Diode Laser Therapy Systems

K220381 · Beijing Lasertell Medical Co., Ltd. · GEX · May 20, 2022 · General, Plastic Surgery

Device Facts

Record IDK220381
Device NameDiode Laser Therapy Systems
ApplicantBeijing Lasertell Medical Co., Ltd.
Product CodeGEX · General, Plastic Surgery
Decision DateMay 20, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Diode Laser Therapy Systems is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

Device Story

Diode Laser Therapy Systems (AlexMED Pro) is a surgical laser device for hair removal; utilizes semiconductor diode emitting 808nm infrared radiation. Principle of operation: photoepilation; laser energy absorbed by melanin in hair follicles; thermal destruction of hair follicle and reproductive stem cells. Device consists of laser source, handpiece, and foot-switch. Operated by clinicians in professional settings. Output is laser energy delivered to treatment area; thermal effect destroys follicles without damaging surrounding epidermis. Benefits include long-term, stable reduction of hair regrowth.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing only, including electrical safety, EMC, laser safety, and biocompatibility testing.

Technological Characteristics

Class IV diode laser; 808nm±2nm wavelength; 1-100J/cm² fluence; 1-10Hz frequency; 1-300ms pulse duration. Materials: 6061 Aluminum, Sapphire, ABS plastic. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1, ISO 10993-5, ISO 10993-10. Foot-switch activated.

Indications for Use

Indicated for hair removal and permanent hair reduction on all skin types (Fitzpatrick I-VI), including tanned skin.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. May 20, 2022 Beijing LaserTell Medical Co., Ltd. Mr. Ray Wang Official Correspondent Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District Beijing, Beijing 102401 China Re: K220381 Trade/Device Name: Diode Laser Therapy Systems Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: April 18, 2022 Received: April 22, 2022 Dear Ray Wang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K220381 Device Name Diode Laser Therapy Systems #### Indications for Use (Describe) The Diode Laser Therapy Systems is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------|---------------------------------------------------------------------------------| | <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ The assigned 510(k) Number: K220381 # 510(k) Summary This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92. - 1. Date of Preparation:2022/05/18 - 2. Sponsor Identification ## Beijing LaserTell Medical Co., Ltd. Block1, No.12 Jingsheng South 2nd Road, JQ Science Park, Tongzhou District, Beijing, China. Contact Person: Zeng Xun Position: General Manager Tel: +86 13810934930 Fax: +86 10 56370036 Email: alex(@lasertell.com - 3. Designated Submission Correspondent Mr. Ray Wang ### Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, Beijing, 102401, China Tel: +86-18910677558 Fax: +86-10-56335780 Email: information(@believe-med.com {4}------------------------------------------------ #### 4. Identification of Proposed Device Trade Name: Diode Laser Therapy Systems Common Name: Powered Laser Surgical Instrument Model(s): AlexMED Pro Regulatory Information: Classification Name: Powered Laser Surgical Instrument Classification: II; Product Code: GEX; Regulation Number: 21 CFR 878.4810; Review Panel: General & Plastic Surgery; Indication For Use Statement: The Diode Laser Therapy Systems is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. #### న్. Device Description Diode Laser Therapy Systems is a surgical device, which is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI); The Diode Laser Therapy Systems utilize a semiconductor diode with invisible infrared radiation as a laser source to emit 808nm wavelength laser which is absorbed by melanin. The laser power is delivered to the treatment area via laser handpiece. The emission laser is activated by a foot-switch. The Diode Laser Therapy Systems utilize the principle of photoepilation for hair removal. The Photoepilation is a technique for removal of unwanted hair by thermal destruction of the hair follicle and its reproductive system (stems cells). The heat is caused by selective absorption of electromagnetic radiation emitted by laser light sources. As melanin is the main chromophore existing in hair follicles, so the melanin could absorb the energy from the laser, which would result in temperature rapid increase, then the hair follicle and its reproductive system (stems cells) would be destroyed by increased high temperature without damage epidermis and the surrounding normal tissue. - 6. Identification of Predicate Device(s) 510(k) Number: K180353 Product Name: Diode Laser Hair Removal Device Manufacturer: Zhengzhou PZ Laser Slim Technology Co., Ltd {5}------------------------------------------------ #### 7. Substantially Equivalent (SE) Comparison | Item | Proposed Device | Predicate Device | Remark | |------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------| | Device Name | Diode Laser Therapy Systems | Diode Laser Hair Removal / Device | | | Classification<br>Regulation | 21 CFR 878.4810 | 21 CFR 878.4810 | SAME | | Classification Panel | General & Plastic<br>Surgery | General & Plastic<br>Surgery | SAME | | Class | II | II | SAME | | Product Code | GEX | GEX | SAME | | Indication for use | The Diode Laser Therapy Systems is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. | The Diode Laser Hair Removal device is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. | SAME | ## Table 7-1 General Comparison Table 7-2 Performance Comparison | ITEM | Proposed Device | Predicate Device | Remark | |-------------------------|------------------------|--------------------------------|------------| | Laser Type | Diode Laser | Diode Laser | SAME | | Laser<br>Classification | Class IV | Class IV | SAME | | Laser<br>Wavelength | 808nm±2nm | 808nm | SAME | | Spot Size | 15mm×15mm | 1.44cm² | Analysis 1 | | Fluence | 1-100J/cm² | 1-100J/cm² | SAME | | Frequency | 1-10Hz | 1-20 Hz | Analysis 2 | | Pulse Duration | 1-300ms | 10~400ms | Analysis 3 | | Power Supply | 110 Va.c.60Hz<br>2000W | AC 110V-230V/50-60Hz<br>2000VA | Analysis 4 | | Dimension | 460mm×390mm×1230mm | 560mm×380mm×1180mm | Analysis 5 | | Weight | 75Kg | 60Kg | Analysis 5 | {6}------------------------------------------------ ### Analvsis : #### Analysis 1: #Spot size The proposed device is different in Spot Size from the predicate, Spot size only affects the area of treatment, not affect the therapeutic effect. Non-clinical tests about the spot size have been conducted on the proposed device and the results show that the proposed device meets the requirements. Therefore, this difference will not affect the substantially equivalency. #### Analysis 2: #Frequency The frequency for the proposed device is different from the predicate device. However, the frequency of the proposed device is within the range of that of the predicate device, which can justify that the difference in the parameter of frequency will not raise new safety issues of the proposed device. And the bench tests conducted on the proposed device is same with the predicate device, the results of which could support the substantially equivalency with predicate device. #### Analysis 3: #Pulse Duration The proposed device only has slight difference in pulse duration with the predicate device. The frequency of the proposed device is within the range of that of the predicate device, which can justify that the difference in the parameter of Pulse duration will not raise new safety issues of the proposed device. So the slight difference is considered to have no effectiveness and safety. And the bench tests conducted on the proposed device is same with the predicate device, the results of which could support the substantially equivalency with predicate device. #### Analysis 4: # Power Supply The Power Supply of proposed device is different with the predicate device, but the proposed device has been conducted as IEC 60601-1 and IEC 60601-1-2, the results shown that, there is no effect the effectiveness and safety. #### Analysis 5: #Dimension/ Weight The proposed device is different in dimension and weight from the predicate device. By complying with IEC 60601-1, the mechanical performance of the proposed device is determined to be accepted, therefore, this difference will not affect the substantially equivalency. | Item | Proposed Device | Predicate Device | Remark | |----------------------------------|-----------------|------------------|--------| | EMC, Electrical and Laser Safety | | | | Table 7-3 Safety Comparison {7}------------------------------------------------ | Electrical Safety | Comply with IEC 60601-1, IEC 60601-2-22 | Comply with IEC 60601-1, IEC 60601-2-22 | SAME | |------------------------------------------------|---------------------------------------------------------------------------------------------|-----------------------------------------|------------| | EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | SAME | | Laser Safety | Comply with IEC 60601-2-22, IEC 60825 | Comply with IEC 60601-2-22, IEC 60825 | SAME | | Patient Contact Materials and Biocompatibility | | | | | Patient Indirect Contact Materials | Tip of Handle (6061 Aluminum & Sapphire)<br>Plastic shell (Acrylonitrile Butadiene Styrene) | Sapphire | Analysis 6 | | Cytotoxicity | No Cytotoxicity | No Cytotoxicity | SAME | | Sensitization | No evidence of sensitization | No evidence of sensitization | SAME | | Irritation | No evidence of irritation | No evidence of irritation | SAME | ### Analysis 6 # Patient Indirect Contact Materials The patient contact material for the proposed device is different from the predicate device. However, biocompatibility test has been conducted on the proposed device and the test result can meet the requirements of ISO 10993 series standard. Therefore, this difference will not affect safety and effectiveness of the proposed device. #### Non-Clinical Test Conclusion 8. Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards: IEC 60601-1-2:2014 Medical Electrical Equipment-Part 1-2: General Requirements For Basic Safety And Essential Performance Collateral Standard: Electromagnetic Compatibility-Requirements And Tests IEC 60601- 1:2005+CORR.1:2006+CORR.2:2007+A1:2012 Medical Electrical Equipment - Part 1: General Requirements for Safety IEC 60601-2-22:2007 + A1:2012 +A2:2019 Medical Electrical Equipment - Part 2: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment. IEC 60825-1:2014 (Third Edition) Safety of laser products - Part 1: Equipment classification, and requirements ISO 10993-5: 2009 Biological Evaluation Of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity ISO 10993-10: 2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization. {8}------------------------------------------------ #### 9. Clinical Testing No clinical study is included in this submission. - Conclusion 10. The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(K) submission, the Diode Laser Therapy Systems (AlexMED Pro) is as safe, as effective, and performs as well as or better than the legally marketed predicate device, Diode Laser Hair Removal Device cleared under K180353.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...