K190936 · Shanghai Apolo Medical Technology Co., Ltd. · GEX · Jun 17, 2019 · General, Plastic Surgery
Device Facts
Record ID
K190936
Device Name
Q-Switched Nd: YAG Laser System
Applicant
Shanghai Apolo Medical Technology Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jun 17, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Q-Switched Nd: YAG Laser System is intended for use in tattoo removal, treatment of benign vascular lesions, treatment of benign pigmented lesions, incision, ablation, vaporization of soft tissue for general dermatology as follows: 532nm wavelength (nominal delivered energy of 585nm and 650mm with optional dye handpiece): - Removal of light ink (red, sky blue, green, purple, and orange) tattoo - Treatment of benign vascular lesions including, but not limited to: telangiectasias, - Treatment of benign epidermal pigmented lesions including, but not limited to: cafe-au-lait, solar lentiginos, senile lentiginos, Becher's, nevi Freckles, Nevus spilus, Seborrheic Keratoses - Treatment of Post Inflammatory Hyper-Pigmentation 1064nm wavelength: - Removal dark ink (black, blue and brown) tattoo - Removal of benign dermal pigmented lesions including, but not limited to: Nevus of OTA, Common Nevi, and Melasma. - Removal or lightening of unwanted hair with or without adjuvant preparation - Skin resurfacing procedures for the treatment of acne scars and wrinkles.
Device Story
Q-Switched Nd:YAG laser system; generates pulsed coherent light at 1064nm and 532nm (expandable to 585nm/650nm via dye handpieces). System includes laser generator, articulated arm, power supply, cooling system, and control interface. Operated by clinicians in dermatology settings. Laser beam directed via articulated arm to target tissue for ablation, vaporization, or pigment/ink destruction. Output affects tissue via photothermal/photomechanical interaction; aids in tattoo removal, lesion treatment, and skin resurfacing. Benefits include non-invasive tissue modification and aesthetic improvement.
Clinical Evidence
No clinical data; bench testing only. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, and IEC 60825-1 verified design specifications and safety.
Indicated for patients requiring tattoo removal (light/dark ink), treatment of benign vascular lesions (e.g., telangiectasias), benign epidermal/dermal pigmented lesions (e.g., cafe-au-lait, Nevus of OTA, Melasma), hair removal/lightening, and skin resurfacing for acne scars/wrinkles.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
SPECTRA Q-Switched Nd:YAG Laser System with Dye Handpieces (K113588)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and name on the right. The Department of Health & Human Services logo is a stylized depiction of a human figure, while the FDA acronym and name are written in blue, with the word "ADMINISTRATION" in a smaller font size below "U.S. FOOD & DRUG".
Shanghai Apolo Medical Technology Co., Ltd. % Felix Li Regulatory Affairs 4F, Building A, No. 388 Yindu Road Xuhui District Shanghai, 200231 China
June 17, 2019
Re: K190936
Trade/Device Name: Q-Switched Nd: Y AG Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument for use in General and Plastic Surgery and in Dermatology Regulatory Class: Class II Product Code: GEX Dated: April 4, 2019 Received: April 10, 2019
Dear Felix Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Neil R.P. Ogden, MS Acting Assistant Director, THT4A3: Light Based Devices Team DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K190936
Device Name
Q-Switched Nd: YAG Laser Systems
#### Indications for Use (Describe)
The Q-Switched Nd: Y AG Laser System is intended for use in tattoo removal, treatment of benign vascular lesions, treatment of benign pigmented lesion, incision, ablation, vaporization of soft tissue for general dermatology as follows:
532nm wavelength (nominal delivered energy of 585nm and 650mm with optional dye handpiece):
- Removal of light ink (red, sky blue, green, purple, and orange) tattoo
- Treatment of benign vascular lesions including, but not limited to: telangiectasias,
- Treatment of benign epidermal pigmented lesions including, but not limited to: cafe-au-lait, solar lentiginos, senile lentiginos, Becher's, nevi Freckles, Nevus spilus, Seborrheic Keratoses
- Treatment of Post Inflammatory Hyper-Pigmentation
1064nm wavelength:
- Removal dark ink (black, blue and brown) tattoo
- Removal of benign dermal pigmented lesions including, but not limited to: Nevus of OTA, Common Nevi, and Melasma.
- Removal or lightening of unwanted hair with or without adjuvant preparation
- Skin resurfacing procedures for the treatment of acne scars and wrinkles.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
## I Submitter
Shanghai Apolo Medical Technology Co., Ltd. 4F, Building A, No.388, Yindu Road, Xuhui District, Shanghai 200231,China
Establishment Registration Number: 3007120647
Contact person: Felix Li Position: Regulatory Affairs Phone: +86-138 4919 0618 Fax: +86-21-34622840 E-mail: liqiang@apolo.com.cn
# II Proposed Device
| Trade Name of Device: | Q-Switched Nd: YAG Laser System |
|-----------------------|-----------------------------------|
| Common name: | Powered Laser Surgical Instrument |
| Regulation Number: | 21 CFR 878.4810 |
| Regulatory Class: | Class II |
| Product code: | GEX |
| Review Panel: | General & Plastic Surgery |
# III Predicate Devices
| 510(k) Number: | K113588 |
|-----------------|---------------------------------------------------------------|
| Trade name: | SPECTRA Q-Switched Nd:YAG Laser<br>System with Dye Handpieces |
| Common name: | Powered Laser Surgical Instrument |
| Classification: | Class II |
| Product Code: | GEX |
| Manufacturer | Lutronic Corporation |
### IV Device description
The Q-Switched Nd: YAG Laser System is based on the Q-Switch Nd: YAG and frequency-double Nd:YAG laser technology. The system is composed of laser generator,
{4}------------------------------------------------
articulated arm, laser power supply, cooling system, display and control system,foot switch and others accessories.
The Q-Switched Nd: YAG Laser System produces a pulsed beam of coherent near infrared (1064nm) and visible (532nm) light. This beam is directed to the treatment zone by means of an articulated arm coupled to a handpiece. In addition, two dye handpieces are available that concert the 532nm wavelength to 585nm and 650nm.
The Q-Switched Nd: YAG Laser System includes two models: HS-290K and HS-290EK.
### V Indication for use
The Q-Switched Nd: YAG Laser System is intended for use in tattoo removal, treatment of benign vascular lesions, treatment of benign pigmented lesions, incision, ablation, vaporization of soft tissue for general dermatology as follows:
# 532nm wavelength (nominal delivered energy of 585nm and 650nm with optional dye handpiece):
- · Removal of light ink (red. sky blue, green, purple, and orange) tattoo,
- · Treatment of benign vascular lesions including, but not limited to: telangiectasias,
- · Treatment of benign epidermal pigmented lesions including. but not limited to: cafe-au-lait, solar lentiginos, senile lentiginos, Becher's, nevi Freckles, Nevus spilus, Seborrheic Keratoses,
- · Treatment of Post Inflammatory Hyper-Pigmentation.
### 1064nm wavelength:
- · Removal dark ink (black, blue and brown) tattoo,
- · Removal of benign dermal pigmented lesions including, but not limited to: Nevus of OTA, Common Nevi, and Melasma,
- · Removal or lightening of unwanted hair with or without adjuvant preparation,
- · Skin resurfacing procedures for the treatment of acne scars and wrinkles.
#### VI Comparison of technological characteristics with the predicate devices
A table comparing the key features of the proposed and predicate devices is provided below.
| Item | Subject device | Predicate device<br>(K113588) |
|--------------|---------------------------------|-----------------------------------------------------------------|
| Product name | Q-Switched Nd: YAG Laser System | SPECTRA Q-Switched<br>Nd:YAG Laser System with<br>Dye Handpiece |
| Product Code | GEX | GEX |
{5}------------------------------------------------
| Regulation<br>No. | 21 CFR 878.4810 | 21 CFR 878.4810 |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Class | Class II | Class II |
| Indication for<br>use | The Q-Switched Nd: YAG Laser<br>System is intended for use in tattoo<br>removal, treatment of benign<br>vascular lesions, treatment of benign<br>pigmented lesions, incision, excision,<br>ablation, vaporization of soft tissue<br>for general dermatology as follows: | The SPECTRA Laser System is<br>indicated for the Incision,<br>excision, ablation, vaporization<br>of soft tissues for general<br>dermatology, dermatologic and<br>general surgical procedure for<br>coagulation and hemostasis. |
| | 532nm wavelength (nominal | |
| | delivered energy of 585nm and | 532nm Wavelength (normal |
| | 650nm with optional dye | delivered energy of 585nm and |
| | handpiece): | 650nm with optional dye<br>handpieces): |
| | Removal of light ink (red, sky<br>blue, green, purple, and orange)<br>tattoo Treatment of benign vascular<br>lesions including, but not limited<br>to: telangiectasias, Treatment of benign epidermal<br>pigmented lesions including, but<br>not limited to: cafe-au-lait, solar<br>lentiginos, senile lentiginos,<br>Becher's, nevi Freckles, Nevus<br>spilus, Seborrheic Keratoses Treatment of Post Inflammatory<br>Hyper-Pigmentation | Tattoo removal: light ink<br>(red, tan, purple, orange,<br>sky blue, green) Removal of Epidermal<br>Pigment Lesions Removal of Minor<br>Vascular Lesions including<br>but not limited to<br>telangiectasias Treatment of Lentigines Treatment of Café-Au-lait Treatment of Post<br>Inflammatory<br>Hyper-Pigmentation Treatment of Becker's<br>Nevi, Freckles and Nevi<br>Spilus |
| | 1064nm wavelength: | 1064nm Wavelength |
| | Removal dark ink (black, blue<br>and brown) tattoo Removal of benign dermal<br>pigmented lesions including, but<br>not limited to: Nevus of OTA,<br>Common Nevi, and Melasma. | Tattoo removal: dark ink<br>(black, blue and brown) Removal of Epidermal<br>Pigment Lesions Removal of Nevus of Ota |
| | Removal or lightening of unwanted hair with or without adjuvant preparation Skin resurfacing procedures for the treatment of acne scars and wrinkles | Removal or lightening of unwanted hair with or without adjuvant preparation Treatment of Common Nevi Skin resurfacing procedure for the treatment of acne scars and wrinkle Treatment of melasma. |
| Anatomical site | Skin and subcutaneous tissue | Skin and subcutaneous tissue |
| Technology | Q-Switched Nd: YAG amd KTP Nd:YAG Laser | Q-Switched Nd: YAG amd KTP Nd:YAG Laser |
| Pump lamp source | Xenon lamp | Xenon lamp |
| Wavelength (nm) | 1064nm/532nm | 1064nm/532nm |
| Aiming beam wavelength | 650nm | 655nm |
| Laser output mode | Q-Switched pulse | Q-Switched pulse |
| Maximum pulse energy | 1064nm wavelength: 1200mJ;<br>532nm wavelength:500mJ<br>SPT mode: 1400mJ | 1064nm wavelength: 1200mJ;<br>532nm wavelength:400mJ<br>SPT mode: 1500mJ |
| Pulse duration<br>Repetition rate | 4~6ns (Q-Switched mode) /300μs (SPT mode)<br>Max,10Hz | 5~10nsQ-Switched mode )/300µs (Spectra mode)<br>Max,10Hz |
| Nominal ocular hazard distance | 2580m | unknown |
| Spot size | Adjustable spot size 1~10mm | 3, 4, 5, 6, 7, 8mm/1, 2, 3, 4, 5, 6, 7mm (option) |
| Beam delivery | Articulated arm light guide | Articulated arm light guide |
| Cooling system | Water cooling, forced-air cooling, Copper Radiator | Closed internal circulating water cooling, outer circulation |
{6}------------------------------------------------
{7}------------------------------------------------
| | | strong wind cooling |
|----------------------------|-----------------------------------------------------------------------------------|-----------------------------|
| System<br>dimension | HS-290K:<br>11.8"(W)×32.7"(L)×34.6"(H)<br>HS-290EK:<br>11.8"(W)×29.5"(L)×34.6"(H) | 11.6"(W)×25.8"(L)×66.93"(H) |
| System<br>weight | HS-290K: 176.4lbs<br>HS-290EK: 167.6 lbs | 194lbs |
| Electrical<br>requirements | AC 110V, 50/60Hz | AC 220-230V, 50/60Hz |
| Maximum<br>power | 200MW | 240MW |
## VII Non-Clinical Testing
A battery of tests were performed to verity that the proposed device met all design specification. The test result demonstrated that the proposed device complies with the following standards:
## Electrical safety and electromagnetic compatibility
IEC 60601-1: 2005+corr.1:2006+Corr.2.2007+A1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
IEC 60601-1-2:2014 Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance-Collateral Standard: Electromagnetic disturbances -Requirements and tests
IEC 60601-2-22:2013 Medical electrical equipment - Part 2-22: Particular requirements for the safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
IEC 60825-1:2007 Safety of Laser products-Part 1:Equipment classification and requirements
# VIII Clinical Testing
It is not applicable.
# IX Conclusion
The proposed device, Q-Switched Nd: YAG Laser System has the intended use as the predicate device. It presents similar technological characteristics as the predicate device including the laser type, wavelengths, device design, pulse width, frequency, spot sizes. The non-clinical testing demonstrates that the device is as safe, as effective and performs as well as the legally marketed device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.