← Product Code [DJG](/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/DJG) · K972571

# QUIKSTRIP ONE STEP OPIATES TEST (K972571)

_Drial Consultants, Inc. · DJG · Aug 20, 1997 · Clinical Toxicology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/DJG/K972571

## Device Facts

- **Applicant:** Drial Consultants, Inc.
- **Product Code:** [DJG](/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/DJG.md)
- **Decision Date:** Aug 20, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.3650
- **Device Class:** Class 2
- **Review Panel:** Clinical Toxicology

## Indications for Use

Syntron's QuikStrip One Step Opiates assay is a rapid, qualitative, competitive binding immunoassay for the determination of Opiates in urine at the cutoff level of 300 ng/ml. The test provides only preliminary data which should be confirmed by other methods such as gas chromatography/mass spectrophotometry (GC/MS). Clinical considerations and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated6. Syntron's QuikStrip One Step Opiates Test is not intended to monitor drug levels, but only to screen urines for the presence of Opiates and its metabolites.

## Device Story

Chromatographic absorbent device for qualitative urine screening of opiates. Input: urine sample. Principle: competitive binding immunoassay; drug/metabolites in sample compete with immobilized drug conjugate on porous membrane for limited antibody sites. Labeled antibody-dye conjugate binds free drug; complex competes with immobilized antigen in positive reaction zone. Result: absence of magenta band indicates drug level >300 ng/ml; presence of magenta band in control zone confirms reagent/device function. Used in medical/forensic settings; operated by laboratory personnel. Output: visual color band. Clinical decision-making: provides preliminary screening data; requires GC/MS confirmation for positive results. Benefit: rapid identification of potential opiate presence.

## Clinical Evidence

Clinical trial of 298 urine samples compared the device against Syva EMIT® II (300 ng/ml cutoff). Results showed 100% relative sensitivity, 100% relative specificity, and 100% accuracy. All positive samples were confirmed by GC/MS. No false positives or false negatives were reported.

## Technological Characteristics

Chromatographic absorbent device; porous membrane support; antibody-dye conjugate; competitive binding immunoassay principle. Qualitative visual readout. Standalone device. No electronic components or software.

## Regulatory Identification

An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.

## Special Controls

*Classification.* Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

## Reference Devices

- Syva EMIT® II

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K972571

# 510k Submission for QuikStrip One Step Opiates Test Syntron Bioresearch, Inc.

AUG 2 0

Page 72 of 72

Revision A 6/18/97 Printed on 7/8/97

# Summary of Safety and Effectiveness

The sponsor, Syntron Bioresearch, Inc. (2774 Loker Ave. West, Carlsbad, California, 92008), has developed, manufactured, and tested under GMP/GLP guidelines a device for the qualitative testing of urine for the presence of Opiates and its metabolites in a screening format.

The trade name of the device is QuikStrip One Step Opiates Test having a designated common name of Opiates Test System and a classification as a Class II device per 21 CFR ¶ 862.3250. This device is intended for the medical/forensic screening of urine.

Syntron's QuikStrip One Step Opiates Test consists of a chromatographic absorbent device in which the drug or drug metabolites in the sample compete with a drug conjugate immobilized on a porous membrane support for the limited antibody sites. As the test sample flows up through the absorbent device, the labeled antibody-dye conjugate binds to the free drug in the specimen forming an antibody:antigen complex. This complex competes with immobilized antigen conjugate in the positive reaction zone and will not produce a magenta color band when the drug is above the detection level of 300 ng/ml. Unbound dye conjugate binds to the reagent in the control zone, producing a magenta color band, demonstrating that the reagents and device are functioning correctly.

In-house testing of Syntron's QuikStrip One Step Opiates Test yielded a relative sensitivity or agreement within positive samples of 1.000 and relative specificity or agreement within negative samples of 1.000 and an accuracy of 100% when tested against Syva EMIT® II on samples documented to be positive by GC/MS. A clinical trial consisting of 298 samples was run and the combined data yielded a relative sensitivity of 100%, a relative specificity of 100% with an accuracy of 100% when compared to Emit II® run at 300 ng/ml.

All positive samples by either screening method were confirmed by GC/MS. The testing performed by both the sponsor and the Clinical Trial site did not find any false positives or false negatives in the samples tested.

Additional information on this submission may be obtained by contacting Dr. Cleve W. Laird, President, Drial Consultants, Inc. at 805-522-6223(Ca) or by fax at 805-522-1526.

{1}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/1/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of three human profiles facing right, stacked on top of each other. The profiles are simple and abstract, with a flowing, curved design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the central image.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Cleve W. Laird, Ph.D. President and CEO Drial Consultants, Inc. 1420 Los Angeles Avenue Suite 201 Simi Valley, California 93065

AUG 20 1951

Re : K972571 Trade Name: QuikStrip One Step Opiate Assay Requlatory Class: II Product Code: DJG Dated: July 8, 1997 Received: July 10, 1997

Dear Dr. Laird:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Requlations.

{2}------------------------------------------------

Paqe 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) Additionally, for questions on the promotion and 594-4588. advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely yours,

Steven Autman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radioloqical Health

Enclosure

{3}------------------------------------------------

#### 510(k) Number (if Known): Has Yet to be assigned

Device Name: Syntron's QuikStrip One Step Opiates assay

### Indications For Use: . . . . . .

Syntron's QuikStrip One Step Opiates assay is a rapid, qualitative, competitive binding immunoassay for the determination of Opiates in urine at the cutoff level of 300 ng/ml. The test provides only preliminary data which should be confirmed by other methods such as gas chromatography/mass spectrophotometry (GC/MS). Clinical considerations and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated6. Syntron's QuikStrip One Step Opiates Test is not intended to monitor drug levels, but only to screen urines for the presence of Opiates and its metabolites.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANDOTHER PAGE IF NEEDED)

Concurance of CDRH, Office of Device Evaluation (ODE)

or

Semore fa Montgomery
(Division Sign Off)

(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number 972521

Perscription Use: _ C (Per 21 CFR 801.109

Over The Counter Use: (Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/DJG/K972571](https://fda.innolitics.com/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/DJG/K972571)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
