K220103 · Shenzhen Leaflife Technology Co., Ltd. · GEX · Apr 12, 2022 · General, Plastic Surgery
Device Facts
Record ID
K220103
Device Name
LED Therapy Device (Planar LED/Planar LED mini)
Applicant
Shenzhen Leaflife Technology Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Apr 12, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LED Therapy Device is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
Device Story
Device uses diode LED (780-850 nm) to perform hair removal via selective photothermolysis; melanin in hair follicles absorbs near-infrared light, causing thermal damage to surrounding hair follicle tissues. Device consists of main unit, handpiece, and foot switch; operated by clinician in dermatological/surgical settings. Output is light energy delivered to treatment area; provider uses device to achieve permanent hair reduction. Benefits include stable, long-term reduction in hair regrowth.
Clinical Evidence
No clinical data; bench testing only. Performance verified via spot size and energy output accuracy testing. Biocompatibility testing (ISO 10993-5, ISO 10993-10) confirmed no cytotoxicity, sensitization, or irritation.
Technological Characteristics
Diode LED light source (780-850 nm); sapphire handpiece contact material; fluence 1-100 J/cm²; pulse duration 3-400 ms; frequency 1-10 Hz. Electrical safety per ANSI/AAMI ES60601-1; EMC per IEC 60601-1-2; photobiological safety per IEC 62471; non-laser light source safety per IEC 60601-2-57.
Indications for Use
Indicated for hair removal and permanent hair reduction in adults of all skin types (Fitzpatrick I-VI), including tanned skin.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
TRIA Laser Hair Removal System, Model TRIA (K120737)
IPL Salon Hair Reduction System, Model F60001 (K181568)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the letters "FDA" followed by "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font size.
April 12, 2022
Shenzhen Leaflife Technology Co., Ltd Cheng Qiang Regulatory registration supervisor 4F,Bldg. C, JMD Industrial Park, No.39 Qingfeng Blvd., Baolong Industrial Area, Longgang Dist., Shenzhen, Guangdong 518116 China
Re: K220103
Trade/Device Name: LED Therapy Device (Planar LED/Planar LED mini) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: January 6, 2022 Received: January 12, 2022
Dear Cheng Qiang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA
has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220103
Device Name
LED Therapy Device (Planar LED/ Planar LED mini)
Indications for Use (Describe)
The LED Therapy Device is intended for hair removal, permanent hair reduction on all skin type I-VI), including tanned skin.
Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
#### SUBMITTER l.
Shenzhen Leaflife Technology Co., Ltd 4F, Bldg. C, JMD Industrial Park, No.39 Qingfeng Blvd., Baolong Industrial Area, Longgang Dist., Shenzhen, RP China Phone: 086-(0)17875910506 Fax: 086-0755-27215592
| Primary Contact Person: | Cheng Qiang |
|-------------------------|---------------------------------------|
| | Regulatory Affairs Manager |
| | Shenzhen Leaflife Technology Co., Ltd |
| Tel: | 086-0755-27216609 |
| Fax: | 086-0755-27215592 |
| Phone: | 086-(0)17875910506 |
| Email: | cq@leaflife.cn |
Date Prepared:
01/06/2022
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for LEAFLIFE. The logo is in blue and green, with the word "LEAFLIFE" in blue, with a green line above the "F" and "E". Above the word is a green arc with a leaf at the top. Below the word is the Chinese translation of the word in gray.
#### II. PROPOSED DEVICE
| Trade Name: | LED Therapy Device |
|----------------------|-----------------------------------------------------|
| Common Name: | Powered Laser Surgical Instrument |
| Classification Name: | Powered Laser Surgical Instrument (21 CFR 878.4810) |
| Regulation Class: | II |
| Product Code: | OHT |
| Review Panel: | General & Plastic Surgery |
# III. PREDICATE DEVICE
| Predicate device (Primary) | Laser Therapy Device, Model LM-L808A: K202980 |
|----------------------------|----------------------------------------------------------|
| Predicate device | TRIA Laser Hair Removal System, Model TRIA: K120737 |
| Predicate device | IPL Salon Hair Reduction System, Model F60001<br>K181568 |
The predicates have not been subject to a design-related recall.
# IV. DEVICE DESCRIPTION
The LED Therapy Device is a surgical device intended for use in dermatologic and general surgical procedure. It utilizes a diode LED as a light source (780-850 nm). The LED light power is delivered to the treatment area via a LED hand piece. The emission LED is activated by a foot switch and a hand piece.
The proposed LED Therapy Device is equipped with a light depilation function that is based on the theory of selective photo thermolysis. There is abundant melanin in the hair follicle and hair stem. The melanin is distributed amidst the cells between the hair ball substrate and it can be transferred to the structure of hair stems (such as medulla, cortex, and hair cuticle). The Near infrared light can precisely target the melanin and apply the depilation treatment selectively. After the melanin absorbs Near infrared light energy, the temperature rises, in this way, the surrounding hair follicle tissues are damaged, and the hairs are removed.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for LEAFLIFE. The logo has the word "LEAFLIFE" in blue, with a green arc above the word. There is a green leaf on the right side of the arc. Below the word "LEAFLIFE" are two Chinese characters in gray.
### V. INTENDED USE
The LED Therapy Device is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin.
Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
# vi. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The LED Therapy Device (LM-LNIRA, LM-LNIRB) is substantially equivalent to the cleared predicate device.
The LED Therapy Device (LM-LNIRA, LM-LNIRB) has the same intended use, and similar technological characteristics such as fluence, wavelength, spot size, and pulse duration, etc. with the predicates. Please refer to the following table for details:
{6}------------------------------------------------
Image /page/6/Figure/0 description: The image shows the logo for LEAFLIFE. The logo is a combination of text and a graphic element. The text "LEAFLIFE" is written in a stylized font, with the top half of the letters in green and the bottom half in blue. Above the text is a green leaf, and to the right of the leaf are some green bubbles. Below the text are some Chinese characters.
| Item | Proposed Device | Predicate Device<br>K202980 (Primary) | Predicate Device<br>K120737 | Predicate Device<br>K181568 | Remark |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|
| Device name | LED Therapy Device<br>(Planar LED/ Planar LED mini) | Laser Therapy Device | TRIA Laser Hair Removal<br>System (TRIA) | IPL Salon Hair Reduction<br>System | / |
| Product model | LM-LNIRA, LM-LNIRB | LM-L808A | TRIA | F60001 | / |
| K number | On pending | K202980 | K120737 | K181568 | / |
| Product code | OHT | GEX | GEX | OHT | Same |
| Classification<br>regulation | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | Same |
| Item | Proposed Device | Predicate Device<br>K202980 (Primary) | Predicate Device<br>K120737 | Predicate Device<br>K181568 | Remark |
| Intended Use | The LED Therapy Device is<br>intended for hair removal,<br>permanent hair reduction on all<br>skin types (Fitzpatrick skin type<br>I-VI), including tanned skin.<br>Permanent hair reduction is<br>defined as the long-term, stable<br>reduction in the number of<br>hairs regrowing when measured<br>at 6, 9, and 12 months after the<br>completion of a treatment<br>regime. | The Laser Therapy Device is<br>intended for hair removal,<br>permanent hair reduction<br>on all skin types (Fitzpatrick<br>skin type I-VI), including<br>tanned skin.<br>Permanent hair reduction is<br>defined as the long-term,<br>stable reduction in the<br>number of hairs regrowing<br>when measured at 6, 9, and<br>12 months after the<br>completion of a treatment<br>regime. | TRIA is an<br>over-the-counter device<br>intended for adjunctive<br>use with shaving for hair<br>removal sustained with<br>periodic treatments. TRIA<br>is also intended for<br>permanent reduction in<br>hair regrowth defined as a<br>long-term, stable<br>reduction in hair counts<br>following a treatment<br>regime. Permanent hair<br>reduction is defined as<br>long term, stable<br>reduction in the number<br>of hairs regrowing when<br>measured out to 6, 9, and<br>12 months after the<br>completion of the<br>treatment regimen. | The IPL Salon Hair<br>Reduction System (Model:<br>F60001) is an over the<br>Counter device intended<br>for the removal of<br>unwanted body and/or<br>facial hair in adults. It is<br>also intended for<br>permanent reduction in<br>unwanted hair.<br>Permanent hair reduction<br>is defined as the long-term<br>stable reduction in the<br>number of hairs<br>re-growing when<br>measured at 6. 9. and 12<br>months after the<br>completion of a treatment<br>regimen. | Same |
| Prescription<br>use or not | Prescription use | Prescription use | Over-The-Counter Use | Over-The-Counter Use | Same |
| Configuration | Main Unit | Main Unit | Main Unit | Main Unit | Same |
| | Hand piece | Handpiece | / | Finger switch | Same |
| Item | Proposed Device | Predicate Device<br>K202980 (Primary) | Predicate Device<br>K120737 | Predicate Device<br>K181568 | Remark |
| | Foot Control | Foot Control | / | / | Same |
| Laser Type | Diode LED | Diode laser | Diode laser | Xenon Arc Flashlamp | Discussion<br>1 |
| Laser<br>Classification | Risk Group 3 | Class IV | Class IV | Exempt group | Discussion<br>1 |
| Laser<br>Wavelength | 780-850 nm | 808 nm | 800nm +/-20%<br>(640nm-960nm) | 475nm~1200nm | Discussion<br>1 |
| Spot size | (17mm*22mm)<br>(LM-LNIRA)<br>(17mm*17mm)<br>(LM-LNIRB) | (12mm*12mm) 1.44cm² | $Φ$ 10mm (Round) | Treatment Area (regular<br>window): 3.025 [cm²]<br>Treatment area (facial<br>adapter): 1.72 [cm²] | Discussion<br>2 |
| Fluence | 1-100J/cm² | 1-100J/cm² | 6J/cm², 10J/cm², 14J/cm²,<br>18J/cm², 22J/cm² | Level 1: 8.62J<br>Level 2: 9.45J<br>Level 3: 10.64J<br>Level 4: 11.48J<br>Level 5: 12.70J | Same |
| Frequency | 1-10Hz | 1-10Hz | / | / | Same |
| Pulse Duration | 3-400ms | 5-400ms | / | 11-12ms | Similar |
| Item | Proposed Device | Predicate Device<br>K202980 (Primary) | Predicate Device<br>K120737 | Predicate Device<br>K181568 | Remark |
| Power Supply | AC 220~240V 50/60 Hz<br>or<br>AC 100~120V 50/60 Hz | AC 220~240V 50/60 Hz<br>or<br>AC 100~120V 50/60 Hz | 100~240V,50/60Hz | 100-240 VAC, 50/60Hz | Discussion<br>3 |
| Dimension | 430mm x 500mm x 870mm<br>(LM-LNIRA)<br>408mm x 460mm x 593mm<br>(LM-LNIRB) | 650mm x 430mm x 315mm | / | 143*69.5*43mm(H*W*D) | Discussion<br>3 |
| Weight | 30kg (LM-LNIRA)<br>24kg (LM-LNIRB) | 22kg | / | 650g | Discussion<br>3 |
{7}------------------------------------------------
Image /page/7/Figure/0 description: The image shows the logo for LEAFLIFE. The logo is in blue and green, with a stylized leaf above the "L". The text "LEAFLIFE" is in blue, with a green line running through the middle of each letter. Below the text is a blue curved line. There is also some text in another language below the English text.
{8}------------------------------------------------
Image /page/8/Figure/0 description: The image shows the logo for LEAFLIFE. The logo features the word "LEAFLIFE" in blue and green, with a green leaf above the "F". There are also some Chinese characters below the word "LEAFLIFE".
{9}------------------------------------------------
Image /page/9/Figure/1 description: The image shows the logo for LEAFLIFE. The logo is blue and green, and it features a stylized leaf and the word "LEAFLIFE" in a modern font. There is also some text in another language below the word LEAFLIFE. The logo is clean and simple, and it conveys a sense of nature and life.
#### Discussion 1
The proposed device is different in Laser Classification, Laser Wavelength, and Laser Safety test standard from the predicate device. Mainly because of these differences caused by the inconsistency of Laser Type, this difference can be proved by non-clinical test There is no security impact. The difference that mainly affects performance may exist in the difference in Laser Wavelength can be verified by the Predicate device (K120737), Predicate device (K181568) and the wavelength range of light absorbed by melanin to prove that there is no effect on performance. Therefore, this difference will not affect the substantial equivalency.
#### Discussion 2
The proposed device is different in Spot Size only affects the area of treatment, not affecting the therapeutic effect, Therefore, this difference will not affect the substantial equivalency.
#### Discussion 3
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for LEAFLIFE. The logo is a combination of text and a graphic. The text "LEAFLIFE" is written in a stylized font, with the top half of the letters in green and the bottom half in blue. Above the text is a green leaf graphic. Below the text is the Chinese translation of the name.
The proposed device is different in dimension and weight from the predicate device. By complying with ANS/AAMI ES60601-1, the Electrical Safety of the proposed device is determined to be accepted, this difference will not affect the substantial equivalency.
| Item | Proposed Device | Predicate Device<br>K202980 (Primary) | Predicate Device<br>K120737 | Predicate Device<br>K181568 | Remark |
|------------------------------|------------------------------------------|------------------------------------------|------------------------------------------|------------------------------------------|---------|
| Patient Contact<br>Materials | Sapphire in handpiece | Sapphire in handpiece | / | / | Same |
| Cytotoxicity | No Cytotoxicity | No Cytotoxicity | / | / | Same |
| Sensitization | No evidence of sensitization | No evidence of<br>sensitization | / | / | Same |
| Irritation | No evidence of irritation | No evidence of irritation | / | / | Same |
| Electrical Safety | Comply with ANSI/AAMI<br>ES60601-1 | Comply with ANSI/AAMI<br>ES60601-1 | Comply with IEC 60601-1. | Comply with IEC<br>60601-1. | Same |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC<br>60601-1-2 | Comply with IEC<br>60601-1-2 | Comply with IEC<br>60601-1-2 | Same |
| Laser Safety | Comply with IEC<br>60601-2-57, IEC 62471 | Comply with IEC<br>60601-2-22, IEC 60825 | Comply with IEC<br>60601-2-22, IEC 60825 | Comply with IEC<br>60601-2-57, IEC 62471 | Similar |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for LEAFLIFE. The logo is composed of the word "LEAFLIFE" in blue and green, with a green leaf above the "L". There is also some text in another language below the word "LEAFLIFE". The logo is simple and modern, and it conveys a sense of nature and health.
## VII. Non-Clinical Test Conclusion
Non-clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
ES60601-1:2005/(R) 2012 and A1:2012, C1:2009/(R)2012 and ANSI/AAMI A2:2010/(R)2012, Medical electrical equipment — Part 1: General requirements for basic safety and essential performance
IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
IEC 60601-2-57:2011, Medical Electrical Equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use.
IEC 62471:2006, Photobiological safety of lamps and lamp systems.
ISO 10993-5 Third Edition 2009-06-01, Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity. (Biocompatibility)
ISO 10993-10 Third Edition 2010-08-01, Biological Evaluation of Medical Devices -Part 10: Tests for Irritation and Skin Sensitization. (Biocompatibility)
Performance Testing for Spot Size Accuracy and Energy Output Accuracy.
VIII. CLINICAL TEST CONCLUSION
No clinical study is included in this submission.
# IX. SUBSTANTIALLY EQUIVALENT (SE) CONCLUSION
Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.