. The test result is positive. The device also provides a built-in control with a different antibody reaction at the control region. This control line should always appear whether or not the drugs or metabolites are present. If the control line does not appear the test cartridge should be discarded. This means that negative urine will produce two colored bands, and positive urine will produce only one band at control region Performance: The product performance was evaluated by correlation study using blind-labeled clinical specimens that have been measured by GC/MS. This study produced > 94% agreement with GC/MS results. In addition, clinical site study was performed at two certified laboratories and demonstrated that Amedica Biotech Drug Screen Methamphetamine Test can be use by professionals to obtain a visual, qualitative detection of drugs of abuse. The results of these study and comparison with Rapid Diagnostics methamphetamine test demonstrated that Amedica Biotech Drug Screen Methamphetamine Test is substantially equivalent to the predicate kit. Manufacturer: Amedica Biotech, Inc. 28301 Industrial Blvd. Suite K Hayward, CA 94545 Phone: (510) 785-5980 {1}------------------------------------------------ Fax: (510) 785-5973 Predicate kit: Rapid Methamphetamine Test Rapid Diagnostics, Inc. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name arranged in a circular fashion around a symbol. The symbol consists of three stylized human profiles facing to the right, stacked on top of each other. ## DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Mr. Jeff Chen President Amedica Biotech Inc. 28301 Industrial Blvd. - Suite K Havward. CA 94545 FC 1 6 2002 k023101 Trade/Device Name: Amedica Drug Screen Methamphetamine Test Regulation Number: 21 CFR 862.3610 Regulation Name: Methamphetamine test system Regulatory Class: Class II Product Code: LAF Dated: November 12, 2002 Received: November 19, 2002 Dear Mr. Chen: Re: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that vour device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820). {3}------------------------------------------------ ## Page 2 This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Steven Sutman Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Page 510(k) Number (if known): K023101 Amedica Drug Screen Methamphetamine Test Device Name: Indications For Use: The Amedica Drug Screen Methamphetamine Test is an in vitro diagnostic test for the rapid detection of methamphetamine in human urine at a cutoff of 1000 ng/ml. This test kit is used to obtain a visual, qualitative result and is intended for professional use. It is not intended for over the counter sale. This assay provides only a preliminary result. A more specific alternative chemical method is needed to obtain a confirmed result. Alan Cooper (Division Sign-Off) Division of Clinical Laboratory 510(k) Number (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) ✓ OR Over-The-Counter Use (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...