FREQUENCY DOUBLED ND:YAG PHOTOCOAGULATOR(VIRIDIS) FREQUENCY DOUBLED ND:YAG PHOTOCOAGULATOR(EVERGREEN 532)

K960867 · B.V. Intl. S.A. · GEX · May 14, 1996 · General, Plastic Surgery

Device Facts

Record IDK960867
Device NameFREQUENCY DOUBLED ND:YAG PHOTOCOAGULATOR(VIRIDIS) FREQUENCY DOUBLED ND:YAG PHOTOCOAGULATOR(EVERGREEN 532)
ApplicantB.V. Intl. S.A.
Product CodeGEX · General, Plastic Surgery
Decision DateMay 14, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Used with a slit lamp adaptor, this system is indicated for use in retinal photocoagulation for the treatment of ophthalmic conditions including : Proliferative Diabetic Retinopathy, Macular Degeneration, and Retinal Detachment.

Device Story

The VIRIDIS (also marketed as EVERGREEN 532) is an ophthalmic laser photocoagulator. It utilizes an Nd:YAG slab laser source pumped by an 810 nm laser diode, with a KTP crystal for frequency doubling to produce a 532 nm treatment beam. A 670 nm red laser diode serves as the aiming beam. The device delivers energy through three interchangeable systems: slit lamp adaptor, indirect ophthalmoscope adaptor, or endocular probe adaptor. It is operated by ophthalmologists in clinical settings to perform retinal photocoagulation. The device provides selectable exposure durations (0.01 to 3.0 seconds) and a maximum output power of 1.5 Watts. By delivering precise laser energy to the retina, the device enables the treatment of conditions such as proliferative diabetic retinopathy, macular degeneration, and retinal detachment, aiming to stabilize or improve patient vision.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Laser source: Nd:YAG slab; Pumping: 810 nm laser diode; Frequency doubling: KTP crystal; Output wavelength: 532 nm; Aiming beam: 670 nm diode laser; Delivery systems: Slit lamp, indirect ophthalmoscope, endocular probe; Cooling: Solid state; Power: Max 1.5 W; Pulse repetition rate: 12,000 Hz; Focal spot size: 50 micrometers; Exposure range: 0.01 to 3.0 seconds.

Indications for Use

Indicated for patients requiring retinal photocoagulation for ophthalmic conditions including Proliferative Diabetic Retinopathy, Macular Degeneration, and Retinal Detachment.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{1} BVI BIOVISION K960867 MAY 14 1996 Page 1/3 510(k) Summary of Safety and Effectiveness The following information is submitted in accordance with the requirements of 21 CFR 807.92 : ## Submitted by : ![img-0.jpeg](img-0.jpeg) Jean Abascal B. V. International 89, Boulevard Etienne CLEMENTEL 63100 Clermont-Ferrand FRANCE Contact person : Jean Abascal Telephone number : (33) 73 25 62 27 FAX number : (33) 73 25 54 11 **Date Prepared :** February 21st 1996 **Device Name :** Classification Name : Ophthalmic Laser Common / Usual Name : Ophthalmic Laser Photocoagulator Proprietary Name : VIRIDIS Frequency Doubled Nd:YAG Photocoagulator Other Proprietary Name : EVERGREEN 532 Frequency Doubled Nd:YAG Photocoagulator **Classification :** Devices Classification Panel : Ophthalmic Product code : 86HQF Ophthalmic lasers are Class II medical devices in accordance with 21 CFR § 886.4390. **Note :** "VIRIDIS" and "EVERGREEN 532" are identical. Only the housing is different. B.V.INTERNATIONAL : 89, Boulevard Etienne Clémentel - 63100 Clermont-Ferrand - Tel : 73 25 62 27 FRANCE Fax : 73 25 54 11 {2} BVI BIOVISION Page 2 / 3 510(k) Summary of Safety and Effectiveness ## Predicate Devices : - An other Frequency Doubled Nd:YAG Photocoagulator : The EMERALD from Crystal Focus - A Continuous-wave Argon Laser.: The NOVUS 2000 from Coherent ## Device Description The laser source is an Nd:YAG Slab, pumped by a Laser Diode (810 nm), with a frequency doubling crystal (KTP). The treatment beam is delivered through a delivery system and is aimed using a Red Laser Diode (670 nm). The 3 possible delivery systems are the following : - Slit lamp adaptor - Indirect ophthalmoscope adaptor - Endocular probe adaptor Output power of all these delivery systems have a maximum of 1.5 Watts. ## Intended Use : Used with a slit lamp adaptor, this system is indicated for use in retinal photocoagulation for the treatment of ophthalmic conditions including : Proliferative Diabetic Retinopathy, Macular Degeneration, and Retinal Detachment. ## Predicate Comparison : A chart comparing the VIRIDIS (or EVERGREEN 532) with the 2 predicate devices, demonstrating substantial equivalence, is attached. {3} Page 3 / 3 BVI BIOVISION 510(k) Summary of Safety and Effectiveness ## Device Comparison | Manufacturer | Coherent Medical | B.V.International | Crystal Focus | | --- | --- | --- | --- | | Model | NOVUS 2000 | VIRIDIS or EVERGREEN 532 | EMERALD | | Same Intended Use | Retinal Photocoagulation for the treatment of Ophthalmic conditions. | | | | Same type of Delivery systems | 1°) Slit Lamp adaptor with Zoom 2°) Indirect Ophthalmoscope adaptor 3°) Endocular Probe | | | | Lasing medium | Argon | ND:YAG | ND:YAG | | Pumping system | High Voltage | Diode Laser 810 nm | Flash | | Doubling Frequency Crystal | --- | KTP | KTP | | Output Wavelength | 488 - 514 nm | 532 nm | 532 nm | | Pulsing system | Continuous | Pseudo-Continuous | Pulsed | | Pulse Duration | --- | 25 micro sec. | 1 micro sec. | | Average Power | 2.5 W | 1.5 W | 2 W | | Pulse Repetition Rate | Continuous | 12 000 Hz | 10 000 Hz | | Mini. Focal Spot Size : Diameter | 50 micro meters | 50 micro meters | 50 micro meters | | Aiming beam : | Coaxial with treatment beam | | | | Aiming beam : type | He-Ne laser | Diode Laser | He-Ne laser | | Aiming beam : Wavelength | 632 nm | 670 nm | 632 nm | | Cooling method | Internally liquid cooled | Solid state cooling | Internally liquid cooled | | Exposure Selections ( t ) | 0.01s to continuous | 0.01 -0.02 -0.05 -0.07 -0.1 -0.2 -0.3 -0.4 -0.5 -0.7 -1.0 -2.0 -3.0 sec. | 0.01- 0.02- 0.05- 0.1-0.2- 0.3- 0.4 -0.5 -0.7 -1.0 and 3.0 sec. | | Power requirements : Voltage | 208 - 240 Vac | 100 to 240 Vac | 208 - 240 Vac | | Power requirements : Current | 45 A | 5 A | 13 A |
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...