K012320 · Linvatec Corp. · HAB · Oct 19, 2001 · General, Plastic Surgery
Device Facts
Record ID
K012320
Device Name
POWERPRO BATTERY SYSTEM
Applicant
Linvatec Corp.
Product Code
HAB · General, Plastic Surgery
Decision Date
Oct 19, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4820
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The PowerPro™ Battery System functions as a powered instrument system consisting of handpieces and accessories to perform cutting of soft tissue and bone. The fields of application include orthopedic procedures such as osteotomies, small and large bone trauma, orthopedic reconstruction, and total joint replacement, neurosurgical fusion and plating procedures, and medial sternotomy. The PowerPro Battery System also includes a sterilization case to enclose the rechargeable batteries during sterilization, and also to maintain sterility of the batteries until used.
Device Story
PowerPro Battery System is a powered surgical instrument system; comprises pistol-grip handpieces, rechargeable batteries, universal battery charger, and sterilization trays. Handpieces accept various attachments including blades, burs, drills, and routers. Used in OR settings by surgeons to perform cutting of soft tissue and bone during orthopedic, neurosurgical, and sternotomy procedures. System includes a sterilization case for battery processing and sterile storage. Device provides mechanical power to surgical attachments; enables precise bone/tissue resection; facilitates surgical workflow in orthopedic and neurosurgical interventions.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Powered surgical instrument system; includes handpieces, rechargeable batteries, charger, and sterilization case. Mechanical actuation for cutting attachments (blades, burs, drills, routers).
Indications for Use
Indicated for patients undergoing orthopedic procedures (osteotomies, bone trauma, reconstruction, joint replacement), neurosurgical fusion and plating, and medial sternotomy requiring soft tissue and bone cutting.
Regulatory Classification
Identification
Surgical instrument motors and accessories are AC-powered, battery-powered, or air-powered devices intended for use during surgical procedures to provide power to operate various accessories or attachments to cut hard tissue or bone and soft tissue. Accessories or attachments may include a bur, chisel (osteotome), dermabrasion brush, dermatome, drill bit, hammerhead, pin driver, and saw blade.
Predicate Devices
PowerPro Electric System (Linvatec Corporation)
Advantage System (Linvatec Corporation)
Synthes Power Drive System (Synthes)
Allegiance Genesis Container System (Allegiance Healthcare Corp.)
Submission Summary (Full Text)
{0}------------------------------------------------
July 20, 2001
# SUMMARY OF SAFETY AND EFFECTIVENESS
In accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92, Linvatec Corporation is hereby submitting the 510(k) Summary of Safety and Effectiveness for the PowerPro™ Battery System, 510(k) Number
## Submitter A.
·
Linvatec Corporation 11311 Concept Boulevard Largo, Florida 33773-4908
## B. Company Contact
Laura D. Seneff, RAC Manager, Regulatory Affairs (727) 399-5234 Telephone (727) 399-5264 FAX
# C. Device Name
Trade Name: PowerPro™ Battery System
Common Name: Battery powered surgical instrument and accessories/attachments
Classification Names: Surgical instrument motors and accessories/attachments 878.4820 Sterilization Wrap 880.6850
Proposed Class/Device: II Product Code HAB, FRG
{1}------------------------------------------------
K012320
Summary of Safety and Effectiveness PowerPro™ Battery System 510(k) # July 20, 2001 Page 2 of 2
# Predicate/Legally Marketed Devices D.
PowerPro™ Electric System, Linvatec Corporation Advantage™ System, Linvatec Corporation Synthes® Power Drive System, Synthes Allegiance Genesis™ Container System, Allegiance Healthcare Corp.
## E. Device Description
The PowerPro Battery System is a powered instrument system consisting of various pistol-grip handpieces, universal battery charger, rechargeable batteries and sterilization trays. The handpieces utilize various attachments, such as blades, burs, drills and routers.
#### ui Intended Use
The PowerPro™ Battery System functions as a powered instrument system consisting of handpieces and accessories to perform cutting of soft tissue and bone. The fields of application include orthopedic procedures such as osteotomies, small and large bone trauma, orthopedic reconstruction, and total joint replacement, neurosurgical fusion and plating procedures, and medial stemotomy.
The PowerPro Battery System also includes a sterilization case to enclose the rechargeable batteries during sterilization, and also to maintain sterility of the batteries until used.
#### G. Substantial Equivalence
The PowerPro Battery System is substantially equivalent in design, function and intended use to the predicate devices named above. The PowerPro Battery System does not raise any new safety or effectiveness issues when compared to these similar devices.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol with three curved lines representing its wings or body. The bird is positioned to the right of a circular text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 1 9 2001
Ms. Laura D. Seneff, RAC Manager, Regulatory Affairs Linvatec Corporation 11311 Concept Boulevard Largo, Florida 33773
Re: K012320 Trade/Device Name: PowerPro™ Battery System Regulation Number: 878.4820, 880.6850, 882.4360 Regulation Name: Surgical instrument motors and accessories/attachments Sterilization wrap Electric cranial drill motor Regulatory Class: II
Product Code: GEY, FRG, HBC Dated: July 20, 2001 Received: July 23, 2001
Dear Ms. Seneff:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave teviewed your becally be device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreativent of the enactment date of the Medical Device Amendments, or to conninered proc to May 20, 1978, are cance with the provisions of the Federal Food, Drug, devices that have ooon require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The 1 ou may, merelore, manel of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it IT your device is classified (600 a0010) als. Existing major regulations affecting your device can
may be subject to such additional controls. Existing major regulations an may oe subject to back additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I loast of advised that I Driversion that your device complies with other requirements of the Act that I Drinas matures and regulations administered by other Federal agencies. You must or any I catales and the Act's requirements, including, but not limited to: registration and listing (21 Comply with an the For 8 requirements) and manufacturing practice requirements as set OF N Far 807), adoling (21 CFR Part 820); and if applicable, the electronic forth in the quality by sections (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Laura D. Seneff, RAC
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Susan Walk, us
Fo
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
July 20, 2001
510(k) Number (if known): Kol2320
Device Name: PowerPro™ Battery System
Indications for Use:
The PowerPro™ Battery System functions as a powered instrument system consisting of handpieces and accessories to perform cutting of soft tissue and bone. The fields of application include orthopedic procedures such as osteotomies, small and large bone trauma, orthopedic reconstruction, and total joint replacement, neurosurgical fusion and plating procedures, and medial sternotomy.
The PowerPro Battery System also includes a sterilization case to enclose the rechargeable batteries during sterilization, and also to maintain sterility of the batteries until used.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE If NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
OR
Over-the-Counter Use
(Per 21 CFR 801.109)
(Optional Format 1-2-96)

(Division Sign-Off) Division of General, Restorative and Neurological Devices
510(k) Number K012320
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.