K203491 · Laseroptek Co., Ltd. · GEX · Apr 7, 2021 · General, Plastic Surgery
Device Facts
Record ID
K203491
Device Name
PicoLO Nd: YAG Picosecond Laser System
Applicant
Laseroptek Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Apr 7, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
PicoLO laser system is intended for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery. <Tattoo Mode> 1064nm The 1064nm wavelength of the PicoLO laser system is indicated for tattoo removal for all skin types (Fitzpatrick I-VI) to treat the following tattoo colors: black, brown, green, blue and purple. 532nm The 532nm wavelength of the PicoLO laser system is indicated for tattoo removal for Fitzpatrick skin types I-III to treat the following tattoo colors: red, yellow and orange. <PH Mode> The PicoLO laser system is also indicated for benign pigmented lesions removal for Fitzpatrick Skin Types I-IV. The 1064 handpiece (1064nm) is also indicated for the treatment of acne scars in Fitzpatrick Skin Types II-V. The PicoLO laser system is also indicated for treatment of wrinkles in Fitzpatrick Skin Types I-IV.
Device Story
Multi-wavelength, pulsed Q-switched Nd:YAG picosecond laser system; delivers thermal energy to epidermis and dermis via focusing optics; used in dermatology, general, and plastic surgery; operated by clinicians. System comprises main body, color touch screen, articulated arm, handpiece, and foot switch. Input: user-selected parameters via interface; Output: laser pulses at 1064nm or 532nm. Device facilitates removal of tattoos, pigmented lesions, acne scars, and wrinkles. Clinical benefit derived from targeted thermal energy delivery to skin tissue.
Clinical Evidence
Bench testing only. Performance evaluated against standards: AAMI/ANSI ES60601-1, IEC 60601-1-2 (EMC), IEC 60601-1-6 (usability), IEC 60601-2-22 (laser safety), IEC 60825-1 (laser classification), ISO 14971 (risk management), and ISO 10993-5/10 (biocompatibility). No clinical data presented.
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services USA logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 7, 2021
Laseroptek Co., Ltd. % Do Hyun Kim CEO BT Solutions, Inc. Unit 904, Eonju-ro 86-gil 5, Gangnam-gu Seoul. Seoul 06210 Korea, South
Re: K203491
Trade/Device Name: PicoLO Nd: YAG Picosecond Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: November 23, 2020 Received: November 27, 2020
Dear Do Hyun Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Purva U. Pandya -S
Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
#### 510(k) Number (if known) K203491
Device Name PicoLO Nd:YAG Picosecond Laser System
#### Indications for Use (Describe)
PicoLO laser system is intended for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery.
#### <Tattoo Mode>
1064nm
The 1064nm wavelength of the PicoLO laser system is indicated for tattoo removal for all skin types (Fitzpatrick I-VI) to treat the following tattoo colors: black, brown, green, blue and purple.
#### 532nm
The 532nm wavelength of the PicoLO laser system is indicated for tattoo removal for Fitzpatrick skin types I-III to treat the following tattoo colors: red, yellow and orange.
<PH Mode>
The PicoLO laser system is also indicated for benign pigmented lesions removal for Fitzpatrick Skin Types I-IV.
The 1064 handpiece (1064nm) is also indicated for the treatment of acne scars in Fitzpatrick Skin Types II-V.
The PicoLO laser system is also indicated for treatment of wrinkles in Fitzpatrick Skin Types I-IV.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-------------------------------------------|
| | For Testing Under 21 CFR 58 Subpart D |
| | For The Control Under 21 CFR 58 Subpart G |
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 5. 510(k) Summary
## 1. General Information
| Applicant/Submitter: | Laseroptek Co., Ltd. |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | #114, #116, #117, #203, #204 Hyundai I Valley 31<br>Galmachi-Ro, 244beon-gil, Jungwon-Gu<br>Seongnam-Si, Gyeonggi-do, 13212<br>Rep. of Korea (South Korea)<br>Tel) +82.31.8023.5150<br>Fax) +82.31.8023.5151 |
| Contact Person: | Do-Hyun Kim, BT Solutions, Inc. |
| Address: | 904, Eonju-ro 86-gil 5,<br>Gangnam-gu, Seoul, 06210, Republic of Korea<br>Tel) +82.2.538.9140<br>Email) ceo@btsolutions.co.kr |
| Preparation Date: | March 5, 2021 |
## 2. Device Name and Code
| Device Trade Name: | PicoLO Nd:YAG Picosecond Laser System |
|----------------------|----------------------------------------------------------------------------------------|
| Common Name: | Nd:YAG Laser |
| Classification Name: | Laser surgical instrument for use in general and plastic<br>surgery and in dermatology |
| Product Code: | GEX |
| Regulation Number: | 878.4810 |
| Classification: | Class II |
| Review Panel: | General & Plastic Surgery (ODE) |
## 3. Predicate Device
PicoLO Nd: YAG Picosecond Laser System is substantially equivalent to the following devices
| Applicant | Device Name | 510(k) Number |
|----------------------|------------------------------------------|---------------|
| Laseroptek Co., Ltd. | PicoLO Nd:YAG Picosecond<br>Laser System | K183392 |
#### Table 5.1 Predicate device
{4}------------------------------------------------
Reference Device: Syneron Candela Corporation's PicoWay Laser System (K170597)
#### 4. Device Description
The PicoLO laser system is a multi-wavelength, pulsed laser system designed for the treatment of benign pigmented lesions. A key feature of the device is its ability to produce multiple laser wavelengths (1064 nm and 532 nm). The PicoLO Nd: YAG Picosecond Laser System consists of a set of O-switched Nd:YAG lasers, controlled by an embedded processor, to be used in dermatology. The laser system uses focusing optics to deliver a pattern of thermal energy to the epidermis and dermis. This system consists of main body, color touch screen, articulated arm, hand piece and foot switch.
#### 5. Indications / Intended Use
PicoLO laser system is intended for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery.
<Tattoo Mode>
1064nm
The 1064nm wavelength of the PicoLO laser system is indicated for tattoo removal for all skin types (Fitzpatrick I-VI) to treat the following tattoo colors: black, brown, green, blue and purple. 532nm
The 532nm wavelength of the PicoLO laser system is indicated for tattoo removal for Fitzpatrick skin types I-III to treat the following tattoo colors: red, yellow and orange.
<PH Mode>
The PicoLO laser system is also indicated for benign pigmented lesions removal for Fitzpatrick Skin Types I-IV.
The 1064 handpiece (1064nm) is also indicated for the treatment of acne scars in Fitzpatrick Skin Types II-V.
The PicoLO laser system is also indicated for treatment of wrinkles in Fitzpatrick Skin Types I-IV.
## 6. Technical Characteristics in Comparison
## 6.1. Technical Characteristics in Comparison to Predicate Device
PicoLO Nd: YAG Picosecond Laser System is substantially equivalent to the following legally marketed predicate device.
Table 5 2 Comparison table hetween Predicate device and Proposed device
| | TWOTE OULLYWOOD WOLF OVER POLL LAWINGS OF LAS WIN LIVE OF LOG AND LIVE<br>Predicate Device | Proposed Device | | |
|--|--------------------------------------------------------------------------------------------|-----------------|--|--|
| | | | | |
{5}------------------------------------------------
## PicoLO Nd:YAG Picosecond Laser System 510(k) Summary
| 510(K) Number | K183392 | K203491 |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | GEX | GEX |
| Classification /<br>Regulation | Class II/878.4810 | Class II/878.4810 |
| Manufacturer | Laseroptek Co.,Ltd. | Laseroptek Co.,Ltd. |
| Device Name | PicoLO Nd:YAG Picosecond Laser System | PicoLO Nd:YAG Picosecond Laser System |
| Clearance Date | 11 Feb 2019 | N/A |
| Intended Use /<br>Indications for<br>Use: | PicoLO laser system is intended for use in<br>surgical and aesthetic applications in the<br>medical specialties of dermatology and<br>general and plastic surgery.<br><br>1064nm<br>The 1064nm wavelength of the PicoLO<br>laser system is indicated for tattoo removal<br>for all skin types (Fitzpatrick I-VI) to treat<br>the following tattoo colors: black, brown,<br>green, blue and purple.<br><br>532nm<br>The 532nm wavelength of the PicoLO laser<br>system is indicated for tattoo removal for<br>Fitzpatrick skin types I-III to treat the<br>following tattoo colors: red, yellow and<br>orange. | PicoLO laser system is intended for use in<br>surgical and aesthetic applications in the<br>medical specialties of dermatology and<br>general and plastic surgery.<br><br><Tattoo Mode><br><br>1064nm<br>The 1064nm wavelength of the PicoLO<br>laser system is indicated for tattoo removal<br>for all skin types (Fitzpatrick I-VI) to treat<br>the following tattoo colors: black, brown,<br>green, blue and purple.<br><br>532nm<br>The 532nm wavelength of the PicoLO laser<br>system is indicated for tattoo removal for<br>Fitzpatrick skin types I-III to treat the<br>following tattoo colors: red, yellow and<br>orange.<br><br><PH Mode><br>The PicoLO laser system is also indicated<br>for benign pigmented lesions removal for<br>Fitzpatrick Skin Types I-IV.<br>The 1064 handpiece (1064nm) is also<br>indicated for the treatment of acne scars in<br>Fitzpatrick Skin Types II-V.<br>The PicoLO laser system is also indicated<br>for treatment of wrinkles in Fitzpatrick Skin<br>Types I-IV. |
| Wavelength | 1064/532 nm (Accuracy ±20%) | 1064/532 nm (Accuracy ±20%) |
| Pulse Duration<br>(Pulse Width) | 450ps (1064nm), (Accuracy ±20%)<br>380ps (532nm), (Accuracy ±20%) | 450ps (1064nm), (Accuracy ±20%)<br>380ps (532nm), (Accuracy ±20%) |
| Pulse Energy<br>(max) | 500mJ (1064nm), (Accuracy ±20%)<br>350mJ (532nm), (Accuracy ±20%) | Tattoo Mode<br>500mJ (1064nm), (Accuracy ±20%)<br>350mJ (532nm), (Accuracy ±20%)<br><br>PH Mode*<br>400mJ (1064nm), (Accuracy ±20%) |
| | | 200mJ (532nm), (Accuracy ±20%) |
| Peak Power<br>(Gigawatts) | 1.1 | 1.1 |
| Aiming Beam | Laser diode, 635nm/ <5mW | Laser diode, 635nm/ <5mW |
| Repetition Rate<br>(Hz) | Single, M3, M5, 1~10Hz (Accuracy: ±<br>20%) | Single, M3, M5, 1~10Hz (Accuracy: ±<br>20%) |
| Spot size (mm) | 1064 (10 mm)<br>532 (7 mm)<br>Collimator (20 mm)<br>Zoom (2~7 mm) | 1064 (10 mm)<br>532 (7 mm)<br>Collimator (20 mm)<br>Zoom (2~7 mm) |
| Laser Type | Q-switched Nd:YAG Laser | Q-switched Nd:YAG Laser |
| Activation | Via foot-switch | Via foot-switch |
| Display | TFT LCD Touch screen | TFT LCD Touch screen |
| Cooling System | Internal water to air heat exchanger | Internal water to air heat exchanger |
| Electrical Power | 220-230VAC, 50/60Hz | 220-230VAC, 50/60Hz |
| Beam Delivery<br>System | Articulated Arm with Handpiece | Articulated Arm with Handpiece |
| System<br>Dimensions(mm) | 350(W) x 1080(L) x 970(H) | 350(W) x 1080(L) x 970(H) |
| System Weight<br>(kg) | 110 kg | 110 kg |
{6}------------------------------------------------
PH mode* is intended for 3 newly added indications, as below;
- 1) benign pigmented lesions removal for Fitzpatrick Skin Types I-IV,
- 2) the treatment of acne scars in Fitzpatrick Skin Types II-V, and
- 3) the treatment of wrinkles in Fitzpatrick Skin Types I-IV.
#### 6.2. Technical Characteristics in Comparison to Reference Device
Table 5.3 Comparison table between Reference device and Proposed device
| | Reference Device | Proposed Device |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(K) Number | K170597 | K203491 |
| Product Code | GEX | GEX |
| Classification /<br>Regulation | Class II/878.4810 | Class II/878.4810 |
| Manufacturer | Syneron Candela Corporation | Laseroptek Co.,Ltd. |
| Device Name | PicoWay Laser System | PicoLO Nd:YAG Picosecond Laser System |
| Clearance Date | 25 May 2017 | N/A |
| Intended Use /<br>Indications for<br>Use: | The PicoWay laser system is indicated for<br>the following at the specified wavelength: | PicoLO laser system is intended for use in<br>surgical and aesthetic applications in the |
| | 532nm: Removal of tattoos for Fitzpatrick<br>skin types I-III to treat the following tattoo<br>colors: red, yellow and orange. | medical specialties of dermatology and<br>general and plastic surgery. |
| | 785nm: Removal of tattoos for Fitzpatrick<br>skin types II-IV to treat the following tattoo<br>colors: green and blue.<br>1064nm: Removal of tattoos for all skin<br>types (Fitzpatrick I-VI) to treat the<br>following tattoo colors: black, brown, | <Tattoo Mode> |
| | | 1064nm |
| | | The 1064nm wavelength of the PicoLO<br>laser system is indicated for tattoo removal<br>for all skin types (Fitzpatrick I-VI) to treat<br>the following tattoo colors: black, brown,<br>green, blue and purple. |
| | green, blue and purple. | 532nm |
| | The PicoWay laser system is also indicated<br>for benign pigmented lesions removal for<br>Fitzpatrick Skin Types I-IV. | The 532nm wavelength of the PicoLO laser<br>system is indicated for tattoo removal for<br>Fitzpatrick skin types I-III to treat the<br>following tattoo colors: red, yellow and |
| | The Resolve handpiece (1064 nm) is also<br>indicated for the treatment of acne scars in<br>Fitzpatrick Skin Types II-V. | orange. |
| | The Resolve handpieces are also indicated | <PH Mode> |
| | for treatment of wrinkles in Fitzpatrick Skin<br>Types I-IV. | The PicoLO laser system is also indicated<br>for benign pigmented lesions removal for<br>Fitzpatrick Skin Types I-IV. |
| | | The 1064 handpiece (1064nm) is also<br>indicated for the treatment of acne scars in<br>Fitzpatrick Skin Types II-V. |
| | | The PicoLO laser system is also indicated<br>for treatment of wrinkles in Fitzpatrick Skin<br>Types I-IV. |
| Wavelength | 532nm, 1064nm<br>785 nm | 1064/532 nm (Accuracy ±20%) |
| Pulse Duration<br>(Pulse Width) | 450ps (1064nm)<br>375ps (532nm) | 450ps (1064nm), (Accuracy ±20%)<br>380ps (532nm), (Accuracy ±20%) |
| Pulse Energy<br>(max) | 400mJ (1064nm)<br>200mJ (532nm) | 500mJ (1064nm), (Accuracy ±20%)<br>350mJ (532nm), (Accuracy ±20%) |
| Peak Power<br>(Gigawatts) | Up to 0.9 | 1.1 |
| Aiming Beam | Unknown | Laser diode, 635nm/ <5mW |
| Repetition Rate<br>(Hz) | Single, 1-10 Hz (1064nm and 532 nm) | Single, M3, M5, 1~10Hz (Accuracy: ±<br>20%) |
| Spot size (mm) | Zoom 2-10 mm (1064nm and 532nm)<br>Fractional and non-fractional 1064 (6x6<br>mm²)<br>Fractional and non-fractional 532 (6x6<br>mm²) | 1064 (10 mm)<br>532 (7 mm)<br>Collimator (20 mm)<br>Zoom (2~7 mm) |
| Laser Type | Q-switched Nd:YAG Laser | Q-switched Nd:YAG Laser |
| Activation | Via foot-switch | Via foot-switch |
| Display | LCD Touch screen | TFT LCD Touch screen |
| Cooling System | Unknown | Internal water to air heat exchanger |
| Electrical Power | 200-240 VAC, 50/60 Hz, 30 A, 4600 VA single | 220-230VAC, 50/60Hz |
| Beam Delivery<br>System | Articulated Arm with Handpiece | Articulated Arm with Handpiece |
| System<br>Dimensions(mm) | 1070 (H) x 460 (W) x 690 (D) | 350(W) x 1080(L) x 970(H) |
| System Weight<br>(kg) | 125 kg | 110 kg |
{7}------------------------------------------------
## 510(k) Summary
{8}------------------------------------------------
#### 7. Performance Data
Non-clinical tests: Testing conducted on the PicoLO Nd:YAG Picosecond Laser System shows that it refers to the relevant mandatory performance standards for laser products 21 CFR 1040.10 and 1040.11. Other performance, such as electromagnetic compliance, etc, were tested using following standards:
- PicoLO Nd:YAG Picosecond Laser System is tested and evaluated according to -AAMI/ANSI ES60601-1:2005 and A1:2012. All the results presented in the submission demonstrate general requirements for basic safety and essential performance.
- -Effect to the device by electromagnetic disturbances were tested and evaluated according to the FDA-recognized consensus standard IEC 60601-1-2: 2007. All the results presented here demonstrated the requirements and tests for electromagnetic disturbances.
- -PicoLO Nd:YAG Picosecond Laser System is tested and evaluated according to FDArecognized consensus standard IEC 60601-1-6:2010/AMD1:2013. All the results presented here demonstrated the General requirements for safety - Collateral Standard: Usability.
- -PicoLO Nd:YAG Picosecond Laser System is tested and evaluated according to FDArecognized consensus standard IEC 60601-2-22: 2007 (Third Edition) + A1:2012. All the results presented here demonstrated the particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment.
- -Safety of laser products is evaluated according IEC 60825-1: 2014. All the results presented here demonstrated the equipment classification and requirements.
- Risk management was recorded according to the FDA-recognized consensus standard ISO -14971: 2012. All the results presented here demonstrated the application of risk management to medical devices.
- -Usability was documented according to the FDA-recognized consensus standard IEC 62366: 2008. All the results presented here demonstrated the application of usability engineering to medical devices.
- Biocompatibility was tested and evaluated according to FDA-recognized consensus standard ISO 10993-5: 2009 and ISO 10993-10: 2010.
{9}------------------------------------------------
#### 8. Substantial Equivalence
The modified PicoLO Nd:YAG Picosecond Laser System, subject of this submission, is a modification of the PicoLO Nd:YAG Picosecond Laser System cleared under K183392. The intended use of the modified device is expanded as compared with the predicate device. This expansion does not affect technological principles between the modified device and the predicate device.
Based upon the predicted overall performance characteristics for the PicoLO Nd: Y AG Picosecond Laser System, Laseroptek Co. Ltd. believes that no sigmificant differences exist in usage of its underlying technological principles between PicoLO Nd:YAG Picosecond Laser System and the predicate device.
#### 9. Conclusions
The technological characteristics of the subject device PicoLO Nd: Y AG Picosecond Laser System are comparable to the predicate device for comparable indications for use. Thus, subject device PicoLO is concluded to be substantially equivalent to the predicates.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.