Applied Biosystems™ 3500 Dx Genetic Analyzer and Applied Biosystems™ 3500xL Dx Genetic Analyzer

K191030 · Life Technologies Corporation · PCA · Feb 21, 2020 · Clinical Chemistry

Device Facts

Record IDK191030
Device NameApplied Biosystems™ 3500 Dx Genetic Analyzer and Applied Biosystems™ 3500xL Dx Genetic Analyzer
ApplicantLife Technologies Corporation
Product CodePCA · Clinical Chemistry
Decision DateFeb 21, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.2570
Device ClassClass 2

Indications for Use

The Applied Biosystems™ 3500 Dx Genetic Analyzer and the Applied Biosystems™ 3500xL Dx Genetic Analyzer are in vitro diagnostic devices intended for detection of fluorescently-labeled human genomic deoxyribonucleic acid (DNA) nucleotides by capillary electrophoresis. The Applied Biosystems™ 3500 Dx Genetic Analyzer and the Applied Biosystems™ 3500xL Dx Genetic Analyzer are indicated for sequencing and fragment analysis using FDA-cleared or approved assays.

Device Story

Genetic analyzer performs capillary electrophoresis to separate fluorescently-labeled human genomic DNA fragments; utilizes 8-capillary (3500 Dx) or 24-capillary (3500xL Dx) arrays. System inputs include PCR products from human genomic DNA; samples are electrokinetically injected into capillaries; high voltage drives migration toward anode; laser excitation triggers fluorescence; CCD camera detects signals. Integrated software (3500 Series Dx Data Collection Software 3 IVD v3.2) converts fluorescence to digital data; performs spatial/spectral calibration, base-calling, and size-calling. Used in clinical laboratory settings by trained personnel. Output data exported for assay-specific downstream analysis. Enables clinical decision-making by providing accurate DNA fragment sizing and sequencing results for diagnostic assays.

Clinical Evidence

Clinical performance evaluated across 3 US clinical laboratory sites using multiple instruments and a representative fragment analysis assay compared against an appropriate method comparison assay. All pre-established performance criteria were met. Non-clinical reproducibility study performed across multiple sites, instruments, and operators also met acceptance criteria.

Technological Characteristics

Capillary electrophoresis system; 8 or 24-capillary array; POP-6/POP-7 polymer separation matrix; laser-induced fluorescence detection via CCD camera. Connectivity includes data transmission to computer workstations. Software-controlled instrument calibration (spatial/spectral). Complies with IEC 60601-1-2 and UL 61010-1.

Indications for Use

Indicated for sequencing and fragment analysis of human genomic DNA using FDA-cleared or approved assays. For prescription use only.

Regulatory Classification

Identification

Instrumentation for clinical multiplex test systems is a device intended to measure and sort multiple signals generated by an assay from a clinical sample. This instrumentation is used with a specific assay to measure multiple similar analytes that establish a single indicator to aid in diagnosis. Such instrumentation may be compatible with more than one specific assay. The device includes a signal reader unit, and may also integrate reagent handling, hybridization, washing, dedicated instrument control, and other hardware components, as well as raw data storage mechanisms, data acquisition software, and software to process detected signals.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9. The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Instrumentation for Clinical Multiplex Test Systems.” See § 862.1(d) for the availability of this guidance document.

The special control is FDA's guidance document entitled "Class II Special Controls Guidance Document: Instrumentation for Clinical Multiplex Test Systems."

Predicate Devices

Submission Summary (Full Text)

{0} Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY # INSTRUMENT ONLY ## I Background Information: A 510(k) Number K191030 B Applicant Life Technologies Corporation C Proprietary and Established Names Applied Biosystems™ 3500 Dx Genetic Analyzer and Applied Biosystems™ 3500xL Dx Genetic Analyzer D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | PCA | Class II Exempt, Meets the limitation of exemptions 862.9(a) | 21 CFR 862.2570 - Instrumentation For Clinical Multiplex Test Systems | CH - Clinical Chemistry | ## II Submission/Device Overview: A Purpose for Submission: Change to the instrument software to provide fragment analysis of DNA B Type of Test: DNA Genetic Analyzer K191030 - Page 1 of 6 {1} K191030 - Page 2 of 6 ## III Intended Use/Indications for Use: ### A Intended Use(s): See Indications for Use below. ### B Indication(s) for Use: The Applied Biosystems™ 3500 Dx Genetic Analyzer and the Applied Biosystems™ 3500xL Dx Genetic Analyzer are in vitro diagnostic devices intended for detection of fluorescently-labeled human genomic deoxyribonucleic acid (DNA) nucleotides by capillary electrophoresis. The Applied Biosystems™ 3500 Dx Genetic Analyzer and the Applied Biosystems™ 3500xL Dx Genetic Analyzer are indicated for sequencing and fragment analysis using FDA-cleared or approved assays. ### C Special Conditions for Use Statement(s): Rx - For Prescription Use Only ## IV Device/System Characteristics: ### A Device Description: **Device and consumables:** The Applied Biosystems™ 3500 Dx Genetic Analyzer and Applied Biosystems™ 3500xL Dx Genetic Analyzer include the following components: - 8-capillary (Applied Biosystems™ 3500 Dx Genetic Analyzer) or 24-capillary (Applied Biosystems™ 3500xL Dx Genetic Analyzer) array and separation matrix to separate DNA fragments by size (POP™ polymer) - Reagents and consumables - Computer workstation and monitor - Integrated software for instrument control, data collection, quality control, base-calling, and size-calling of samples The following consumables are required to operate the instrument: - 50cm Capillary Array: enables the labeled DNA fragments to migrate from the cathode toward the anode for sequencing detection - POP-6™ Polymer: used as a separation matrix to separate DNA fragments by size during electrophoresis for sequencing - POP-7™ Polymer: used as a separation matrix for separating DNA fragments by size during electrophoresis for fragment analysis - Hi-Di™ Formamide: sample re-suspension solution used for electro-kinetic injection and denaturing the DNA - Sequencing Standard v1.1: used for spectral calibration of the instrument and instrument performance check - Cathode Buffer Container: pre-filled with running buffer which maintains a source of ions and the correct pH for electrophoresis {2} - Anode Buffer Container: pre-filled with running buffer which maintains a source of ions and the correct pH for electrophoresis - Conditioning Reagent: pre-filled pouch used for priming the polymer pump, washing the pump between polymer type changes, and during instrument shutdown - DS-30 Matrix Standard - Dx: used for spectral calibration - DS-33 Matrix Standard - Dx: used for spectral calibration - DS-33 GeneScan™ Install Kit - Dx: used for instrument operational qualification - GeneScan™ 600 LIZ® Size Standard v2.0 - Dx: used as a ladder for sizing DNA fragments - Other accessories (e.g. sample plate holders, plate retainers, septa) Software: The Applied Biosystems™ 3500 Dx Series Data Collection Software 3 IVD v3.2 provides control and monitoring of the instrument, collection of data during an injection, and primary analysis of injection data. The software provides three main functions: 1. Facilitates user entry of specific plate information, including plate and sample IDs, that are used by the instrument to perform the injection, to map the assay to wells on a plate, and to perform data reduction. 2. Communicates with the instrument to provide injection parameters and retrieve both raw instrument data and general instrument status. 3. Provides data reduction of instrument data for storage in sample files. These files can then be exported for analysis using a secondary, assay-specific software. Principle of Operation: DNA, labeled with fluorescent dyes, is transferred to a capillary electrophoresis plate that is used for injection into the instrument. During capillary electrophoresis, fluorescently labeled PCR products are electrokinetically injected into capillaries. High voltage is applied to move DNA fragments toward the positive electrode. Shortly before reaching the positive electrode, the fluorescently labeled DNA fragments, separated by size, move through the path of a laser beam. The laser beam causes the dyes on the fragments to fluoresce. A charge coupled device (CCD) camera detects the fluorescence. The Applied Biosystems™ 3500 Dx Series Data Collection Software 3 IVD v3.2 converts the fluorescence signal to digital data and performs size calling operations on the data. The raw data files are then transferred to a computer with the assay-specific software for downstream analyses. The Applied Biosystems™ 3500 Dx Genetic Analyzer and the Applied Biosystems™ 3500xL Dx Genetic Analyzer were cleared for in vitro diagnostic sequence analysis use in BK110039. B Instrument Description Information: | Modes of Operation | Yes | No | | --- | --- | --- | | Does the applicant's device contain the ability to transmit data to a computer, webserver, or mobile device? | ☑ | ☐ | | Does the applicant's device transmit data to a computer, webserver, or mobile device using wireless transmission? | ☑ | ☐ | | Software | | | | FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types. | ☑ | ☐ | K191030 - Page 3 of 6 {3} K191030 - Page 4 of 6 1. **Instrument Name:** Applied Biosystems™ 3500 Dx Genetic Analyzer and Applied Biosystems™ 3500xL Dx Genetic Analyzer 2. **Specimen Identification:** The user can use either use a handheld barcode reader or can manually enter the barcodes identified on the test plates. The user manually enters individual sample IDs per well on a plate. 3. **Specimen Sampling and Handling:** Human genomic DNA, from EDTA-anticoagulated whole human blood (fresh or stored under validated storage conditions) may be isolated using laboratory-validated sample preparation methodologies (precipitation, silica column, or magnetic beads). PCR products from FDA- cleared or approved devices intended to be used with the Applied Biosystems™ 3500 Dx Genetic Analyzer and Applied Biosystems™ 3500xL Dx Genetic Analyzer may be used immediately or stored under validated storage conditions (protected from light) prior to capillary electrophoresis analysis. The PCR product is plated onto a 96-well plate for use with the analyzer. 4. **Calibration:** Two calibrations are required to prepare the instrument for sample runs: 1. **Spatial calibration:** The software uses images collected during the spatial calibration to establish a relationship between the signal emitted by each capillary and the position where that signal is detected by the CCD camera. Spatial calibrations should be performed after removal or replacement of the capillary array. The instrument requires spatial calibration when it detects a newly installed capillary array. During the calibration, the software performs quality checks to ensure that sequence and fragment analysis attributes do not exceed the set thresholds. 2. **Spectral calibration:** The spectral calibration creates a de-convolution matrix for each capillary that compensates for dye overlap (reduces raw data from the instrument) in the dye data stored in each sample file and this information is used to convert the 20-color data into 4-, 5-, or 6-dye data. Spectral calibration should be performed when a new dye is used, when the capillary is removed or replaced, or when the user observes a decrease in spectral separation in the data peaks of raw or analyzed data. The instrument must be calibrated for the detection of each fluorescent dye used with the assay. 5. **Quality Control:** Size standards are present in each sample (internal control) and are labeled with the fluorescent dye, LIZ, for quality control. The assay may also include external positive and negative controls in each run for the assay software to interpret sample results correctly. {4} V Substantial Equivalence Information: A Predicate Device Name(s): 3500 Dx/3500xL Dx Genetic Analyzer CS2 and 3500 Dx Series Software B Predicate 510(k) Number(s): BK110039 C Comparison with Predicate(s): | Similarities and Differences | | | | --- | --- | --- | | Device & Predicate Device(s): | K191030 | BK110039 | | Intended Use | For detection of fluorescently labeled DNA by capillary electrophoresis | Same | | Analysis Type | Sequencing and fragment analysis | Sequencing analysis | | Software | 3500 Series Dx Data Collection Software 3 IVD v3.2 | 3500 Dx Data Collection Software 2011 v1.01 | | Environment of Use | Prescription use | Same | | Specimen Type | Whole blood | Same | | Input Sample | Human genomic DNA | Same | | Technology | Capillary electrophoresis, 8- and 24-capillary configurations | Same | VI Standards/Guidance Documents Referenced: - IEC 60601-1-2: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests - UL 61010-1: Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use - Part 1: General Requirements - Class II Special Controls Guidance Document: Instrumentation for Clinical Multiplex Test Systems, 2005 VII Performance Characteristics (if/when applicable): A Analytical Performance: The Applied Biosystems™ 3500 Dx Genetic Analyzer and Applied Biosystems™ 3500xL Dx Genetic Analyzer were previously legally marketed for sequencing analysis (see BK110039). The performance information referenced below (AmplideX® Fragile X Dx & Carrier Screen Kit, approved under DEN190023) supports the performance of the Applied Biosystems™ 3500 Dx K191030 - Page 5 of 6 {5} Genetic Analyzer and Applied Biosystems™ 3500xL Dx Genetic Analyzer for DNA fragment analysis. 1. Precision/Reproducibility: See DEN190023. 2. Linearity: Not applicable. 3. Analytical Specificity/Interference: See DEN190023. 4. Accuracy (Instrument): See DEN190023. 5. Carry-Over: See DEN190023. B Other Supportive Instrument Performance Characteristics Data: DNA input, DNA extraction equivalence, and specimen handling stability studies were also conducted using the assay. See DEN190023. VIII Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. IX Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. K191030 - Page 6 of 6
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...