QMS LIDOCAINE

K090282 · Thermo Fisher Scientific, Inc. · KLR · May 29, 2009 · Clinical Toxicology

Device Facts

Record IDK090282
Device NameQMS LIDOCAINE
ApplicantThermo Fisher Scientific, Inc.
Product CodeKLR · Clinical Toxicology
Decision DateMay 29, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3555
Device ClassClass 2

Indications for Use

The QMS Lidocaine Immunoassay is intended for the quantitative determination of lidocaine in human serum or plasma on automated clinical chemistry analyzers. The results obtained are used in the diagnosis and treatment of lidocaine overdose and in monitoring levels of lidocaine to help ensure appropriate therapy.

Device Story

QMS Lidocaine Immunoassay; homogeneous particle-enhanced turbidimetric immunoassay. Input: human serum or plasma samples. Principle: antibody-binding competition between drug in sample and microparticle-bound drug; antibody binding causes microparticle agglutination. Output: rate of absorbance measured spectrophotometrically, proportional to agglutination rate, indicating lidocaine concentration. Used in clinical chemistry laboratories on automated analyzers (e.g., Hitachi 717). Healthcare providers use quantitative results to adjust lidocaine therapy or manage overdose. Benefits: enables precise therapeutic drug monitoring to ensure efficacy and safety.

Clinical Evidence

Bench testing only. Precision evaluated per CLSI EP5-A2 (N=80 per level, CV 5.06-8.13%). Linearity/reportable range 0.8-10 μg/mL (R²=0.9968). Analytical specificity tested against metabolites (MEGX, GX) and 36 common drugs; cross-reactivity <1% for most, except Mepivacaine (63%) and Bupivacaine (20%). Method comparison with predicate (N=52) showed R=0.993, slope 1.076. Matrix comparison confirmed compatibility with various collection tubes.

Technological Characteristics

Homogeneous particle-enhanced turbidimetric immunoassay. Reagents: R1 (anti-lidocaine mouse monoclonal antibody, <0.1% sodium azide), R2 (lidocaine-coated microparticles, <0.1% sodium azide). Energy source: spectrophotometric measurement. Form factor: liquid reagents for automated clinical chemistry analyzers. Connectivity: integrated into automated analyzer workflow. Sterilization: N/A (reagents).

Indications for Use

Indicated for patients requiring monitoring of lidocaine levels for therapeutic management or diagnosis of lidocaine overdose. No specific age or gender restrictions provided.

Regulatory Classification

Identification

A lidocaine test system is a device intended to measure lidocaine, an antiarrythmic and anticonvulsant drug, in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of lidocaine overdose or in monitoring levels of lidocaine to ensure appropriate therapy.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE A. 510(k) Number: k090282 B. Analyte: Lidocaine C. Type of Test: Turbidimetric immunoassay D. Applicant: Thermo Fisher Scientific Inc., E. Proprietary and Established Names: QMS Lidocaine F. Regulatory Information: 1. Regulation section: 21 CFR 862.3555 2. Classification: Class II 3. Product code: KLR 4. Panel: Toxicology H. Intended Use: 1. Intended use(s): The QMS Lidocaine Immunoassay is intended for the quantitative determination of lidocaine in human serum or plasma on automated clinical chemistry analyzers. The results obtained are used in the diagnosis and treatment of lidocaine overdose and in monitoring levels of lidocaine to help ensure appropriate therapy. 2. Indication(s) for use: See Intended use above. 3. Special conditions for use statement(s): For prescription use only. 4. Special instrument requirements: The assay has been validated by the manufacturer on the Hitachi 717. {2} Page 2 of 8 I. Device Description: The QMS Lidocaine Immunoassay assay consists of the packaged reagents (R1 and R2): The R1 Antibody Reagent includes Anti-lidocaine Mouse Monoclonal Antibody and &lt;0.1% sodium azide as preservative. The R2 Microparticles include Lidocaine-coated Microparticles and &lt;0.1% sodium azide as preservative. This product contains human sourced and/or potentially infectious components. Components sourced from human blood have been tested by FDA-approved methods and found to be nonreactive for HBsAg, anti-HIV 1/2, and anti-HCV. J. Substantial Equivalence Information: 1. Predicate device name(s): Roche ONLINE TDM Lidocaine Assay 2. Predicate 510(k) number(s): k032334 3. Comparison with predicate: The assays have the same indications for use, and use the same technology and reagents. Both assays are for use on automated analyzers. K. Standard/Guidance Document Referenced (if applicable): CLSI document: “Evaluation of Precision Performance of Quantitative Measurement Methods, Approved Guideline-Second Edition (EP05-A2). L. Test Principle: The assay is a homogeneous particle-enhanced turbidimetric assay, based on antibody-binding competition between drug in the sample and microparticle-bound drug. Antibody binding causes microparticle agglutination. The rate of absorbance, measured spectrophotometrically, is proportional to the rate of agglutination. M. Performance Characteristics (if/when applicable): Performance was validated on the Hitachi 717 instrument, using the Roche Preciset TDM calibrators (A-F) (k031856) and the Roche TDM Control Set (k060429). 1. Analytical performance: 2 {3} Page 3 of 8 a. Precision Precision was performed in-house, using the protocol described in CLSI EP5-A2. Commercially available tri-level human serum control set (k060429) containing lidocaine was assayed in duplicate, twice a day for 20 non-consecutive days on one analyzer, by one operator, using one reagent and calibrator lot. Each control sample was tested for a total of 80 replicates. Calibration was performed initially and re-calibration was performed at the day 10 data point. Results are summarized below. | | | | Within Run | | Between Day | | Total | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Control | N | Mean (μg/ml) | SD (ug/mL) | %CV | SD (ug/mL) | %CV | SD (ug/mL) | %CV | | Low | 80 | 1.76 | 0.10 | 5.66 | 0.10 | 5.41 | 0.14 | 8.13 | | Mid | 80 | 4.42 | 0.15 | 3.48 | 0.11 | 2.44 | 0.22 | 5.06 | | High | 80 | 8.83 | 0.24 | 2.73 | 0.19 | 2.17 | 0.46 | 5.27 | b. Linearity/assay reportable range: Recovery was determined by spiking lidocaine into human serum negative for the drug to achieve concentrations across the assay range. The theoretical concentration of lidocaine (in the table below) was determined gravimetrically. The average of replicates measured with the QMS reagents were used to determine mean recovery. Results are shown below. | Theoretical Conc. (μg/mL) | Mean Recovered Conc. (μg/mL) | % Recovery | | --- | --- | --- | | 9.33 | 8.57 | 91.9 | | 4.67 | 4.29 | 91.9 | | 2.33 | 2.12 | 91.0 | Additional linearity and recovery studies across the claimed assay range (0.75 to 10 μg/mL) were performed by diluting commercially available serum based calibrator (up to 10.0 μg/mL) with drug-free or lower concentration calibrators. These studies and the limit of quantitation study below supported the claimed assay range of 0.8 to 10 μg/mL. The percent recovery across the range of the assay observed in these studies was 93 - 99%. Results of linear regression analysis were: Slope: 0.961 Intercept: -0.04 R² = 0.9968 3 {4} Page 4 of 8 c. Traceability, Stability, Expected values (controls, calibrators, or methods): Controls and calibrators were previously cleared (k060429 and k031856) and are sold separately. d. Detection limit: A patient pool with a lidocaine value of 0.93 µg/mL was diluted with normal human serum negative for lidocaine to give the following theoretical values: 0.154, 0.309, 0.463, 0.617, and 0.771 µg/mL. Each level was assayed twice a day for 5 days. The mean concentration, the standard deviation and the coefficient of variations (CV) were calculated. In addition, recovery studies with calibrator-derived samples containing concentrations of lidocaine near the assay lower limit, demonstrated recovery within 93-95% relative to theoretical calibrator-derived concentrations. The limit of quantitation was determined to be 0.8 ug/mL, where the assay has a CV less than 20% and recovery is within +/- 10% (see table below for data): | mean measured sample concentration (µg/mL) | 0.79 | 0.93 | | --- | --- | --- | | SD (µg/mL) | 0.107 | 0.094 | | % CV | 13.58% | 10.17% | e. Analytical specificity: Cross-Reactivity: The assay was evaluated for potential interference from the N-ethylglycyl-2,6-xylidide (MEGX) and glycyl-2,6-xylidide (GX), the two major in vivo metabolites produced by the liver. Metabolites of &gt;99% purity were spiked at 10 times the highest known concentrations of each into a human serum control containing lidocaine at a concentration of approximately 4 µg/ml. Control samples contained the same concentration of lidocaine, with no metabolite. Percent cross reactivity for metabolites and drugs were defined as [(Da-Dt)/C] x 100, where Dt = the measured concentration of the control analyte, Da = measured concentration of the control analyte + cross-reactant and C = known concentration cross reactant. Calculated cross-reactivities are based on the median of triplicate determinations. Less than 1% cross-reactivity was observed for the MEGX and GX. Endogenous Compounds: To evaluate interference from endogenous compounds, a series of dilutions containing varying levels of the endogenous compounds was prepared from spiked and negative serum pools. Testing was done in the presence of 5-7 4 {5} Page 5 of 8 $\mu \mathrm{g} / \mathrm{ml}$ lidocaine (except for human anti-mouse antibodies (HAMA) testing, in which samples contained $\sim 5~\mu \mathrm{g} / \mathrm{ml}$ lidocaine). Percent recovery was calculated relative to control samples containing lidocaine without spiked endogenous compounds. The mean observed concentration was used in calculation of recovery. The following is a summary of the obtained data. | Endogenous Substance tested | Endogenous substance Concentration | N | Target (No Interferent) μg/mL | Mean Observed Recovery μg/mL | % Recovery | | --- | --- | --- | --- | --- | --- | | Bilirubin | 15 mg/dL | 2 | 7.25 | 6.63 | 91.4 | | Hemoglobin | 10 g/L | 2 | 5.28 | 5.37 | 102.0 | | Triglyceride | 2000 mg/dL | 3 | 5.24 | 5.12 | 97.84 | | Rheumatoid Factor | 1500 IU | 3 | 5.21 | 5.13 | 98.53 | | Total protein | 1-12 g/dL | 3 | 5.90 | 6.00 | 103 | A normal human serum pool (control), and HAMA type 1 and HAMA type 2 samples was spiked with the same amounts of lidocaine. Each of the samples was assayed in duplicate. Results are shown in the table below. As with any assay employing mouse antibodies, the possibility exists for interference by HAMA in the sample, which could cause falsely elevated results. | | Rep 1 (μg/mL) | Rep 2 (μg/mL) | Mean Recovery (μg/mL) | SD | CV | % Recovery | | --- | --- | --- | --- | --- | --- | --- | | Control | 4.47 | 4.24 | 4.36 | 0.163 | 3.7 | - | | HAMA Type-1 | 4.08 | 3.98 | 4.03 | 0.071 | 1.8 | 92.0 | | HAMA Type-2 | 4.02 | 4.03 | 4.03 | 0.007 | 0.2 | 92.0 | Thirty-six common drugs and compounds were tested for interference. Results are shown below. ND indicates not detectable. Cross-reactivity of $&lt; 1\%$ was considered ND. 5 {6} Page 6 of 8 | Compound | Concentration Tested (μg/mL) | % Cross-Reactivity | | --- | --- | --- | | Mepivacaine | 10 | 63 | | Bupivacaine | 40 | 20 | | Acetaminophen | 200 | ND | | Acetyl Cysteine | 1660 | ND | | Acetylsalicylic acid | 650 | ND | | Ampicillin-Na | 53 | ND | | Asorbic Acid | 60 | ND | | Cefoxitin | 660 | ND | | Cyclosporine | 0.5 | ND | | Digoxin | 0.01 | ND | | Disopyramide | 10 | ND | | d-Methamphetamine | 10 | ND | | Ephedrine | 0.1 | ND | | Flecainide | 10 | ND | | Furosemide | 60 | ND | | Hydrochlorothiazide | 6 | ND | | Ibuprofen | 500 | ND | | Isoproterenol | 0.01 | ND | | Levodopa | 80 | ND | | Lidocaine-N-ethyl Bromide | 100 | ND | | Methylodopa+1,5 | 15 | ND | | Metronidazole | 120 | ND | | Mexiletine | 100 | ND | | Phenybutazone | 100 | ND | | Phenytoin (DPH) | 50 | ND | | PPX (L-Pipecolic acid-2,6-xylidide) | 10 | 3.4 | | Procainamide | 24 | ND | | Propanolol | 2.0 | ND | | Quinidine | 12 | ND | | Rifampicin | 64 | ND | | Tetracycline | 15 | ND | | Theophylline | 40 | ND | | Tocainide | 100 | 1.0 | ND = Not Detectable f. Assay cut-off: Not applicable 6 {7} Page 7 of 8 2. Comparison studies: (outliers, repeat measurements, graphs) a. Method comparison with predicate device: Results obtained with the QMS Lidocaine Assay were compared to those obtained with the same serum and plasma samples using the Roche Lidocaine assay. No other specific criteria (in terms of patient demographics) were applied in sample selection. 52 samples were evaluated by the QMS assay on a Hitachi 717 analyzer in accordance with the procedure described in the proposed package insert. Single measurements for each method were used in the analysis. The samples ranged in concentration on the Hitachi 717 from 1.11 µg/mL to 9.38 µg/mL. Results of Passing-Bablok analysis were: Slope: 1.076 [1.044 to 1.104] Intercept: -0.096 [-0.153 to -0.024] R = 0.993 R² = 0.986 b. Matrix comparison: Recovery studies for serum and plasma samples were performed using blood in various collection tubes from healthy volunteers, including: serum/glass tube; no additive, serum with clot activator, lithium heparin tube, sodium heparin tube and K2 EDTA tube. The serum or plasma was spiked with four different concentrations of lidocaine (1.8 µg/mL, 5 µg/mL, 8 µg/mL and 10 µg/mL). Samples were analyzed on day 0, 3 and 7. The percent recovery of lidocaine in serum and plasma was within +/-97.50 or better. In addition, samples containing high concentration of anticoagulants were spiked and tested with lidocaine at 1.8, 5 and 8 µg/mL. The results showed no interference with the QMS Lidocaine assay. 3. Clinical studies: a. Clinical Sensitivity: Not applicable. b. Clinical specificity: Not Applicable. c. Other clinical supportive data (when a. and b. are not applicable): Not Applicable. 4. Clinical cut-off: Not Applicable. 5. Expected values/Reference range: The manufacturer includes the following in the package insert: 7 {8} Page 8 of 8 "The therapeutic range for total concentration is stated as 1.5 to 6 ug/mL."* "Each laboratory should investigate the transferability of the expected values to its own patient population and if necessary determine its own reference ranges." * Burtis CA, et. al. Tietz Textbook of Clinical Chemistry and Molecular Diagnostics. 4th ed. Elsevier Saunders; 2006:1259. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports substantial equivalence decision. 8
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...