← Product Code [DJR](/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/DJR) · K983703

# MODIFICATION TO ABUSCREEN ONLINE FOR METHADONE (K983703)

_Roche Diagnostic Systems, Inc. · DJR · Dec 11, 1998 · Clinical Toxicology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/DJR/K983703

## Device Facts

- **Applicant:** Roche Diagnostic Systems, Inc.
- **Product Code:** [DJR](/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/DJR.md)
- **Decision Date:** Dec 11, 1998
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 862.3620
- **Device Class:** Class 2
- **Review Panel:** Clinical Toxicology

## Indications for Use

Abuscreen ONLINE Methadone is an in vitro diagnostic test for the qualitative and semiquantitative detection of methadone in human urine on automated clinical chemistry analyzers at a cutoff concentration of 300 ng/mL. Semiquantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program. Measurements obtained by this device are used in the diagnosis of methadone use or abuse.

## Device Story

Abuscreen ONLINE Methadone is an in vitro diagnostic reagent kit for detecting methadone in human urine. It utilizes a kinetic interaction of microparticles in solution (KIMS) assay principle, where light transmission changes based on antibody-antigen binding. The device is designed for use on the Hitachi 917 automated clinical chemistry analyzer. The system processes patient samples in barcode-labeled vessels, performs photometric measurements, and transmits results to a computer unit. It is operated by laboratory personnel in a clinical setting. The output provides qualitative or semiquantitative methadone concentrations, which clinicians use to diagnose methadone use or abuse. The modified kit includes a surfactant in the diluent to optimize performance on the Hitachi 917 platform.

## Clinical Evidence

Bench testing only. Performance evaluated via precision (within-run and day-to-day) and accuracy studies. Precision studies at 150, 225, 300, 375, and 450 ng/mL showed CVs ranging from 0.8% to 4.4%. Accuracy was assessed against confirmed positive samples (N=50), demonstrating 100% agreement with confirmed results. Limit of detection was determined to be 19 ng/mL.

## Technological Characteristics

In vitro diagnostic reagent kit; KIMS (Kinetic Interaction of Microparticles in Solution) methodology; measures changes in light transmission. Reagents: methadone polyclonal (goat) antibody, conjugated methadone derivative microparticles, and buffer with surfactant. Designed for automated clinical chemistry analyzers (Hitachi 917).

## Regulatory Identification

A methadone test system is a device intended to measure methadone, an addictive narcotic pain-relieving drug, in serum and urine. Measurements obtained by this device are used in the diagnosis and treatment of methadone use or overdose and to determine compliance with regulations in methadone maintenance treatment.

## Special Controls

*Classification.* Class II (special controls). A methadone test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

## Predicate Devices

- Abuscreen ONLINE for Methadone ([K930928](/device/K930928.md))

## Reference Devices

- Hitachi 917 Analyzer System ([K953239](/device/K953239.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the Roche logo. The logo consists of the word "Roche" written in black font inside of a hexagon. The hexagon is outlined in black.

# 510(k) Summary

## Abuscreen ONLINE® Methadone

In accordance with the Safe Medical Devices Act of 1990, a 510(k) summary is provided as outlined in 21 CFR 807.92.

The assigned 510(k) number is: _______________________________________________________________________________________________________________________________________________

#### Identification of 510(k) Sponsor: I.

Roche Diagnostic Systems, Inc. a subsidiary of Hoffmann-La Roche, Inc. Branchburg Township 1080 U.S. Highway 202 Somerville, New Jersey 08876-3771

510(k) Submission dated October 20, 1998

Rita Smith Contact: Senior Regulatory Affairs Associate Phone: (908) 253-7545 (908) 253-7547 Fax:

Branchburg Township 1080 U.S. Highway 202 Somerville, New Jersey 08876-3771 

{1}------------------------------------------------

#### II. Device Name:

The device name, including both the trade/proprietary name and the classification name are provided in the table below.

### Table 1

| Product Name                      | Classification<br>Name           | Product<br>Code | CFR<br>Number and<br>Regulatory Class |
|-----------------------------------|----------------------------------|-----------------|---------------------------------------|
| Abuscreen ONLINE for<br>Methadone | Enzyme Immunoassay,<br>Methadone | DJR             | 862.3620<br>Class II                  |

#### III. Identification of the legally marketed device to which the 510(k) sponsor claims equivalence:

The following table identifies the legally marketed devices to which Roche Diagnostic Systems, Inc. claims equivalence.

| C<br>able<br>(<br>2<br>11 |  |
|---------------------------|--|
|---------------------------|--|

| Product Name                      | Predicate Product<br>Name         | 510(k) Number and Date<br>Predicate Cleared |
|-----------------------------------|-----------------------------------|---------------------------------------------|
| Abuscreen ONLINE for<br>Methadone | Abuscreen ONLINE for<br>Methadone | K930928 /24/93                              |

{2}------------------------------------------------

#### Description of the Device/Statement of Intended Use: IV.

Abuscreen ONLINE Methadone is an in vitro diagnostic test for the qualitative and semiquantitative detection of methadone in human urine on automated clinical chemistry analyzers at a cutoff concentration of 300 ng/mL. Semiquantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program. Measurements obtained by this device are used in the diagnosis of methadone use or abuse.

The proposed Abuscreen ONLINE Methadone test kit is specifically intended for use on the Hitachi 917 Analyzer and future similar analyzer models. It was adapted from the currently marketed Abuscreen ONLINE Methadone test kit. The labeling and packaging have been changed for use on the Hitachi 917 Analyzer as well as an addition of a surfactant to the diluent. This modified test kit is not a replacement to the currently marketed kit.

The Hitachi 917 Analyzer System is a fully automatic, computer-controlled system for clinical chemistry. It was conceived for both quantitative and qualitative in vitro determination using a large variety of tests for analysis, e.g. in serum and urine. Integrated in the system is an ion-selective unit for determination of electrolytes. The throughput per hour is 800 tests for clinical chemistry (1200 with electrolytes). The system consists of the analyzer which performs all functions required for fully automatic sample and test processing. Beginning with the automatic recording of patient samples - provided that they are supplied in barcode-labeled vessels - up to the photometric measurement and results transmission to the computer unit. Additional detailed information about the Hitachi 917 Analyzer is contained in volume II of the premarket notification (K953239) cleared on September 25, 1995.

#### Summary of the technological characteristics of the new device in comparison to V. those of the predicate.

Tables 3 outlines the technological characteristics (methodologies) of the Abuscreen ONLINE Methadone in comparison to that of the legally marketed predicate product.

{3}------------------------------------------------

#### Brief discussion of the clinical and nonclinical tests relied on for a determination of VI. substantial equivalence:

Tables 3 demonstrates the results of clinical and nonclinical studies performed using the Abuscreen ONLINE Methadone. The significant performance characteristics relied upon for a determination of substantial equivalence are summarized in this chart. This information concludes that the performance of this device is essentially equivalent to the legally marketed predicate device.

{4}------------------------------------------------

## Abuscreen ONLINE Methadone for Hitachi 917 Table 3

|                                              | Proposed:                                                                                                                                                                                |     |                                         | Previously Cleared: (K930928)                                                                                                                                                            |     |
|----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----|
|                                              | Abuscreen ONLINE Methadone                                                                                                                                                               |     |                                         | Abuscreen ONLINE Methadone                                                                                                                                                               |     |
|                                              | for Hitachi 917                                                                                                                                                                          |     |                                         | (1000 Test Kit)                                                                                                                                                                          |     |
| Methodology                                  | Kinetic interaction of<br>microparticles in a solution as<br>measured by changes in light<br>transmission                                                                                |     |                                         | Kinetic interaction of<br>microparticles in a solution as<br>measured by changes in light<br>transmission                                                                                |     |
| Sample type                                  | urine                                                                                                                                                                                    |     |                                         | urine                                                                                                                                                                                    |     |
| Intended Use                                 | qualitative and semiquantitative<br>detection of methadone                                                                                                                               |     |                                         | qualitative and semiquantitative<br>detection of methadone                                                                                                                               |     |
| Calibrator                                   | Abuscreen ONLINE Calibration<br>Pack or Abuscreen ONLINE<br>Calibrator Level 3                                                                                                           |     |                                         | Abuscreen ONLINE Calibration<br>Pack or Abuscreen ONLINE<br>Calibrator Level 3                                                                                                           |     |
| Cutoff(s)                                    | 300 ng/mL                                                                                                                                                                                |     |                                         | 300 ng/mL                                                                                                                                                                                |     |
| Reagent (active<br>ingredients)              | 1. Ab reagent: methadone<br>polyclonal (goat) antibody in<br>buffer<br>2. Microparticle reagent:<br>Conjugated methadone<br>derivative microparticles in<br>buffer<br>3. Diluent: Buffer |     |                                         | 1. Ab reagent: methadone<br>polyclonal (goat) antibody in<br>buffer<br>2. Microparticle reagent:<br>Conjugated methadone<br>derivative microparticles in<br>buffer<br>3. Diluent: Buffer |     |
| Performance Characteristics:                 |                                                                                                                                                                                          |     |                                         |                                                                                                                                                                                          |     |
| Precision Qualitative (300 ng/mL Cutoff):    |                                                                                                                                                                                          |     |                                         |                                                                                                                                                                                          |     |
|                                              | >95% negative at 225 ng/mL                                                                                                                                                               |     |                                         | >95% negative at 240 ng/mL                                                                                                                                                               |     |
|                                              | >95% positive at 375 ng/mL                                                                                                                                                               |     |                                         | >95% positive at 360 ng/mLL                                                                                                                                                              |     |
| Within Run                                   | Mean (OD)                                                                                                                                                                                | CV% |                                         |                                                                                                                                                                                          |     |
| 150 ng/mL                                    | 5661                                                                                                                                                                                     | 1.1 |                                         |                                                                                                                                                                                          |     |
| 225 ng/mL                                    | 4437                                                                                                                                                                                     | 1.9 |                                         |                                                                                                                                                                                          |     |
| 300 ng/mL                                    | 3286                                                                                                                                                                                     | 1.4 |                                         |                                                                                                                                                                                          |     |
| 375 ng/mL                                    | 2205                                                                                                                                                                                     | 1.9 |                                         |                                                                                                                                                                                          |     |
| 450 ng/mL                                    | 1780                                                                                                                                                                                     | 2.7 |                                         |                                                                                                                                                                                          |     |
| Day-to-Day                                   | Mean (OD)                                                                                                                                                                                | CV% |                                         |                                                                                                                                                                                          |     |
| 150 ng/mL                                    | 5541                                                                                                                                                                                     | 3.2 |                                         |                                                                                                                                                                                          |     |
| 225 ng/mL                                    | 4308                                                                                                                                                                                     | 4.1 |                                         |                                                                                                                                                                                          |     |
| 300 ng/mL                                    | 3212                                                                                                                                                                                     | 3.7 |                                         |                                                                                                                                                                                          |     |
| 375 ng/mL                                    | 2146                                                                                                                                                                                     | 3.6 |                                         |                                                                                                                                                                                          |     |
| 450 ng/mL                                    | 1749                                                                                                                                                                                     | 4.4 |                                         |                                                                                                                                                                                          |     |
| Proposed:                                    |                                                                                                                                                                                          |     | Previously Cleared: (K930928)           |                                                                                                                                                                                          |     |
| Abuscreen ONLINE                             |                                                                                                                                                                                          |     | Abuscreen ONLINE Methadone              |                                                                                                                                                                                          |     |
| Methadone for Hitachi 917                    |                                                                                                                                                                                          |     | (1000 Test Kit)                         |                                                                                                                                                                                          |     |
| Precision Quantitative ( 300 ng/mL Cutoff ): |                                                                                                                                                                                          |     |                                         |                                                                                                                                                                                          |     |
| Within Run                                   | Mean (ng/mL)                                                                                                                                                                             | CV% | Conc. (ng/mL)                           | Mean (ng/mL)                                                                                                                                                                             | CV% |
| 150 ng/mL                                    | 148                                                                                                                                                                                      | 2.4 | 150                                     | 152                                                                                                                                                                                      | 4   |
| 225 ng/mL                                    | 220                                                                                                                                                                                      | 3.2 | 240                                     | 235                                                                                                                                                                                      | 3   |
| 300 ng/mL                                    | 299                                                                                                                                                                                      | 1.5 | 300                                     | 301                                                                                                                                                                                      | 1   |
| 375 ng/mL                                    | 411                                                                                                                                                                                      | 1.0 | 340                                     | 340                                                                                                                                                                                      | 1   |
| 450 ng/mL                                    | 444                                                                                                                                                                                      | 0.8 |                                         |                                                                                                                                                                                          |     |
| Day-to-Day                                   | Mean (ng/mL)                                                                                                                                                                             | CV% | Conc. (ng/mL)                           | Mean (ng/mL)                                                                                                                                                                             | CV% |
| 150 ng/mL                                    | 151                                                                                                                                                                                      | 4.2 | 150                                     | 151                                                                                                                                                                                      | 4   |
| 225 ng/mL                                    | 222                                                                                                                                                                                      | 3.7 | 240                                     | 236                                                                                                                                                                                      | 3   |
| 300 ng/mL                                    | 303                                                                                                                                                                                      | 2.2 | 300                                     | 297                                                                                                                                                                                      | 3   |
| 375 ng/mL                                    | 407                                                                                                                                                                                      | 2.0 | 340                                     | 340                                                                                                                                                                                      | 1   |
| 450 ng/mL                                    | 452                                                                                                                                                                                      | 1.5 |                                         |                                                                                                                                                                                          |     |
| Accuracy                                     |                                                                                                                                                                                          |     |                                         |                                                                                                                                                                                          |     |
| 300 ng/mL Cutoff                             | N= 50 Confirmed Pos.<br>50 Pos. 0 Neg.                                                                                                                                                   |     | N = 73 Confirmed Pos.<br>73 Pos. 0 Neg. |                                                                                                                                                                                          |     |
| Limit of Detection                           | 19 ng/mL                                                                                                                                                                                 |     | < 15 ng/mL                              |                                                                                                                                                                                          |     |

{5}------------------------------------------------

## Abuscreen ONLINE Methadone for Hitachi 917 Table 3 (Continued)

{6}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

### DEC 1 1 1998

Ms. Rita Smith Senior Requlatory Affairs Associate Roche Diagnostic Systems, Inc. A Subsidiary of Hoffmann-La Roche, Inc. Branchburq Township 1080 U.S. Highway 202 Somerville, New Jersey 08876-3771

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Re: K983703 Abuscreen ONLINE® Methadone Assay Trade Name: Regulatory Class: İI Product Code: DJR Dated: October 20, 1998 Received: October 21, 1998

Dear Ms. Smith:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to leqally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{7}------------------------------------------------

#### Page 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

of Page 1

510(k) Number (if known)

Device Name: Abuscreen ONLINE® Methadone

Indications for Use:

Abuscreen ONLINE Methadone is an in vitro diagnostic test for the qualitative and semiquantitative detection of methadone in human urine on automated clinical chemistry analyzers at a cutoff concentration of 300 ng/mL. Semiquantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program. Measurements obtained by this device are used in the diagnosis of methadone use or abuse.

### (PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use (Optional Format 1-2-96)

(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number. K983703

---

**Source:** [https://fda.innolitics.com/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/DJR/K983703](https://fda.innolitics.com/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/DJR/K983703)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
