Browse hierarchy: [Clinical Toxicology (TX)](/submissions/TX) → [Subpart D — Clinical Toxicology Test Systems](/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems) → [21 CFR 862.3250](/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/862.3250) → DIR — Free Radical Assay, Cocaine

# DIR · Free Radical Assay, Cocaine

_Clinical Toxicology · 21 CFR 862.3250 · Class 2_

**Canonical URL:** https://fda.innolitics.com/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/DIR

## Overview

- **Product Code:** DIR
- **Device Name:** Free Radical Assay, Cocaine
- **Regulation:** [21 CFR 862.3250](/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/862.3250)
- **Device Class:** 2
- **Review Panel:** [Clinical Toxicology](/submissions/TX)
- **3rd-party reviewable:** yes

## Identification

A cocaine and cocaine metabolite test system is a device intended to measure cocaine and a cocaine metabolite (benzoylecgonine) in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of cocaine use or overdose.

## Classification Rationale

Class II (special controls). A cocaine and cocaine metabolite test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

## Special Controls

*Classification.* Class II (special controls). A cocaine and cocaine metabolite test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

## Recent Cleared Devices (1 of 1)

| Record | Device Name | Applicant | Decision Date | Decision |
| --- | --- | --- | --- | --- |
| [K863888](https://fda.innolitics.com/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/DIR/K863888.md) | DU PONT PHENYTOIN ASSAY AND CALIBRATOR | E.I. Dupont DE Nemours & Co., Inc. | Nov 5, 1986 | SESE |

## Top Applicants

- E.I. Dupont DE Nemours & Co., Inc. — 1 clearance

---

**Source:** [https://fda.innolitics.com/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/DIR](https://fda.innolitics.com/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/DIR)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com