K210634 · Asclepion Laser Technologies GmbH · GEX · Apr 13, 2022 · General, Plastic Surgery
Device Facts
Record ID
K210634
Device Name
MCL 31 Dermablate System
Applicant
Asclepion Laser Technologies GmbH
Product Code
GEX · General, Plastic Surgery
Decision Date
Apr 13, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MCL 31 Dermablate System with its accessories is indicated for coagulation, vaporization, ablation, or cutting of soft tissue (skin) in Dermatology, Plastic Surgery, Oral Surgery, ENT, Gynecology, General Surgery, Podiatry, and Ophthalmology (skin around the eyes). The MCL 31 Dermablate System, when used with its micro beam handpieces, is intended for use in Dermatological procedures and Skin resurfacing procedures.
Device Story
MCL 31 Dermablate System is a pulsed Er:YAG laser (2940 nm) for soft tissue surgery. System includes main console, detachable handpieces (VarioTEAM, MicroSpot, MY JULIET), and footswitch. Operates via articulated arm delivery; includes suction unit for laser plume removal. Used by physicians in clinical settings (Dermatology, Plastic Surgery, etc.) for coagulation, vaporization, ablation, or cutting. Physician controls laser emission via footswitch; output parameters (fluence, pulse duration, repetition rate) are adjusted based on procedure. Device enables precise tissue interaction for resurfacing and surgical applications; benefits include controlled tissue removal with minimal thermal damage.
Clinical Evidence
Bench testing only. Compliance with EN 60601-1, IEC 60601-1-2, IEC 60601-2-22, and EN/IEC 60825-1. Biocompatibility testing for MY JULIET handpiece per ISO 10993-5, ISO 10993-10, and ISO 10993-23 (cytotoxicity, irritation, sensitization). Software verification and validation performed per IEC 62304.
Technological Characteristics
Pulsed Er:YAG laser, 2940 nm wavelength. Articulated arm delivery. Max fluence 250 J/cm² (VarioTEAM). Pulse duration 100–1000 μs. Repetition rate up to 20 Hz. Biocompatible materials per ISO 10993. Software lifecycle per IEC 62304. System includes suction unit for plume removal.
Indications for Use
Indicated for coagulation, vaporization, ablation, or cutting of soft tissue (skin) in patients undergoing procedures in Dermatology, Plastic Surgery, Oral Surgery, ENT, Gynecology, General Surgery, Podiatry, and Ophthalmology (skin around the eyes). Micro beam handpieces are indicated for dermatological and skin resurfacing procedures.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 13, 2022
Asclepion Laser Technologies GmbH Dania Paolo Regulatory Affairs Manager Bruesseler Strasse 10 Jena, Thuringia 07747 Germany
Re: K210634
Trade/Device Name: MCL 31 Dermablate System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: February 25, 2021 Received: March 3, 2021
Dear Dania Paolo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya, D.Eng. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210634
Device Name MCL 31 Dermablate System
#### Indications for Use (Describe)
The MCL 31 Dermablate System with its accessories is indicated for coagulation, vaporization, or cutting of soft tissue (skin) in Dermatology, Plastic Surgery, Oral Surgery, ENT, Gynecology, General Surgery, Podiatry, and Ophthalmology (skin around the eves).
The MCL 31 Dermablate System, when used with its micro beam handpieces, is intended for use in Dermatological procedures and Skin resurfacing procedures.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Asclepion Laser Technologies. The logo features the word "Asclepion" in a gray, sans-serif font, with the words "Laser Technologies" in a smaller font size underneath. To the right of the text is a cluster of blue circles of varying sizes, arranged in a scattered pattern.
| Submitter: | ASCLEPION LASER TECHNOLOGIES GmbH<br>Bruesseler Str. 10<br>07747 Jena, Germany |
|----------------------|---------------------------------------------------------------------------------|
| Contact Person: | Tom Gruender<br>Regulatory Affairs Manager<br>ASCLEPION LASER TECHNOLOGIES GmbH |
| Phone: | +49 3641 77 00 201 |
| Fax: | +49 3641 77 00 302 |
| e-mail: | tom.gruender@asclepion.com |
| Preparation Date: | April 12, 2022 |
| Device Trade Name: | MCL 31 Dermablate System |
| Common Name: | Er:YAG laser |
| Classification Name: | Powered Laser Surgical Instrument |
| Regulation Number: | 21 CFR 878.4810 |
| Regulatory Class: | Class II |
| Product Code: | GEX |
| Submission Type: | Traditional 510(K) |
#### Legally Marketed Predicate Devices:
- . MCL 31 Dermablate (K150140) - Asclepion Laser Technologies
- . MicroSpot (K152153) - Asclepion Laser Technologies
#### Device Description:
The MCL 31 Dermablate System is a pulsed Er:YAG laser emitting a wavelength of 2940 nm. The system comprises a main console unit, detachable handpieces, and a footswitch. The MCL 31 Dermablate System can be operated with micro beam and non-micro beam handpieces. The system incorporates a suction unit for the safe removal of laser plume. The laser is fired with a foot-operated switch (footswitch).
#### Intended Use/Indications for Use:
The MCL 31 Dermablate System with its accessories is indicated for coagulation, vaporization, ablation, or cutting of soft tissue (skin) in Dermatology, Plastic Surgery, ENT, Gynecology, General Surgery, Podiatry, and Ophthalmology (skin around the eyes).
The MCL 31 Dermablate System, when used with its micro beam handpieces, is intended for use in Dermatological procedures and Skin resurfacing procedures.
#### Summary of Technical Characteristics:
The MCL 31 Dermablate System that is the subject of this submission has identical technical characteristics as the previously cleared MCL 31 Dermablate System (K150140) and its MicroSpot handpiece (K152153).
The MY JULIET handpiece has similar technical characteristics, design, and functional features as the MicroSpot handpiece (K152153) for the MCL 31 Dermablate System, except for the differences shown
{4}------------------------------------------------
# K210634 510(k) Summary
Image /page/4/Picture/1 description: The image shows the logo for Asclepion Laser Technologies. The logo features the word "Asclepion" in a gray, sans-serif font, with the words "Laser Technologies" in a smaller font below it. Above and to the right of the word "Asclepion" is a cluster of blue dots of varying sizes, arranged in a circular pattern.
in the table below. The minor differences in technical characteristics do not raise any new or different questions of safety or effectiveness.
## Comparison with Predicate Devices:
The MCL 31 Dermablate System is substantially equivalent to the predicates with the same intended use and similar indications for use and technological characteristics including wavelength, spot size, and energy delivery.
| Specification | This Application | Predicate Devices | |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|
| Trade/Device Name | MCL 31 Dermablate | MCL 31 Dermablate | MicroSpot Handpiece |
| Submitter | Asclepion Laser<br>Technologies GmbH | Asclepion Laser<br>Technologies GmbH | Asclepion Laser<br>Technologies GmbH |
| 510(k) number | K210634 | K150140 | K152153 |
| Indications for use | The MCL 31 Dermablate System<br>with its accessories is indicated for<br>coagulation, vaporization,<br>ablation, or cutting of soft tissue<br>(skin) in Dermatology, Plastic<br>Surgery, Oral Surgery, ENT,<br>Gynecology, General Surgery,<br>Podiatry, and Ophthalmology (skin<br>around the eyes).<br>The MCL 31 Dermablate System,<br>when used with its micro beam<br>handpieces, is intended for use in<br>Dermatological procedures and<br>Skin resurfacing procedures. | Coagulation,<br>vaporization, ablation<br>or cutting of soft tissue<br>(skin) in Dermatology,<br>Plastic Surgery, Oral<br>Surgery, ENT,<br>Gynecology, General<br>Surgery, Podiatry and<br>Ophthalmology (skin<br>around the eyes) | The MicroSpot Handpiece<br>is intended for<br>Dermatological<br>procedures and Skin<br>resurfacing procedures. |
| Laser medium | Er:YAG | Er:YAG | Er:YAG |
| Wavelength | 2940 nm | 2940 nm | 2940 nm |
| Fluence, max. | VarioTEAM Handpiece:<br>250 J/cm²<br>MicroSpot Handpiece:<br>10 J/cm² (Up to 150 J/cm²<br>with stacking pulses)<br>MY JULIET Handpiece:<br>5 J/cm² (Up to 35 J/cm²<br>with stacking pulses) | VarioTEAM Handpiece:<br>250 J/cm² | MicroSpot Handpiece:<br>10 J/cm² (Up to 150 J/cm²<br>with stacking pulses) |
| Energy, max | 2.5 J<br>12 mJ/microbeam<br>(for MicroSpot Handpiece)<br>7 mJ/microbeam (for My Juliet ) | 2.5 J | 12 mJ/microbeam |
| Pulse Duration | 100 – 1000 μs | 100 – 1000 μs | 100 – 1000 μs |
| Repetition rate, max | 20 Hz | 20 Hz | 20 Hz |
| Treatment area / Spot size | VarioTEAM Handpiece:<br>1 – 12 mm<br>MicroSpot Handpiece:<br>350 μm - 600 µm spot,<br>MY JULIET Handpiece:<br>425 μm & 600 µm spot, | VarioTEAM Handpiece:<br>1 – 12 mm | MicroSpot Handpiece:<br>350 μm - 600 µm spot, |
| WLAN module | No | Yes | Yes |
| Delivery system | Articulated Arm | Articulated Arm | Articulated Arm |
{5}------------------------------------------------
# K210634 510(k) Summary
Image /page/5/Picture/1 description: The image shows the logo for Asclepion Laser Technologies. The logo features the word "Asclepion" in gray, with the words "Laser Technologies" in a smaller font size underneath. To the right of the text is a cluster of blue circles of varying sizes, arranged in a pattern that resembles a constellation or a stylized representation of laser beams.
## Nonclinical Performance Data:
The following performance data were applied in support of the substantial equivalence determination:
- . EN 60601-1: Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- . IEC 60601-1-2: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests
- IEC 60601-2-22: Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
- EN/IEC 60825-1: Safety of laser products - Part 1: Equipment classification, and requirements
- IEC 62304: Medical Device Software - Software life cycle processes
- ISO 14971: Medical devices - Applications of risk management to medical devices
Biocompatibility testing was also conducted for the MY JULIET handpiece in accordance with:
- EN/ISO 10993-1: Biological evaluation of medical devices Part 1: Evaluation and testing ● within a risk management process
- ISO 10993-5: Biological evaluation of medical devices - Part 5: Tests for in vitro cytoxicity
- ISO 10993-10: Biological evaluation of medical devices - Part 10 Tests for irritation and skin sensitization
- . ISO 10993-23: Biological evaluation of medical devices - Part 23 Tests for irritation.
Software verification and validation testing was conducted, and documentation is provided in this submission, as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
Biocompatibility: The biocompatibility of MCL 31 Dermablate System and compatible handpieces with the extended indications is established based on the legally marketed predicate devices. For the MY JULIET handpiece testing was conducted regarding Cytotoxicity, Sensitivity and Irritation.
The MCL 31 Dermablate System passed all required testing and is in compliance with all applicable sections of the above-mentioned performance standards.
#### Conclusion:
The MCL 31 Dermablate System has the same intended use and similar indications for use, technological characteristics, and principles of operation as the predicate devices. The minor differences in indications for use and technical characteristics do not raise different questions of safety or effectiveness. Thus, the MCL 31 Dermablate System is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.