K203799 · Innovaquartz, LLC · GEX · Sep 20, 2021 · General, Plastic Surgery
Device Facts
Record ID
K203799
Device Name
ProFlex CO2 Laser Fiber
Applicant
Innovaquartz, LLC
Product Code
GEX · General, Plastic Surgery
Decision Date
Sep 20, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
ProFlex CO2 Laser Fiber is indicated for the ablation, coagulation, incision, and vaporization of soft tissue in open, endoscopic, and laparoscopic surgical procedures.
Device Story
ProFlex CO2 Laser Fiber is a fiber optic energy delivery device; transmits CO2 laser energy (10600 nm) from a laser source to target tissue. Device consists of stainless steel SMA 905 connector, hollow glass tube with silver/silver iodide dielectric coating, fluoropolymer cladding, and ETFE protective jacket. Used in surgical settings (open, endoscopic, laparoscopic) by surgeons to perform soft tissue ablation, coagulation, incision, and vaporization. Fiber is supplied in a coiled HDPE carrier within sterile packaging. Healthcare providers connect the fiber to a compatible CO2 laser system; energy delivery is controlled via the laser console. Output is laser energy at the distal tip; enables precise tissue interaction. Benefits include controlled soft tissue management during surgery.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including design verification, packaging validation, and sterilization validation per ISO 11135:2014 and ISO 10993-7:2008.
Technological Characteristics
Hollow glass fiber with silver/silver iodide dielectric coating; fluoropolymer cladding; ETFE buffer; stainless steel SMA 905 connector. Max power: 40W (CW), 15W (super-pulse). Biocompatible (USP Class VI). Sterilization: Ethylene Oxide (EtO).
Indications for Use
Indicated for ablation, coagulation, incision, and vaporization of soft tissue in open, endoscopic, and laparoscopic surgical procedures. Prescription use only.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 20, 2021
InnovaQuartz LLC Stephen Griffin Chief Technology Officer 23030 N. 15th Ave Phoenix, Arizona 85027
Re: K203799
Trade/Device Name: ProFlex CO2 Laser Fiber (Model: S-COF500) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: August 18, 2021 Received: August 20, 2021
Dear Stephen Griffin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Purva U. Pandya -S
Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K203799
Device Name ProFlex CO2 Laser Fiber (Model: S-COF500)
Indications for Use (Describe)
ProFlex CO2 Laser Fiber is indicated for the ablation, coagulation, incision, and vaporization of soft tissue in open, endoscopic, and laparoscopic surgical procedures.
Type of Use (Select one or both, as applicable)
| <span style="font-size:10px">☒</span> Prescription Use (Part 21 CFR 201 Subpart D) | <span style="font-size:10px">☐</span> Over-The-Counter Use (21 CFR 201 Subpart C) |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of 21CFR Section 807.92.
The assigned 510(k) Number: K203799
| 1. Date of Preparation | 09/17/2021 |
|------------------------|------------|
|------------------------|------------|
| 2. Applicant | |
|--------------|------------------------------------------|
| Name: | InnovaQuartz, LLC. |
| Address: | 23030 N 15th Ave, Phoenix, AZ 85027-1315 |
| Phone: | 623-434-1895 |
| Registration #: | 3010933841 |
|-----------------|------------|
|-----------------|------------|
| Contact Person: | Stephen Griffin, CTO |
|-----------------|-------------------------------------------------|
| Telephone: | 623-434-1895 (main) x101, 623-229-5174 (mobile) |
| Email: | steveg@innovaquartz.com |
- 3. Identification of the Proposed Device
| Trade Name: | ProFlex CO2 Laser Fiber |
|----------------------|-------------------------------------------------------------------------------------|
| Common Name: | Laser Fiber |
| Model(s): | S-COF500 |
| Classification Name: | Laser surgical instrument for use in general and plastic surgery and in dermatology |
| Classification: | II |
| Product Code: | GEX |
| Regulation Number: | 21 CFR 878.4810 |
| Review Panel: | General & Plastic Surgery |
- Identification of Predicate Device 4.
| 510(k) Number: | K100384 |
|----------------|-----------------------------------------------------------------------------------------|
| Product Name: | CO2 Laser WaveGuide |
| Manufacturer: | Lumenis Ltd., 13 Hayetzira St. (POB 240), Yokneam Industrial Park Yokneam, 20692 Israel |
- 5. Device Description
ProFlex CO2 Laser Fibers are fiber optic energy delivery devices consisting of a stainless steel laser connector, e.g. SMA 905, an anodized aluminum expansion nut, strain relief, a hollow glass tube having a dielectric coating (silver/silver iodide) in the bore and a fluoropolymer outer coating and an ethylene tetrafluoroethylene copolymer (ETFE, Tefzel™) protective jacket.
ProFlex CO2 Laser Fibers are packaged in a coiled and tubular, high-density polyethylene (HDPE) carrier providing improved ease of dispensing within the surgical field while maintaining sterility. Coils are contained within non-woven/impermeable polymer, e.g. Tyvek/Mylar, sterile pouches, validated for three-year shelf life, and protected with an outer, nonsterile fiberboard carton. Both sterile pouch and carton are labeled.
All ProFlex CO2 Laser Fiber materials of construction are USP Class VI biocompatible and are compatible with ethylene oxide (EtO) sterilization.
ProFlex CO2 Laser Fibers are for prescription use only.
- 6. Indications for Use
ProFlex CO2 Laser Fiber is indicated for the ablation, excision, incision, incision, and vaporization of soft tissue in open, endoscopic, and laparoscopic surgical procedures.
{4}------------------------------------------------
#### 7. Substantially Equivalent (SE) Comparison
| ITEM | Proposed Device | Predicate Device K161926 | Remark |
|----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Product Code | GEX | GEX | SE |
| Regulation No. | 21 CFR 878.4810 | 21 CFR 878.4810 | SE |
| Class | 2 | 2 | SE |
| Intended Use | ProFlex CO2 Laser Fiber is indicated for<br>the ablation, coagulation, excision,<br>incision, and vaporization of soft tissue in<br>open, endoscopic, and laparoscopic<br>surgical procedures. | The fiberlase CO2 laser waveguide is<br>intended for use in surgical procedures<br>requiring ablation, vaporization, excision,<br>incision and coagulation of soft tissue. The<br>fiberlase CO2 laser waveguide is indicated<br>for use in open surgical procedures such as<br>ent surgery and laparoscopy and endoscopic<br>procedures. The device is limited to carbon<br>dioxide lasers having a sma-905 connector. | SE |
#### Tab 1 General Comparison
The technical characteristics of the subject and predicate devices are compared in the table below.
| Device & Predicate Device(s): | K203799<br>(Subject Device) | K100384<br>(Primary Predicate) |
|----------------------------------------------|----------------------------------------------------|----------------------------------------------------|
| General Device Characteristics | | |
| Laser Type and wavelength (nm) | CO2, 10600 | CO2, 10600 |
| Inner diameter (µm) | 500 | 500 |
| Outer diameter (µm) | 1040 | 1040 |
| Length (m) | 2 | 2 |
| Maximum Input Power (W, CW) | 40 | 40 |
| Maximum Input Power (W, super-pulse<br>mode) | 15 | 15 |
| Operative Bending | 40mm radius @ 45°, 40 W<br>40mm radius @ 90°, 30 W | 40mm radius @ 45°, 40 W<br>40mm radius @ 90°, 30 W |
| Transmission Efficiency (Loss) | ≥ 60% (< 40%) | > 60% (< 40 %) |
The materials used in the construction of the subject and predicate devices are compared below.
| Description:<br>Component | Subject Device:<br>ProFlex CO2 Laser Fiber | Predicate Device:<br>FiberLase CO2 Laser WaveGuide<br>K100384 |
|---------------------------|--------------------------------------------|---------------------------------------------------------------|
| Package | HDPE tube with clips | Paper Card |
| X-Nut | Blue anodized aluminum | Nylon |
| Capillary cladding | Fluoropolymer | Fluoropolymer |
| Capillary buffer | Clear ETFE | Clear ETFE |
| Capillary base | Fused silica | Fused silica |
| Dielectric coating | Silver/Silver Iodide | Silver/Silver Iodide |
| Laser connector | Stainless Steel 905 SMA | Stainless Steel 905 SMA |
| Strain Relief | Black Santoprene rubber boot | Rubber boot |
| Proximal Cap | White Santoprene vented cap | Vented dust cap |
| Distal Cap | Silicone cap | Silicone cap |
| Bend Limiter | Polyolefin heat shrink | Heat shrink |
| Capillary Adhesive | USP VI Epoxy | Epoxy |
| Sterile package | Tyvek/Mylar film chevron pouch | Tyvek/Mylar film chevron pouch |
{5}------------------------------------------------
#### Non-Clinical Testing 8.
Non clinical tests were conducted to verify that the proposed device met all design specifications and was Substantially Equivalent (SE) to the predicate device. The following tests were conducted:
- > Bench Testing to support labeling and verify the performance (e.g., ProFlex CO2 Laser Fiber Instructions for use and labels, Fiber Design Verification, Packaging Design Verification)
- > Performance Qualification of ProFlex CO2 with Hoop Insert Packaging to Demonstrate 10t6 Sterility per ISO 11135:2014 and Ethylene Oxide sterilization residuals per ISO 10993-7:2008
#### 9. Clinical Testing
No clinical study is included in this submission.
- 10. Conclusion
Based on the comparison and analysis above, the proposed subject device is determined to be Substantially Equivalent (SE) to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.