DETECTABUSE LIQUID CONTROL, DETECTABUSE STAT-SKREEN LIQUID CONTROL
Applicant
Biochemical Diagnostic, Inc.
Product Code
DIF · Clinical Toxicology
Decision Date
Jul 3, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3280
Device Class
Class 1
Indications for Use
The Detectabuse® Liquid control is intended for use as quality control urine to monitor the precision of laboratory urine toxicology testing procedures for the analytes listed in the package insert. For prescription use only
Device Story
Detectabuse® Liquid Control is a stabilized human-based urine product used as a quality control material in clinical toxicology laboratories. Positive controls are gravimetrically spiked with reference drug standards and metabolites; negative controls are certified via immunoassay, GC/MS, and/or LC/MS. The product contains sodium azide or a proprietary preservative. It is used by laboratory personnel to monitor the precision of urine toxicology testing procedures. The device provides known concentrations of analytes, allowing clinicians to verify the accuracy and reliability of their testing platforms. It does not perform diagnostic tests itself but serves as a reference material to ensure the validity of laboratory results.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Stabilized human-based urine matrix. Preservatives: sodium azide (<1%) or proprietary alternative. Storage: -10 to -20°C or 2-8°C. Analytes include drugs of abuse, metabolites, creatinine, pH, and specific gravity. No electronic or software components.
Indications for Use
Indicated for use as a quality control urine to monitor the precision of laboratory urine toxicology testing procedures.
Regulatory Classification
Identification
A clinical toxicology control material is a device intended to provide an estimation of the precision of a device test system and to detect and monitor systematic deviations from accuracy resulting from reagent or instrument defects. This generic type of device includes various single, and multi-analyte control materials.
Related Devices
K033404 — LIQUICHEK URINE TOXICOLOGY CONTROL (CONFIRMATORY SERIES) · Bio-Rad · Dec 12, 2003
K050682 — LIQUICHEK URINE TOXICOLOGY CONTROL (LEVEL C1) · Bio-Rad Laboratories, Inc. · Apr 29, 2005
K153474 — Detectabuse Liquid Control Urine, Detectabuse Stat-Skreen Liquid Control Urine, Detectabuse Liquid Control Urine, AU/NZ, Detectabuse Liquid Control Urine, Immunoassay Series, Detectabuse Liquid Control Urine, GC/MS and Confirm series · Biochemical Diagnostics, Inc. · Feb 23, 2016
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE
A. 510(k) Number:
k121122
B. Purpose for Submission:
New Device
C. Analyte:
Quality control materials for Delta-9-THC-COOH, Benzoylecgonine, Phencyclidine (PCP), Codeine, Propoxyphene, Methaqualone, Morphine -3-glucuronide, 6-Monoacetylmorphine, Morphine, d-Amphetamine, d-Methamphetamine, Secobarbital, Butalbital, Phenobarbital, Oxazepam, Methadone, MDMA, MDA, MDEA, Oxycodone, Buprenorphine, EDDP, Nordiazepam, Norpropoxyphene, Cotinine, Fentanyl, Ethanol, Nortriptyline, Creatinine, pH, Specific Gravity
D. Type of Test:
Not applicable
E. Applicant:
Biochemical Diagnostics, Inc.
F. Proprietary and Established Names:
Detectabuse® Liquid Control
G. Regulatory Information:
1. Regulation section:
21 CFR 862.3280, Clinical Toxicology Control Material
2. Classification:
Class I, reserved
3. Product Code:
DIF
4. Panel:
Toxicology (91)
H. Intended Use:
1. Intended use(s):
See indications for use below
2. Indication(s) for use:
The Detectabuse® Liquid control is intended for use as quality control urine to monitor the precision of laboratory urine toxicology testing procedures for the analytes listed in the package insert.
{1}
3. Special condition for use statement(s):
For prescription use only
4. Special instrument Requirements:
Not applicable.
I. Device Description:
Detectabuse® Liquid Control
Each bottle contains stabilized human based urine. Positive control urines have been gravimetrically spiked with reference drug standards and/or appropriate metabolites. Negative control urines are certified negative by combination of immunoassay, GC/MS and/or LC/MS for the constituents listed on our target sheets. The products contain less than 1% sodium azide as a preservative. For assays sensitive to sodium azide such as ELISA we substitute a proprietary preservative approved by the manufacturers and DEA.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Detectabuse® Liquid Control Urine
2. Predicate K number(s):
k925586
3. Comparison with predicates:
| Device | Predicate Device (K925586) | Modified Device (K121122) |
| --- | --- | --- |
| Intended Use | Same | Is intended for use as quality control urine to monitor the precision of laboratory urine toxicology testing procedures |
| Form | Same | Liquid |
| Matrix | Same | Human Urine |
| Storage | Closed vial: The controls are stable until the expiration date when stored at -10 to -20°C and protected from light.
The controls are stable until the expiration date when stored at 2-8°C.
Open vial: not available | Closed vial: The controls are stable until (1) 36 months when stored at 2-8°C with the exception of oxazepam; (2) 48 months when stored at -10 to -20°C and protected from light with the exception of oxazepam.
Open vial: The controls are stable 30 days when stored at 2-8°C. |
{2}
3
| Analytes | Delta-9-THC-COOH, Benzoylecgonine, Phencyclidine (PCP), Codeine, Propoxyphene
Methaqualone, Morphine -3-glucuronide, 6-Monoacetylmorphine, Morphine, d-Amphetamine, d-Methamphetamine, Secobarbital, Butalbital, Phenobarbital, Oxazepam, Methadone | Same analytes as predicate device A with the additional claims:
MDMA, MDA, MDEA, Oxycodone, Buprenorphine, EDDP, Nordiazepam, Norpropoxyphene, Cotinine, Fentanyl, Ethanol, Nortriptyline, Creatinine, pH, Specific Gravity |
| --- | --- | --- |
K. Standard/Guidance Document Referenced (if applicable):
None were referenced.
L. Test Principle:
Not applicable.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Not applicable.
b. Linearity/assay reportable range:
Not applicable.
c. Traceability (controls, calibrators, or method):
a) Traceability
The controls are gravimetrically spiked with reference standards purchased from commercial vendors. The drug values printed within the package insert are validated by using GC/MS, LC/MS, HPLC, or Immunoassay screening.
b.) Stability
Stability studies have been performed to determine the open vial stability and shelf life stability of the Detectabuse® Liquid control.
Product claims are as follows: open vial stability is 30 days when stored tightly capped at 2-8°C; closed vial stability is 10 months at 2-8°C and 4 years at -10°C to -20°C.
d. Detection limit:
Not applicable.
e. Analytical specificity:
Not applicable.
f. Assay cut-off:
Not applicable.
{3}
2. Comparison studies:
a. Method comparison with predicate device:
Not applicable.
b. Matrix comparison:
Not applicable.
3. Clinical studies:
a. Clinical sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a and b are not applicable):
Not available.
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
Not applicable.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports substantial equivalence decision.
4
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.