SYVA EMIT II PLUS ETHYL ALCOHOL ASSAY, MODELS 9K309UL/9K409UL
Applicant
Syva Co.
Product Code
DML · Clinical Toxicology
Decision Date
Feb 1, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3040
Device Class
Class 2
Indications for Use
Emit® II Plus Ethyl Alcohol Assay is an enzyme assay. The assay is intended for use in the quantitative analysis of ethyl alcohol in human serum or plasma. The Emit® II Plus Ethyl Alcohol Assay is an enzyme assay intended for use in the quantitative rne Emit - 11 Tub any) Aconol in human urine, plasma, and serum. The Emit® II Plus Ethyl Alcohol Assay is designed for use with most clinical chemistry analyzers.
Device Story
Emit® II Plus Ethyl Alcohol Assay is an in vitro diagnostic enzyme assay for quantitative ethanol measurement in human urine, serum, or plasma. Used in clinical laboratory settings with standard clinical chemistry analyzers; operated by laboratory personnel. The assay utilizes enzymatic reactions to detect ethanol concentrations; results are provided to clinicians to assist in patient assessment. Benefits include rapid, accurate quantification of alcohol levels in biological samples to support clinical decision-making.
Clinical Evidence
Bench testing only. Performance validated through precision studies (within-run %CV 1.39-1.55%; total %CV 1.61-1.78%), linearity testing, and spike recovery (95.7-101.5%). Interference studies evaluated bilirubin, hemoglobin, triglycerides, and LDH/Lactate. Sensitivity reported as <3.0 mg/dL. Correlation to predicate method showed r=0.986 (serum) and r=0.997 (urine).
Technological Characteristics
Enzyme-based in vitro diagnostic reagent kit. Designed for use on clinical chemistry analyzers. Quantitative range 10-600 mg/dL. Sensitivity <3.0 mg/dL. Compatible with various anticoagulants (EDTA, Sodium Citrate, Sodium Heparin, Oxalate/Fluoride).
Indications for Use
Indicated for quantitative analysis of ethyl alcohol in human urine, serum, or plasma using clinical chemistry analyzers.
Regulatory Classification
Identification
An alcohol test system is a device intented to measure alcohol (e.g., ethanol, methanol, isopropanol, etc.) in human body fluids (e.g., serum, whole blood, and urine). Measurements obtained by this device are used in the diagnosis and treatment of alcohol intoxication and poisoning.
{1}------------------------------------------------
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS For Emit® Ethyl Alcohol Assay
#### 1. Manufacturer and Contact Information:
| Manufacturer: | Syva Company |
|---------------|-------------------------|
| | 3403 Yerba Buena Rd. |
| | P.O. Box 49013 |
| | San Jose, CA 95161-9013 |
Paul Rogers Contact Information: Syva Company - Dade Behring Inc. 3403 Yerba Buena Road San Jose, CA 95161-9013 Tel: 408-239-2000
### 2. Device Classification Name:
The Clinical Chemistry and Clinical Toxicology Devices Panel have classified "Ethyl Alcohol System" as Class II.
#### 3. Intended Use:
Emit® II Plus Ethyl Alcohol Assay is an enzyme assay. The assay is intended for use in the quantitative analysis of ethyl alcohol in human serum or plasma.
#### 4. Device Description and Characteristics:
This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of SMDA 1990.
Syva Company - Dade Behring Inc. is submitting the Premarket Notification, 510(k) with the intention to commercially market an in vitro diagnostic reagent test kit for the analysis of ethyl alcohol in human urine, serum, or plasma.
The Syva Emit® II Plus Ethyl Alcohol Assay is an enzyme assay intended for use in quantitative analysis of ethyl alcohol in human urine, serum, or plasma. The Syva Emit® II Plus Ethyl Alcohol Assay has been found to be equivalent to the predicate device: Emit® Ethyl Alcohol Assay (K903153) with regard to intended use, assay sample, and overall performance characteristics.
Comparative Analysis: The Syva Emit® II Plus Ethyl Alcohol Assay showed excellent correlation to the predicate method. The comparative analysis to the predicate method resulted in a correlation of 0.986 with serum and 0.997 with urine, with a slope value of 1.000 with both urine and serum.
{2}------------------------------------------------
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS For Emit® Ethyl Alcohol Assay (cont.)
Precision: A Precision study was performed and the Syva Emit® II Plus Ethyl Alcohol Assay demonstrated acceptable within-run precision with coefficients of variation (%CV) ranging from 1.39% to 1.55% and acceptable total precision with coefficients of variation (%CV) ranging from 1.61% to 1.78%.
Specificity: A study was performed on potentially cross-reacting compounds and the Syva Emit® II Plus Ethyl Alcohol Assay demonstrated acceptable levels of cross-reactivity, ranging from <0.1 -- 14.2%.
Sensitivity: The sensitivity level of the Emit® II Plus Ethyl Alcohol Assay is less than 3.0 mg/dL. This level represents the lowest concentration of amphetamines that can be distinquished from 0 ng/mL with a confidence level of 95%.
Endogenous Interference: Endogenous interference in serum, due to either bilirubin, hemoglobin, or trigvcerides, in the Syva Emit® II Plus Ethyl Alcohol Assay are 99, 100, and 97% recovery compared to controls. The presence of LDH/Lactate does not cause a significant interference in the Svva Emit® II Plus Ethyl Alcohol Assay (0.4 mg/dL), A series of potentially interfering endogenous substances in urine were tested by the Syva Emit® II Plus Ethyl Alcohol Assav. The recoveries for the interfering substances range from 99 - 104%.
Spiked Sample Recovery: Spike recovery of ethanol in human serum and urine by the Syva Emit® II Plus Ethyl Alcohol Assay was performed across the quantitative range of between 10 and 600 mg/dL. Recovery of ethanol spikes ranged between 95.7 - 101.5%.
High Sample Dilution: Dilution of high serum and urine samples by water the Negative Calibrator were performed in the Syva Emit® II Plus Ethyl Alcohol Assav. Recovery from the normal concentrations in both types of samples diluted with either water or Negative Calibrator ranged from 97.1 - 100.9%.
Anticoagulants: The performance of the anticoagulants EDTA (K3), Sodium Citrate, Sodium Heparin, and Oxalate/Fluoride, as compared to serum, was assayed by the Syva Emit® II Plus Ethyl Alcohol Assay. Average recovery versus the serum control ranged from95.5 – 102.3%.
Linearity: The linearity of diluted serum and urine samples were tested in the Syva Emit® II Plus Ethyl Alcohol Assay. Both serum and urine samples are considered linear according to the acceptance criteria of the CAP linearity survey.
Urine Preservatives: the effect of urine preservatives was tested by the Syva Emit® II Plus Ethyl Alcohol Assay. Recoveries of urine preservatives compared to controls ranged from 97.5 -- 99.1%.
{3}------------------------------------------------
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS For Emit® Ethyl Alcohol Assay (cont.)
Calibration Stability: The calibration stability of the Syva Emit® II Plus Ethyl Alcohol Assay on the SYVA- 30R Biochemical System was determined to be 25 days.
## 5. Substantial Equivalence:
In conclusion, Syva Company considers the Emit® II Plus Ethyl Alcohol Assay to be substantially equivalent to the Ethyl Alcohol Assay (K903153) with regard to intended use, assay sample, and overall performance characteristics.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a bird in flight, composed of three curved lines.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
FEB 1 2000
Mr. Paul L. Rogers Jr. Senior Manager, Regulatory Affairs Syva Company - Dade Behring Inc. P.O. Box 49013 3403 Yerba Buena Road San Jose, California 95161-9013
Re: K993980
> Trade Name: Emit® II Plus Ethyl Alcohol Assay Regulatory Class: II Product Code: DML Dated: November 23, 1999 Received: November 24, 1999
Dear Mr. Rogers:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Autman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) Number (If known):
Emit® II Plus Ethyl Alcohol Assay Device Name:
## Indications for Use:
The Emit® II Plus Ethyl Alcohol Assay is an enzyme assay intended for use in the quantitative rne Emit - 11 Tub any) Aconol in human urine, plasma, and serum. The Emit® II Plus Ethyl Alcohol Assay is designed for use with most clinical chemistry analyzers.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use_
(Optional Format 1-2-96)
Device Name: Emit® II Plus Ethyl Alcohol Assay
Patricia Bernhardt for S.D.C.
Division of Clinical Laboratory Devices 510(k) Number.
く.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.