K082030 · Siemens Healthcare Diagnostics, Inc. · DLJ · Aug 22, 2008 · Clinical Toxicology
Device Facts
Record ID
K082030
Device Name
DIMENSION VISTA CSAE CALIBRATOR, MODEL KC440
Applicant
Siemens Healthcare Diagnostics, Inc.
Product Code
DLJ · Clinical Toxicology
Decision Date
Aug 22, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3200
Device Class
Class 2
Indications for Use
The CSAE CAL is an in vitro diagnostic product for the calibration of CSAE method on the Dimension Vista® system.
Device Story
The Dimension Vista® Cyclosporine Extended Range Calibrator is an in vitro diagnostic product used to calibrate the CSAE method on the Dimension Vista® clinical chemistry system. The kit consists of two levels of cyclosporine in a preserved whole blood hemolysate (200 ng/mL and 2000 ng/mL). The system automatically prepares intermediate calibrator levels (400, 800, and 1400 ng/mL) on-board the analyzer. The device is used in clinical laboratory settings by trained laboratory personnel. The calibration process ensures the accuracy of cyclosporine concentration measurements, which are critical for therapeutic drug monitoring in patients. The output of the calibration process allows the analyzer to provide precise quantitative results for patient samples, aiding clinicians in adjusting immunosuppressive therapy.
Clinical Evidence
No clinical data. Bench testing only. Stability testing confirmed 12-month shelf life at -20°C and 14-day stability for thawed/open/on-board vials. Traceability established via LC/MS/MS and USP Cyclosporine A reference material.
Technological Characteristics
Matrix: preserved whole blood hemolysate. Analyte: Cyclosporine. Form factor: two-level liquid calibrator kit. On-board processing: Dimension Vista® analyzer performs automated dilution to generate intermediate concentrations (400, 800, 1400 ng/mL) from two primary levels (200, 2000 ng/mL). Traceability: USP cyclosporine A via LC/MS/MS.
Indications for Use
Indicated for the calibration of the CSAE method on the Dimension Vista® system for in vitro diagnostic use.
Regulatory Classification
Identification
A clinical toxicology calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. A clinical toxicology calibrator can be a mixture of drugs or a specific material for a particular drug (e.g., ethanol, lidocaine, etc.). (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
Dimension® CSAE Cyclosporine Extended Range Calibrator (K061503)
Related Devices
K053108 — DIMENSION CSAE CYCLOSPORINE EXTENDED RANGE CALIBRATOR, MODEL DC108 · Dade Behring, Inc. · Feb 16, 2006
K061503 — DIMENSION CYCLOSPORINE EXTENDED RANGE CALIBRATOR (DC108A) · Dade Behring, Inc. · Jun 22, 2006
K052015 — DIMENSION EXTENDED RANGE CYCLOSPORINE CALIBRATOR, CATALOG # DC108 · Dade Behring, Inc. · Sep 12, 2005
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k082030
B. Purpose for Submission:
New device
C. Measurand:
Calibrator for Cyclosporine
D. Type of Test:
Calibrator Materials
E. Applicant:
Siemens Healthcare Diagnostics Inc.
F. Proprietary and Established Names:
Dimension Vista Cyclosporine Extended Range Calibrator (p/n KC440)
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| DLJ | Class II | 21 CFR§ 862.3200 | Toxicology |
H. Intended Use:
1. Intended use(s):
See Indications for Use below.
2. Indication(s) for use:
The CSAE CAL is an in vitro diagnostic product for the calibration of CSAE method on the Dimension Vista system.
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
Dimension Vista system
I. Device Description:
The Dimension Vista Cyclosporine Extended Range Calibrator contains cyclosporine in a preserved whole blood hemolysate. The kit consists of three vials of Calibrator A (200 ng/mL) and three vials of Calibrator B (2000 ng/mL). The intermediate calibrator levels (400, 800 and 1400 ng/mL) are prepared on-board the Dimension® Vista analyzer.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Dimension CSAE Cyclosporine Extended Range Calibrator
2. Predicate 510(k) number(s):
k061503
{1}
3. Comparison with predicate:
The Dimension Vista CSA Extended Range Calibrator has the same analytes, form, matrix, stability and traceability as the predicate device. The predicate device is used for calibration of the Dimension clinical chemistry system and the Syva Emit 2000 Cyclosporine Specific Assay whereas the new device is used for the calibration of the Dimension Vista system. Another difference is that the new device has two levels (200 and 2000 ng/mL), with the intermediate levels (400, 800 and 1400 ng/mL) prepared onboard the Dimension Vista analyzer, whereas the predicate has five levels (200, 400, 800, 1400 and 2000 ng/mL).
K. Standard/Guidance Document Referenced (if applicable):
None referenced.
L. Test Principle:
Not Applicable
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Not Applicable
b. Linearity/assay reportable range:
Not Applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability:
The calibrator is traceable to an internal master pool containing USP cyclosporine A. Values are confirmed by LC/MS/MS.
Stability:
Stability testing protocols and acceptance criteria were described and found to be acceptable. The product stored at -20°C is stable for 12 months. An unopened but thawed (2 - 8 °C) product is stable for 14 days. A punctured vial onboard the Dimension Vista system is stable for 14 days, as is an open (non-punctured) vial refrigerated at 2 - 8 °C. The intermediate calibrators that are prepared onboard the Dimension Vista system are prepared fresh each time they are needed, and then disposed of immediately afterwards.
Value Assignment:
Values are assigned to commercial stocks by using a Dimension or Dimension Vista instrument calibrated with master pools. The master pools are created by gravimetric addition of Cyclosporine USP Reference Material into drug free whole blood hemolysate and are assigned values using Dimension instruments calibrated with anchor pools. The anchor pools are prepared through the same method but are assigned values using LC/MS/MS.
{2}
d. Detection limit:
Not Applicable
e. Analytical specificity:
Not Applicable
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
Not Applicable
b. Matrix comparison:
Not Applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Not Applicable
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.