Stellar M22 for Intense Pulsed Light (IPL) and Laser System
K193500 · Lumenis, Ltd. · GEX · Jan 16, 2020 · General, Plastic Surgery
Device Facts
Record ID
K193500
Device Name
Stellar M22 for Intense Pulsed Light (IPL) and Laser System
Applicant
Lumenis, Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jan 16, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Lumenis Stellar M22 has connection capability with the following available treatment handpieces, for multi-application treatment options. All handpieces are designed for aesthetic and dermatological skin procedure applications, as follows: - The Stellar Intense Pulsed Light (IPL) handpiece with a spectrum of 400-1200 nm (with 9 different filters) is indicated for: - Benign epidermal lesions, including dyschromia, hyperpigmentation, melasma, and ephelides (freckles) - Cutaneous lesions, including warts, scars and striae - Benign cutaneous vascular lesions, including port wine stains, hemoangiomas, facial, truncal and leg telangiectasias, erythema of rosacea, angiomas and spider angiomas, poikiloderma of Civatte, leg veins and venous malformations - Removal of unwanted hair and to effect stable long term, or permanent* hair reduction in skin types I-V through selective targeting of melanin in hair follicles - Mild to moderate inflammatory Acne (Acne vulgaris) - The Nd:YAG Laser handpiece with a wavelength of 1064 nm (Multi-Spot Nd:YAG) is indicated for: - The coagulation and hemostasis of vascular lesions and soft tissue, including the treatment and clearance of superficial and deep telangiectasias (venulectasias) and reticular veins (0.1-4.0 mm. diameter) of the leg - The removal of unwanted hair from all skin types, and to effect table long term, or permanent* hair reduction in skin types I-V through selective targeting of melanin in hair follicles - The non-ablative treatment of facial wrinkles - ResurFX module and handpiece, with wavelength of 1565 nm, is indicated for: - Use in dermatological procedures requiring fractional skin resurfacing and coagulation of soft tissue - The Q-Switched Nd:YAG Laser Handpiece with a wavelength of 1064 nm is indicated for: - Removal of dark tattoos - Treatment of pigmented lesions
Device Story
Multi-application platform using selective photothermolysis for aesthetic/dermatological procedures. Inputs: user-selected treatment parameters via console. Outputs: IPL (400-1200nm), Nd:YAG laser (1064nm), or Er:Glass fiber-laser (1565nm) energy delivered via handpieces. Used in clinical settings by trained professionals. Device features sapphire cooling, scanner-assisted fractional resurfacing, and multiple sequential pulsing (AOPT). Output affects tissue via thermal coagulation/hemostasis or melanin targeting. Benefits include hair reduction, lesion clearance, and skin resurfacing. System includes lightguide recognition and remote access for internal company use.
Clinical Evidence
Bench testing only. Evidence includes design verification/validation, risk analysis (ISO 14971), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), IPL compatibility (IEC 60601-2-57), and software verification/validation.
Technological Characteristics
Multi-application platform; energy sources: IPL (400-1200nm), Nd:YAG laser (1064nm), Er:Glass fiber-laser (1565nm). Features sapphire cooling, dual-axis scanner, and lightguide recognition. Connectivity: Windows 10 Embedded OS, remote access capability. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-57.
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Lumenis Ltd. % Jonathan Kahan Regulatory Counsel Hogan Lovells US LLP 555 Thirteenth Street NW Washington, District of Columbia 20004
Re: K193500
Trade/Device Name: Stellar M22 for Intense Pulsed Light (IPL) and Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX, ONF, ONG Dated: December 17, 2019 Received: December 17, 2019
Dear Jonathan Kahan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for
combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jessica Mavadia-Shukla Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| Indications for Use Statement | |
|-------------------------------|--|
|-------------------------------|--|
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
| Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement on last page |
|---------------------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------------|
| 510(k) Number (if known) | K193500 |
|--------------------------------|-------------------------------------------------------------|
| Device Name | Stellar M22 for Intense Pulsed Light (IPL) and Laser system |
| Indications for Use (Describe) | |
The subject Stellar M22 has connection capability with the following available treatment handpieces, for multi-application treatment options. All handpieces are designed for aesthetic and dermatological skin procedure applications, as follows:
- The Intense Pulsed Light (IPL) handpiece with a spectrum of 400-1200 nm (with 9 different filters) is indicated for:
- Benign epidermal lesions, including dyschromia, hyperpigmentation, melasma, and o ephelides (freckles)
- Cutaneous lesions, including warts, scars and striae O
- Benign cutaneous vascular lesions, including port wine stains, hemoangiomas, facial, O truncal and leg telangiectasias, erythema of rosacea, angiomas and spider angiomas, poikiloderma of Civatte, leg veins and venous malformations
- Removal of unwanted hair and to effect stable long term, or permanent* hair O reduction in skin types I-V through selective targeting of melanin in hair follicles
- Mild to moderate inflammatory Acne (Acne vulgaris) o
- The Nd:YAG Laser handpiece with a wavelength of 1064 nm (Multi-Spot Nd:YAG) is ● indicated for:
- The coagulation and hemostasis of vascular lesions and soft tissue, including the O treatment and clearance of superficial and deep telangiectasias (venulectasias) and reticular veins (0.1-4.0 mm. diameter) of the leg
- The removal of unwanted hair and to effect table long term, or permanent* hair O reduction in skin types I-V through selective targeting of melanin in hair follicles
- The non-ablative treatment of facial wrinkles o
- . ResurFX module and handpiece, with wavelength of 1565 nm, is indicated for:
- Use in dermatological procedures requiring fractional skin resurfacing and O
{3}------------------------------------------------
coagulation of soft tissue
#### The Q-Switched Nd:YAG Laser Handpiece with a wavelength of 1064 nm is indicated ● for:
- Removal of dark tattoos O
- o Treatment of pigmented lesions
*Note
Permanent hair reduction is defined as long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after completion of treatment regime.
C)
Type of Use (Select one or both, as applicable)
区 Prescription Use (Part 21 CFR 801 Subpart D)
[ ] Over-The-Counter Use (21 CFR 801 Subpart
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
Page 1 of 1 FDA
PSC Publishing Services (301) 443-6
{4}------------------------------------------------
# 510(k) SUMMARY
# Stellar M22 for Intense Pulsed Light (IPL) and Laser System
### Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Lumenis Ltd. 6 Hakidma Street PO Box 240 Yokneam Industrial Park Yokneam 2069204, Israel
Phone: +972-4-9599000 Facsimile: +972-4-9599230
Contact Person: Shlomit Segman, RA Senior Manager
Date Prepared: December 17, 2019
### Name of the device:
Stellar M22 for Intense Pulsed Light (IPL) and Laser system
### Classification Name:
Powered Laser Surgical Instrument
### Regulation Class:
Class II
# Product Code:
GEX (21 C.F.R. § 878.4810)
#### Subsequent Product Codes:
ONF, ONG
#### Predicate Device:
M22 System (K170060)
#### Intended Use
The Lumenis Stellar M22 has connection capability with the following available treatment handpieces, for multi-application treatment options. All handpieces are designed for aesthetic and dermatological skin procedure applications, as follows:
{5}------------------------------------------------
- The Stellar Intense Pulsed Light (IPL) handpiece with a spectrum of 400-1200 nm (with 9 different filters) is indicated for:
- Benign epidermal lesions, including dyschromia, hyperpigmentation, melasma, and o ephelides (freckles)
- Cutaneous lesions, including warts, scars and striae O
- Benign cutaneous vascular lesions, including port wine stains, hemoangiomas, facial, O truncal and leg telangiectasias, erythema of rosacea, angiomas and spider angiomas, poikiloderma of Civatte, leg veins and venous malformations
- Removal of unwanted hair and to effect stable long term, or permanent* hair O reduction in skin types I-V through selective targeting of melanin in hair follicles
- Mild to moderate inflammatory Acne (Acne vulgaris) o
- The Nd:YAG Laser handpiece with a wavelength of 1064 nm (Multi-Spot Nd:YAG) is ● indicated for:
- The coagulation and hemostasis of vascular lesions and soft tissue, including the O treatment and clearance of superficial and deep telangiectasias (venulectasias) and reticular veins (0.1-4.0 mm. diameter) of the leg
- The removal of unwanted hair from all skin types, and to effect table long term, or o permanent* hair reduction in skin types I-V through selective targeting of melanin in hair follicles
- The non-ablative treatment of facial wrinkles o
- ResurFX module and handpiece, with wavelength of 1565 nm, is indicated for: ●
- Use in dermatological procedures requiring fractional skin resurfacing and o coagulation of soft tissue
- . The Q-Switched Nd:YAG Laser Handpiece with a wavelength of 1064 nm is indicated for:
- Removal of dark tattoos O
- Treatment of pigmented lesions o
*Note: Permanent hair reduction is defined as long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after completion of treatment regime.
{6}------------------------------------------------
# Device Description
The subject Stellar M22 for Intense Pulsed Light (IPL) and Laser System is a multi-application, multitechnology platform with four (4) available treatment handpieces:
- Stellar Intense Pulsed Light (IPL) handpiece; ●
- . Multi-Spot Nd:YAG laser handpiece;
- ResurFX non-ablative laser handpiece: ●
- Q-Switched Nd:YAG laser handpiece. ●
The following accessories are provided with and/or may be purchased independently for each of the four (4) available treatment handpieces of the Stellar M22 for Intense Pulsed Light (IPL) and Laser Systems:
- The Stellar IPL handpiece has nine (9) different filters available: Cut-off filters of 515, . 560, 590,615, 640, 695 and 755 nm, Notch Filters of 400-600 & 800-1200 nm and 530-650& 900-1200 nm. Further, the IPL handpiece has three (3) Sapphire Cool LightGuides available with sizes of: 15mm x 35mm, 8mm x 15 mm, 6 mm diameter.
- The Multi-Spot Nd:YAG handpiece has three (3) different LightGuides available in sizes . of: 2mm x 4mm, 6 mm, and 9 mm.
- The ResurFX handpiece has two (2) different treatment tips available: SapphireCool and Precision tips
- . The Q-Switched Nd:YAG handpiece has both disposable and gold plated metal treatment tips available. The disposable treatment tips and metal treatment tips are available in seven (7) different sizes of: 2, 2.5, 3.5, 4, 5, 6 and 8 mm.
### Technological Characteristics
The intended use and indications for use of the Stellar M22 for Intense Pulsed Light (IPL) and Laser Systems are the same as the predicate device. In addition, the same technological characteristics and principles of operation apply for the Stellar M22 for Intense Pulsed Light (IPL) and Laser Systems and the predicate device.
Comparison table of the technological characteristics of the Stellar M22 for Intense Pulsed Light (IPL) and Laser System compared to the predicate device is provided below:
| | Stellar M22<br>(subject device) | Cleared M22 and<br>ResurFX Systems<br>(K170060) | Comparison |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Principle of Operation | Selective<br>photothermolysis | Selective<br>photothermolysis | Same as predicate |
| Device Components | System console<br>Four (4) available<br>treatment handpieces:<br>• Stellar Intense<br>Pulsed Light<br>(IPL) | System console<br>Four (4) available<br>treatment handpieces:<br>• Universal<br>Intense Pulsed<br>Light (IPL) | Same as predicate |
| Monitor Display | Multi-Spot<br>Nd:YAG laser<br>handpiece Q-Switched<br>Nd:YAG laser<br>handpiece ResurFX non-<br>ablative laser<br>module and<br>handpiece 12" | Multi-Spot<br>Nd:YAG laser<br>handpiece Q-Switched<br>Nd:YAG laser<br>handpiece ResurFX non-<br>ablative laser<br>module and<br>handpiece 8.4" | Substantially<br>equivalent |
| Software operating<br>system | Windows 10 Embedded | Windows CE 5 | Substantially<br>equivalent |
| IPL Handpiece | | | |
| Wavelength | 400-1200 nm | 400-1200 nm | Same as predicate |
| Pulse duration<br>(msec) | Up to 20 msec – single<br>pulse | Up to 20 msec – single<br>pulse | Same as predicate |
| Operational<br>wavelengths | The IPL handpiece<br>comes with 9 filters for<br>various wavelengths: Cut-off filters: 515,<br>560, 590, 615, 640,<br>695, 755 nm Acne Filter (Notch<br>filter 400-600 and<br>800-1200) Vascular Filter<br>(Notch filter 530-650<br>& 900-1200) | The IPL handpiece<br>comes with 10 filters for<br>various wavelengths: Cut-off filters: 515,<br>560, 590, 615, 640,<br>695, 755 nm Acne Filter (Notch<br>filter 400-600 and<br>800-1200) Vascular Filter<br>(Notch filter 530-650<br>& 900-1200) KTP filter (525-585<br>nm) | Same as predicate<br>but without the<br>KTP filter |
| SapphireCool<br>LightGuides | Height: 55 mm | Height: 40 mm | Substantially<br>equivalent – minor<br>modification to<br>dimension |
| Spot sizes (cm²) | 8x15 mm (1.20 cm2)<br>15x35 mm (5.25 cm2)<br>6 mm round (0.3 cm2) | 8x15 mm (1.20 cm2)<br>15x35 mm (5.25 cm2)<br>6 mm round (0.3 cm2) | Same as predicate |
| Maximum Fluence | Up to 35 or 56 J/cm²,<br>upon tip size | Up to 35 or 56 J/cm²,<br>upon tip size | Same as predicate |
| Pulse Rate [Hz] | Up to 1 Hz | Up to 1 Hz | Same as predicate |
| Multiple Sequential<br>Pulsing | Yes, 1, 2, and 3 pulses,<br>varying fluence per<br>pulse (AOPT mode) | Yes, 1, 2, and 3 pulses,<br>varying fluence per<br>pulse (AOPT mode) | Same as predicate |
| Contact sapphire<br>cooling temperature | 0 to10 °C (Rectangular)<br>0 to 15°C (Round) | 4 to 10°C | Substantially<br>Equivalent - |
| | | | |
| Treatment screens | Treatment present with<br>additional basic<br>treatment presets | Treatment present | Substantially<br>equivalent - Minor<br>software update to<br>platform |
| LightGuide<br>recognition | Yes | N/A | Substantially<br>Equivalent -<br>Addition of minor<br>Hardware and<br>software features<br>to recognize<br>connected fibers |
| Multi-Spot Nd:YAG Handpiece | | | |
| Operational<br>Wavelength | 1064 nm | 1064 nm | Same as predicate |
| Spot sizes (mm) | 2x4, 6, 9 | 2x4, 1.5, 6, 9 | Substantially<br>Equivalent |
| Maximum Fluence | Up to 225 J/cm² upon tip<br>size (due to removal of<br>the 1.5 mm tip) | Up to 600 J/cm² upon tip<br>size | Substantially<br>Equivalent |
| Pulse rate | Up to 1 Hz | Up to 1 Hz | Same as predicate |
| Multiple Sequential<br>Pulsing | 1, 2, and 3 | 1, 2, and 3 | Same as predicate |
| Contact sapphire<br>cooling temperature | -4 to 10 °C | 4 to 10°C | Substantially<br>Equivalent |
| ResurFX Handpiece | | | |
| Operational<br>Wavelength | 1565 nm | 1565 nm | Same as predicate |
| Maximum Energy | Up to 40 or 70 mJ per<br>micro-beam, upon tip | Up to 40 or 70 mJ per<br>micro-beam, upon tip | Same as predicate |
| Type of Laser | Er:Glass Fiber-laser with<br>scanner | Er:Glass Fiber-laser with<br>scanner | Same as predicate |
| Tip treatment width | 18 mm Sapphire Cool<br>Tip<br>18 mm Precision Tip | 18 mm Sapphire Cool<br>Tip<br>18 mm Precision Tip | Same as predicate |
| Scanner | Dual axis scanner | Dual axis scanner | Same as predicate |
| Scanner shapes | Line, square, rectangle,<br>circle, donut, hexagon,<br>vertical line, and vertical<br>rectangle | Line, square, rectangle,<br>circle, donut, hexagon,<br>vertical line, and vertical<br>rectangle | Same as predicate |
| Treatment screens | Treatment present with<br>additional basic<br>treatment presets | Treatment present | Substantially<br>equivalent - Minor<br>software update to<br>platform |
| Q-Switched Nd:YAG Handpiece | | | |
| Operational<br>Wavelength | 1064 nm | 1064 nm | Same as predicate |
| Spot sizes (mm, diameter) | Disposable tips: 2.0, 2.5, 3.5 and 5.0 mm<br>Gold plated metal tips:<br>2.0, 2.5, 3.5, 4.0, 5.0, 6.0 and 8.0 mm | Disposable tips: 2.0, 2.5, 3.5 and 5.0 mm<br>Gold plated metal tips:<br>2.0, 2.5, 3.5, 4.0, 5.0, 6.0 and 8.0 mm | Same as predicate |
| Maximum Fluence | Up to 14 J/cm², upon tip size | Up to 14 J/cm², upon tip size | Same as predicate |
| Pulse Duration (nsec) | 6-8 | 6-8 | Same as predicate |
| Pulse Rate (Hz) | 0.5-5.0 | 0.5-5.0 | Same as predicate |
| Software | | | |
| Remote access capabilities | Yes | N/A | Substantially<br>Equivalent -<br>Addition of minor<br>software features<br>for internal<br>company use |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
# Performance Data
The following tests were performed to validate the modifications to the device:
- Design verification and validation testing .
- Risk analysis activities in compliance with ISO 14971 ●
- Electrical safety in accordance with IEC 60601-1 and EMC in accordance with IEC 60601-1-2
- . IPL compatibility testing as required to conform with IEC 60601-2-57
- . Software verification and validation
### Conclusions
Stellar M22 for Intense Pulsed Light (IPL) and Laser system has the same intended use and similar indications, principles of operation, and technological characteristics as M22 System. The minor differences in the Stellar M22 for Intense Pulsed Light (IPL) and Laser system's technological characteristics do not raise any new questions of safety or effectiveness. Performance data demonstrates that the Stellar M22 for Intense Pulsed Light (IPL) and Laser system is as safe and effective as M22 System. Thus, the Stellar M22 for Intense Pulsed Light (IPL) and Laser system is substantially equivalent to its predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.