K190742 · Neogenix, LLC Dba Ogenix · KPJ · Dec 13, 2019 · General, Plastic Surgery
Device Facts
Record ID
K190742
Device Name
EPIFLO-28
Applicant
Neogenix, LLC Dba Ogenix
Product Code
KPJ · General, Plastic Surgery
Decision Date
Dec 13, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5650
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The EPIFLO-28 is intended to provide topical oxygen to treat 1) skin ulcerations due to diabetes, venous stasis, post-surgical infections and gangrenous lesions, 2) pressure ulcers, 3) amputations/infected stumps, 4) skin grafts, 5) burns, and 6) frostbite.
Device Story
EPIFLO-28 is a single-patient, disposable, battery-operated oxygen concentrator; delivers 98-100% oxygen continuously at ~3.0 ml/hour for 28 days. Oxygen travels via sterile cannula to wound site beneath bandage; operates at ambient pressure. Used in home or clinical settings; patient or caregiver operates. Device features green/red LEDs for status/error monitoring. Provides topical oxygen therapy to promote healing; reduces pro-inflammatory cytokines, proteases, and macrophages in wound tissue. Benefits include extended 28-day therapy duration compared to predicate.
Clinical Evidence
Clinical study (n=61) evaluated predicate device usability over 12 weeks. Subject device usability study included lay users (n=17) and healthcare professionals (n=18). Clinical validation study evaluated wound response to delivery cannula via weekly measurements, wound fluid, and tissue punch biopsies over 4 weeks. Results demonstrated reduction in pro-inflammatory cytokines, proteases, and macrophages at 4-week follow-up.
Technological Characteristics
Single-patient, disposable, battery-operated oxygen concentrator. Uses Membrane Electrode Assembly for oxygen generation. Delivers ~3.0 ml/hour at ambient pressure. Enclosure contacts intact skin; cannula contacts breached/compromised skin. Biocompatibility testing per ISO 10993 standards (cytotoxicity, sensitization, irritation, pyrogenicity). Includes LED status indicators. No complex software algorithms; device is a mechanical/electrochemical oxygen delivery system.
Indications for Use
Indicated for patients with skin ulcerations (diabetic, venous stasis, post-surgical, gangrenous), pressure ulcers, amputations/infected stumps, skin grafts, burns, and frostbite.
Regulatory Classification
Identification
A topical oxygen chamber for extremities is a device that is intended to surround a patient's limb and apply humidified oxygen topically at a pressure slightly greater than atmospheric pressure to aid healing of chronic skin ulcers such as bedsores.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance: Topical Oxygen Chamber for Extremities.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
December 13, 2019
Neogenix, LLC dba Ogenix Srinivasan Sarangapani President 3401 Enterprise Pkwy, Suite 340 Beachwood, Ohio 44122
Re: K190742
Trade/Device Name: EPIFLO-28 Regulation Number: 21 CFR 878.5650 Regulation Name: Topical Oxygen Chamber For Extremities Regulatory Class: Class II Product Code: KPJ Dated: November 12, 2019 Received: November 12, 2019
Dear Srinivasan Sarangapani:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Kimberly M. Ferlin, Ph.D. Assistant Director (acting) DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190742
Device Name EPIFLO-28
Indications for Use (Describe)
The EPIFLO-28 is intended to provide topical oxygen to treat 1) skin ulcerations due to diabetes, venous stasis, post-surgical infections and gangrenous lesions, 2) amputations/infected stumps, 4) Skin grafts, 5) burns, and 6) frostbite
Type of Use (Select one or both, as applicable)
| <div> <span> <b> </b> Prescription Use (Part 21 CER 801 Subpart D) </span> </div> |
|-----------------------------------------------------------------------------------------------------------------------|
| <div> <span> <b> </b> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) SUMMARY FOR EPIFLO-28
#### 1. SPONSOR
Neogenix, LLC DBA Ogenix 3401 Enterprise Pkwy, Suite 340 Beachwood, OH 44122
Contact Person: Srinivasan (Sarang) Sarangapani, President 781-702-6732 Telephone:
Date Prepared: December 13, 2019
#### 2. Device Name
| Proprietary Name: | EPIFLO-28 |
|----------------------|-----------------------------------------------|
| Common/Usual Name: | Transdermal continuous Oxygen Therapy System |
| Classification Name: | Topical oxygen chamber for extremities |
| | 21 CFR 878.5650 – Class II (Special Controls) |
Product Code: KPJ
#### 3. PREDICATE DEVICES
EPIFLO System - K120764
#### 4. Device Description
EPIFLO-28 is an oxygen delivery device that incorporates a disposable oxygen concentrator. It consists of (1) the oxygen concentrator, and (2) the sterile cannula. The oxygen concentrator is a single patient, single use, disposable, battery-operated device that is capable of delivering 98 to 100 percent oxygen continuously for twenty eight days at a rate of ~3.0 ml/hour. The cannula conveys the oxygen from the oxygen concentrator to the area beneath the bandage overlying the wound.
#### 5. INDICATIONS FOR USE
The EPIFLO-28 is intended to provide topical oxygen to treat 1) skin ulcerations due to diabetes, venous stasis, post-surgical infections and gangrenous lesions, 2) pressure ulcers, 3) amputations/infected stumps, 4) skin grafts, 5) burns, and 6) frostbite.
{4}------------------------------------------------
#### 6. Technological Characteristics and Substantial Equivalence
The predicate device EPIFLO System (7 & 15 day models) as well as the current model EPIFLO-28 delivers oxygen to the wound at ambient pressure (typically 1 atmosphere). The predicate devices as well as the EPIFLO-28 deliver near 100 percent pure oxygen to the wound site at a rate of ~ 3 mL/hr). Both the Predicate EPIFLO 7 & 15 day models and the EPIFLO-28 are single patient, single use disposable devices. The Membrane Electrode Assembly in the EPIFLO-28 is the same as used in the predicate device EPIFLO System (7&15 day models).
The predicate device provides continuous oxygen for 15 days, whereas EPIFLO-28 provides continuous oxygen for 28 days. The predicate device uses different blinking rates of the red indicator light to indicate normal operation and end of life. EPIFLO-28 has two indicator LEDs (green and red) to indicate normal operation and error state and/or end of life. There is no error state indication in the predicate device. EPIFLO-28 has the same footprint as the predicate device, but is taller by quarter of an inch. The delivery cannula was changed from the predicate.
#### 7. Performance Testing
Testing was conducted to validate that the EPIFLO-28 performed according to its specifications. These tests included performance testing for oxygen delivery per the specification for labeled durations, LED operation verification and EMC compatibility. Test methods were the same or more as used in the predicate device.
Biocompatibility testing for EPIFLO-28 enclosure (permanent contact with intact skin) included: cytotoxicity, sensitization, and irritation; for the oxygen delivery cannula (permanent contact with breached/compromised skin) the tests included: cytotoxicity, sensitization, irritation, and material mediated pyrogenicity. In addition, the cannula was subjected to an extractable/leachable analysis. A toxicological risk assessment was performed on the chemical compounds identified in the extractables/leachables testing; no chemical compound was detected above the conservative threshold of toxicological concern. To address the local tissue response to the delivery cannula, a clinical study evaluated wounds exposed to the delivery cannula throuqh wound measurements, wound fluid, and tissue punch biopsies collected weekly for 4 weeks. The wound fluid and punch biopsies were evaluated for pro-inflammatory cytokines, proteases, and macrophaqes and the results demonstrated a reduction in pro-inflammatory cytokines, proteases, and macrophages at the Week 4 follow up.
To address usability in the home environment, patients (n = 61) in a clinical study used the predicate device for up to 12 weeks. At each visit, patients answered questions in the case report forms about any issues with using the device. A usability study was also conducted with the subject device to evaluate lay user (n = 17) and healthcare professional (n = 18) user experience with the EPIFLO-28.
{5}------------------------------------------------
#### CONCLUSIONS 8.
EPIFLO-28 delivers the same oxygen therapy as the predicate device, but does so for a longer period before replacement is required. It also has the same risk/benefit profile as the predicate. Based on these facts, and based on the results of the performance tests and clinical validation performed by Ogenix, it is concluded that EPIFLO-28 is as safe and as effective as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.