K170331 · Quanta System Spa · GEX · Mar 3, 2017 · General, Plastic Surgery
Device Facts
Record ID
K170331
Device Name
Cyber Ho
Applicant
Quanta System Spa
Product Code
GEX · General, Plastic Surgery
Decision Date
Mar 3, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Cyber Ho laser system, including a fiber optic delivery system, is indicated for use in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue and in Lithotripsy of stones. It is indicated in medical specialties including, but not limited to: Urology Gastroenterology Arthroscopy Neurosurgery Pulmonary Gynecology ENT Dermatology Plastic Surgery General Surgery
Device Story
Cyber Ho is a Holmium laser surgical system; utilizes fiber optic delivery system to transmit laser energy to target tissue or stones. Operates via incision, excision, resection, ablation, vaporization, coagulation, and hemostasis of soft tissue; performs lithotripsy. Used in OR or clinical settings by physicians (urologists, surgeons, etc.). Device provides controlled laser emission; parameters (power, energy, frequency) adjusted by operator to achieve desired surgical effect. Benefits include precise tissue management and stone fragmentation. Modifications from predicate Litho include hardware/software changes and expanded emission parameter ranges.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), usability (EN 60601-1-6), laser-specific performance (IEC 60601-2-22), and laser safety (IEC 60825-1).
Technological Characteristics
Holmium laser source; fiber optic delivery. Complies with IEC 60601-1, IEC 60601-1-2, EN 60601-1-6, IEC 60601-2-22, and IEC 60825-1. Hardware and software modified from predicate; expanded power, energy, and frequency ranges.
Indications for Use
Indicated for patients requiring surgical soft tissue procedures (incision, excision, resection, ablation, vaporization, coagulation, hemostasis) or lithotripsy across various specialties including urology, gastroenterology, arthroscopy, neurosurgery, pulmonary, gynecology, ENT, dermatology, plastic, and general surgery.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 3, 2017
Quanta System Spa Francesco Dell'Antonio Vice President Regulatory Affairs and QA Via Acquedotto 109 Samarate (va), 21017 IT
Re: K170331
Trade/Device Name: Cyber Ho Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: February 1, 2017 Received: February 2, 2017
Dear Mr. Dell'Antonio:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. |
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| 510(k) Number (if known) | K170331 |
|-----------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Cyber Ho |
| Indications for Use (Describe) | The Cyber Ho laser system, including a fiber optic delivery system, is indicated for use in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue and in Lithotripsy of stones. |
| It is indicated in medical specialties including, but not limited to: | <ul><li>Urology</li><li>Gastroenterology</li><li>Arthroscopy</li><li>Neurosurgery</li><li>Pulmonary</li><li>Gynecology</li><li>ENT</li><li>Dermatology</li><li>Plastic Surgery</li><li>General Surgery</li></ul> |
| Type of Use (Select one or both, as applicable) | <div> <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> </div> |
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
| FORM FDA 3881 (1/14) | Page 1 of 2 | PSC Publishing Services (301) 443-6740 EF |
|----------------------|-------------|-------------------------------------------|
|----------------------|-------------|-------------------------------------------|
{3}------------------------------------------------
. WOJER SSERICA JIAMS FRATS ASS 3HT OT MAOG GETTY MAON AUDY CHAS TOM OD .588 to to to to to to to wrogen and to stremeriupes of your selligation in the series
ələlər və bəbən əlab ətir virin bir nərədən sinə mərkəzi varilərindən ilə növlərindən məlumatı və məlumatının mən :ວ່າ ,ກອນນd ຣ່າປຸ ຄູນເວນນອາ ເດີຍ ຂອງຍຸມຂອງຍຸມຂອງຍຸມຂອງຫຼວງບາງ əti birinci keçirilər. Bu sinod et ədən və bir növü ilə qalınmış və mənisə və və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və b
vop.sqy.ebj@jjejSAAA Poperwork Reduction Act (PRA) Stat Department of Health and Human Services
to notice a co produes of bailupes you si uostro e pre , iosenods no toubros tou vem vonege nA.,
no noito e los produes of beilupes you si uostrogs no toubros tour vous vous
{4}------------------------------------------------
## 5. Special 510(K) SUMMARY - Device Modifications
## Introduction:
This document contains the 510(k) Summary for the device Cyber Ho. The basis of this submission is Modifications to a Device already cleared. The content of this summary is based on the requirements of 21 CFR 807.92(c).
| Applicant /<br>Manufacturer<br>Name and Address: | Quanta System SPA<br>Via Acquedotto, 109<br>Samarate (VA)<br>Italy, 21017 |
|--------------------------------------------------|-------------------------------------------------------------------------------------------------|
| 510(k) Contact Person: | Francesco Dell'Antonio<br>Vice President Regulatory Affairs and QA<br>Quanta System SPA |
| | Email: francesco.dellantonio@quantasystem.com<br>Phone: +39-0331-376797<br>Fax: +39-0331-367815 |
| Date Prepared: | February 1st 2017 |
| Device Name: | Cyber Ho |
| Classification: | Class II |
| Classification Name: | Laser surgical instrument for use in general and plastic<br>surgery and in dermatology. |
| Regulation Number: | 21 CFR 878.4810 |
| Product Code: | GEX |
| Basis for Submission: | Device modifications |
| Legally Marketed Device | Litho (K163009) - Quanta System SPA<br>Lumenis Pulse 120h (K140388) - Lumenis LTD |
## Performance Standards:
There are no mandatory performance standards for this device.
{5}------------------------------------------------
## Description of the modifications:
This Special 510(k) of the modified device Cyber Ho is submitted due to Device Modifications of the already cleared device Litho (K163009) due to hardware and software change, together with a broadening of the range of some laser emission parameters such as power, energy and frequency.
The modified device has the same intended use of the unmodified device. Moreover the the intended use of the modified device, as described in its labeling, has not changed as a result of the modifications.
Based on the nature of the changes implemented, the device underwent and successfully passed electrical safety, EMC, performance testing and software verifications and validation according to the relevant standards.
## Intended Use/Indications for Use
The modified device Cyber Ho has the same intended use of the unmodified device, as follows:
The Cyber Ho laser system, including a fiber optic delivery system, is indicated for use in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, resection, ablation, vaporization, coagulation and haemostasis of soft tissue and in Lithotripsy of stones.
It is indicated in medical specialties including, but not limited to:
- . Urology
- Gastroenterology
- Arthroscopy .
- . Neurosurgery
- Pulmonary .
- . Gynecology
- 0 ENT
- Dermatology 0
- Plastic Surgery ●
- General Surgery .
{6}------------------------------------------------
#### Substantial Equivalence:
The modified and unmodified devices have the same intended use and the same fundamental scientific technology, based on Holmium laser sources.
The modified device Cyber Ho has a broader range of emission parameters compared to the unmodified device Litho, therefore another predicate device, Lumenis Pulse 120h (K140388), has been considered in order to demonstrate the safety and effectiveness of the parameters combinations not included in the unmodified device.
Thus the modified device Cyber Ho is substantially equivalent to its identified predicate devices.
## Performace testing
The modified device Cyber Ho was subjected to performance testing in accordance with the following recognized consensus standards related to electromagnetic compatibility, electrical safety and performances:
> IEC 60601-1: 2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
> IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
> EN 60601-1-6:2010 + A1:2013 Medical electrical equipment -- Part 1-6: General requirements for basic safety and essential performance - Collateral Standard: Usability
> IEC 60601-2-22:2007+ A1:2012 Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
> IEC 60825-1 Ed. 2.0 (2007) Safety of laser products - Part 1: Equipment classification and requirements
The modified device Cyber Ho passed all the required testing and is in compliance will all applicable sections of the above mentioned performance standards.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.