K202848 · Lightsense Technologies, Ltd. · GEX · Dec 14, 2021 · General, Plastic Surgery
Device Facts
Record ID
K202848
Device Name
S2 Pigment Removal System
Applicant
Lightsense Technologies, Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Dec 14, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K202848 · Dec 14, 2021
S2 Pigment Removal System
Lightsense Technologies, Ltd.
Commercial clinical practice records
Real-world data from 957 patients were used to support the safety and performance of the S2 Pigment Removal System across different wavelengths, tattoo colors, and skin types, including specific safety outcomes for Fitzpatrick V and VI skin types.
Safety and performance across wavelengths, tattoo colors, and skin types; incidence of post-treatment reactions
Indications for Use
S2 Pigment Removal System is intended for use in removal of tattoo ink. The S2 Pigment Removal System is indicated for the removal of tattoo ink according to the following wavelengths: 800 nm: The LightSense S2 800 nm Ti:Sapphire laser system is indicated for tattoo removal for Fitzpatrick Skin Types I-III. 532 nm: The LightSense S2 532 nm Nd-YAG laser system is indicated for tattoo removal for Fitzpatrick Skin Types I - III. 1064 nm: The LightSense S2 1064 nm Nd-YAG laser system is indicated for tattoo removal for Fitzpatrick Skin Types I-VI.
Device Story
S2 Pigment Removal System uses high-intensity laser pulses to remove tattoo ink. System components: two treatment lasers (800nm Ti:Sapphire; 1064/532nm Nd:YAG), aiming beam, galvanometric scanner, articulating arm, hand-piece with adjustable spacer, foot pedal. Operated by clinician in clinical setting. Device delivers laser energy to target tattoo pigment; scanner/spacer facilitate precise delivery. Output: laser energy for photomechanical/photothermal ink fragmentation. Clinical benefit: effective tattoo clearance across various skin types and ink colors. Safety features include emergency stop, door interlock, motion sensor, and real-time monitoring. Healthcare providers use device to perform multi-session tattoo removal; output affects treatment planning and session frequency based on patient response and skin type.
Clinical Evidence
Clinical study of 58 subjects (Fitzpatrick I-IV) with colored tattoos. Primary endpoint: tattoo clearing at 8 weeks post-final treatment (max 16 sessions) via blinded photographic assessment. Results: 100% of subjects achieved >50% clearance. Real-world data from 957 patients supported safety/efficacy across skin types I-VI. No serious adverse effects reported; minor reactions (swelling, redness) were temporary. Bench and animal studies confirmed output parameters and optimal treatment protocols.
Indicated for removal of tattoo ink in patients Fitzpatrick Skin Types I-VI, depending on laser wavelength (800nm: I-III; 532nm: I-III; 1064nm: I-VI).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 14, 2021
Lightsense Technologies Ltd % John Smith Partner Hogan Lovells US LLP 555 Thirteenth Street NW Washington, District of Columbia 20004
Re: K202848
Trade/Device Name: S2 Pigment Removal System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: November 1, 2021 Received: November 1, 2021
Dear John Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
510(k) Number (if known)
#### K202848
Device Name
#### S2 Pigment Removal System
Indications for Use (Describe)
S2 Pigment Removal System is intended for use in removal of tattoo ink. The S2 Pigment Removal System is indicated for the removal of tattoo ink according to the following wavelengths:
800 nm:
The LightSense S2 800 nm Ti:Sapphire laser system is indicated for Fitzpatrick Skin Types I-II.
532 nm:
The LightSense S2 532 nm Nd-YAG laser system is indicated for tattoo removal for Fitzpatrick Skin Types I - III.
1064 nm:
The LightSense S2 1064 nm Nd-YAG laser system is indicated for tattoo removal for Fitzpatrick Skin Types I-VI.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
{4}------------------------------------------------
# 510(k) SUMMARY LIGHTSENSE TECHNOLOGIES LTD's S2 Pigment Removal System K202848
#### Submitter
LIGHTSENSE TECHNOLOGIES LTD 109 Great Portland Street London W1W 6QG United Kingdom
Phone: +44 20 7031 3471
Contact Person: Dr. Deganit Barak
Date Prepared: December 13, 2021
Name of Device: S2 Pigment Removal System
Common or Usual Name: Powered Laser Surgical Instrument
Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology
Regulatory Class: Class II
Product Code: GEX
Predicate Devices: Cynosure, PicoSure Workstation (K160480)
#### Device Description:
The S2 Pigment Removal Laser System (S2) uses high intensity laser pulses to remove tattoo ink from skin. The S2 includes the two treatment lasers, an aiming beam, a small galvanometric scanner, an articulating arm, a hand-piece with an adjustable spacer, and a foot pedal. The S2 uses two lasers to deliver three working wavelengths: (1) 800nm, ultra-short pulsed Ti:Sapphire laser, and (2) 1064 and 532nm Nd:YAG short pulse laser. The combined lasers are designed to remove tattoos.
#### Intended Use / Indications for Use
S2 Pigment Removal System is intended for use in removal of tattoo ink. The S2 Pigment Removal System is indicated for the removal of tattoo ink according to the following wavelengths:
800 nm:
The LightSense S2 800 nm Ti:Sapphire laser system is indicated for tattoo removal for Fitzpatrick Skin Types I-III.
{5}------------------------------------------------
532 nm:
The LightSense S2 532 nm Nd-YAG laser system is indicated for tattoo removal for Fitzpatrick Skin Types I - III.
1064 nm:
The LightSense S2 1064 nm Nd-YAG laser system is indicated for tattoo removal for Fitzpatrick Skin Types I-VI.
# Summary of Technological Characteristics:
Laser energy emitted at different wavelengths is the technological principle for both the subject and predicate devices. At a high level, the subject and predicate devices are based on the following same technological elements:
Both systems include multiple lasers which operate at comparable wavelengths, allowing the operator to select the best wavelength for the patient based on the tattoo which the patient wishes to have removed. The laser systems also both have treatment hand-pieces connected to articulated arms that are connected to the main console where the user interface is located. The following technological differences exist between the subject and predicate devices:
A table comparing the key features of the subject and predicate devices is provided below.
| | S2 Pigment Removal<br>System | PicoSure™ | | |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|----------|-------------|
| Components | Laser systems, three<br>wavelengths, treatment hand<br>pieces, articulated arms,<br>main console, user interface,<br>switch pedal | Laser systems, three<br>wavelengths, treatment hand<br>pieces, articulated arms, main<br>console, user interface, switch<br>pedal | | |
| Accessories | Spacer (connected to the<br>treatment hand piece) | interchangeable handpieces<br>including spacers and treatment<br>tips | | |
| Power Source | 100-240V, 50-60Hz | 200-240V, 50/60Hz | | |
| | Nd:YAG | Ti:Sapph | Nd:YVO4 | Alexandrite |
| Wavelength (nm) | 1064, 532 | 800 | 1064,532 | 755 |
| pulse energy (mJ) | 1-8 mJ | 1-5 mJ | 450 | 450 |
| pulse width (ps_ | 400-700 | 8 | 450-900 | 450-900 |
| spot size (diam)/ treatment<br>field (mm) | 0.4-1 / 1-10 | 0.4-1 / 1-10 | 4 | 3 |
| pulse repetition rate (Hz) | 500 | 1000 | 1-10 | 1-10 |
{6}------------------------------------------------
| Fluence (J/cm²) | 0.5-8 J/cm² | 0.5-4 J/cm² | 3.6/1.5 | 6.4 |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----|
| Safety Features | Emergency stop button, Door interlock, Emission Indication LED, Outline Mode Before Lasing, Outline Scan, Motion Sensor, Spacer Feedback, Realtime Monitor, Built-in test | | Key Switch, Emergency laser stop, Standby Mode, Delayed Ready Mode, Automatic Shutdown Feature, Remote Interlock, Audible Tone, Laser Danger Sign, Locking Casters, Device Labels | |
| Biocompatibility | Yes (spacer) | | Yes (treatment tip) | |
| Software | Yes | | Yes | |
| Sterilization | No (Cleaning and disinfection of the spacer) | | Cleaning and disinfection of the treatment tip, following by autoclave sterilization | |
| Standards with which the Device Complies | IEC 60601-1, IEC 60601-2, ANSI Z-136. 3; CFR Part 1000 [parts 1040.10 and 1040.11]; 12-273 IEC 60825-1 Edition 2.0 2007-03 | | IEC 60601-1, IEC 60601-2, ANSI Z-136. 3; CFR Part 1000 [parts 1040.10 and 1040.11]; 12-273 IEC 60825-1 Edition 2.0 2007-03 | |
## Performance Data
Safety and performance data demonstrates that the S2 Pigment Removal System complies with the following FDA recognized consensus standards:
- . AAMI/ANSI ES60601-1:2005(R) 2012 and A1:2012 Electrical safety
- IEC 60601-1-2:2014 Electromagnetic compatibility including radiated emissions and immunity
- IEC 60601-1-6:2010 General requirements for basic safety and essential performance -● Collateral standard: Usability
- IEC 60601-2-22: 2007 (Third Edition) + A1:2012 Essential performance and safety for laser products
- IEC 60825-1:2014 Safety of laser products
- ISO 10993-5:2009 and ISO 10993-10:2010 Biocompatibility to address cytotoxicity, sensitization, and irritation of skin contacting components
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" consistent with "moderate" level of concern. Each element of the SRS was tested and found to meet the requirements.
{7}------------------------------------------------
In addition, bench testing was performed to address verification of the device output parameters, including wavelength and fluence transmission. Moreover, two animal preclinical studies were conducted. One study was conducted to validate findings from the bench testing and to confirm optimal parameters for tattoo removal. The other study was conducted to establish optimal treatment protocols. In all instances, the S2 Pigment Removal System functioned as intended and consistent with the predicate device.
A clinical study was conducted in 58 subjects with colored tattoos and various skin types (Fitzpatrick I-IV) and at various depths within the skin. Subjects received a maximum number of 16 treatments, where a maximum of 2 treatments were applied at each study visit, spaced 2 weeks apart. Treatments were conducted with the three wavelengths, using the specific wavelength relevant to the specific tattoo color, in order to validate the safety and efficacy of the tattoo removal by the device. Some subjects were treated by more than one laser wavelength depending on the coloration of the tattoo. The study primary endpoint was an evaluation of tattoo clearing as rated by a panel of blinded reviewers based on a photographic assessment performed 8 weeks post-final treatment based on a 5 point global assessment scale.
The primary and secondary endpoints were both reached successfully in terms of efficacy and safety. No serious adverse device effects were reported and the few minor observed adverse effects (AE) were minor with no observed long-term effects.
Analyzing the clearing of tattoos at the 8 week follow up visit after the final treatment session, when analyzed by tattoo color segment, results show 100% percent of subjects in the analysis population with tattoo clearance of more than 50% compared with baseline photographs. Results by wavelength show 100% responders regardless of the wavelength used during treatment.
The safety and performance of the device is further supported by real world data gathered in commercial clinical practice. Data from a total of 957 patients undergoing treatment with the device for tattoo removal support the device safety and performance across wavelengths, tattoo colors, and skin types.
In addition to the data discuss above from the clinical study,136 patients have been treated with the 532nm wavelength for the clearance of tattoos with red, yellow and orange ink. Tattoo clearance using 532nm has shown to achieve equivalent or better clearance by session compared to the 1064nm laser wavelength for the removal black ink. Other less-common colors have similarly obtained favorable results. Specifically, purple ink treated with 800nm (n=13) has obtained greater than 50% clearance (as assessed immediately following treatment) in fewer treatments than was necessary to obtain the same level of clearance for red, yellow, orange inks or black ink.
For persons with darker skin types (Fitzpatrick V), data from 109 patients from 109 patients from commercial clinical use supports the safety and performance of the 1064nm laser wavelength using a modified treatment protocol with additional time between treatment sessions. Efficacy is comparable for all patients removing black ink with the 1064nm laser wavelength, regardless of skin type. The available real world data shows a lower overall incidence rate of post-treatment reactions in Fitzpatrick V and VI patients (7.34%) than Fitzpatrick I-IV patients (9.67%). Regarding the classification of these in-treatment reactions; persons with darker skin tones are more likely to observe temporary swelling (73.1%) and less likely to observe redness (79.5%) when compared to Fitzpatrick 1-4 patient groups.
{8}------------------------------------------------
Based on the clinical performance as documented in the pivotal clinical study and the real world data collection, the S2 Pigment Removal System has a safety and effectiveness profile that is similar to the predicate device.
## Conclusions:
The S2 Pigment Removal System has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. The minor differences in indications do not alter the intended therapeutic use of the device and do not alter its safety or effectiveness when used as instructed in the product labeling. In addition, the minor technological differences between the S2 Pigment Removal System and its predicate device raise no new issues of safety or effectiveness. Moreover, performance data, animal data, and clinical study data demonstrate that the S2 Pigment Removal System is as safe and effective as the PicoSure predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.