K180593 · Syneron-Candela · GEX · Jun 5, 2018 · General, Plastic Surgery
Device Facts
Record ID
K180593
Device Name
Vbeam Prima Laser System
Applicant
Syneron-Candela
Product Code
GEX · General, Plastic Surgery
Decision Date
Jun 5, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
1,064nm: The Vbeam Prima Laser System is intended for the coagulation and hemostasis of benign vascular lesions such as, but not limited to, port wine stains, hemangiomas, warts, telangiectasia, rosacea, venus lakes, leg veins, spider veins, and poikiloderma of civatte and treatment of benign cutaneous lesions such as, but not limited to lentigos (age spots), solar lentigos (sun spots), cafe au lait manules, seborrheic keratoses, nevi, chloasma, verrucea, skin tags, and keratoses. The laser is also indicated for the treatment of wrinkles such as, but not limited to, periocular and perioral wrinkles. 595 nm: General Surgery: Photocoagulation of benign cutaneous vascular lesions and benign cutaneous lesions. Dermatology/Plastic Surgery: For treatment of benign cutaneous vascular lesions, such as facial and leg telangiectasia, rosacea, port wine stains, hemangiomas, angioma, spider angioma, Poikiloderma of Civatte, and benign cutaneous lesions, such as warts, scars, striae and Psoriasis and the treatment of wrinkles. Treatment of Benign Epidermal Pigmented Lesions. Treatment of Inflammatory Acne Vulgaris. Gynecology: Photocoagulation of benign cutaneous lesions and benign vascular lesions in gynecology. Podiatry: Treatment of benign cutaneous lesions, such as warts.
Device Story
Transportable Class IV laser system; delivers 595 nm (pulse-dye) or 1064 nm (Nd:YAG) energy. Comprised of main console, power supply, electronics, firmware, software, cooling system, and handpieces. User selects treatment parameters (fluence, pulse count, spot size) via color touch screen. Handpieces incorporate Dynamic Cooling Device (DCD) or optional EverCool Delivery System to cool skin surface during treatment via cryogen spray or cooling mechanism. Used by clinicians in clinical settings for photocoagulation, hemostasis, and lesion treatment. Output energy affects target tissue; cooling protects epidermis. Benefits include precise lesion removal and vascular coagulation.
Clinical Evidence
No clinical studies were conducted. Substantial equivalence established via bench testing, electrical safety/EMC testing (IEC 60601 series, IEC 60825-1), and biocompatibility testing (ISO 10993-5, ISO 10993-10). Bench study confirmed cooling performance of DCD and EverCool handpieces.
Technological Characteristics
Class IV laser; 595 nm (pulse-dye) and 1064 nm (Nd:YAG) wavelengths. Features internal calibration port/meter, DCD (cryogen spray) or EverCool handpieces. Spot sizes: 3-7mm, 7-11mm, 11-15mm, 3x10mm. Compliance: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1. Biocompatibility per ISO 10993. Software-controlled via touch screen interface.
Indications for Use
Indicated for patients requiring treatment of benign vascular lesions (e.g., port wine stains, hemangiomas, telangiectasia, rosacea, leg veins), benign cutaneous lesions (e.g., warts, lentigos, nevi, scars, striae, psoriasis, skin tags), inflammatory acne vulgaris, and wrinkles. Applicable to general surgery, dermatology, plastic surgery, gynecology, and podiatry.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Candela Family of Pulsed Dye Laser Systems (K050673)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 5, 2018
Syneron-Candela Sharon Timberlake VP, Global Regulatory and Clinical Affairs 530 Boston Post Road Wayland, Massachusetts 01778
Re: K180593
Trade/Device Name: Vbeam Prima Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: March 2, 2018 Received: March 6, 2018
Dear Sharon Timberlake:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820);
{1}------------------------------------------------
### Page 2 - Sharon Timberlake
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jennifer R. Stevenson -S3
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K180593
Device Name Vbeam Prima Laser System
Indications for Use (Describe)
1,064nm:
The Vbeam Prima Laser System is intended for the coagulation and hemostasis of benign vascular lesions such as, but not limited to, port wine stains, hemangiomas, warts, telangiectasia, rosacea, venus lakes, leg veins, and poikiloderma of civatte and treatment of benign cutaneous lesions such as, but not limited to lentigos (age spots), solar lentigos (sun spots), cafe au lait manules, sebortheic keratoses, nevi, chloasma, verrucea, skin tags, and keratoses. The laser is also indicated for the treatment of wrinkles such as, but not limited to, periocular and perioral wrinkles.
595 nm:
General Surgery: Photocoagulation of benign cutaneous vascular lesions and benign cutaneous lesions.
Dermatology/Plastic Surgery: For treatment of benign cutaneous vascular lesions, such as facial and leg telangiectasia. rosacea, port wine stains, hemangioma, spider angioma, poikiloderma of civatte, and benign cutaneous lesions, such as warts, scars, striae, psoriasis and the treatment of benign epidermal pigmented lesions. Treatment of inflammatory acne vulgaris.
Gynecology: Photocoagulation of benign cutaneous lesions and benign vascular lesions in gynecology.
Podiatry: Treatment of benign cutaneous lesions, such as warts.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial, sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary Vbeam® Prima Laser System
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.92
#### 1. DATE PREPARED
March 1, 2018
#### 2. APPLICANT NAME
Candela Corporation 530 Boston Post Road Wayland, MA 01778 P: (508) 358-0304 M: (617) 957-1434
#### 3. OFFICIAL CORRESPONDENT
Sharon Timberlake, MSHS, RAC, CCRA VP, Global Clinical & Regulatory Affairs Candela Corporation 530 Boston Post Road Wayland, MA 01778 P: (508) 358-0304 M: (617) 957-1434 sharont@syneron-candela.com
#### 4. DEVICE INFORMATION
| Proprietary Name: | Vbeam® Prima Laser System |
|------------------------|------------------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Pulse Dye and Nd: YAG Laser |
| Classification Name: | Laser surgical instrument for use in General and Plastic<br>surgery and in dermatology (21 CFR Section 878.4810) |
| Product Code: | GEX |
| Device Classification: | Class 2 |
| Laser Classification: | Class IV |
#### 5. PREDICATE DEVICE
Candela Family of Pulsed Dye Laser Systems (K050673), primary predicate Cynosure Cynergy Multiplex Laser (K050779), additional predicate
{4}------------------------------------------------
#### 6. INTENDED USE
### 1,064 nm:
The Vbeam® Prima Laser System is intended for the coagulation and hemostasis of benign vascular lesions such as, but not limited to, port wine stains, hemangiomas, warts, telangiectasia, rosacea, venus lakes, leg veins, spider veins, and poikiloderma of civatte and treatment of benign cutaneous lesions such as, but not limited to lentigos (age spots), solar lentigos (sun spots), cafe au lait manules, seborrheic keratoses, nevi, chloasma, verrucea, skin tags, and keratoses.
The laser is also indicated for the treatment of wrinkles such as, but not limited to, periocular and perioral wrinkles.
### 595 nm:
General Surgery: Photocoagulation of benign cutaneous vascular lesions and benign cutaneous lesions.
Dermatology/Plastic Surgery: For treatment of benign cutaneous vascular lesions, such as facial and leg telangiectasia, rosacea, port wine stains, hemangiomas, angioma, spider angioma, Poikiloderma of Civatte, and benign cutaneous lesions, such as warts, scars, striae and Psoriasis and the treatment of wrinkles. Treatment of Benign Epidermal Pigmented Lesions.
Treatment of Inflammatory Acne Vulgaris.
Gynecology: Photocoagulation of benign cutaneous lesions and benign vascular lesions in gynecology.
Podiatry: Treatment of benign cutaneous lesions, such as warts.
#### 7. DEVICE DESCRIPTION
The Vbeam Prima Laser system is designed as a transportable device that delivers laser energy at wavelengths of 595 nm (pulse-dye) or 1064 nm (Nd: YAG). The device is comprised of a main console, power supply, controlled electronics, firmware, software, cooling system, and handpieces. There is also an internal calibration port with an internal meter located in the system. The handpiece offers a set of distance gauges available in the following spot ranges: 3-7 mm, 7-11 mm, 11-15 mm, and 3 X 10 mm. The user selects treatment settings, such as fluence, number of pulses, and spot size selection via a color touch screen. The handpiece incorporates a Dynamic Cooling Device (DCD) into its handpiece that is designed to cool the skin surface during treatment, which consists of an electrically controlled spray nozzle, cryogen reservoir canister and an electronic control circuitry. The Vbeam Prima Laser System also offers an optional EverCool Delivery System Handpiece for skin cooling during treatment.
{5}------------------------------------------------
#### 8. TECHNOLOGICAL CHARACTERISTICS
The Vbeam Prima Laser System has the same design, technological characteristics, operating principles, and intended use as the Candela Family of Pulsed Dye Laser Systems (K050673) and the Cynosure Cynergy Multiplex Laser (K050779). The devices all share the same technical features, such as calibration port, wavelengths, laser medium, delivery systems, power supply, cooling system, electronics, firmware, software and user display screen. The devices all share the same operating principles, such as energy, repetition rate, and offer a range of spot sizes. Any minor differences do not raise any new types of safety or effectiveness questions because the Vbeam Prima Laser System parameters are the same as the predicates.
#### 9. PERFORMANCE DATA
The following performance data was included in the 510(k) Premarket Notification to support the substantial equivalence determination:
### Electrical Safety and Electromagnetic Compatibility:
Electrical safety and electromagnetic compatibility testing of Epic Pro 940 was conducted according to the recognized standards. The device passed the required testing and is in full compliance with the standards:
- . IEC 60825-1: Safety of Laser Products – Part 1: Equipment classification and requirements
- . IEC 60601-1-2: Medical electrical equipment – Part 1-2: General requirements for safety - collateral standard: electromagnetic compatibility (EMC)requirements and test
- . IEC 60601-1: Medical electrical equipment – Part 1: General requirements for basic safety and essential performance
- . IEC 60601-2-22: Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
### Biocompatibility
The biocompatibility of the Vbeam Prima Laser System has been established based on the predicate devices and the results of ISO 10993-5 and ISO 10993-10 series of testing.
### Software Verification & Validation
Software verification and validation testing was conducted and results demonstrated that testing results were found acceptable for software release. Software testing was performed per FDA's guidance document "Guidance for the Content of Premarket Submission for Software Contained in Medical Devices".
{6}------------------------------------------------
### Bench Testing
A bench study was conducted to confirm that the DCD handpiece and Evercool handpiece can produce similar cooling results.
### Clinical Testing
Based on the similarities of the device specifications, intended use, indications for use between the Vbeam Prima Laser System and its predicate devices, no clinical studies were needed to support this 510(k) Premarket Notification.
#### 10. STATEMENT OF SAFETY AND EFFECTIVENESS
The Vbeam Prima Laser System shares a similar design and intended use. Additionally, technological characteristics, including wavelengths, laser mediums, pulse width, spot sizes, energy, cooling systems, and repetition rate are similar between the Vbeam Prima Laser System and its predicate devices. The data presented in this 510(k) Premarket Notification further supports this conclusion since the results did not raise any new safety or efficacy concerns. Therefore, the Vbeam Prima Laser System is substantially equivalent to its predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.