K210535 · Rohrer Aesthetics, LLC · GEX · May 24, 2021 · General, Plastic Surgery
Device Facts
Record ID
K210535
Device Name
UltraLight LED System
Applicant
Rohrer Aesthetics, LLC
Product Code
GEX · General, Plastic Surgery
Decision Date
May 24, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
LED red light: The UltraLight" is indicated for use as a non-invasive dermatological aesthetic treatment for the reduction of circumference of hips, waist, and thighs. The UltraLight" is also indicated for use in dermatology for the treatment of superficial, benign vascular, and pigmented lesions. LED blue light: The UltraLight" is indicated to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris. LED green light: The UltraLight "green Light" is indicated for use as a non-invasive dermatological aestheic treatment for the reduction of circumference of hips, waist, and thighs.
Device Story
UltraLight LED System is a low-level light therapy device for aesthetic dermatology. Device emits three wavelengths: red (633 nm), blue (415 nm), and green (532 nm). System utilizes a 28.8 cm x 38.0 cm irradiation board to deliver light directly to target areas. Operated by clinician via LCD control panel. Red and green light modes target circumference reduction; red light also treats vascular/pigmented lesions; blue light treats moderate inflammatory acne vulgaris. Forced air ventilation provides cooling. Device is intended for professional use in clinical settings. Output is light energy; clinical decision-making relies on provider assessment of patient skin/aesthetic needs. Benefits include non-invasive treatment options for specified dermatological conditions.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing and comparison of technical specifications to predicate devices. Bench testing included IEC 60601-1 (general safety/performance) and IEC 60601-1-2 (electromagnetic compatibility).
Technological Characteristics
Low-level LED system; wavelengths: 633nm (red), 415nm (blue), 532nm (green). Irradiation board: 28.8cm x 38.0cm. Power: 100-120VAC, 50/60Hz. Cooling: Forced air ventilation. LCD control panel interface. Class II device.
Indications for Use
Indicated for patients seeking non-invasive dermatological aesthetic treatment for circumference reduction of hips, waist, and thighs; treatment of superficial, benign vascular, and pigmented lesions; and treatment of moderate inflammatory acne vulgaris.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.
May 24, 2021
Rohrer Aesthetics, LLC Mark Rohrer President 105 Citation Court Birmingham, Alabama 35209
Re: K210535
Trade/Device Name: UltraLight LED System Regulation Number: 21 CFR 878.5400 Regulation Name: Low Level Laser System For Aesthetic Use Regulatory Class: Class II Product Code: OLI Dated: April 29, 2021 Received: May 3, 2021
Dear Mark Rohrer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210535
Device Name
UltraLight LED System
Indications for Use (Describe)
LED red light: The UltraLight" is indicated for use as a non-invasive dermatological aesthetic treatment for the reduction of circumference of hips, waist, and thighs. The UltraLight" is also indicated for use in dermatology for the treatment of superficial, benign vascular, and pigmented lesions.
LED blue light: The UltraLight" is indicated to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris.
LED green light: The UltraLight "green Light" is indicated for use as a non-invasive dermatological aestheic treatment for the reduction of circumference of hips, waist, and thighs.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## K210535
## 510(K) Summary
This 510(K) Summary for the UltraLight LED System is prepared in accordance with the requirements of 21 CFR 807.92.
| Applicant: | Rohrer Aesthetics, LLC |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | Rohrer Aesthetics, LLC<br>105 Citation Court<br>Birmingham, AL 35209 |
| Contact Person: | Mr. Mark Rohrer |
| Telephone: | 205-356-1172 – phone<br>mrohrer@rohreraesthetics.com |
| Preparation Date: | May 21, 2021 |
| Device Trade Name: | UltraLight LED System |
| Common Name: | Low Level Laser System for Aesthetic Use |
| Regulation Name: | Low Level Laser System for Aesthetic Use |
| Regulation Number: | 21 CFR 878.5400 |
| Product Code: | OLI |
| Predicate Devices: | K180338 (Cellulize), K160880 (Photonica Professional), and<br>K030883 (Omnilux Blue) |
| Regulatory Class: | Class II |
| Device Description: | The UltraLight LED System is a low level laser system for<br>aesthetic use. The device can emit three different<br>wavelengths (i.e., red light @ 633 nm, blue light @ 415 nm,<br>and green light @ 532 nm) to achieve its intended purpose.<br>All three wavelengths have different indications for use. The<br>device works by illuminating the skin with a single<br>wavelength of red, blue, or green light. The device utilizes a<br>28.8cm X 38.0 cm irradiation board to deliver light directly to<br>the desired target. The device is activated and controlled<br>through the LCD control panel. |
| Indications for Use: | LED red light: The UltraLight "red light" is indicated for use<br>as a non-invasive dermatological aesthetic treatment for the<br>reduction of circumference of hips, waist, and thighs. The<br>UltraLight "red light" is also indicated for use in dermatology<br>for the treatment of superficial, benign vascular, and<br>pigmented lesions.<br><br>LED blue light: The UltraLight "blue light" is indicated to treat<br>dermatological conditions and specifically indicated to treat |
moderate inflammatory acne vulgaris.
{4}------------------------------------------------
LED green light: The UltraLight "green Light" is indicated for
use as a non-invasive dermatological aesthetic treatment for
the reduction of circumference of hips, waist, an
Comparison of Technical Specifications:
| Characteristic - Red Light Mode | UltraLight LED System | Photonica Professional |
|----------------------------------|-------------------------------|-----------------------------|
| | (510K Pending) | (K160880) |
| Wavelength | 633nm±6nm | 635±2nm |
| Bandwidth | 10nm | 10nm |
| Output Intensity/Irradiance | 105mW/cm² | 105mW/cm² |
| Recommended Treatment Time | 8 or 20 min | 8 or 20 min |
| Standard Energy | 126 J/cm² | 126 J/cm² |
| Console Dimension | 35cm×71cm×122cm | 167cm×62cm×61cm |
| Weight | 54kg | 52kg |
| Power Source | 100 - 120vac, 3 amps, 50/60Hz | 110vac, 3 amps, 60Hz |
| Operating Temperature | +5C - +35C | +5C - +35C |
| Cooling Mechanism | Forced Air Ventilation | Forced Air Ventilation |
| Characteristic – Blue Light Mode | UltraLight LED System | Ominlux Blue |
| | (510K Pending) | (K030883) |
| Wavelength | 415nm±5nm | 415±5nm |
| Bandwidth | 22nm | 22nm±3nm |
| Output Intensity/Irradiance | 40mW/cm² | 33.33 - 47.62 mW/cm² |
| Recommended Treatment Time | 20 min | 20 min |
| Max. Energy | 48 J/cm² | 40 J/cm² |
| Console Dimension | 35cm×71cm×122cm | 37cm×18cm×49cm |
| Weight | 54kg | 12kg |
| Power Source | 100 - 120vac, 3 amps, 50/60Hz | 100-120vac, 3 amps, 50/60Hz |
| Operating Temperature | +5C - +35C | +5C - +35C |
| Cooling Mechanism | Forced Air Ventilation | Forced Air Ventilation |
| Characteristic - Green Light | UltraLight LED System | Cellulize |
| Mode | (510K Pending) | (K180338) |
| Wavelength | 532nm±3nm | 532±3nm |
| Bandwidth | 10nm | 10nm |
| Output Intensity/Irradiance | 95.14 mW/cm² | 95.14 - 105 mW/cm² |
| Recommended Treatment Time | 8 or 20 min | 8 or 20 min |
| Max. Energy | 45.7 J/cm² in 8 min | 50 J/cm² in 8 min |
| | 114 J/cm²in 20 min | 126 J/cm² in 20 min |
| Console Dimension | 35cm×71cm×122cm | 183.2cm×62.2cm×61cm |
| Weight | 54kg | 52kg |
| Power Source | 100 - 120vac, 3 amps, 50/60Hz | 100-120vac, 3 amps, 50/60Hz |
| Operating Temperature | +5C - +35C | +5C - +35C |
| Cooling Mechanism | Forced Air Ventilation | Forced Air Ventilation |
{5}------------------------------------------------
#### Comparison of Indications for Use:
| | UltraLight LED System<br>(510K Pending) | Photonica Professional (K160880) |
|----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Red Light Mode | For use as a non-invasive<br>dermatological aesthetic treatment for<br>the reduction of circumference of hips,<br>waist, and thighs. The UltraLight "red<br>light" is also indicated for use in<br>dermatology for the treatment of<br>superficial, benign vascular, and<br>pigmented lesions. | For use as a non-invasive<br>dermatological aesthetic treatment for<br>the reduction of circumference of<br>hips, waist, and thighs. Photonica<br>Professional is also indicated for use<br>in dermatology for the treatment of<br>superficial, benign vascular, and<br>pigmented lesions. |
| | UltraLight LED System<br>(510K Pending) | Ominlux Blue<br>(K030883) |
|-----------------|-----------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Blue Light Mode | Is indicated to treat dermatological<br>conditions and specifically indicated<br>to treat moderate inflammatory acne<br>vulgaris. | Is indicated to treat dermatological<br>conditions and specifically indicated to<br>treat moderate inflammatory acne<br>vulgaris. |
| | UltraLight LED System<br>(510K Pending) | Cellulize<br>(K180338) |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Green Light<br>Mode | Is indicated for use as a non-invasive<br>dermatological aesthetic treatment for<br>the reduction of circumference of<br>hips, waist, and thighs. | Is indicated for use as a non-invasive<br>dermatological aesthetic treatment for<br>the reduction of circumference of<br>hips, waist, and thighs. |
Performance Data: The following performance data were provided in support of (Non-Clinical Testing) the substantial equivalence determination: IEC 60601-1 Test for Medical Electrical equipment was performed for General Requirements for basic safety and essential performance IEC 60601-1-2 Test for Medical Equipment for General Requirements for basic safety and essential performance: electromagnetic compatibility Bench Performance Data to support technical specifications of the device. Results of Clinical Study: A human clinical study was not required for this device since the technical specifications of the device are comparable to the technical specifications of the predicates. Conclusion: The UltraLight LED System is substantially equivalent to the Photonic Professional (K160880), the Omnilux Blue (K030883) and the Cellulize (K180338).
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.