K172818 · Lameditech Co. , Ltd. · GEX · Mar 7, 2018 · General, Plastic Surgery
Device Facts
Record ID
K172818
Device Name
LMT-3000
Applicant
Lameditech Co. , Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Mar 7, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Indications for Use
The LMT-3000 is a portable, battery-operated laser lancing device indicated for obtaining capillary blood samples.
Device Story
LMT-3000 is a portable, battery-operated laser lancing device; uses Er:YAG laser technology (2940 nm) to ablate epidermis and papillary dermis; creates small hole in fingertip to access capillary blood. Operated by user in clinic or home; user sets power level, positions fingertip, and presses discharge button. Output is capillary blood sample for subsequent analysis. Benefits include reduced pain compared to steel lancets. Device is non-sterile, reusable, and rechargeable.
Clinical Evidence
Randomized study of 50 patients comparing LMT-3000 to standard steel lancets. Subjects underwent sampling on both hands with both devices across two visits. Primary endpoint: glucose value comparison. Results showed no significant difference in glucose values between laser and steel lancet methods. No dropouts reported. Patients reported less pain with LMT-3000.
Technological Characteristics
Er:YAG laser (2940 nm ± 10%); Class 4 laser (IEC 60825-1); 3.7V Li-Polymer rechargeable battery; polycarbonate housing; IP22 ingress protection; BF Type applied parts. Complies with ISO 10993-1, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-11, IEC 60601-2-22, and IEC 62366.
Indications for Use
Indicated for obtaining capillary blood samples in institutional and at-home settings for patients requiring blood sampling.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. Food & Drug Administration' in blue text.
March 7, 2018
Lameditech Co., Ltd. % Ken Pilgrim Senior Consultant, RA Emergo Global Consulting, LLC 2500 Bee Cave Road, Building 1, Suite 300 Austin, Texas 78746
Re: K172818
Trade/Device Name: LMT-3000 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument for use in General and Plastic Surgery and in Dermatology Regulatory Class: Class II Product Code: GEX Dated: January 31, 2018 Received: February 6, 2018
Dear Ken Pilgrim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Jennifer R. Stevenson -23 For Binita S. Ashar, M.D., M.B.A., F.A.C.S.
Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K172818
Device Name LMT-3000
Indications for Use (Describe)
The LMT-3000 is a portable, battery-operated laser lancing device indicated for obtaining capillary blood samples in both institutional and at-home settings.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# LMT-3000
# K172818
## 1. Submission Sponsor
Lameditech Co., Ltd.
#1002, ACE High End Tower 6th
234, Beotkkot-ro, Geumcheon-gu, Seoul
The Republic of Korea
Contact: Jong Seuk Choi
Title: CEO
#### 2. Submission Correspondent
Emergo Global Consulting, LLC
2500 Bee Cave Road
Building 1, Suite 300
Austin, TX 78746 USA
Office Phone: (512) 327.9997
Contact: Ken Pilgrim
Title: Senior RA/QA Consultant
#### 3. Date Prepared
October 2, 2017
#### 4. Device Identification
| Trade/Proprietary Name: | LMT-3000 |
|-------------------------|----------------------------------------------------------------------------------------|
| Common/Usual Name: | Laser Lancing Device |
| Classification Name: | Laser surgical instrument for use in general and plastic surgery and in<br>dermatology |
| Regulation Number: | 21 CFR § 878.4810 |
{4}------------------------------------------------
Product Code: GEX, Powered Laser Surgical Instrument
Device Class: Class II
Classification Panel: General & Plastic Surgery
# 5. Legally Marketed Predicate Device(s)
K013021 Innotech USA, Inc. LightLance™ Laser Skin Perforator
### 6. Indication for Use Statement
The LMT-3000 is a portable, battery-operated laser lancing device indicated for obtaining capillary blood samples.
### 7. Device Description
The LMT-3000 lancing device uses a battery-powered laser beam (Er:YAG laser technology), which makes a tiny hole in the skin to obtain a capillary blood sample. The LMT-3000 ablates the epidermis and a thin layer of papillary dermis with minimal photothermal and photomechanical damage to the surrounding tissue.
### 8. Substantial Equivalence Discussion
The following table compares the LMT-3000 to the predicate device with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.
| Manufacturer | Lameditech Co., Ltd. | Innotech USA, Inc. |
|---------------------|---------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | LMT-3000 | LightLance Laser Perforator |
| 510(k) Number | K172818 | K013021 |
| Product Code | GEX | GEX |
| Regulation Number | 21 CFR § 878.4810 | 21 CFR § 878.4810 |
| Regulation Name | Powered Laser Surgical Instrument | Powered Laser Surgical Instrument |
| Indications for Use | The LMT-3000 is a portable, battery-<br>operated laser lancing device indicated<br>for obtaining capillary blood samples. | The LightLance™ Laser Skin Perforator is<br>intended to be used for ablation of skin<br>tissue to establish capillary blood<br>access. The LightLance™ Laser Skin<br>Perforator general indication for use is<br>for the perforation of skin to draw<br>capillary blood for screening purpose.<br>The device is specifically indicated for |
| Table 5-A - Comparison of Characteristics | | |
|------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Lameditech Co., Ltd. | Innotech USA, Inc. |
| Trade Name | LMT-3000 | LightLance Laser Perforator |
| | | obtaining capillary blood samples for<br>subsequent analysis of blood glucose<br>concentration in both institutional and<br>home settings |
| Mechanism of Action | The LMT-3000 Laser Skin Perforator is a<br>pulsed erbium doped yttrium-<br>aluminum-garnet(Er:YAG) laser. It<br>outputs a single pulse of laser light with<br>a wavelength of 2940 nm, which ablates<br>a small hole in the skin to a depth<br>sufficient to access capillary blood. | The LightLance Laser Skin Perforator is a<br>pulsed erbium doped yttrium-<br>aluminum-garnet (Er:YAG) laser. It<br>outputs a single pulse of laser light with<br>a wavelength of 2940 nm, which ablates<br>a small hole in the skin to a depth<br>sufficient to access capillary blood. |
| Technology Overview | The wavelength of this laser is 2.94 µm,<br>which matches water's absorption<br>peak. Taking in to account that the<br>water is the major constituent of<br>organic tissue, it's possible to explain<br>the strong absorption of tissue for this<br>wavelength.<br>On the other hand, due this strong<br>absorption in water, this laser exhibits a<br>little penetration depth | The wavelength of this laser is 2.94 µm,<br>which matches water's absorption<br>peak. Taking in to account that the<br>water is the major constituent of<br>organic tissue, it's possible to explain<br>the strong absorption of tissue for this<br>wavelength.<br>On the other hand, due this strong<br>absorption in water, this laser exhibits a<br>little penetration depth |
| Anatomical Location | Fingertip | Fingertip |
| Conditions of Use | 1. Set the Power Level<br>2. Position the Fingertip<br>3. Press the Discharge button.<br>4. Collect the Blood Sample | 1. Set the Power Level<br>2. Position the Fingertip<br>3. Press the Discharge button.<br>4. Collect the Blood Sample |
| Material | Polycarbonate | Unknown |
| Battery Operated | Rechargeable battery (3.7 Vdc) | Yes |
| Power Supply | 100~240Vac, 50-60 Hz | Not known |
| Output | 5 Vdc, 1.0 A | Not known |
| Applied Parts | BF Type | Not known |
| Laser Source | Er:YAG | Er:YAG |
| Laser Class | Class 4 (IEC 60825-1) | Class 4 (IEC 60825-1) |
| Manufacturer | Lameditech Co., Ltd. | Innotech USA, Inc. |
| Trade Name | LMT-3000 | LightLance Laser Perforator |
| Laser Pulse Repetition<br>Rate | Single | Single |
| Laser wavelength<br>spectrum | 2940 nm ± 10% | 2940 nm ± 10% |
| Degree of protection<br>against ingress of<br>water or particulate<br>matter | IP22 | Not known |
| Unit Dimensions | 34.9 (W) X 102.45 (D) X 132.2 mm (H) | Not known |
| Weight | 170 g | Not known |
| Operating<br>Temperature | Ambient temperature +5 to +40°C | Not known |
| Operating Humidity | Relative humidity 5 to 90%,<br>noncondensing | Not known |
| Storage Temperature | Ambient temperature -10 to +60°C | Not known |
| Storage Humidity | Relative humidity 5 to 90%,<br>noncondensing | Not known |
| Single-Use | No | No |
| Battery Type | 3.7 Li-Polymer Battery 700 mAh,<br>rechargeable, replaceable (Type ICR-<br>18350) | Not known |
| Sterile | No | No |
| Complies with ISO<br>10993-1 | Yes | Yes |
| Electrical Safety<br>Testing Passed | Testing passed | Testing passed |
| Conditions of Use | Rx Only | Rx Only |
{5}------------------------------------------------
{6}------------------------------------------------
# 5. Non-Clinical Performance Data
As part of demonstrating safety and effectiveness of LMT-3000 and in showing substantial equivalence to the predicate devices that are subject to this 510(k) submission, Lameditech Co., Ltd. completed a number of non-clinical performance tests. The LMT-3000 meets all the requirements for overall design,
{7}------------------------------------------------
biocompatibility, and electrical safety results confirming that the design output meets the design inputs and specifications for the device.
The LMT-3000 passed all the testing in accordance with internal requirements, national standards, and international standards shown below to support the safety and efficacy of the subject device.
- . Biocompatibility testing per ISO 10993-1
- . Electrical safety testing per IEC 60601-1
- . Electromagnetic Compatibility (EMC) testing per IEC 60601-1-2
- . Laser Safety per IEC 60825-1
- . Usability per IEC 60601-1-6 and IEC 62366
- . Performance of Home Health Care Equipment per IEC 60601-1-11
- Performance of surgical, cosmetic, therapeutic and diagnostic equipment per IEC 60601-2-22
- . Software verification and validation testing per IEC 62304/FDA Guidance
- Cleaning and Sterilization Testing - pass
- . Storage and Transport Testing - pass
# 6. Clinical Performance Data
To validate the performance, safety, and effectiveness of the LMT-3000 and to demonstrate substantial equivalence to the predicate device, a randomized study was conducted comparing the LMT-3000 to standard steel lancets.
A total of 50 patients were evaluated between two visits, one week apart, for both the LMT-3000 and steel lancets. The sampling was done with the right hand first and the left hand subsequently for all subjects. During the first visit, half (25) of 50 subjects were sampled with the needle lancets on the right hand and with laser lancet on left hand. Another half of 50 subjects were sampled with the laser lancet on the right hand and with the needle lancet on left hand. During the second visit, the same procedures were done in the opposite order.
Analysis on the resultant glucose values indicated no significant difference in the glucose value between the needle lancet and laser lancet. There were no dropouts from either arm of the study group.
Some patients in the study reported less pain using the LMT-3000 compared to using the steel lancet for obtaining blood samples on some occasions.
# 7. Statement of Substantial Equivalence
lt has been shown in this 510(k) submission that the differences, between the LMT-3000 device and the predicate device listed above, do not raise new types of questions regarding safety and effectiveness. Further, the LMT-3000 device utilizes the same type of technology as the predicate device. The LMT-3000, as designed and manufactured, is determined to be substantially equivalent to the above-listed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.