DEKA SMARTXIDE Family (Smartxide Touch, Smartxide Punto)
K180193 · El.En Electronic Engineering Spa · GEX · Feb 21, 2018 · General, Plastic Surgery
Device Facts
Record ID
K180193
Device Name
DEKA SMARTXIDE Family (Smartxide Touch, Smartxide Punto)
Applicant
El.En Electronic Engineering Spa
Product Code
GEX · General, Plastic Surgery
Decision Date
Feb 21, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Incision, excision, ablation, vaporization, and coagulation of body soft tissue including intraoral tissue, in medical specialities including aesthetic (dermatology and plastic surgery), otolaryngology (ENT), gynaecology, neurosurgery, dental and oral surgery and genitourinary surgery. The use with the scanning unit is indicated for ablative skin resurfacing.
Device Story
DEKA Smartxide family (Smartxide Touch, Smartxide Punto) are CO2 laser systems (10.6 μm) for soft tissue surgery. Device utilizes 30W, 50W, or 60W CO2 laser source; energy delivered via articulated arm with seven-mirror optical assembly to treatment area. Scanning unit enables ablative skin resurfacing. Operated by clinicians in medical settings (dermatology, plastic surgery, ENT, gynecology, neurosurgery, dental/oral surgery). System modifications include varied power outputs and updated external covers for Punto models. Clinicians use device to perform precise tissue ablation/vaporization; output affects clinical decision-making by enabling targeted surgical intervention. Benefits include controlled tissue removal and resurfacing capabilities.
Clinical Evidence
No clinical data. Bench testing only: verification of maximum laser output for 30W, 50W, and 60W models; electrical safety and EMC compliance testing for 60W configuration; and verification of new cover design specifications.
Technological Characteristics
CO2 laser source (10.6μm); max power 30W, 50W, 60W. Articulated arm delivery system with seven-mirror optical assembly. Pulse duration 0.02-70 ms; pulse repetition rate 5-100 Hz. Scanner spot size 350μm; focal length 100mm. Handpiece spot sizes 0.2mm, 0.4mm.
Indications for Use
Indicated for patients requiring incision, excision, ablation, vaporization, or coagulation of soft tissue across various specialties including dermatology, plastic surgery, ENT, gynecology, neurosurgery, and dental/oral surgery. Scanning unit indicated for ablative skin resurfacing.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo is a stylized representation of a human figure, while the FDA part includes the acronym "FDA" in blue, followed by "U.S. FOOD & DRUG" in a larger font size and "ADMINISTRATION" in a smaller font size below.
February 21, 2018
El.En Electronic Engineering S.p.A. Paolo Peruzzi Regulatory Affairs Manager Via Baldanzese 17 Calenzano, 50041 It
Re: K180193
Trade/Device Name: DEKA SMARTXIDE Family (Smartxide Touch, Smartxide Punto) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: January 22, 2018 Received: January 24, 2018
Dear Paolo Peruzzi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely. Jennifer R. Stevenson -23 For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180193
#### Device Name
DEKA Smartxide family (Smartxide Touch, Smartxide Punto)
Indications for Use (Describe)
Incision, excision, ablation, vaporization of body soft tissue including intraoral tissue, in medical specialities including aesthetic (dermatology and plastic surgery), otolaryngology (ENT), gynaecology, neurosurgery, dental and oral surgery and genitourinary surgery. The use with the scanning unit is indicated for ablative skin resurfacing.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) Summary
# DEKA Smartxide family of Laser Systems - Special 510(k)
### Submitter:
El.En. S.p.A. Via Baldanzese, 17 50041 Calenzano (FI), Italy
### Contact:
Paolo Peruzzi Regulatory Affairs Manager & Official Correspondent Phone: +39.055.8826807 E-mail: p.peruzzi@elen.it
#### Date Summary Prepared:
January 22, 2018
### Device Trade Name:
Deka Smartxide family (Smartxide Touch, Smartxide Punto)
#### Common Name:
Medical Laser system
#### Classification Name:
Laser Surgical Instrument for use in General and Plastic Surgery and in Dermatology (GEX)
#### Classification Number:
21 CFR 878.4810
#### Equivalent Devices:
DEKA Smartxide Touch (K172362)
#### Device Description:
The DEKA Smartxide Family of laser system are medical devices equipped with a 10.600 nm Carbon Dioxide (CO2) laser source having a maximum power of 30W/50W/60W. The laser energy is delivered to the treatment area via an articulated arm and a delivery accessory connected to its distal end. The articulated arm is an optical assembly that delivers free beam laser radiation. It is made up of seven mirrors placed on rotating knuckles: the mechanical accuracy of the articulated arm allows the CO2 laser beam to travel inside it and along its axis regardless of the arm orientation. The DEKA Smartxide Family laser is equipped with a Scanning Unit that allows to perform skin resurfacing treatments.
{4}------------------------------------------------
The modifications to the devices are
- 1) different maximum available output power: 60W, 30W and 50W and
- 2) different external system and scanner covers for DEKA SmartXide Punto models.
The intended use of the modified devices, as described in the labeling, has not changed as a result of the modifications.
## Intended Use:
Incision, excision, ablation, vaporization, and coagulation of body soft tissue including intraoral tissue, in medical specialities including aesthetic (dermatology and plastic surgery), otolaryngology (ENT), gynaecology, neurosurgery, dental and oral surgery and genitourinary surgery. The use with the scanning unit is indicated for ablative skin resurfacing.
## Substantial equivalence discussion:
The DEKA Smartxide family of laser systems is substantially equivalent to the DEKA Smartxide Touch (K172362).
| Device Trade<br>Name | DEKA SMARTXIDE<br>FAMILY OF LASER<br>SYSTEMS | PREDICATE DEVICE<br>DEKA SMARTXIDE TOUCH<br>(K172362) |
|-------------------------------|-----------------------------------------------------|-------------------------------------------------------|
| Laser Type | CO2 | CO2 |
| Wavelength | 10.6μm | 10.6μm |
| Max Power | 30W, 50W (Smartxide Punto)<br>60W (Smartxide Touch) | 40W |
| Handpieces Spot<br>Sizes | 0.2 mm, 0.4 mm | 0.2 mm, 0.4 mm |
| Pulse Duration | 0.02 to 70 ms | 0.02 to 70 ms |
| Pulse Rep Rate | 5 to 100 Hz | 5 to 100 Hz |
| Scanner Spot<br>size | 350μm | 350μm |
| Scanner Focal<br>length (EFL) | 100mm | 100mm |
The DEKA Smartxide family of laser systems has the same indications for use as the abovementioned predicate device, with same principle of operation and essentially the same performances.
## Clinical Performance Data:
None
## Non-Clinical Performance Data:
The following verification and validation activities have been performed on the modified devices:
- Verification of actual maximum laser output for each model (30W, 50W, 60W). The test methods, acceptance criteria and test results are documented in the Design History File of Smartxide Punto and Smartxide Touch.
{5}------------------------------------------------
- Verification of electric safety and EMC compliance for the 60W model as worst case electric configuration. The test methods, acceptance criteria and test results are documented in the Design History File of Smartxide Touch.
- Verification of new covers design specifications for Smartxide Punto. -
# Conclusion:
We can conclude that the DEKA Smartxide family of laser systems is substantially equivalent to the predicate device.
# Additional Information:
None
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.