A/C ENZYMATIC HOMOCYSTEINE ASSAY

K030754 · Anticancer, Inc. · LPS · Jul 11, 2003 · Clinical Chemistry

Device Facts

Record IDK030754
Device NameA/C ENZYMATIC HOMOCYSTEINE ASSAY
ApplicantAnticancer, Inc.
Product CodeLPS · Clinical Chemistry
Decision DateJul 11, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1377
Device ClassClass 2

Indications for Use

The A/C Enzymatic Homocysteine Assay is intended for the quantitative determination of total homocysteine (tHCY) in human plasma or serum. The device can assist in the diagnosis and treatment of patients suspected of having hyperhomocysteinemia. The A/C Enzymatic Homocysteine Assay Kit is only for in vitro diagnostic use.

Device Story

A/C Enzymatic Homocysteine Assay is an in vitro diagnostic test for measuring total homocysteine (tHCY) in human plasma or serum. The assay uses recombinant homocysteinase (rHCYase) to convert tHCY into hydrogen sulfide (H2S), which reacts with N,N-dibutyl phenylene diamine (DBPDA) to form a measurable chromophore. The assay is performed on the Hitachi 912 Automatic Analyzer, which automates the four-step reaction process: DTT reduction, rHCYase reaction, chromophore formation, and oxidation. The analyzer measures absorbance at 700 and 660 nm. Results are used by clinicians to diagnose and manage hyperhomocysteinemia. The assay provides a throughput of 360 tests per hour, aiding clinical decision-making by providing quantitative tHCY levels.

Clinical Evidence

Bench testing only. Performance evaluated on the Hitachi 912 Automatic Analyzer. Precision: within-run CVs 3.0-4.8%; between-assay CVs 4.9-7.8%. Linearity up to 80 umol/L. Detection limit 1.5 umol/L. Interference testing showed <10% interference from L-cysteine and no interference from L-methionine at physiological concentrations. Method comparison with Bio-Rad HPLC (n=121) yielded r=0.977, y=0.98x+1.90, and mean difference of -1.62 umol/L (SD=2.33).

Technological Characteristics

Enzymatic assay using recombinant homocysteinase (rHCYase) and DBPDA chromophore. Operates on Hitachi 912 Automatic Analyzer. Four-step reaction: DTT reduction, rHCYase reaction, chromophore formation, and potassium ferricyanide oxidation. Measures absorbance at 700/660 nm. Linear range up to 80 umol/L. In vitro diagnostic use.

Indications for Use

Indicated for the quantitative determination of total homocysteine (tHCY) in human plasma or serum to assist in the diagnosis and treatment of patients suspected of having hyperhomocysteinemia.

Regulatory Classification

Identification

A urinary homocystine (nonquantitative) test system is a device intended to identify homocystine (an analogue of the amino acid cystine) in urine. The identification of urinary homocystine is used in the diagnosis and treatment of homocystinuria (homosystine in urine), a heritable metabolic disorder which may cause mental retardation.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## Summary of Safety and Effectiveness Information Supporting a Substantially Equivalent Determination #### Submitter Information 1. A/C Diagnostics Division AntiCancer Inc. 7917 Ostrow Str. San Diego, California 92111 Phone: (858)654-2555 FAX: (858)268-4175 e-mail: all(@anticancer.com Contact Person: Submitter: Yuying Tan, M.D. Research Manager AntiCancer Inc. Date of Summary Preparation: February 18, 2003 #### 2. Device Information Device Name: ### A/C Enzymatic Homocysteine Assay Classification Name: Class: Product Code: Homocysteine Assay II LPS #### 3. Predicate Device Information | Device Name: | Bio-Rad Homocysteine by HPLC<br>Bio-Rad Laboratories<br>Diagnostics Group<br>4000 Alfred Nobel Drive<br>Hercules, California 94547 | |--------------|------------------------------------------------------------------------------------------------------------------------------------| |--------------|------------------------------------------------------------------------------------------------------------------------------------| 510(k) Number: K993107 K030754 {1}------------------------------------------------ #### Information of Manufacturer 4. | Manufacturer: | |---------------| |---------------| Bioserv Corporation 5340 Eastgate Mall San Diego, CA 92121 Telephone: (858) 450-3123 (858) 450-0785 FAX: FDA establishment registration number: US FDA 2027352 Contact Person : Mary Richardson Quality Assurance Manager Bioserv Corporation #### న. Statement of Intended Use - The A/C Enzymatic Homocysteine Assay is intended for the quantitative determination . of total homocysteine (tHCY) in human plasma or serum. - The device can assist in the diagnosis and treatment of patients suspected of having . hyperhomocysteinemia. - The A/C Enzymatic Homocysteine Assay Kit is only for in vitro diagnostic use. . #### Description of Device 6. The A/C Enzymatic Homocysteine Assay is calibrated with A/C Enzymatic Homocysteine Assay Calibrators. The A/C Enzymatic Homocysteine Assay is assayed for the verification of the accuracy and precision on the Hitachi 912 Automatic Analyzer. The A/C Enzymatic Homocysteine Assay measures tHCY. The principle of the assay is that recombinant homocysteinase (rHCYase) produces hydrogen sulfide (H2S) from tHCY, which is quantified by use of N,N-dibutyl phenylene diamine (DBPDA), the combination of which forms a chromophore. The A/C Enzymatic Homocysteine Assay on the Hitachi 912 Automatic Analyzer used four reagents, a number compatible with implementation on the Hitachi 912 Automatic Analyzer. We used 30 uL of EDTA plasma in a dithiothreitol (DTT) reduction reaction (1 mmol/L DTT, 0.2% Triton X-100 in 40 mmol/L sodium phosphate buffer [pH 8.3]) for 1.5 minutes to release bound homocysteine. The rHCYase reaction (0.05 mg/ml in 40 mmol/L sodium phosphate buffer [pH 8.3] with 20 umol/L pyridoxal 5-phosphate [PLP]) is then run for 3.5 minutes. The DBPDA chromophore (12.5 mmol/L DBPDA in 1.5 N H2SO4) is then added and 5 minutes later, an oxidant, potassium ferricyanide (5 mmol/L K3Fe(CN), in 10 mmol/L sodium phosphate {2}------------------------------------------------ buffer [pH 7.6]), is added. Five minutes after addition of oxidant, the end-points are read at absorbances of 700 and 660 nm. As the assay is based on an increase in absorbance over baseline, no blank without enzyme was used. The detection limit of the assay is 1.5 umol/L defined by quantification of a serial dilution of a plasma sample of tHCY diluted to 0.77 umol/L. The limit of quantification is defined as the lowest concentration measured having a CV <20%. The linear range extends to at least 80 umol tHCY/L as determined by measuring varying amounts of homocysteine in phosphate buffered saline. The within-run imprecisions (CV) run over 8 repeats were 4.8% at 8.9 umol/L tHCY; 3.0% for 14.9 umol/L tHCY; and 4.5% for 25 umol/L tHCY. Between-assay imprecisions (CV) over 10 days were 7.8% for 8.8 umol/L tHCY: 5.9% for 15 umol/L tHCY; and 4.9% for 25 umol/L tHCY. These imprecisions are with in ranges reported for currently-used assays including the Bio-Rad HCY assay on HPLC, which is used as a comparison method in the present study. The results showed that L-cysteine (L-CYS) in the physiological concentrations (0-200 µmol/L) had less than 10% interference and L-methionine (L-MET) in the physiological concentrations (0-200 umol/L) had no interference in the A/C Enzymatic Homocysteine Assay on the Hitachi 912 Automatic Analyzer. The A/C Enzymatic Homocysteine Assay on the Hitachi 912 Automatic Analyzer is a four-step reaction. The total assay takes 15 minutes, and the through-put is 360 tests per hour. #### 7. Method Comparison To establish equivalence to an existing device, and thus establish the safety and effectiveness. the A/C Enzymatic Homocysteine Assay on Hitachi 912 Automatic Analyzer is compared to the Bio-Rad Homocysteine by HPLC (k 993107). We assayed 121 plasma samples with the A/C Enzymatic Homocysteine Assay on the Hitachi 912 Automatic Analyzer (y) and with the Bio-Rad Homocysteine Assay by HPLC (x). The regression equation was y = 0.98 + 1.90 (r = 0.977). The mean difference (SD) between the A/C Enzymatic Homocysteine Assay and the Bio-Rad Homocysteine by HPLC was -1.62 umol/L (SD=2.33). Differences were not significantly correlated with homocysteine concentration (Pearson r = 0.12, p = 0.185). {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 JUL 1 1 2003 Yuying Tan, M.D. Research Manager AntiCancer Inc. A/C Diagnostics Division 7917 Ostrow Street San Diego, CA 92111 Re: k030754 Trade/Device Name: A/C Enzymatic Homocysteine Assay Regulation Number: 21 CFR 862.1377 Regulation Name: Urinary homocystine (non-quantitative) test system Regulatory Class: Class II Product Code: LPS Dated: May 15, 2003 Received: May 16, 2003 Dear Dr. Tan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {4}------------------------------------------------ Page 2 - . This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Steven Putman Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Appendix-6 K 030754 # Indications for Use: - Enzymatic Homocysteine Assay is intended A/C for the ● The quantitative determination of total homocysteine (tHCY) in human plasma or serum. - The device can assist in the diagnosis and treatment of patients . suspected of having hyperhomocysteinemia. - The A/C Enzymatic Homocysteine Assay is only for in vitro diagnostic ● use. \/ prescription use Albert Juh Division Sign-Off for Jean Cooper Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) 050754
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...