← Product Code [DJG](/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/DJG) · K172416

# Synermed Opiate Enzyme Immunoassay (K172416)

_Infrared Laboratory Systems, LLC (Dba Synermed) · DJG · Apr 19, 2018 · Clinical Toxicology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/DJG/K172416

## Device Facts

- **Applicant:** Infrared Laboratory Systems, LLC (Dba Synermed)
- **Product Code:** [DJG](/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/DJG.md)
- **Decision Date:** Apr 19, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.3650
- **Device Class:** Class 2
- **Review Panel:** Clinical Toxicology

## Indications for Use

The Synermed Enzyme Immunoassay is intended for the qualitative and semi-quantitative determination of opiates in human urine at a cutoff value of 300 ng/mL when calibrated against morphine. The assay is designed for professional use with a number of automated clinical chemistry analyzers. This assay is for prescription use only. The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmatory method such as GCMS or permitting laboratories to establish quality control procedures. The assay provides only a preliminary analytical result. A more specific alternative analytical chemistry method must be used in order to obtain a confirmed analytical result. Gas or Liquid Chromatography/Mass Spectrometry (GC/MS or LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.

## Device Story

Synermed Opiate Enzyme Immunoassay; competitive enzyme immunoassay for urine opiate detection. Input: human urine samples. Principle: competition between drug in sample and morphine-labeled G6PDH enzyme for fixed antibody; enzyme activity inversely proportional to drug concentration. Output: spectrophotometric absorbance change at 340 nm; qualitative or semi-quantitative results. Used in clinical laboratories; operated by trained laboratory personnel on Synermed IR-500 Chemistry Analyzer. Output used by clinicians to identify need for confirmatory testing (GC/MS or LC/MS) and clinical decision-making regarding drug use. Benefits: rapid preliminary screening for opiates.

## Clinical Evidence

Bench testing only. Precision/reproducibility evaluated per CLSI EP05-A3 and EP12-A using 11 urine concentrations (80 measurements each). Linearity assessed per NCCLS EP06-A across 10 concentrations. Analytical specificity tested against 10 endogenous compounds at +/- 25% of cutoff. Method comparison performed on 100 samples against Hitachi 717 and LC-MS/MS; demonstrated 91% agreement with LC-MS/MS for positive samples and 100% for negative samples.

## Technological Characteristics

Enzyme immunoassay reagent kit; contains mouse monoclonal anti-morphine antibody, G6PDH enzyme-drug conjugate, NAD, and sodium azide preservative. Operates via competitive binding; absorbance measured on automated clinical chemistry analyzers (e.g., IR-500). Storage at 2-8°C. No specific ASTM materials cited.

## Regulatory Identification

An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.

## Special Controls

*Classification.* Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

## Predicate Devices

- Lin-Zhi International Enzyme Immunoassay, Opiate ([K110298](/device/K110298.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
k172416

B. Purpose for Submission:
Adding a previously cleared test to the Synermed IR-500 Chemistry Analyzer

C. Measurand:
Opiates

D. Type of Test:
Competitive enzyme immunoassay; qualitative and semi-quantitative

E. Applicant:
Infrared Laboratory Systems, LLC

F. Proprietary and Established Names:
Synermed Opiate Enzyme Immunoassay

G. Regulatory Information:
|  Product Code | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  DJG | Class II | 21 CFR 862.3650, Opiate test system | Toxicology (91)  |

H. Intended Use:

1. Intended use(s):
Refer to Indications for Use below.

2. Indication(s) for use:
The Synermed Enzyme Immunoassay is intended for the qualitative and semi-quantitative determination of opiates in human urine at a cutoff value of 300 ng/mL when calibrated against morphine. The assay is designed for professional use with a number of automated clinical chemistry analyzers. This assay is for

{1}

prescription use only. The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmatory method such as GCMS or permitting laboratories to establish quality control procedures.

The assay provides only a preliminary analytical result. A more specific alternative analytical chemistry method must be used in order to obtain a confirmed analytical result. Gas or Liquid Chromatography/Mass Spectrometry (GC/MS or LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.

3. Special conditions for use statement(s):

For prescription use only.

For in vitro diagnostic use only.

4. Special instrument requirements:

Synermed IR-500 Chemistry Analyzer

I. Device Description:

The Synermed Opiate Enzyme Immunoassay is ready to use and consists of the following components:

- Antibody/Substrate Reagent (R1): Contains mouse monoclonal anti-morphine antibody, opiate-6-phosphate (G6P), nicotinamide adenine dinucleotide (NAD), stabilizers, and sodium azide (0.09 %) as a preservative.
- Enzyme-drug Conjugate Reagent (R2): Contains morphine-labeled opiate-6-phosphate dehydrogenase (G6PDH) in buffer with sodium azide (0.09 %) as a preservative.

J. Substantial Equivalence Information:

1. Predicate device name(s):

LZI Opiate Enzyme Immunoassay on Hitachi 717 Chemistry Analyzer

2. Predicate 510(k) number(s):

k110298

{2}

3

3. Comparison with predicate:

Opiate Enzyme Immunoassay:

|  Similarities and Differences  |   |   |
| --- | --- | --- |
|  Item | Candidate Device: Synermed Opiate Enzyme Immunoassay | Predicate Device: LZI Opiate Enzyme Immunoassay on Hitachi 717 Chemistry Analyzer (k110298)  |
|  Intended Use | For the qualitative and semi-quantitative determination of opiates in human urine | Same  |
|  Test Principle or Method | Competitive enzyme immunoassay | Same  |
|  Sample Type | Urine | Same  |
|  Cutoff | 300 ng/mL | Same  |
|  Instrument use for | Synermed IR-500 analyzer | Hitachi 717 analyzer  |

K. Standard/Guidance Document Referenced (if applicable):

CLSI EP5-A3: “Evaluation of Precision of Quantitative Measurement Procedures: Approved Guideline-Third Edition”

CLSI EP12-A2: “User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline-Second Edition”

CLSI EP17-A2: “Evaluation of Detection Capability for Clinical Laboratory Measurement”

L. Test Principle:

The Opiate Enzyme Immunoassay is a competitive enzyme immunoassay based on competition between drug in the sample and drug labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for a fixed amount of antibody in the reagent. Enzyme activity decreases upon binding to the antibody, and the drug concentration in the sample is measured in terms of enzyme activity. In the absence of drug in the sample, morphine-labeled G6PDH conjugate is bound to antibody and enzyme activity is inhibited. When free drug is present in the sample antibody binds to the free drug, the unbound morphine-labeled G6PDH then exhibits its maximum enzyme activity. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH resulting in an absorbance change measured spectrophotometrically at 340 nm.

M. Performance Characteristics:

{3}

4

1. Analytical performance:

a. Precision/Reproducibility:

The precision studies were performed on the Synermed IR-500 by measuring eleven levels of human urine pools for opiate (0, 75, 150, 225, 300, 375, 450, 525, 600, 800, and 1000 ng/mL). Each sample was run in duplicate twice a day for twenty days for a total of 80 measurements at each concentration in both qualitative and semi-quantitative modes. The data are summarized in the following tables:

Semi quantitative analysis (300 ng/mL cutoff)

|  Concentration as a % of the Cut-off Level | Target Concentration (ng/mL) | Results  |
| --- | --- | --- |
|  0 | 0 | 80 Negative / 0 Positive  |
|  25 | 75 | 80 Negative / 0 Positive  |
|  50 | 150 | 80 Negative / 0 Positive  |
|  75 | 225 | 80 Negative / 0 Positive  |
|  100 | 300 | 7 Negative / 73 Positive  |
|  125 | 375 | 80 Positive / 0 Negative  |
|  150 | 450 | 80 Positive / 0 Negative  |
|  175 | 525 | 80 Positive / 0 Negative  |
|  200 | 600 | 80 Positive / 0 Negative  |
|  267 | 800 | 80 Positive / 0 Negative  |
|  333 | 1000 | 80 Positive / 0 Negative  |

Qualitative Analysis

|  Concentration as a % of the Cut-off Level | Target Concentration (ng/mL) | Results  |
| --- | --- | --- |
|  0 | 0 | 80 Negative / 0 Positive  |
|  25 | 75 | 80 Negative / 0 Positive  |
|  50 | 150 | 80 Negative / 0 Positive  |
|  75 | 225 | 80 Negative / 0 Positive  |
|  100 | 300 | 0 Negative / 80 Positive  |
|  125 | 375 | 80 Positive / 0 Negative  |
|  150 | 450 | 80 Positive / 0 Negative  |
|  175 | 525 | 80 Positive / 0 Negative  |
|  200 | 600 | 80 Positive / 0 Negative  |

{4}

|  Concentration as a % of the Cut-off Level | Target Concentration (ng/mL) | Results  |
| --- | --- | --- |
|  267 | 800 | 80 Positive / 0 Negative  |
|  333 | 1000 | 80 Positive / 0 Negative  |

b. Linearity/assay reportable range:

To demonstrate linearity, a drug recovery study in the semi-quantitative mode was conducted on the Synermed IR-500 by measuring twelve levels of human urine pools (Samples were prepared by intermixing a high urine pool with a low urine pool to obtain twelve concentrations across the measuring range: 0, 25, 75, 150, 225, 300, 450, 600, 800, 1000, 1100, and 1250 ng/mL), using four replicates at each level. Each level of pooled human urine was confirmed by UHPLC-MSMS. The results confirmed acceptable recovery of samples tested.

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Traceability

The calibrator to be used with this device were previously cleared under k020769.

d. Detection limit:

See section M.1.a. for device performance around the cutoff.

e. Analytical specificity:

Cross reactivity from structurally related compounds, as well as interference of exogenous and endogenous substances was evaluated in k110298.

f. Assay cut-off:

Analytical performance of the device around the claimed cutoff is described in the precision section above.

2. Comparison studies:

a. Method comparison with predicate device:

A method comparison study was performed using unaltered clinical urine samples. A total of 100 samples were analyzed in singlicate in both Qualitative and Semi-quantitative modes, and the results were compared to UHPLC-MSMS. The results from the study in the two modes were identical and are summarized in the table below.

{5}

6

|  Candidate Device Results | Negative | <50% of cutoff by LC/MS (< 150ng/mL) | Near Cutoff Negative (Between 50% below the cutoff and the cutoff by LC/MS) (150 ~ 299 ng/mL) | Near Cutoff Positive (Between the cutoff and 50% above the cutoff by LC/MS) (300 ~ 450 ng/mL) | High Positive (Greater than 50% above the cutoff by LC/MS) (> 450 ng/mL)  |
| --- | --- | --- | --- | --- | --- |
|  Positive | 0 | 0 | 0 | 11 | 31  |
|  Negative | 5 | 19 | 30 | 4* | 0  |

*Discordant Results Table

|  Sample # | LC-MS/MS (ng/mL) | Semi-Quantitative Assay Result (ng/mL) | Qualitative Assay Result  |
| --- | --- | --- | --- |
|  50 | 321.82 (Positive) | 265 (Negative) | Negative  |
|  53 | 306.20 (Positive) | 267 (Negative) | Negative  |
|  57 | 314.80 (Positive) | 297 (Negative) | Negative  |
|  58 | 327.67 (Positive) | 294 (Negative) | Negative  |

b. Matrix comparison:

Not applicable. This device is intended to be used with urine samples only.

3. Clinical studies:

a. Clinical Sensitivity:

Not applicable.

b. Clinical specificity:

Not applicable.

c. Other clinical supportive data (when a. and b. are not applicable):

Not applicable.

4. Clinical cut-off:

Not applicable.

5. Expected values/Reference range:

Not applicable.

{6}

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

7

---

**Source:** [https://fda.innolitics.com/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/DJG/K172416](https://fda.innolitics.com/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/DJG/K172416)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
