DILKES LASER/SUCTION CANNULAE

K981143 · Exmoor Plastics , Ltd. · GEX · Sep 14, 1998 · General, Plastic Surgery

Device Facts

Record IDK981143
Device NameDILKES LASER/SUCTION CANNULAE
ApplicantExmoor Plastics , Ltd.
Product CodeGEX · General, Plastic Surgery
Decision DateSep 14, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Any surgical procedure within the nasal cavity, the oral cavity or the oropharynx, which requires laser surgery and suction capability to remove smoke.

Device Story

Dilkes Laser/Suction Cannulae is a single-use surgical instrument designed for laser surgery and smoke evacuation in the nasal, oral, or oropharyngeal cavities. The device consists of an Exmoor single-use suction tube, stainless steel needle tubing, and silicone rubber tubing to maintain alignment. The device connects to external suction via a proximal ball feature. A laser fiber is inserted through the stainless steel tubing, emerging at the distal end for surgical use. The device is operated by a surgeon during procedures to simultaneously provide suction and laser delivery. By removing surgical smoke, the device improves visibility and safety for the clinician and patient.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Materials: stainless steel needle tubing, silicone rubber tubing, polyethylene handle. Design: single-use, multi-channel suction/laser delivery cannula. Energy: passive suction (external source) and laser fiber delivery. Sterilization: not specified.

Indications for Use

Indicated for patients undergoing surgical procedures within the nasal cavity, oral cavity, or oropharynx requiring laser surgery and smoke evacuation.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 2. SEP 1 4 1998 K981143 ( lontract: Margaret Blackmore 4 September 1998. Ref: K981143 SUMMARY OF SAFETY AND EFFECTIVENESS | Trade Name: | Dilkes Laser/Suction Cannulae | |------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Common Name: | Twin Channel Suction Tube | | Classification Name: | Unknown | | Predicate Devices: | 1. House Suction/Irrigation Tube 55-0207, manufactured by Richards,<br>2. Microstat hand-piece from Laserscope | | Description of Device: | This device is constructed from an Exmoor Single Use Suction Tube (MP17291), 510K 892504, a piece of stainless steel needle tubing (MP170288) identical to that used in MP170291, covered by the same 510K 892504, and a piece of silicone rubber tubing (MP180163) which is used to hold the two pieces of stainless steel tubing together in the correct position. It uses locally supplied suction tubing to connect the device to suction by pushing the tubing over the ball feature at the proximal end of the handle. The laser fibre is passed down the exposed end of the stainless steel tubing adjacent to the handle until it emerges from the distal end. The laser fibre protruding from the distal end of the instrument should be stripped and cleaved and adjusted to the required length. | | Intended Use: | Any surgical procedure within the nasal cavity, the oral cavity or the oropharynx, which requires laser surgery and suction capability to remove smoke. | | Comparison with Predicate Devices: | The size of the House Suction/Irrigation Tube is very similar to the Dilkes Laser/Suction Cannulae.<br><br>The predicate device is currently being used for suction/Laser delivery.<br><br>The materials are similar in that stainless steel is used in both cases. The handle of the subject device is polyethylene, as this is a single - use device. The predicate device is re-usable. | {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other, with flowing lines suggesting movement or connection. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP 1 4 1998 Ms. Margaret Blackmore Regulatory Affairs Administrator Exmoor Plastics Limited Lisieux Way Taunton, TA1 2LB, U.K. Re: K981143 Trade Name: Dilkes Laser/Suction Cannulae Regulatory Class: II Product Code: GEX Dated: August 5, 1998 Received: August 11, 1998 Dear Ms. Blackmore: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {2}------------------------------------------------ ## Page 2 - Ms. Margaret Blackmore This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Celia M. Witten, Ph.D., M.D. Director Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ 510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________ Device Name:___________DILKES LASER/SUCTION CANNULAE_ Indications for Use: Any surgical procedure within the nasal cavity, the oral cavity or the Any surgical procedure within the nasar surgery and suction capability to remove smoke. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | (Division Sign-Off) | |---------------------| |---------------------| Division of General Restorative Devices K98511(Division Sign-Om) Division of General Restorative Devices K981143 510(k) Number Prescription Use_ (Per 21 CFR 801.109) Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________ (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...