K991632 · Fotona D.D. · GEX · Aug 10, 1999 · General, Plastic Surgery
Device Facts
Record ID
K991632
Device Name
FOTONA NOVALIS ER: YAG LASER SYSTEM
Applicant
Fotona D.D.
Product Code
GEX · General, Plastic Surgery
Decision Date
Aug 10, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Fotona Novalis Er: Y AG Laser System and Accessories are intended for surgical incision/excision, cutting, ablation, vaporization, and coagulation of soft tissue. All soft tissue is included, such as, skin, cutaneous tissue, subcutaneous tissue, striated and smooth tissue, meniscus, mucous membrane, lymph vessels and nodes, organs, and glands. DermatologyPlastic Surgery : epidermal nevi, telangiectasia, spider veins, actinic chellitis, keloids, verrucae, skin tags, anal tags, keratoses, scar revision (including acne scars), debulking benign tumors and cysts, skin resurfacing, superficial skin lesions, and performing diagnostic biopsies. General Surgery : surgical incision/excision, cutting, ablation, vaporization, and coagulation of soft tissue where skin incision, tissue dissection, excision of external tumors and lesions, complete or partial resection of internal organs, tumors and lesions, tissue ablation and/or vessel coagulation may be indicated. Genitourinary : lesions of the external genitalia, urethra and anus, penis, scrotum and urethra, vulvar lesions, polyps and familial polyps of the colon. Gynecology : cerivecal intraepithelial neoplasia (CIN), herpes simplex, endometrial adhesions, cysts and condyloma. Oral/Maxillofacial : benign oral tumors, oral and glossal lesions and gingivectomy. Otorhinolaryngology/Head and Neck (ENT) : ear, nose and throat lesions, polyps, cysts, lyperkeratosis, excision of carcinogenic tissue and oral leukoplakia. Ophthalmology : soft tissue surrounding the eye and orbit anterior capsulotomy. Podiatry : warts, plantar vernicae, large mosaic verrucae and matrixectorny.
Device Story
Microprocessor-controlled Er:YAG laser system; generates 2940 nm pulsed light. Integrated as an accessory to the Fotona Novalis Ruby laser system; shares touchscreen interface. Comprises optical delivery system (articulated arm, handpiece), electronic power supply, and optical chamber (laser rod/cavity). Operated by clinicians in surgical/clinical settings. Laser energy delivered to soft tissue for incision, excision, ablation, vaporization, and coagulation. Provides precise tissue interaction for various surgical specialties; benefits include controlled tissue removal and hemostasis.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and intended use comparison to predicate devices.
Technological Characteristics
Flashlamp-pumped Er:YAG laser; 2940 nm wavelength. Microprocessor-controlled. Optical delivery via articulated arm and focusing/collimated handpiece. Electronic power supply and interface circuitry. Accessory: Fotona SkinScan. Class II device.
Indications for Use
Indicated for surgical incision, excision, cutting, ablation, vaporization, and coagulation of soft tissue across dermatology, plastic surgery, general surgery, genitourinary, gynecology, oral/maxillofacial, ENT, ophthalmology, and podiatry specialties. Applicable to patients requiring soft tissue management including tumors, lesions, cysts, and various dermatological conditions.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
AUG 10 199
$$\ll \approx \ll 3\prime$$
#### Summary of Safety and Effectiveness Data Appendix E u
# General Information and Description
The Fotona Novalis Er. YAG system ia a microprocessor controlled device which generates lascr light with a wavelongth of 2940 nm when used in conjunction with a host Novelis R system. When combined, the Er. YAG accessory and the host Ruby system constitute the Novalis RE laser system.
The Fotona Er. Y AG system is designed as an accessory for use with the Fotona Novalis R laser system "The Fir Y A ( suh-system is fincrionally integrated to the host laser system. When integrated, the host laser system recognizes the presence of the accessory and permits activation of the 2940 nm pulsed light via the same touchscreen as the host Ruby
The Novalis Er. YAG system is designed with 3 major sub-systems:
d) An optical dclivery system, interfacing the cncrgy from the lascr to the patient via an articulated arm and a focusing or collimated handpiece.
b) An electronic power supply and interfsce circuitry.
c) An optical chamber containing laser rod and laser cavity optics.
Accessories available for use with the Fotona Novalis Er: Y AG:
- Fotona SkinScan Scanning Device .
### Summary of Substantial Equivalence
Fotona believes that its Novalis system is substantially equivalcnt to the Lascrscope Venus Erbium Laser System (EL Laser System) and to the Continuum Biomedical (Con-Bio) CB Erbium/2.94 Er: YAG Laser.
Both lasers are cleared for surgical incision, cutting, ablation, vaportzation, and coagulation of soft tissue All soft tissue is included, such as, skin, cutaneous tissue, subcutaneous tissue, striated and smooth tissue, muscle, cartilage meniscus, mucous membrane, lymph vessels and nodes, organs, and glands.
Technologically, the predicate devices have identical characteristics to the Novalis Er: Y AG laser, all three comprising an electronic control module and a flashlamp pumped Er. YAG laser rod generating light at a wavelength of 2.94 um, which is subsequently delivered to the patient via an articulated delivery arm and focusing or collimated handpiece.
The risk and benefits for the Fotona Novalis are comparable to the Laserscope Venus Laser System and Continuum Biomedical (Con-Bio) CB Erbium/2.94 Er: YAG Laser System when used for similar clinical applications.
It is therefore believed that there are no new questions of Safety or Effectiveness raised by the introduction of this device.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 10 1999
Ms. Mojca Valjavec, Dipl. Eng. Product Manager Lasers Division Fontona d.d. Stegne 7, 1210 Ljubljana, Slovenia
K991632 Re: K991634 Trade Name: Fontona NovalisR Ruby Laser System Fontona Novalis Er: YAG Laser System Regulatory Class: II Product Code: GEX Dated: May 3, 1999 Received: May 12, 1999
Dear Ms. Valjavec:
We have reviewed your Section 510(k) notification of intent to market the devices referenced above, and we have determined the devices are substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the devices, subject to the general control provisions of the Act. The general control provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your devices are classified (see above) into either class II (Special Controls) or class III (Premarket Approval), they may be subject to such additional controls: Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General Regulation (21 CFR Part 820), and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition. FDA may publish further announcements concerning your devices in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{2}------------------------------------------------
# Page 2 – Ms. Mojca Valjavec, Dipl. Eng.
This letter will allow you to begin marketing your devices as described in your 510(k) premarket notification. The FDA, finding of substantial equivalence of your devices to legally marketed predicate devices, results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your devices, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification (21 CFR 807.97)." Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html."
Sincerely yours,
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{3}------------------------------------------------
## APPENDIX F
Page I of 1
# INDICATIONS FOR USE STATEMENT
510(k) Number (if known):
Device Name:
. 201
# FOTONA NOVALIS Er: YAG LASER SYSTEM
Indications For Use:
The Fotona Novalis Er: Y AG Laser System and Accessories are intended for surgical incision/excision, cutting, ablation, vaporization, and coagulation of soft tissue. All soft tissue is included, such as, skin, cutaneous tissue, subcutaneous tissue, striated and smooth tissue, meniscus, mucous membrane, lymph vessels and nodes, organs, and glands
DermatologyPlastic Surgery : epidermal nevi, telangiectasia, spider veins, actinic chellitis, keloids, verrucae, skin tags, anal tags, keratoses, scar revision (including acne scars), debulking benign tumors and cysts, skin resurfacing, superficial skin lesions, and performing diagnostic biopsies.
General Surgery : surgical incision/excision, cutting, ablation, vaporization, and coagulation of soft tissue where skin incision, tissue dissection, excision of external tumors and lesions, complete or partial resection of internal organs, tumors and lesions, tissue ablation and/or vessel coagulation may be indicated
Genitourinary : Jesions of the external genitalia, urethra and anus, penis, scrotum and urethra, vulvar lesions, polyps and familial polyps of the colon
Gynecology : cerivecal intraepithelial neoplasia (CIN), herpes simplex, endometrial adhesions, cysts and condyloma
Oral/Maxillofacial : benign oral tumors, oral and glossal lesions and gingivectomy
Otorhinolaryngology/Head and Neck (ENT) : ear, nose and throat lesions, polyps, cysts, lyperkeratosis, excision of carcinogenic tissue and oral leukoplakia
Ophthalmology : soft tissue surrounding the eye and orbit anterior capsulotomy
Podiatry : warts, plantar vernicae, large mosaic verrucae and matrixectorny
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use<br>(Per 21 CFR 801.109) | <div style="text-align:left;">X</div> |
|------------------------------------------|---------------------------------------|
|------------------------------------------|---------------------------------------|
OR
| (Division Sign-Off) | |
|-----------------------------------------|---------|
| Division of General Restorative Devices | |
| 510(k) Number | K991632 |
| Over-The-Counter Use | |
Premarket Notification for Fotona Novalis Fri VAC. I asar Custom
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.