K191962 · Quantel Medical · GEX · Dec 5, 2019 · General, Plastic Surgery
Device Facts
Record ID
K191962
Device Name
Vitra 2
Applicant
Quantel Medical
Product Code
GEX · General, Plastic Surgery
Decision Date
Dec 5, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Vitra 2 is for use in the treatment of ocular pathology of anterior and posterior segments including retinal photocoagulation and pan retinal photocoagulation of vascular and structural abnormalities of the retina and choroids including: - Proliferative and nonproliferative diabetic retinopathy - Choroidal neovascularization - Branch retinal vein occlusion - Treatment of choroidal neovascularization associated with wet Age-related macular degeneration - Retinal tears and detachments - Macular edema
Device Story
Vitra 2 is a 532 nm frequency-doubled Nd:YAG laser system for ophthalmic photocoagulation. It delivers laser energy to ocular tissues via slit lamp, indirect ophthalmoscope, operating microscope, or endocular probe. A scanning laser delivery system allows for automated delivery of predetermined spot patterns (square, circle, triple arcs). The device is operated by trained medical personnel in a clinical setting. The user initiates treatment via footswitch, with the ability to abort at any time. The system includes a resume function. By photocoagulating vascular and structural abnormalities, the device treats conditions like diabetic retinopathy and macular edema, potentially preserving or improving patient vision.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including compliance with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-1-6 (usability), IEC 60601-2-22 (laser safety), and IEC 60825-1 (laser product safety), alongside hardware and software validation activities.
Technological Characteristics
Frequency-doubled Nd:YAG laser; 532 nm wavelength; 1500 mW max power; 10 ms to continuous pulse duration; 635-650 nm aiming beam; Peltier cooling. Delivery via fiber optic to slit lamp, indirect ophthalmoscope, operating microscope, or endocular probe. Includes scanning motor system for pattern delivery. Complies with IEC 60601-1, 60601-1-2, 60601-1-6, 60601-2-22, and IEC 60825-1.
Indications for Use
Indicated for patients with ocular pathology of anterior and posterior segments, including proliferative/nonproliferative diabetic retinopathy, choroidal neovascularization (including wet AMD), branch retinal vein occlusion, retinal tears/detachments, and macular edema.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 5, 2019
Quantel Medical % Maureen O'connell President O'Connell Regulatory Consultants, Inc. 44 Oak Street Stoneham, MA 02180
Re: K191962
Trade/Device Name: Vitra 2 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument for Use in General and Plastic Surgery and in Dermatology Regulatory Class: Class II Product Code: GEX Dated: October 31, 2019 Received: November 4, 2019
Dear Maureen O'Connell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general control's provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidanceregulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-devicesafety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally. you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-andconsumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Tieuvi Nguyen, Ph.D. Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191962
Device Name Vitra 2®
Indications for Use (Describe)
The Vitra 2 is for use in the treatment of ocular pathology of anterior segments including retinal photocoagulation and pan retinal photocoagulation of vascular and structural abnormalities of the retina and choroids including:
- · Proliferative and nonproliferative diabetic retinopathy
- Choroidal neovascularization
- · Branch retinal vein occlusion
- · Treatment of choroidal neovascularization associated with wet Age-related macular degeneration
- · Retinal tears and detachments
- · Macular edema
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) SUMMARY K191962
## Quantel Medical Vitra 2®
### 510(k) Owner
Quantel Medical 11 rue Bois Joli CS40015 63808 Cournon D'Auvergne-Cedex France 33 04 73 745 732 Phone: Contact Person: Bruno Pagès
### Submission Correspondent:
Maureen O'Connell O'Connell Regulatory Consultants, Inc. 44 Oak Street Stoneham, MA 02180 Phone: 978-207-1245
Date Prepared: November 25, 2019
## Trade Name of Device
Vitra 2 ®
### Common or Usual Name
Ophthalmic Lasers
### Classification Name
Lasers, Ophthalmic; 21 C.F.R. 878.4810 Class II Product Code: GEX
### Predicate Device
Quantel Medical Vitra Multispot (K122251)
### Device Description
The Vitra 2 is a laser system which emits a treatment beam at 532 nm and is intended for use in photocoagulation of ocular tissues in the treatment of diseases of the eye.Laser energy is delivered to opaque structures within the eye by means of delivery systems including slit lamp, indirect ophthalmoscope, operating microscope and endocular probe. The standard delivery system includes a lens system to focus the laser energy and vary the size of the laser
{4}------------------------------------------------
spot in the plane of observation of the slit lamp, for example. It also includes a Multispot delivery system made by scanner motors which can deliver several spots to the retina in a pattern. The laser energy is delivery system by the means of a flexible fiber optic. For most procedures, a laser contact lens is used to direct the laser energy to the part of the eye being treated. The contact lens also helps to hold the eye open and still so that the laser energy can be delivered effectively.
The Quantel Medical VITRA 2 includes a Scanning Laser Delivery System adaptor with scanner controls that may be coupled to a slit lamp. Once activated by the user, the VITRA 2 delivers a predetermined pattern by sequentially scanning the placement of the laser spots and the emission of the individual pulses of laser light. Treatment is initiated by pressing the footswitch and may be aborted by releasing the footswitch.
## Intended Use / Indications for Use
The Vitra 2 is for use in the treatment of ocular pathology of anterior and posterior segments including retinal photocoagulation and pan retinal photocoagulation of vascular and structural abnormalities of the retina and choroids including:
- · Proliferative and nonproliferative diabetic retinopathy
- · Choroidal neovascularization
- · Branch retinal vein occlusion
- · Treatment of choroidal neovascularization associated with wet Age-related macular degeneration
- · Retinal tears and detachments
- Macular edema
## Substantial Equivalence
Ouantel Medical believes that the Vitra 2 described in this notification and for use under the conditions of the proposed labeling is substantially equivalent to a legally marketed predicate device that is a Class II medical device regulated under product code GEX (21 C.F.R. 878.4810) . The predicate device is the Quantel Medical Vitra Multispot laser system cleared by FDA in K122251 and intended for use in photocoagulation of the posterior and anterior segments of the eye.
Both the Vitra 2 and the Vitra Multispot devices are intended for use in photocoagulation of the posterior and anterior segments. Additionally, both devices have the identical indications for use statement. Both devices are prescription devices which are intended to be used by trained medical personnel.
As shown in the table below, the Vitra 2 and the Vitra Multispot have the same technological characteristics. Both devices are frequency doubled Nd: YAG lasers with a wavelength of 532 nm. The maximum power at the tissue is 1500 mW for both systems with a pulse duration of 10 ms to continuous for both systems and the identical pulse repeat interval. Both the Vitra 2 and the Vitra Multispot use an aiming laser with a wavelength of 635-650 nm. Both systems are cooled by Peltier effect.
{5}------------------------------------------------
For the Single Spot Slit Lamp Adapters, the emission modes available in both the Vitra 2 and the Vitra Multispot are single, repeat, painting and continuous with an identical spot size of 50 to 500 microns. For the MultiSpot Slit Lamp Adapters in the Multispot Mode, the emission modes available in both the Vitra 2 and the Vitra Multispot are square, circle, triple arcs and single spot. For the MultiSpot Slit Lamp Adapters in the Monospot Mode, the spot modes available in both the Vitra 2 and the Vitra Multispot are single, repeat, painting and continuous. For the MultiSpot Slit Lamp Adapters, both the Vitra Multispot are compatible with the same slit lamps and laser systems except the Vitra 2 is also compatible with the CSO 9800. Both systems can be used with the Keeler Vantage Plus laser indirect ophthalmoscope. The Vitra 2 can also be used with a commercially available sterile laser probe.
Therefore, as the Vitra 2 and the Vitra Multispot have the same intended use and similar technological characteristics, the Vitra 2 is substantially equivalent to the Vitra Multispot.
| SPECIFICATIONS | | |
|------------------------------------------------|---------------------------------------------------------------------------------|--------------------------------------------------------------|
| MANUFACTURER | QUANTEL MEDICAL | QUANTEL MEDICAL |
| 510(K) | - | K122251 |
| TYPE OF DELIVERY SYSTEM | Slit lamp<br>Indirect ophthalmoscope<br>Endocular probe<br>Operating microscope | Slit lamp<br>Indirect ophthalmoscope<br>Operating microscope |
| PULSING SYSTEM | Continuous | Continuous |
| LASER SOURCE | Frequency doubled Nd:YAG | Frequency doubled Nd:YAG |
| WAVELENGTH | 532 nm | 532 nm |
| MAXIMUM POWER | 1500mW (at tissue) | 1500mW (at tissue) |
| PULSE DURATION | 10 ms to continuous | 10 ms to continuous |
| REPEAT INTERVAL | 0.1 - 0.2 - 0.3 - 0.5 - 0.7s | 0.1 - 0.2 - 0.3 - 0.5 - 0.7s |
| AIMING BEAM WAVELENGTH | 635 – 650 nm | 635 – 650 nm |
| Delivery System-Single Spot Slit Lamp Adapters | | |
| SPOT MODE | Single, repeat, painting, continuous | Single, repeat, painting, continuous |
| SPOT SIZE | 50 µm to 500 µm | 50 µm to 500 µm |
| Delivery System-MultiSpot Slit Lamp Adapters | | |
| MONOSPOT MODE | | |
| EMISSION MODES | Single, repeat, painting, continuous | Single, repeat, painting, continuous |
| SPOT SIZE | 50 µm to 400 µm | 50 µm to 500 µm |
| MULTISPOT MODE | | |
| EMISSION MODES | Square, circle, triple arcs, single spot | Square, circle, triple arcs, single spot |
| SPOT SIZE | 100 µm to 400 µm | 100 µm to 500 µm |
| RESUME FUNCTION | Yes | Yes |
{6}------------------------------------------------
## Performance Data
Performance testing was conducted in order to demonstrate compliance with recognized consensus standards and to demonstrate substantial equivalence:
- IEC 60601-1:2005 + Corr. 1:2006+Corr. 2:2007+A1:2012 Medical electrical ● equipment-Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2:2014: Medical electrical equipment-Part 1-2: General requirements for ● basic safety and essential performance-Collateral standard: Electromagnetic compatibility-Requirements and tests
- IEC 60601-1-6:2010+A1:2013 Medical electrical equipment-Part 1-6: General . requirements for safety-Collateral Standard Usability
- . IEC 60601-2-22: 2007 (Third Edition) + A1:2012 Medical Electrical Equipment Part 2: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
- IEC 60825-1 Safety of Laser Products-Part 1: Equipment Classification and . Requirements
Additionally, hardware and software validation activities were performed to ensure the device performed as intended and software documentation appropriate for the level of concern was provided.
In summary, the performance testing provided shows that the device performs as intended and is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.