ACON PPX ONE STEP PROPOXYPHENE TEST STRIP AND ACON PPX STEP PROPOXYPHENE TEST DEVICE
K040445 · ACON Laboratories, Inc. · JXN · May 19, 2004 · Clinical Toxicology
Device Facts
Record ID
K040445
Device Name
ACON PPX ONE STEP PROPOXYPHENE TEST STRIP AND ACON PPX STEP PROPOXYPHENE TEST DEVICE
Applicant
ACON Laboratories, Inc.
Product Code
JXN · Clinical Toxicology
Decision Date
May 19, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3700
Device Class
Class 2
Indications for Use
The ACON® One Step Propoxyphene Test Device and Test Strip are enzyme immunoassays for the qualitative determination of propoxyphene in human urine. The assay has a cutoff concentration of 300 ng/mL. The device is for in vitro diagnostic use. The device is for prescription use. The ACON® One Step Propoxyphene Test Device and Test Strip provide only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/Mass spectrometry is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used. The assay is intended for use in point-of-care settings.
Device Story
Lateral flow immunochromatographic assay for qualitative screening of Propoxyphene in urine; utilizes mouse monoclonal antibodies for competitive binding. Input: urine sample; Output: visual colored line in test region (negative) or absence of line (positive). Procedural control line confirms proper wicking. Used at point-of-care sites by healthcare professionals. Provides preliminary analytical results; requires confirmation via GC/MS. Benefits patient by enabling rapid, on-site drug screening to inform clinical decision-making.
Clinical Evidence
No clinical studies performed. Performance established via bench testing: precision (3 days, 3 operators, 3 lots), analytical specificity (cross-reactivity and interference testing), and method comparison (n=352 clinical samples). Method comparison against GC/MS showed 94% positive agreement and 99% negative agreement.
Technological Characteristics
Lateral flow immunochromatographic assay; competitive binding principle; mouse monoclonal antibody-based detection. Form factor: test strip or test device. No instrumentation required. Qualitative visual readout. Cutoff: 300 ng/mL.
Indications for Use
Indicated for qualitative detection of propoxyphene in human urine at 300 ng/mL cutoff. Intended for prescription, point-of-care professional use as a preliminary screening test.
Regulatory Classification
Identification
A propoxyphene test system is a device intended to measure propoxyphene, a pain-relieving drug, in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of propoxyphene use or overdose or in monitoring levels of propoxyphene to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). A propoxyphene test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Submission Summary (Full Text)
{0}
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE
A. 510(k) Number:
K040445
B. Analyte:
Propoxyphene
C. Type of Test:
Qualitative immunoassay
D. Applicant:
Acon Laboratories, Inc.
E. Proprietary and Established Names:
ACON® One Step Propoxyphene Test Device and Test Strip
F. Regulatory Information:
1. Regulation section:
21 CFR § 862.3700
2. Classification:
Class II
3. Product Code:
JXN
4. Panel:
Toxicology (91)
G. Intended Use:
1. Intended use(s):
Refer to Indications for use.
2. Indication(s) for use:
The ACON® One Step Propoxyphene Test Device and Test Strip
are enzyme immunoassays for the qualitative determination of propoxyphene
in human urine. The assay has a cutoff concentration of 300 ng/mL.
The device is for in vitro diagnostic use.
The device is for prescription use.
3. Special condition for use statement(s):
The ACON® One Step Propoxyphene Test Device and Test Strip provide only
a preliminary analytical test result. A more specific alternative chemical
method must be used to obtain a confirmed analytical result. Gas
{1}
Page 2 of 16
chromatography/Mass spectrometry is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.
The assay is intended for use in point-of-care settings.
4. Special instrument Requirements:
Not applicable. The device is a visually read single-use device.
H. Device Description:
This submission includes both a single-use dipstick (referred to as a test strip) and a cassette (referred to as a test device). With the dipstick, operators dip the test strip into the urine and the reaction is initiated by movement of the sample through the test strip. With the cassette, operators add several drops of the sample to the sample well. The test reaction is initiated by movement of the sample through the test strip.
I. Substantial Equivalence Information:
1. Predicate device name(s):
Instant-View® Propoxyphene (PPX) Urine Test Cassette
2. Predicate K number(s):
K022915
3. Comparison with predicate:
Both devices are for the qualitative determination of the same analyte in the same matrix, and utilize the same cutoff concentration. Both are visually-read single use devices.
| Similarities | | |
| --- | --- | --- |
| Item | Devices (Dipstick and Cassette) | Predicate |
| Methodology | Same | Lateral flow
immunochromatographic |
| Type of assay | Same | Qualitative |
| Matrix | Same | Urine |
| Cutoff | Same | 300 ng/mL |
| Intended user | Same | Professional use |
| Endpoint | Same | Colored lines |
| Differences | | |
| Item | Device (Dipstick and Cassette) | Predicate |
| Read time | 5 minutes | 4 – 7 minutes |
{2}
Page 3 of 16
## J. Standard/Guidance Document Referenced (if applicable):
The sponsor referenced the following guidance document(s) in their submission:
Premarket Submission and Labeling Recommendations for Drugs of Abuse Screening Tests - Draft Guidance for Industry and FDA Staff, published December 2003
The sponsor indicated deviation from this guidance in regards to interference testing.
## K. Test Principle:
The test employs lateral flow immunochromatographic technology.
Drug in the sample and drug-labeled conjugate (containing a chromagen) compete for antibody binding sites in the test area of the test strip. Binding of drug in the sample causes the absence of a line at the test area, i.e., a positive result. When drug is not present in the sample, the drug-labeled conjugate binds at the test line, resulting in formation of a line, i.e., a negative result. The absence or presence of the line is determined visually by the operator.
The test antibody consists of monoclonal mouse antibody against propoxyphene.
The internal process control consists of goat anti-rabbit IgG antibody as the stationary phase and rabbit IgG-colloidal gold conjugate as the mobile phase.
## L. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Specimen description: drug free urine spiked with propoxyphene
Number of days: three
Replicates per day: ten
Lots of product used: three
Number of operators: three
Operator: laboratorian
Testing Facility: manufacturer's laboratory
Results of the study are presented below:
{3}
Page 4 of 16
Propoxyphene Precision Study Results, Lot 1, Day 1 – Test Strip
| Concentration of sample, ng/mL | Number of determinations | Results # Neg/ #Pos |
| --- | --- | --- |
| 0 | 30 | 30/0 |
| 150 | 30 | 30/0 |
| 225 | 30 | 24/6 |
| 375 | 30 | 7/23 |
| 450 | 30 | 0/30 |
| 600 | 30 | 0/30 |
Propoxyphene Precision Study Results, Lot 2, Day 2 – Test Strip
| Concentration of sample, ng/mL | Number of determinations | Results # Neg/ #Pos |
| --- | --- | --- |
| 0 | 30 | 30/0 |
| 150 | 30 | 30/0 |
| 225 | 30 | 25/5 |
| 375 | 30 | 8/22 |
| 450 | 30 | 0/30 |
| 600 | 30 | 0/30 |
Propoxyphene Precision Study Results, Lot 3, Day 3 – Test Strip
| Concentration of sample, ng/mL | Number of determinations | Results # Neg/ #Pos |
| --- | --- | --- |
| 0 | 30 | 30/0 |
| 150 | 30 | 30/0 |
| 225 | 30 | 26/4 |
| 375 | 30 | 7/23 |
| 450 | 30 | 0/30 |
| 600 | 30 | 0/30 |
Propoxyphene Precision Study Results, Lot 1, Day 1 – Test Device
| Concentration of sample, ng/mL | Number of determinations | Results # Neg/ #Pos |
| --- | --- | --- |
| 0 | 30 | 30/0 |
| 150 | 30 | 30/0 |
| 225 | 30 | 26/4 |
| 375 | 30 | 8/22 |
| 450 | 30 | 0/30 |
| 600 | 30 | 0/30 |
{4}
Page 5 of 16
Propoxyphene Precision Study Results, Lot 2, Day 2 – Test Device
| Concentration of sample, ng/mL | Number of determinations | Results # Neg/ #Pos |
| --- | --- | --- |
| 0 | 30 | 30/0 |
| 150 | 30 | 30/0 |
| 225 | 30 | 26/4 |
| 375 | 30 | 7/23 |
| 450 | 30 | 0/30 |
| 600 | 30 | 0/30 |
Propoxyphene Precision Study Results, Lot 3, Day 3 – Test Device
| Concentration of sample, ng/mL | Number of determinations | Results # Neg/ #Pos |
| --- | --- | --- |
| 0 | 30 | 30/0 |
| 150 | 30 | 30/0 |
| 225 | 30 | 27/3 |
| 375 | 30 | 7/23 |
| 450 | 30 | 0/30 |
| 600 | 30 | 0/30 |
b. Linearity/assay reportable range:
Not applicable. The assay is intended for qualitative use.
c. Traceability (controls, calibrators, or method):
Control materials are required but are not specifically identified in the labeling. The sponsor’s Package Insert directs the user to follow local, state, and federal guidelines for testing QC materials and to process the controls in the same manner as unknown samples
The device has an internal process control which confirms sufficient specimen volume, adequate membrane wicking, and correct procedural technique.
d. Detection limit:
Sensitivity of this assay is characterized by validating performance around the claimed cutoff concentration of the assay, including a determination of the lowest concentration of drug that is capable of producing a positive result.
{5}
Page 6 of 16
Drug-free urine specimens were spiked with propoxyphene to concentrations of 0, 150, 225, 300, 375, 450, and 600 ng/mL. The spiked samples were then tested with both the test strip and test device, with the following results:
Test Strip
| Concentration of sample, ng/mL | % 0f cutoff | Number of determinations | Results # Neg/ #Pos |
| --- | --- | --- | --- |
| 0 | 0 | 30 | 30/0 |
| 150 | -50% | 30 | 30/0 |
| 225 | -25% | 30 | 24/6 |
| 300 | Cutoff | 30 | 17/13 |
| 375 | +25% | 30 | 7/23 |
| 450 | +50% | 30 | 0/30 |
| 600 | +100% | 30 | 0/30 |
Test Device
| Concentration of sample, ng/mL | % 0f cutoff | Number of determinations | Results # Neg/ #Pos |
| --- | --- | --- | --- |
| 0 | 0 | 30 | 30/0 |
| 150 | -50% | 30 | 30/0 |
| 225 | -25% | 30 | 26/4 |
| 300 | Cutoff | 30 | 19/11 |
| 375 | +25% | 30 | 8/22 |
| 450 | +50% | 30 | 0/30 |
| 600 | +100% | 30 | 0/30 |
e. Analytical specificity:
Cross-reactivity was established by spiking various concentrations of similarly structured drug compounds into drug-free urine. By analyzing various concentration of each compound the sponsor determined the concentration of the drug that produced a response approximately equivalent to the cutoff concentration of the assay. Results of those studies appear in the table below:
Propoxyphene
| Drug Compound | Response equivalent to cutoff in ng/mL |
| --- | --- |
| D-norpropoxyphene | 300 |
{6}
Page 7 of 16
The following compounds were evaluated for potential positive and/or negative interference with the assay. To evaluate for interference the sponsor prepared three urine pools with propoxyphene concentrations of 0, 150, and 450 ng/mL. To aliquots of these pools the sponsor added the potential interferent at a concentration of 100 µg/mL. There were no deviations from the expected results:
| Compounds Tested at 100 µg/mL Conc. | ACON PPX One Step Propoxyphene Test Strip | | | ACON PPX One Step Propoxyphene Test Device | | |
| --- | --- | --- | --- | --- | --- | --- |
| | 0 ng/mL PPX | 150 ng/mL PPX | 450 ng/mL PPX | 0 ng/mL PPX | 150 ng/mL PPX | 450 ng/mL PPX |
| 4-Acetaminophenol | Negative | Negative | Positive | Negative | Negative | Positive |
| Acetone | Negative | Negative | Positive | Negative | Negative | Positive |
| Acetophenetidin | Negative | Negative | Positive | Negative | Negative | Positive |
| N-Acetylprocainamide | Negative | Negative | Positive | Negative | Negative | Positive |
| Acetylsalicylic acid | Negative | Negative | Positive | Negative | Negative | Positive |
| Albumin | Negative | Negative | Positive | Negative | Negative | Positive |
| Amitryptyline | Negative | Negative | Positive | Negative | Negative | Positive |
| Amobarbital | Negative | Negative | Positive | Negative | Negative | Positive |
| Amoxapine | Negative | Negative | Positive | Negative | Negative | Positive |
| Amoxicillin | Negative | Negative | Positive | Negative | Negative | Positive |
| Ampicillin | Negative | Negative | Positive | Negative | Negative | Positive |
| Ascorbic acid | Negative | Negative | Positive | Negative | Negative | Positive |
| DL-Amphetamine | Negative | Negative | Positive | Negative | Negative | Positive |
| Apomorphine | Negative | Negative | Positive | Negative | Negative | Positive |
| Aspartame | Negative | Negative | Positive | Negative | Negative | Positive |
| Atropine | Negative | Negative | Positive | Negative | Negative | Positive |
| Benzilic acid | Negative | Negative | Positive | Negative | Negative | Positive |
| Benzoic acid | Negative | Negative | Positive | Negative | Negative | Positive |
| Benzoylecgonine | Negative | Negative | Positive | Negative | Negative | Positive |
| Benzphetamine | Negative | Negative | Positive | Negative | Negative | Positive |
| Bilirubin | Negative | Negative | Positive | Negative | Negative | Positive |
| (±)- Brompheniramine | Negative | Negative | Positive | Negative | Negative | Positive |
| Buspirone | Negative | Negative | Positive | Negative | Negative | Positive |
| Caffeine | Negative | Negative | Positive | Negative | Negative | Positive |
| Cannabidiol | Negative | Negative | Positive | Negative | Negative | Positive |
| Cannabinol | Negative | Negative | Positive | Negative | Negative | Positive |
| Chloralhydrate | Negative | Negative | Positive | Negative | Negative | Positive |
| Chloramphenicol | Negative | Negative | Positive | Negative | Negative | Positive |
| Chlordiazepoxide | Negative | Negative | Positive | Negative | Negative | Positive |
| Chloroquine | Negative | Negative | Positive | Negative | Negative | Positive |
| Chlorothiazide | Negative | Negative | Positive | Negative | Negative | Positive |
{7}
Page 8 of 16
| Compounds Tested at 100 µg/mL Conc. | ACON PPX One Step Propoxyphene Test Strip | | | ACON PPX One Step Propoxyphene Test Device | | |
| --- | --- | --- | --- | --- | --- | --- |
| | 0 ng/mL PPX | 150 ng/mL PPX | 450 ng/mL PPX | 0 ng/mL PPX | 150 ng/mL PPX | 450 ng/mL PPX |
| (±) - Chlorpheniramine | Negative | Negative | Positive | Negative | Negative | Positive |
| Chlorpromazine | Negative | Negative | Positive | Negative | Negative | Positive |
| Chlorprothixene | Negative | Negative | Positive | Negative | Negative | Positive |
| Cholesterol | Negative | Negative | Positive | Negative | Negative | Positive |
| Clomipramine | Negative | Negative | Positive | Negative | Negative | Positive |
| Clonidine | Negative | Negative | Positive | Negative | Negative | Positive |
| Cocaine HCl | Negative | Negative | Positive | Negative | Negative | Positive |
| Cortisone | Negative | Negative | Positive | Negative | Negative | Positive |
| (-) Cotinine | Negative | Negative | Positive | Negative | Negative | Positive |
| Creatinine | Negative | Negative | Positive | Negative | Negative | Positive |
| Cyclobenzaprine | Negative | Negative | Positive | Negative | Negative | Positive |
| Deoxycorticosterone | Negative | Negative | Positive | Negative | Negative | Positive |
| (-) Deoxyephedrine | Negative | Negative | Positive | Negative | Negative | Positive |
| R(-) Deprenyl HCl | Negative | Negative | Positive | Negative | Negative | Positive |
| Dextromethorphan | Negative | Negative | Positive | Negative | Negative | Positive |
| Diclofenac | Negative | Negative | Positive | Negative | Negative | Positive |
| Diflunisal | Negative | Negative | Positive | Negative | Negative | Positive |
| Digoxin | Negative | Negative | Positive | Negative | Negative | Positive |
| 4-Dimethylaminoantipyrine | Negative | Negative | Positive | Negative | Negative | Positive |
| Diphenhydramine | Negative | Negative | Positive | Negative | Negative | Positive |
| Dicylamine | Negative | Negative | Positive | Negative | Negative | Positive |
| 5,5 - Diphenylhydantoin | Negative | Negative | Positive | Negative | Negative | Positive |
| Doxylamine | Negative | Negative | Positive | Negative | Negative | Positive |
| Ecgonine HCl | Negative | Negative | Positive | Negative | Negative | Positive |
| Ecgonine Methylester | Negative | Negative | Positive | Negative | Negative | Positive |
| EDDP | Negative | Negative | Positive | Negative | Negative | Positive |
| EMDP | Negative | Negative | Positive | Negative | Negative | Positive |
| Ephedrine | Negative | Negative | Positive | Negative | Negative | Positive |
| (-) -Ψ-Ephedrine | Negative | Negative | Positive | Negative | Negative | Positive |
| [1R,2S] (-) Ephedrine | Negative | Negative | Positive | Negative | Negative | Positive |
| (+/-) Epinephrine | Negative | Negative | Positive | Negative | Negative | Positive |
| Erythromycin | Negative | Negative | Positive | Negative | Negative | Positive |
| β-Estradiol | Negative | Negative | Positive | Negative | Negative | Positive |
| Estrone-3-sulfate | Negative | Negative | Positive | Negative | Negative | Positive |
| Ethanol | Negative | Negative | Positive | Negative | Negative | Positive |
| Ethyl-p-aminobenzoate | Negative | Negative | Positive | Negative | Negative | Positive |
| Fenfluramine | Negative | Negative | Positive | Negative | Negative | Positive |
| Fenoprofen | Negative | Negative | Positive | Negative | Negative | Positive |
| Furosemide | Negative | Negative | Positive | Negative | Negative | Positive |
{8}
Page 9 of 16
| Compounds Tested at 100 µg/mL Conc. | ACON PPX One Step Propoxyphene Test Strip | | | ACON PPX One Step Propoxyphene Test Device | | |
| --- | --- | --- | --- | --- | --- | --- |
| | 0 ng/mL PPX | 150 ng/mL PPX | 450 ng/mL PPX | 0 ng/mL PPX | 150 ng/mL PPX | 450 ng/mL PPX |
| Gentisic acid | Negative | Negative | Positive | Negative | Negative | Positive |
| D-Glucose | Negative | Negative | Positive | Negative | Negative | Positive |
| Guaiacol Glyceryl Ether | Negative | Negative | Positive | Negative | Negative | Positive |
| Hemoglobin | Negative | Negative | Positive | Negative | Negative | Positive |
| Hydralazine | Negative | Negative | Positive | Negative | Negative | Positive |
| Hydrochlorothiazide | Negative | Negative | Positive | Negative | Negative | Positive |
| Hydrocodone | Negative | Negative | Positive | Negative | Negative | Positive |
| Hydrocortisone | Negative | Negative | Positive | Negative | Negative | Positive |
| O-Hydroxyhippuric acid | Negative | Negative | Positive | Negative | Negative | Positive |
| p -Hydroxyamphetamine | Negative | Negative | Positive | Negative | Negative | Positive |
| p-Hydroxymethamphetamine | Negative | Negative | Positive | Negative | Negative | Positive |
| p-Hydroxynorephedrine | Negative | Negative | Positive | Negative | Negative | Positive |
| 3-Hydroxytyramine | Negative | Negative | Positive | Negative | Negative | Positive |
| Hydroxyzine | Negative | Negative | Positive | Negative | Negative | Positive |
| Ibuprofen | Negative | Negative | Positive | Negative | Negative | Positive |
| Imipramine | Negative | Negative | Positive | Negative | Negative | Positive |
| Iproniazid | Negative | Negative | Positive | Negative | Negative | Positive |
| (-) Isoproterenol | Negative | Negative | Positive | Negative | Negative | Positive |
| Isoxsuprine | Negative | Negative | Positive | Negative | Negative | Positive |
| Ketoprofen | Negative | Negative | Positive | Negative | Negative | Positive |
| Labetalol | Negative | Negative | Positive | Negative | Negative | Positive |
| Lidocaine | Negative | Negative | Positive | Negative | Negative | Positive |
| Lithium Carbonate | Negative | Negative | Positive | Negative | Negative | Positive |
| Loperamide | Negative | Negative | Positive | Negative | Negative | Positive |
| Maprotiline | Negative | Negative | Positive | Negative | Negative | Positive |
| Meperidine | Negative | Negative | Positive | Negative | Negative | Positive |
| Mephentamine | Negative | Negative | Positive | Negative | Negative | Positive |
| Meprobamate | Negative | Negative | Positive | Negative | Negative | Positive |
| Methamphetamine | Negative | Negative | Positive | Negative | Negative | Positive |
| L-Methamphetamine | Negative | Negative | Positive | Negative | Negative | Positive |
| Methaqualone | Negative | Negative | Positive | Negative | Negative | Positive |
| Methadol | Negative | Negative | Positive | Negative | Negative | Positive |
| Methoxyphenamine | Negative | Negative | Positive | Negative | Negative | Positive |
| Methylphenidate | Negative | Negative | Positive | Negative | Negative | Positive |
| (+/-)3,4-Methylenedioxyamphetamine | Negative | Negative | Positive | Negative | Negative | Positive |
| Methyprylon | Negative | Negative | Positive | Negative | Negative | Positive |
| Nalidixic acid | Negative | Negative | Positive | Negative | Negative | Positive |
| Nalorphine | Negative | Negative | Positive | Negative | Negative | Positive |
{9}
| Compounds Tested at 100 μg/mL Conc. | ACON PPX One Step Propoxyphene Test Strip | | | ACON PPX One Step Propoxyphene Test Device | | |
| --- | --- | --- | --- | --- | --- | --- |
| | 0 ng/mL PPX | 150 ng/mL PPX | 450 ng/mL PPX | 0 ng/mL PPX | 150 ng/mL PPX | 450 ng/mL PPX |
| Naloxone | Negative | Negative | Positive | Negative | Negative | Positive |
| Naltrexone | Negative | Negative | Positive | Negative | Negative | Positive |
| α-Naphthaleneacetic acid | Negative | Negative | Positive | Negative | Negative | Positive |
| Naproxen | Negative | Negative | Positive | Negative | Negative | Positive |
| Niacinamide | Negative | Negative | Positive | Negative | Negative | Positive |
| Nifedipine | Negative | Negative | Positive | Negative | Negative | Positive |
| NorMOR | Negative | Negative | Positive | Negative | Negative | Positive |
| Nimesulide | Negative | Negative | Positive | Negative | Negative | Positive |
| Norethindrone | Negative | Negative | Positive | Negative | Negative | Positive |
| Noscapine | Negative | Negative | Positive | Negative | Negative | Positive |
| D,L-Octopamine | Negative | Negative | Positive | Negative | Negative | Positive |
| Orphenadrine | Negative | Negative | Positive | Negative | Negative | Positive |
| Oxalic acid | Negative | Negative | Positive | Negative | Negative | Positive |
| Oxazepam | Negative | Negative | Positive | Negative | Negative | Positive |
| Oxolinic acid | Negative | Negative | Positive | Negative | Negative | Positive |
| Oxycodone | Negative | Negative | Positive | Negative | Negative | Positive |
| Oxymetazoline | Negative | Negative | Positive | Negative | Negative | Positive |
| OxyMOR | Negative | Negative | Positive | Negative | Negative | Positive |
| Penicillin-G | Negative | Negative | Positive | Negative | Negative | Positive |
| Pentazocine | Negative | Negative | Positive | Negative | Negative | Positive |
| Pentobarbital | Negative | Negative | Positive | Negative | Negative | Positive |
| Perphenazine | Negative | Negative | Positive | Negative | Negative | Positive |
| Phencyclidine | Negative | Negative | Positive | Negative | Negative | Positive |
| Phenelzine | Negative | Negative | Positive | Negative | Negative | Positive |
| Pheniramine | Negative | Negative | Positive | Negative | Negative | Positive |
| Phenobarbital | Negative | Negative | Positive | Negative | Negative | Positive |
| Phenothiazine | Negative | Negative | Positive | Negative | Negative | Positive |
| Phentermine | Negative | Negative | Positive | Negative | Negative | Positive |
| L-Phenylephrine | Negative | Negative | Positive | Negative | Negative | Positive |
| β-Phenylethylamine | Negative | Negative | Positive | Negative | Negative | Positive |
| Phenylpropanolamine | Negative | Negative | Positive | Negative | Negative | Positive |
| Prednisolone | Negative | Negative | Positive | Negative | Negative | Positive |
| Prednisone | Negative | Negative | Positive | Negative | Negative | Positive |
| Procaine | Negative | Negative | Positive | Negative | Negative | Positive |
| Promazine | Negative | Negative | Positive | Negative | Negative | Positive |
| Promethazine | Negative | Negative | Positive | Negative | Negative | Positive |
| DL-Propranolol | Negative | Negative | Positive | Negative | Negative | Positive |
| D-Pseudoephedrine | Negative | Negative | Positive | Negative | Negative | Positive |
| Quinacrine | Negative | Negative | Positive | Negative | Negative | Positive |
{10}
| Compounds Tested at 100 μg/mL Conc. | ACON PPX One Step Propoxyphene Test Strip | | | ACON PPX One Step Propoxyphene Test Device | | |
| --- | --- | --- | --- | --- | --- | --- |
| | 0 ng/mL PPX | 150 ng/mL PPX | 450 ng/mL PPX | 0 ng/mL PPX | 150 ng/mL PPX | 450 ng/mL PPX |
| Quinidine | Negative | Negative | Positive | Negative | Negative | Positive |
| Quinine | Negative | Negative | Positive | Negative | Negative | Positive |
| Ranitidine | Negative | Negative | Positive | Negative | Negative | Positive |
| Riboflavin | Negative | Negative | Positive | Negative | Negative | Positive |
| Salicylic acid | Negative | Negative | Positive | Negative | Negative | Positive |
| Secobarbital | Negative | Negative | Positive | Negative | Negative | Positive |
| Serotonin (5-Hydroxytyramine) | Negative | Negative | Positive | Negative | Negative | Positive |
| Sodium Chloride | Negative | Negative | Positive | Negative | Negative | Positive |
| Sulfamethazine | Negative | Negative | Positive | Negative | Negative | Positive |
| Sulindac | Negative | Negative | Positive | Negative | Negative | Positive |
| Sustiva | Negative | Negative | Positive | Negative | Negative | Positive |
| Tetracycline | Negative | Negative | Positive | Negative | Negative | Positive |
| Tetrahydrocortexolone | Negative | Negative | Positive | Negative | Negative | Positive |
| Tetrahydrocortisone, 3-acetate | Negative | Negative | Positive | Negative | Negative | Positive |
| Tetrahydrocortisone 3-(β-D-glucuronide) | Negative | Negative | Positive | Negative | Negative | Positive |
| Tetrahydrozoline | Negative | Negative | Positive | Negative | Negative | Positive |
| Theophylline | Negative | Negative | Positive | Negative | Negative | Positive |
| Thiamine | Negative | Negative | Positive | Negative | Negative | Positive |
| Thioridazine | Negative | Negative | Positive | Negative | Negative | Positive |
| L-Thyroxine | Negative | Negative | Positive | Negative | Negative | Positive |
| Tolbutamide | Negative | Negative | Positive | Negative | Negative | Positive |
| Trans-2-phenylcyclopropylamine | Negative | Negative | Positive | Negative | Negative | Positive |
| Trazodone | Negative | Negative | Positive | Negative | Negative | Positive |
| Trimethobenzamide | Negative | Negative | Positive | Negative | Negative | Positive |
| Triamterene | Negative | Negative | Positive | Negative | Negative | Positive |
| Trifluoperazine | Negative | Negative | Positive | Negative | Negative | Positive |
| Trimethoprim | Negative | Negative | Positive | Negative | Negative | Positive |
| Trimipramine | Negative | Negative | Positive | Negative | Negative | Positive |
| Tryptamine | Negative | Negative | Positive | Negative | Negative | Positive |
| DL-Tryptophan | Negative | Negative | Positive | Negative | Negative | Positive |
| Tyramine | Negative | Negative | Positive | Negative | Negative | Positive |
| DL-Tyrosine | Negative | Negative | Positive | Negative | Negative | Positive |
| Uric acid | Negative | Negative | Positive | Negative | Negative | Positive |
| Verapamil | Negative | Negative | Positive | Negative | Negative | Positive |
| Zomepirac | Negative | Negative | Positive | Negative | Negative | Positive |
There is the possibility that other substances and/or factors not listed above may interfere with the test and cause false results, e.g., technical or procedural errors.
{11}
To test for potential positive/and or negative interference from varying pH conditions, the sponsor prepared four sample pools with propoxyphene concentrations of 0, 150, 450, and $600\mathrm{ng / mL}$ . Aliquots of these samples were then adjusted to a $\mathsf{pH}$ of 5 to 9 with HCl or NaOH. Results are reported below:
Effect of Urinary pH - Test Strip
| pH | 0 ng/mL Propoxyphene | 150 ng/mL Propoxyphene | 450 ng/mL Propoxyphene | 600 ng/mL Propoxyphene |
| --- | --- | --- | --- | --- |
| 5 | - | - | + | + |
| 5 | - | - | + | + |
| 6 | - | - | + | + |
| 6 | - | - | + | + |
| 7 | - | - | + | + |
| 7 | - | - | + | + |
| 8 | - | - | + | + |
| 8 | - | - | + | + |
| 9 | - | - | + | + |
| 9 | - | - | + | + |
Effect of Urinary pH - Test Device
| pH | 0 ng/mL Propoxyphene | 150 ng/mL Propoxyphene | 450 ng/mL Propoxyphene | 600 ng/mL Propoxyphene |
| --- | --- | --- | --- | --- |
| 5 | - | - | + | + |
| 5 | - | - | + | + |
| 6 | - | - | + | + |
| 6 | - | - | + | + |
| 7 | - | - | + | + |
| 7 | - | - | + | + |
| 8 | - | - | + | + |
| 8 | - | - | + | + |
| 9 | - | - | + | + |
| 9 | - | - | + | + |
To test for potential positive/and or negative interference from varying specific gravity, the sponsor selected fifteen drug-free samples with a specific gravity from 1.007 to 1.029. Each sample was then split into three aliquots and spiked to propoxyphene concentrations of 0, 150, and $450\mathrm{ng / mL}$ . Results are shown below:
{12}
Page 13 of 16
## Effects of Urinary Specific Gravity - Test Strip
| SG | n | 0 ng/mL | 150 ng/mL | 450 ng/mL |
| --- | --- | --- | --- | --- |
| 1.007 | 2 | - | - | + |
| 1.012 | 2 | - | - | + |
| 1.013 | 2 | - | - | + |
| 1.014 | 2 | - | - | + |
| 1.014 | 2 | - | - | + |
| 1.017 | 2 | - | - | + |
| 1.019 | 2 | - | - | + |
| 1.021 | 2 | - | - | + |
| 1.022 | 2 | - | - | + |
| 1.024 | 2 | - | - | + |
| 1.025 | 2 | - | - | + |
| 1.026 | 2 | - | - | + |
| 1.027 | 2 | - | - | + |
| 1.028 | 2 | - | - | + |
| 1.029 | 2 | - | - | + |
## Effects of Urinary Specific Gravity - Test Device
| SG | n | 0 ng/mL | 150 ng/mL | 450 ng/mL |
| --- | --- | --- | --- | --- |
| 1.007 | 2 | - | - | + |
| 1.012 | 2 | - | - | + |
| 1.013 | 2 | - | - | + |
| 1.014 | 2 | - | - | + |
| 1.014 | 2 | - | - | + |
| 1.017 | 2 | - | - | + |
| 1.019 | 2 | - | - | + |
| 1.021 | 2 | - | - | + |
| 1.022 | 2 | - | - | + |
| 1.024 | 2 | - | - | + |
| 1.025 | 2 | - | - | + |
| 1.026 | 2 | - | - | + |
| 1.027 | 2 | - | - | + |
| 1.028 | 2 | - | - | + |
| 1.029 | 2 | - | - | + |
The sponsor did not evaluate the effects of albumin on the assay.
{13}
Page 14 of 16
f. Assay cut-off:
The Substance Abuse and Mental Health Services Administration (SAMHSA) has not recommended a cutoff concentration for propoxyphene.
Characterization of how the device performs analytically around the claimed cutoff concentration appears in the precision section, above.
2. Comparison studies:
a. Method comparison with predicate device:
A total of 352 samples (177 negative and 175 positive) were evaluated by the candidate device and by GC/MS and/or the predicate device.
Sample description: Unaltered clinical urine samples were evaluated. Fifty-one additional diluted samples were also included in the study. The samples were prepared by diluting clinical samples with high drug concentrations with drug-free urine. This was done in order to obtain samples near the cutoff concentration of the assay, because the sponsor was not able to obtain unaltered samples near the cutoff.
Sample selection: A portion of the negative samples were chosen for the study based on whether they screened positive or negative by the predicate device. Another portion of the negative samples having drug concentrations that were below the cutoff concentration of the assay were analyzed by GC-MS. All of the positive samples with concentrations above the cutoff were evaluated by GC-MS.
The study included an adequate number of samples that contained drugs near to the cutoff concentration of the assay. Approximately 10% of the study samples are evenly distributed between plus and minus 50% of the claimed cutoff concentration.
Number of study sites: one
Type of study site(s): Manufacturer's facility
Operator description: Manufacturer's laboratory staff
{14}
Page 15 of 16
# Candidate Device Results vs. Predicate Device Results – Test Strip
| | Positive by Predicate Device | Negative by Predicate Device |
| --- | --- | --- |
| Positive by Candidate Device | 157 | 0 |
| Negative by Candidate Device | 0 | 157 |
% Agreement among positives is 100%
% Agreement among negatives is 100%
# Candidate Device Results vs. Predicate Device Results – Test Device
| | Positive by Predicate Device | Negative by Predicate Device |
| --- | --- | --- |
| Positive by Candidate Device | 157 | 0 |
| Negative by Candidate Device | 0 | 157 |
% Agreement among positives is 100%
% Agreement among negatives is 100%
# Candidate Device Results vs. stratified GC/MS Values – Test Strip
| Candidate Device Results | Negative by the predicate device or less than half the cutoff concentration by GC/MS analysis | Near Cutoff Negative (Between 50% below the cutoff and the cutoff concentration) | Near Cutoff Positive (Between the cutoff and 50% above the cutoff concentration) | High Positive (greater than 50% above the cutoff concentration) |
| --- | --- | --- | --- | --- |
| Positive | 0 | 2 | 7 | 158 |
| Negative | 157 | 18 | 10 | 0 |
GC/MS values used to categorize samples in this table are based on the sum of the concentrations of propoxyphene and norpropoxyphene found in the sample.
% Agreement among positives is 94%
% Agreement among negatives is 99%
{15}
Page 16 of 16
Candidate Device Results vs. stratified GC/MS Values – Test Device
| Candidate Device Results | Negative by the predicate device or less than half the cutoff concentration by GC/MS analysis | Near Cutoff Negative (Between 50% below the cutoff and the cutoff concentration) | Near Cutoff Positive (Between the cutoff and 50% above the cutoff concentration) | High Positive (greater than 50% above the cutoff concentration) |
| --- | --- | --- | --- | --- |
| Positive | 0 | 2 | 7 | 158 |
| Negative | 157 | 18 | 10 | 0 |
GC/MS values used to categorize samples in this table are based on the sum of the concentrations of propoxyphene and norpropoxyphene found in the sample.
% Agreement among positives is 94%
% Agreement among negatives is 99%
b. Matrix comparison:
Not applicable. The assay is intended for only one sample matrix.
3. Clinical studies:
a. Clinical sensitivity:
Not applicable. Clinical studies are not typically submitted for this device type.
b. Clinical specificity:
Not applicable. Clinical studies are not typically submitted for this device type.
c. Other clinical supportive data (when a and b are not applicable):
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
Propoxyphene or its metabolites should not be present in the urine of persons who have not ingested propoxyphene.
M. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.