RAPIDSENSE DRUGS OF ABUSE COCAINE (COC) 300 DEVICE, MODEL 900-0052
K080436 · Quantrx Biomedical Corporation · DIO · Feb 11, 2009 · Clinical Toxicology
Device Facts
Record ID
K080436
Device Name
RAPIDSENSE DRUGS OF ABUSE COCAINE (COC) 300 DEVICE, MODEL 900-0052
Applicant
Quantrx Biomedical Corporation
Product Code
DIO · Clinical Toxicology
Decision Date
Feb 11, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3250
Device Class
Class 2
Indications for Use
The RapidSense™ Drugs of Abuse Cocaine (COC) 300 Device is a lateral flow competitive immunoassay intended for the qualitative detection of the cocaine metabolite, benzoylecgonine, in human urine at a cut-off concentration of 300 ng/mL. The assay is intended for use in professional laboratories by healthcare professionals. For In Vitro Diagnostic Use. This assay provides only a preliminary result. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. To obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas chromatography/mass spectroscopy (GC/MS) is the recommended confirmatory method.
Device Story
RapidSense™ Drugs of Abuse Cocaine (COC) 300 Device is a lateral flow competitive immunoassay for qualitative detection of benzoylecgonine in human urine. Device consists of a plastic cassette housing a nitrocellulose lateral flow strip. Principle of operation: differential migration mechanism; free analyte competes with blue-colored latex-labeled antigen conjugates for antibody binding in a primary capture zone. If analyte is at/above 300 ng/mL, less antibody is available for conjugate binding; excess conjugate migrates to a secondary capture zone, forming a visible positive test line. If analyte is absent/below cutoff, conjugate binds in primary zone, resulting in no visible test line. Used in professional laboratories by healthcare professionals; visually read. Provides preliminary screening results; requires GC/MS confirmation. Benefits include rapid preliminary assessment of drug presence to guide clinical decision-making.
Clinical Evidence
Comparison study of 85 clinical urine specimens against GC/MS reference method. Results: 91.1% positive agreement (41/45), 95.2% negative agreement (40/42), and 93.1% total agreement (81/87).
Technological Characteristics
Lateral flow immunochromatographic assay. Materials: nitrocellulose and cellulose strip in plastic cassette. Dimensions: 5 x 76 mm strip; 2" x 4" cassette. Principle: competitive immunoassay with differential migration. Visual readout. No external energy source or instrumentation required. Stability testing per CEN 13640.
Indications for Use
Indicated for the qualitative detection of cocaine metabolite (benzoylecgonine) in human urine at a 300 ng/mL cut-off concentration. Intended for use by healthcare professionals in professional laboratory settings for preliminary drug screening.
Regulatory Classification
Identification
A cocaine and cocaine metabolite test system is a device intended to measure cocaine and a cocaine metabolite (benzoylecgonine) in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of cocaine use or overdose.
Special Controls
*Classification.* Class II (special controls). A cocaine and cocaine metabolite test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Predicate Devices
ACON COC One Step Test Strip & Test Device (k010841)
Related Devices
K994398 — INSTANT-VIEW COCAINE (BENZYLECGONINE) URINE CASSETTE TEST · Alfa Scientific Designs, Inc. · Jul 27, 2000
K990822 — RAPIDONE - COCAINE TEST · American Bio Medica Corp. · Jul 15, 1999
K994403 — INSTANT-VIEW COCAINE (BENZYLECGONINE) URINE DIP STRIP TEST · Alfa Scientific Designs, Inc. · Aug 14, 2000
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k080436
B. Purpose for Submission:
New Device
C. Measurand:
Cocaine
D. Type of Test:
Qualitative
E. Applicant:
Quantrx Biomedical Corporation
F. Proprietary and Established Names:
RapidSense™ Drugs of Abuse Cocaine (COC) 300 Device
G. Regulatory Information:
1. Regulation section:
21 CFR § 862.3250, Cocaine and cocaine metabolite test system
2. Classification:
Class II
3. Product code:
DIO - Cocaine and cocaine metabolite test system
4. Panel:
91 (Toxicology)
H. Intended Use:
1. Intended use(s):
See Indications for use below.
2. Indication(s) for use:
The RapidSense™ Drugs of Abuse Cocaine (COC) 300 Device is a lateral flow competitive immunoassay intended for the qualitative detection of the cocaine metabolite, benzoylecgonine, in human urine at a cut-off concentration of 300 ng/mL. The assay is intended for use in professional laboratories by healthcare professionals. For In Vitro Diagnostic Use.
This assay provides only a preliminary result. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. To obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas chromatography/mass spectroscopy (GC/MS) is the recommended confirmatory method.
{1}
2
3. Special conditions for use statement(s):
This assay provides only a preliminary result. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. To obtain a confirmed analytical result, a more specific alternative chemical method is needed. Gas chromatography/Mass spectroscopy (GC/MS) is the recommended confirmatory method.
For prescription use only.
The assay is not intended for Point of Care use.
4. Special instrument requirements:
Not applicable. The device is a visually read single-use device.
I. Device Description:
The RapidSense™ Drugs of Abuse Assay is in a plastic cassette approximately 2" x 4" with a disposable pipette inserted. The cassette houses a 5 x 76 mm lateral flow strip made of nitrocellulose and other cellulose materials.
Each cassette is individually packaged in a moisture barrier foil pouch with desiccant. Each kit contains 25 individually pouched devices.
J. Substantial Equivalence Information:
1. Predicate device name(s):
ACON COC One Step Test Strip & Test Device
2. Predicate 510(k) number(s):
k010841
3. Comparison with predicate:
The new and predicate devices have the same indications for use and same cut-off of 300 ng/mL. Both assays are lateral flow competitive immunoassays which provide a visual qualitative end point. Both assays are intended as a screening method that provides a preliminary test result.
The devices differ in interpretation of the test line. The presence of the test line indicates a negative result for the predicate and a positive result for the new device.
K. Standard/Guidance Document Referenced (if applicable):
CEN 13640, Stability Testing of In Vitro Diagnostic Reagents
{2}
3
L. Test Principle:
The test employs lateral flow immunochromatographic technology.
The QuantRx assay is a positive line read, competitive immunoassay test with a differential migration mechanism. The free analyte runs ahead of and continuous with the blue colored latex labeled antigen conjugates and reacts with the antibody immobilized in the primary capture zone. The primary capture zone is located under the top of the cassette, not visible to the user. When analyte is at or above the cutoff concentration, less antibody in the primary capture zone is available to bind with the latex-analyte conjugate. Thus more latex-analyte conjugate migrates beyond the primary capture zone to the secondary capture zone where it forms a test line visible to the reader. When the drug analyte is absent or below the cut off, the latex-conjugate binds with the antibody in the primary capture zone and no line forms at the secondary capture zone (T) resulting in a negative result (no test line).
The internal process control indicates that an adequate volume of sample has been added and that the immunochromatographic strip is intact.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
The sponsor conducted precision studies over ten days using six levels of commercially available drugs of abuse controls containing benzoylecgonine. The tests were run in replicates of two and using different lots of product by different operators at the manufacturer site.
Results of the study are presented below:
COC Precision Study Results:
| Sample Concentration ng/mL | Percent Cut-off | # Observations | Results # Neg/ #Pos | |
| --- | --- | --- | --- | --- |
| | | | # Negative | # Positives |
| 0 | 0 | 20 | 20 | 0 |
| 150 | -50% | 20 | 20 | 0 |
| 225 | -25% | 20 | 14 | 6 |
| 300 | Cut-off | 20 | 7 | 13 |
| 375 | +25% | 20 | 2 | 18 |
| 450 | +50% | 20 | 0 | 20 |
b. Linearity/assay reportable range:
Not applicable. The assay is intended for qualitative use.
{3}
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Control materials are required but are not specifically identified in the labeling. Users are instructed to follow government regulations when determining when to run external controls.
The device has an internal process control.
d. Detection limit:
Sensitivity of this assay is characterized by validating performance around the claimed cutoff concentration of the assay, including a determination of the lowest concentration of drug that is capable of producing a positive result. This information appears in the precision section 1.a., above.
e. Analytical specificity:
Cross-reactivity was established by spiking various concentrations of similarly structured drug compounds into drug-free urine. By analyzing various concentrations of each compound the sponsor determined the concentration of the drug that produced a response approximately equivalent to the cutoff concentration of the assay. Results of those studies appear in the table below:
| Compound | Response equivalent to cutoff in ng/mL |
| --- | --- |
| Cocaine | 5,000 |
| Benzoylecgonine | 300 |
| Ecgonine Methyl Ester | 100,000 |
| Ecgonine | >100,000 |
The sponsor conducted studies to determine the potential interference of the test with variety of endogenous and pharmaceutical substances. A positive (450 ng/mL, +50% of cut-off) and a negative urine pool (150 ng/mL, -50% of cut-off) were used and the potentially interfering compounds were spiked to a concentration of 100 μg/mL into each urine pool. The following compounds showed no interference when tested with the RapidSense™ COC Assay at a concentration of 100μg/mL.
{4}
| Acetaminophen | Chlorpheniramine | Hemoglobin |
| --- | --- | --- |
| Acetone | Creatine | Imipramine |
| Albumin | Dextromethorphan | Isoproterenol |
| Amitryptyline | 4-dimethylaminoantipyrine | Lidocaine |
| Aspartame | Erythromycin | Penicillin G |
| Aspirin | Ethanol | Pheniramine |
| Atropine | Furosemide | Quinidine |
| Benzocaine | Glucose | Sulindac |
| Caffeine | Guaiacol Glycerol Ether | Vitamin C |
There is the possibility that other substances and/or factors not listed above may interfere with the test and cause false results
To test for possible positive and/or negative interference from specific gravity, the sponsor prepared two study control samples. The control samples consisted of drug-free urine spiked with 150 ng/mL and 450 ng/mL of benzoylecgonine. Aliquots of the control samples were then altered to span the specific gravity range of 1.007 to 1.031. No positive or negative interference due to specific gravity was observed.
To test for potential negative interference from pH the sponsor prepared a study control sample consisting of drug-free urine spiked with 450 ng/mL of benzoylecgonine. Aliquots of the control samples were then added to urine samples with a pH range of 4 - 8 and analyzed. No negative interference due to pH was observed.
f. Assay cut-off: Not Applicable.
Characterization of how the device performs analytically around the claimed cutoff concentration appears in the precision Section 1.a above.
2. Comparison studies:
a. Method comparison with predicate device:
A total of 87 samples (42 negative and 45 positive) were evaluated by the new device and by GC/MS.
Sample description: Unaltered clinical urine samples were evaluated.
Sample selection: Samples were selected based on previous GC/MS values.
The study included an adequate number of samples that contained drugs near to the cutoff concentration of the assay. Approximately 10% of the study samples are
{5}
evenly distributed between each plus and minus 50% of the claimed cutoff concentration.
The study was performed at a single study site by laboratory professionals. The results are summarized in the table below:
Number of study sites: one
Type of study site: clinical setting
Operator description: clinical site staff
New Device Results vs. stratified GC/MS Values
| New Device Results | Less than half the cutoff concentration by GC/MS analysis | Near Cutoff Negative (Between 50% below the cutoff and the cutoff concentration) | Near Cutoff Positive (Between the cutoff and 50% above the cutoff concentration) | High Positive (greater than 50% above the cutoff concentration) |
| --- | --- | --- | --- | --- |
| Positive | 0 | 2 | 15 | 26 |
| Negative | 26 | 14 | 4 | 0 |
GC/MS values used to categorize samples in this table are based on the concentration of benzoylecgonine found in the sample.
% Agreement among positives is 91%
% Agreement among negatives is 95%
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable.
b. Clinical specificity:
Not Applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not Applicable.
4. Clinical cut-off:
Not Applicable
{6}
5. Expected values/Reference range:
Not Applicable
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
7
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.