K181019 · Guangzhou Huafei Tongda Technology Co., Ltd. · GEX · Jul 16, 2018 · General, Plastic Surgery
Device Facts
Record ID
K181019
Device Name
Diode Laser System
Applicant
Guangzhou Huafei Tongda Technology Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jul 16, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Diode Laser System is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
Device Story
Diode Laser System (HF-108) is a surgical laser device for hair removal. Input: operator-selected laser energy parameters via control panel. Operation: microcontroller-based system manages constant current board to drive semiconductor laser; cooling system (condenser, cold plate, fans) maintains temperature/humidity; handpiece (sapphire tip) emits 808 nm laser energy. Output: laser energy delivered to patient skin. Used in clinical settings by trained operators. Healthcare provider uses LCD interface to set parameters and monitor system state (temperature, humidity, flow). Device provides permanent hair reduction by targeting hair follicles; cooling components protect skin and prevent device overheating.
Clinical Evidence
Bench testing only. No clinical data included in this submission.
Technological Characteristics
Diode laser (808 nm, Class IV). Handpiece materials: sapphire and aluminum alloy. Cooling: condenser, cold plate, water circulation, fans. Connectivity: standalone. Standards: IEC 60601-1, IEC 60601-2-22, IEC 60825-1, IEC 60601-1-2, ISO 10993-5, ISO 10993-10. Microcontroller-based control system for energy delivery and safety monitoring (over-current, over-voltage, over-temperature).
Indications for Use
Indicated for hair removal and permanent hair reduction on all skin types (Fitzpatrick I-VI), including tanned skin.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.
July 16, 2018
Guangzhou Huafei Tongda Technology Co., Ltd. % Ray Wang General Manager Beijing Believe-Med Technology Services Co., Ltd. Rm. 912, Building #15, Xi Yue Hui, No.5, Yi He North Rd. Fang Shan District Beijing, 102401 Cn
Re: K181019
Trade/Device Name: Diode Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: April 13, 2018 Received: April 17, 2018
Dear Ray Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entiled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely. Jennifer R. Stevenson -53 For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K181019
Device Name
Diode Laser System
#### Indications for Use (Describe)
The Diode Laser System is intended for hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
Type of Use (Select one or both, as applicable)
| <span> <svg class="bi bi-check" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M13.854 3.646a.5.5 0 0 1 0 .708l-7 7a.5.5 0 0 1-.708 0l-3.5-3.5a.5.5 0 1 1 .708-.708L6.5 10.293l6.646-6.647a.5.5 0 0 1 .708 0z" fill-rule="evenodd"></path> </svg> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Tab #7 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: K181019
- 1. Date of Preparation
06/19/2018
- Sponsor 2.
### Guangzhou Huafei Tongda Technology Co., Ltd.
4/F, Building B Industrial and Commercial Trading Park, Hainan Town, Huadi Dadao Nan Road, Fangcun, Liwan District, Guangzhou 510388, P.R. China
Contact Person: Liu Pengqi Position: General Manager Tel: +86-13302285919 Fax: +86-20-81653065 Email: 179330082@qq.com
- 3. Submission Correspondent
Mr. Ray Wang
Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, BeiJing, China 102401
Tel: +86-18910677558 Fax: +86-10-56335780 Email: ray.wang@believe-med.com
{4}------------------------------------------------
#### 4. Identification of Proposed Device
Trade Name: Diode Laser System Common Name: Powered Laser Surgical Instrument Model(s): HF-108
Regulatory Information: Classification Name: Powered Laser Surgical Instrument Classification: II; Product Code: GEX ; Regulation Number: 21 CFR 878.4810; Review Panel: General & Plastic Surgery:
#### Intended Use:
The Diode Laser System is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin.
Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
#### 5. Device Description
The proposed device, Diode Laser System, is a surgical device, which is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI);
#### Function module description
a. Control Panel
The module uses the microcontroller as the heart, utilizes the LCD screen to display all prompt information and the system state information to complete the human-machine interaction function, and realizes the device parameters settings and accurate control of the output laser energy by the operator.
#### b. Main Control Module
The module uses the microcontroller as the heart, receives the laser energy parameters and work instructions from the control panel and detects the footswitch state; Utilizes the sensors of temperature, humidity, liquid level and flow to detect the parameters such as temperature, humidity and water flow during system working, and according to the detected values to calculate the dew point temperature; Controls and detects the work state of constant current board module as well as the temperature and humidity control system; Uploads the state data and alarm information of water circulation system, cooling system, handpiece module and constant current board module during system working.
c. Constant current board module
{5}------------------------------------------------
The module uses the high-power MOS as the heart, receives the laser energy parameters from the main control module, supplies the semiconductor laser with constant drive current which corresponding to the received laser energy parameters to drive the semiconductor laser to emit light. The module also has the detection function of over-current, overvoltage, over-temperature and handpiece state, and uploads the detected data to the control module.
#### d. Temperature and humidity control system
The system mainly includes the condenser, cold plate, water circulation subsystem and fans. The microcontroller of main control module according to the temperature, humidity parameter detected by the sensors to control the working state of the condenser, cold plate and cooling fan to meet the temperature and humidity requirements during the semiconductor laser working.
#### e. Handpiece module
Handpiece module is the heart of the device, which is the execution unit of the device and completes the laser emission function. The module is mainly composed of semiconductor laser, sapphire, temperature and humidity sensor, data storage chips, cooling components and water flow path. The semiconductor laser emits light to output energy, temperature and humidity sensors detects the temperature and humidity parameters during handpiece working, the cooling components and water flow path take away the heat of the semiconductor laser to prevent it from being damaged caused by over-temperature, so prolongs the service life of the semiconductor laser.
{6}------------------------------------------------
#### 6. Identification of Predicate Device
510(k) Number: K162659 Product Name: Diode Laser Hair Removal System Manufacturer: Shandong Huamei Technology Co.,ltd.
510(k) Number: K142845 Product Name: SILKPRO Laser Hair Removal System Manufacturer: Wuhan Lotuxs Technology Co., Ltd.
510(k) Number: K112031 Product Name: Alma Lasers Modified Diode Laser Module with SHR Treatment Mode for use with Family of Soprano XL Multi-Application Platforms Manufacturer: Alma Lasers, Inc.
510(k) Number: K153718 Product Name: Spirit Hair Removal Laser Family. Spirit-918/cFactor -918/mFactor-918 Manufacturer: Active Optical Systems Ltd.
#### 7. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- A IEC 60601-1:2012 Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance:
- > IEC 60601-2-22 Edition 3.1 2012-10, Medical Electrical Equipment - Part 2-22: Particular Requirements For Basic Safety And Essential Performance Of Surgical, Cosmetic, Therapeutic And Diagnostic Laser Equipment.
- > IEC 60825-1:2014, Safety of laser products - Part 1: Equipment classification and requirements.
- > IEC 60601-1-2:2014 , Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests
- > ISO 10993-5 Third Edition 2009-06-01, Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity. (Biocompatibility)
- > ISO 10993-10 Third Edition 2010-08-01, Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization. (Biocompatibility)
- > Performance Testing for Spot Size Accuracy and Energy Output Accuracy.
- 8. Clinical Test Conclusion
{7}------------------------------------------------
No clinical study is included in this submission.
{8}------------------------------------------------
# 9. Substantially Equivalent (SE) Comparison
| Item | Proposed Device | Predicate Device<br>K162659 | Predicate Device<br>K142845 | Predicate Device<br>K112031 | Predicate Device<br>K153718 | Remark |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Product Code | GEX | GEX | OHT | GEX | GEX | SE |
| Regulation<br>Number | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | SE |
| Intended Use | The Diode Laser System is<br>intended for hair removal,<br>permanent hair reduction on<br>all skin types (Fitzpatrick<br>skin type I-VI), including<br>tanned skin.<br>Permanent hair reduction is<br>defined as the long-term,<br>stable reduction in the<br>number of hairs regrowing<br>when measured at 6, 9, and<br>12 months after the<br>completion of a treatment<br>regime. | The Diode Laser System is<br>intended for hair removal,<br>permanent hair reduction on<br>all skin types (Fitzpatrick skin<br>type I-VI), including tanned<br>skin.<br>Permanent hair reduction is<br>defined as the long-term,<br>stable reduction in the number<br>of hairs regrowing when<br>measured at 6, 9, and 12<br>months after the completion of<br>a treatment regime. | The SILKPRO is an<br>over-the-counter device intended<br>for adjunctive use with shaving<br>for hair removal sustained with<br>periodic treatments. SILKPRO is<br>also intended for permanent<br>reduction in hair regrowth<br>defined as a long-term, stable<br>reduction in hair counts<br>following a treatment regime.<br>permanent hair reduction is<br>defined as long term, stable<br>reduction in the number of hairs<br>regrowing when measured out to<br>6, 9, and 12 months after the<br>completion of the treatment<br>regimen | The Alma Lasers Family of Soprano<br>XL Multi-Application Platforms is<br>intended for use in dermatologic and<br>general surgical procedures.<br>The Alma Lasers Modified Diode<br>Laser Module (used with Alma<br>Lasers Family of Soprano XL<br>Multi-Application Platforms):<br>The HR modes is intended for hair<br>removal, permanent hair reduction.<br>The SHR Mode is intended for hair<br>removal, permanent hair reduction.<br>The LaserBlanche Mode is intended<br>for the treatment of benign vascular<br>and pigmented lesions.<br>The Alma Lasers Family of Soprano<br>XL Multi-Application Platforms is<br>intended for use on all skin types | The Spirit Hair Removal Laser<br>Family is generally intended for<br>dermatological use. The devices are<br>specifically indicated for hair<br>removal, permanent hair reduction<br>by using selective laser energy.<br>The Spirit Hair Removal Laser<br>Family is intended for use on all<br>skin types (Fitzpatrick skin types<br>I-VI), including tanned<br>skin.Permanent reduction in hair<br>regrowth is defined as long term,<br>stable reduction in the number of<br>hairs regrowing when measured out<br>to 6, 9, and 12 months after the<br>completion of the treatment regimen. | SE |
Table 7-1 General Comparison
{9}------------------------------------------------
| | | | (Fitzpatrick skin types I-VI),<br>including tanned skin. | | | |
|---------------------------|--------------|--------------|----------------------------------------------------------|--------------|-------------|----|
| | Main Unit | Main Unit | / | Main Unit | / | SE |
| Configuration | Handpiece | Handpiece | Handpiece | Handpiece | Handpiece | SE |
| | Foot Control | Foot Control | / | Foot Control | / | SE |
| Principle of<br>Operation | Diode Laser | Diode Laser | Diode Laser | Diode Laser | Diode Laser | SE |
{10}------------------------------------------------
| Item | Proposed Device | Predicate Device<br>K162659 | Predicate Device<br>K142845 | Predicate Device<br>K112031 | Predicate Device<br>K153718 | Remark |
|-------------------------|------------------------|-----------------------------|-----------------------------|-----------------------------------------------|-----------------------------------|------------|
| Laser Type | Diode Laser | Diode Laser | Diode Laser | Diode Laser | Diode Laser | SE |
| Laser<br>Classification | Class IV | Class IV | Class IV | Class IV | Class IV | SE |
| Laser<br>Wavelength | 808 nm | 808 nm | 810 nm | 810 nm | 810 nm | SE |
| Spot Size | 1.20cm² | 1.44 cm² | 9 x 9 mm (0.81 cm²) | 1.20cm² | 12 x 16 mm (1.92 cm²) (Model 918) | Discussion |
| Pulse<br>Duration | 30ms-200ms | 5-400ms | / | 5-200ms (HR Mode); ≤ 20ms (SHR Mode) | Up to 310 ms (Model 918) | Discussion |
| Fluence | 5-40 J/cm2 | 1-120J/ cm² | 5 - 25J/ cm2 | 1-120J/ cm² (HR Mode); ≤ 10 J/ cm² (SHR Mode) | 6-90 J/cm² (Model 918) | Discussion |
| Frequency | 1-5Hz | 0.5-15Hz | / | ≤ 3 Hz (HR Mode); ≤ 10 Hz (SHR Mode) | ≤ 10 Hz (Model 918) | Discussion |
| Power Supply | 100-240V~ 50/60Hz | AC 110V/60Hz | / | / | / | Discussion |
| Dimension | 510mm×600mm×<br>1000mm | 450mm×<br>550mm×380mm | / | / | / | Discussion |
| Weight | 50Kg | 52Kg | / | / | / | Discussion |
Table 7-2 Performance Comparison
Discussion
The proposed device has same indication for use with 4 predicate devices, such as spot size, fillerce, frequency range, pulse duration.
For these output parameters, the Fluence and Pulse the parameters which may decides that how much and how long time the energy will deliver to
{11}------------------------------------------------
510(k) Summary
the patient's skin, it may effects the safety (too long action time) may burn patient's skin, and effectiveness (too low energy and or too short action time) may make the device could not achieve it's indication for use.
From the comparison above, the fluence of proposed device is 5-40 June, the finence of predicate device are 1-1200 cm², They are not exactly same, but the fluence of proposed or limited the fluence of predicate device. Because the maximum fluence of proposed are lover than some predicate device (120 J/cm²), it can be considered as the proposed device has acceptable safety; and the maximum fluence of proposed are higher than some rredicate device (25// on) , 10 J/ cm²) it can be croposed device has capable to achieve it's indication for use. The Pulse duration of proposed device is 30 ms -200 ms, which is similar with the pulse duration of predicate device, which is longer or same than some predicate device (200 ms, 400 ms), and less than some predicate device (20 ms), so same reason as fluence, the pulse device has acceptable safety and has capable to achieve it's indication for use.
As the justification above, the difference of pulse ranges does not affect the safety or effectiveness of proposed device for it's indications for use.
{12}------------------------------------------------
# Table 7-3 Safety Comparison
| Item | Proposed Device | Predicate Device<br>K162659 | Predicate Device<br>K142845 | Predicate Device<br>K112031 | Predicate Device<br>K153718 | Remark |
|------------------------------------------------|---------------------------------------------|-----------------------------------------|-----------------------------|-----------------------------|-----------------------------|------------|
| Patient Contact Materials and Biocompatibility | | | | | | |
| Patient Contact<br>Materials | Sapphire and aluminum alloy in<br>handpiece | Sapphire in handpiece | | | | Discussion |
| Cytotoxicity | No Cytotoxicity | No Cytotoxicity | | | | |
| Sensitization | No evidence of sensitization | No evidence of sensitization | | | | SE |
| Irritation | No evidence of irritation | No evidence of irritation | | | | |
| EMC, Electrical and Laser Safety | | | | | | |
| Electrical Safety | Comply with IEC 60601-1, IEC<br>60601-2-22 | Comply with IEC 60601-1, IEC 60601-2-22 | | | | SE |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | | | | SE |
| Laser Safety | Comply with IEC 60601-2-22, IEC<br>60825 | Comply with IEC 60601-2-22, IEC 60825 | | | | SE |
# Discussion
The proposed device is different in Patient Contacte device. but both the predicate device and the proposed device has passed the 1SO 1099 series test, we believe these differences will not affectiveness and safety compared with the predicate is deternined to be substantially equivalency with predicate device.
{13}------------------------------------------------
## 10. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.