K971539 · Xanova Corp. · GEX · Jun 3, 1997 · General, Plastic Surgery
Device Facts
Record ID
K971539
Device Name
NOVATIP FAMILY OF LASER HANDPIECE TIPS
Applicant
Xanova Corp.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jun 3, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the Xanova tips is to comprise the distal end element of a surgical CO2 laser beam delivery system for the incision, excision, ablation, or photocoagulation of soft tissue.
Device Story
Novatip NT family consists of disposable distal end tips for hollow fiber CO2 surgical laser delivery systems. Tips function as optical waveguides to direct laser energy for soft tissue vaporization, ablation, incision, or photocoagulation. Used in clinical settings (gynecology, dermatology, dentistry, general surgery, ENT, podiatry) by physicians. Device receives laser energy from the CO2 laser system and outputs a focused beam to the target tissue. Benefits include precise tissue interaction for various surgical procedures. Xanova tips are designed without toxic lead fluoride, unlike predicate devices.
Clinical Evidence
No clinical data were presented. Substantial equivalence is based on bench testing and technological comparison to the predicate device.
Technological Characteristics
Disposable hollow fiber laser handpiece tips. Operates as an optical waveguide for CO2 laser energy. Materials are similar to predicate but exclude lead fluoride. Maintains same mode structure, spot size, and power density as predicate. Non-electronic, passive optical component.
Indications for Use
Indicated for soft tissue incision, excision, ablation, or photocoagulation in gynecology (condyloma, neoplasia, cysts), dermatology (hemangioma, rhinophyma, warts, carcinoma, dermabrasion), dentistry/oral surgery (gingivectomy, frenum release, soft tissue tumors), general surgery (hemorrhoids, skin tags, cysts, ulcers, mastectomy, biopsy), otorhinolaryngology (lymphangioma, turbinectomy), and podiatry (warts, fungal nails, matrixectomy, neuroma).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
K771J39
Image /page/0/Picture/1 description: The image shows the logo for Xanova Corporation. The logo consists of a black square with a white starburst pattern inside, followed by the word "Xanova" in large, bold, black letters. Below the word "Xanova" is the word "CORPORATION" in smaller, sans-serif, black letters.
JUN - 3 1997
# 9.0 510(k) Summary
Page 1 of 2. -
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.
#### 1 . Submitter's name, address, telephone, and contact person:
Xanova Corporation 15018 N.E. 177th Drive Woodinville, WA 98072 206/481-6223 Michael Levy, President
2. Date summary prepared: April 11, 1997
3. Product trade or proprietary name: Novatip NT family of laser handpiece tips
4. Product common name: Laser Handpiece Tips
#### 5. Product classification name:
21CFR § 878.4810 Lasers in general and plastic surgery and in dermatology 21CFR § 874.4500 Lasers for use in ENT 21CFR § 884.4550 Gynecologic surgical laser
### 6. Legally marketed predicate devices used for equivalency:
Luxar laser handpiece tips models LXT-C5, LXT-C10, LXT-005TSS, LXT-120ST, LXNS-T3, as cleared under K960475 Luxar Modified CO2 Surgical Laser System
7 . Descriptions The Xanova Novatips constitute the disposable end element of the hollow fiber deliverv system of a CO2 surgical laser.
8. Statement of intended use: The intended use of the Xanova tips is identical to that of the Luxar tips, namely to constitute the disposable end element of the delivery system of the Luxar laser used for the vaporization, ablation, incision, or photocoagulation of soft tissue in the surgical specialties of gynecology, dermatology, dentistry and oral surgery, general surgery, otorhinolaryngology, and podiatry. No new indications were sought in the premarket notification and no clinical data were presented.
15018 N.E. 177th Drive
Woodinville, WA 98072
g
Phone/Fax (206)481-6229
{1}------------------------------------------------
Page 2 of 2
#### 9. Technological characteristics:
The Xanova tips function in the same manner, and according to the same optical principles as the predicate Luxar tips. The Xanova tips operate within the same range of spot size, power, and power density as the Luxar tips, and have similar transmission efficiencies. The output beam of the Xanova tips has the same mode structure as the output beam of the Luxar tips, and the effect on tissue is the same. The Xanova tips are constructed of materials that are also found in the Luxar tips, although these materials are incorporated in different ways in some cases. The Xanova tips however, do not use toxic lead fluoride, as do the Luxar tips. Xanova believes that the minor differences in performance and differences in construction techniques (where they exist) do not raise any new issues of safety and effectiveness.
Advisory: This information was prepared for the sole purpose of compliance with the Safe Medical Devices Act of 1990. It does not imply that the procedures described herein can be performed with the equipment described without substantial risk of personal injury or death to patients or operators due to operator error or in high-risk procedures.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal for the Department of Health & Human Services USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN - 3 1997
Mr. Michael Levy President Xanova Corporation 15018 N.E. 177th Drive Woodinville, Washington 98072
Re: K971539
> Trade Name: Novatip Family of Laser Handpiece Tips Regulatory Class: II Product Code: GEX Dated: April 11, 1997 Received: April 28, 1997
Dear Mr. Levy:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Fractice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 - Mr. Michael Levy
,
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 -and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
KG)134
## 5.2 Intended Use and Indications for Use
The intended use of the Xanova tips is to comprise the distal end element of a surgical CO2 laser beam delivery system for the incision, excision, ablation, or photocoagulation of soft tissue.
Representative examples of indications for use in various specialties are listed below.
Gynecology-excision and vaporization of cervical, vulvar, and perineal condyloma; ablation. of vaginal and vulvar intraepithelial neoplasia; he pes vaporization of urethral caruncle; I&D Bartholin's and nubothian cysts.
Dermatology-port wine hemangioma removal, rhinophyma reduction; telangiectasia removal; wart removal; basal squamous cell carcinoma removal; blepharoplasty; xanthalasma removal; removal of neurofibromas, hemangiomas, nevi, and tricoeptiheliomas; dermabrasion for lentigos, keratoses, actinic keratosis and actinic chleilitis.
Dentistry/Oral Surgery-gingivectomy; frenum release; gingivoplasty; removal of soft tissue, cysts, and tumors.
General Surgery-hemorrhoid removal; skin tag vaporization; pilodidal cyst removal and repair; debridement of decubitus ulcers and stasis ulcers; mastectomy; breast biopsy; reduction mammoplasty; cytoreduction for metastatic disease; many dermatological procedures.
Otorhinolaryngology- - lymphangioma removal; turbinectomy.
Podiatry-plantar wart vaporization; fungal nail treatment; partial and complete matrixectomy; porokeratoma ablation; Morton's neuroma removal; ingrown toenail treatment.
pcoolfs
Prescription Use
(Per 21 CFR 801.109)
10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.