K020845 · Dade Behring, Inc. · DLJ · May 8, 2002 · Clinical Toxicology
Device Facts
Record ID
K020845
Device Name
EMIT 2000 VANCOMYCIN CALIBRATORS
Applicant
Dade Behring, Inc.
Product Code
DLJ · Clinical Toxicology
Decision Date
May 8, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3200
Device Class
Class 2
Indications for Use
The Emit® 2000 Vancomycin Calibrators are intended for use with the Emit® 2000 Vancomycin Assay as a reference for use in determining vancomycin levels in human serum or plasma.
Device Story
Emit® 2000 Vancomycin Calibrators are liquid, six-level calibrators used in clinical laboratories to establish reference points for the Emit® 2000 Vancomycin Assay. They consist of vancomycin, buffer, and preservatives in a synthetic matrix. The calibrators are used by laboratory technicians to calibrate the assay, ensuring accurate quantification of vancomycin levels in patient serum or plasma samples. This calibration process is essential for healthcare providers to monitor therapeutic drug levels and make informed clinical decisions regarding vancomycin dosing.
Clinical Evidence
No clinical data. Bench testing only. Stability was evaluated by comparing calibrators stored at 9°C versus -20°C, supporting a shelf life of at least 13 months. Calibrator lots are validated by testing 20 replicates at each level and comparing values to a Master Lot to ensure consistency.
Technological Characteristics
Six-level liquid calibrator. Composition: vancomycin, buffer, and preservatives in a synthetic matrix. Preparation: standard gravimetric procedures. Calibration: lot-specific adjustment based on 20-replicate testing against a Master Lot. Stability: 13-month shelf life supported by real-time data.
Indications for Use
Indicated for use as a reference standard for determining vancomycin levels in human serum or plasma samples.
Regulatory Classification
Identification
A clinical toxicology calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. A clinical toxicology calibrator can be a mixture of drugs or a specific material for a particular drug (e.g., ethanol, lidocaine, etc.). (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
K964587 — VANCOMYCIN ASSAY OR THE TECHNICON IMMUNO 1 SYSTEM · Bayer Corp. · Jan 13, 1997
K983280 — VANCOMYCIN EIA ASSAY · Diagnostic Reagents, Inc. · Nov 6, 1998
K090900 — ADVIA CHEMISTRY DRUG CALIBRATOR II, MODEL 10376771 · Siemens Healthcare Diagnostics · Jun 1, 2009
K043386 — VITROS CHEMISTRY PRODUCTS VANC REAGENT, CALIBRATOR KIT 11 AND TDM PERFORMANCE VERIFIERS I, II AND III · Ortho-Clinical Diagnostics, Inc. · Feb 7, 2005
K013076 — SYNCHRON SYSTEMS VANCOMYCIN REAGENT; SYNCHRON SYSTEMS VANCOMYCIN CALIBRATOR · Beckman Coulter, Inc. · Dec 12, 2001
Submission Summary (Full Text)
{0}------------------------------------------------
**MAY 08 2002**
K020845
Dade Behring Inc. Emit® 2000 Vancomycin Calibrators 510(k) Notification March 14, 2002
# 510(k) Summary For Emit® 2000 Vancomycin Calibrators
### Manufacturer's Name, Address, Telephone, and Contact Person, Date of 1. Preparation
Manufacturer: Dade Behring Inc Syva Company 20400 Mariani Avenue Cupertino, CA 95014
Contact Information: Donna A. Wolf Dade Behring Inc. P. O. Box 6101 Newark, DE 19714 Phone: 302-631-0384 Fax: 302-631-6299 E-mail: wolfda@dadebehring.com
Device Name / Classification: 2. Emit® 2000 Vancomycin Calibrator: Class II (862.3200) Classification Number:
Calibrator, drug specific
### Identification of the Legally Marketed Device: 3. Emit® Vancomycin Calibrator (K881040)
#### Device Description: 4.
Emit® 2000 Vancomycin Calibrators are liquid, six-level calibrators prepared from vancomycin, buffer and preservatives. The calibrators are intended for use with the Emit® 2000 Vancomycin Assay as a reference for use in determining vancomycin levels in human serum or plasma.
#### Device Intended Use: 5.
The Emit® 2000 Vancomycin Calibrators are intended for use with the Emit® 2000 Vancomycin Assay as a reference for use in determining vancomycin levels in human serum or plasma.
Medical device to which equivalence is claimed and comparison information: 6. The Emit® 2000 Vancomycin Calibrators are substantially equivalent in intended use to the Emit® Vancomycin Calibrators (K881040) currently marketed. The modified Emit® 2000 Vancomycin Calibrators contain the same vancomycin concentrations as the currently marketed device. Both devices are intended for use in the calibration of human serum or plasma in conjunction with a Vancomycin Reagent.
#### 7. Device Performance Characteristics:
## Stability:
Shelf life stability was evaluated by testing the calibrators at a minimum in duplicate, stored at 9°C, with comparison to calibrator sets stored at -20°C. The assignment of shelf life will be based on real time data from at least 3 lots of calibrators. To date, stability studies support no significant change in recovery for at least 13 months.
### Preparation:
Emit® 2000 Vancomycin Calibrators (modified) are prepared by standard gravimetric procedures using a synthetic matrix with buffers and preservative which has been supplemented with commercially available vancomycin.
Each new calibrator lot is tested in bulk and final container stages with 20 replicates at each calibrator level. These values are compared to a Master Lot. Calibrator levels are adjusted as needed to match Emit® 2000 responses of the Master Lot calibrators.
{1}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of an eagle with three lines extending from its head.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
# MAY 0 8 2002
Ms. Donna A. Wolf Sr. Regulatory Affairs Specialist Dade Behring Inc. 514 GBC Drive Newark, DE 19702
Re: k020845
> Trade/Device Name: Emit® 2000 Vancomycin Calibrators Regulation Number: 21 CFR 862.3200 Regulation Name: Clinical toxicology calibrator Regulatory Class: Class II Product Code: DLJ Dated: March 14, 2002 Received: March 15, 2002
Dear Ms. Wolf:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your 510(k) premarket I his leation. The FDA finding of substantial equivalence of your device to a legally marketed nouticate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, (301) 59 . 1500 . 1500 . 1500 . 16301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small mormation on your respond and Consumer Assistance at its toll-free number (800) 638-2041 or 1301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Dade Behring Inc. Emit® 2000 Vancomycin Calibrators 510(k) Notification March 14, 2002
# Indications For Use Statement
Device Name: Emit® 2000 Vancomycin Calibrators
## Indications for Use:
The Emit® 2000 Vancomycin Calibrators are intended for use with the Emit® 2000 The Ellines 2005 as a reference for use in determining vancomycin levels in human serum or plasma.
Sean Cooper
(Division Sign-Off)
Division of Clinical Laboratory Levices 1020 વ્યવસ 510(k) Number ..
Donna A. Wolf Sr. Regulatory Affairs Specialist March 14, 2002
(Optional format 1-2-96)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109) OR Over-the-counter Use***_***_
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.