K170215 · Patient Pocket, LLC · GEX · Apr 19, 2017 · General, Plastic Surgery
Device Facts
Record ID
K170215
Device Name
LaserDock, HolmiumDock, CO2Dock,OmniDock, KTPDock
Applicant
Patient Pocket, LLC
Product Code
GEX · General, Plastic Surgery
Decision Date
Apr 19, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Indications for Use
LaserDock is indicated to hold laser fibers and prevent any unwanted laser emissions from escaping LaserDock. The LaserDock is designed for use with Ho:YAG ("HolmiumDock"), Nd:YAG, CO2 ("CO2Dock", OmniDock"), and KTP ("KTPDock") laser systems when laser fibers are deployed.
Device Story
Single-use sterile holster for surgical laser fibers; prevents accidental laser discharge, patient burns, and surgical fires. Comprises medical-grade ABS clam-shell housing, internal medical-grade silicone tube, and twist-clamp mechanism. Used in surgical field by clinicians; laser fiber inserted into device when not in use. Silicone tube acts as heat sink to dissipate thermal energy during accidental activation; housing prevents laser energy escape. Validated against ISO 11810-1/11810-2 for ignition and penetration resistance. Protects sterile fiber from contamination and physical damage (crimping/crushing).
Clinical Evidence
Bench testing only. No clinical data. Testing included pull force validation (1-5 lb/f), sterility/shelf-life validation, and laser containment testing (10-second exposure at maximum power for Ho:YAG, Nd:YAG, KTP, and CO2 lasers). Primary/secondary ignition and penetration testing performed per ISO 11810-1:2005 and ISO 11810-2:2007; all samples classified as Class I1 (no ignition) and Class P1 (no penetration).
Technological Characteristics
Materials: Medical-grade ABS housing, internal medical-grade silicone tube. Form factor: Clam-shell holster with twist-clamp. Sterilization: Gamma radiation (overkill method, SAL 10^-6). Standards: ISO 11810-1:2005, ISO 11810-2:2007, ANSI Z136.3. Connectivity: None (mechanical device).
Indications for Use
Indicated for use as a sterile holster for surgical laser fibers (Ho:YAG, Nd:YAG, CO2, KTP) during surgical procedures to secure the fiber and prevent accidental laser emission, patient burns, or surgical fires.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image is a circular logo for the U.S. Department of Health & Human Services. The logo features the department's symbol, which is a stylized representation of three human profiles facing right, stacked on top of each other. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper half of the circle, following its curvature.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 19, 2017
Patient Pocket LLC % Ray Kelly Consultant Licensale, Inc. 68 Southwoods Terrace Southbury, Connecticut 06488
Re: K170215
Trade/Device Name: LaserDock, HolmiumDock, CO2Dock, OmniDock, KTPDock Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument for use in General and Plastic Surgery and in Dermatology Regulatory Class: Class II Product Code: GEX Dated: January 15, 2017 Received: January 24, 2017
Dear Ray Kelly:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K170215
Device Name
LaserDock (HolmiumDock, CO2Dock, OmniDock, KTPDock)
#### Indications for Use (Describe)
LaserDock is indicated to hold laser fibers and prevent any unwanted laser emissions from escaping LaserDock. The LaserDock is designed for use with Ho:YAG ("HolmiumDock"), Nd:YAG, CO2 ("CO2Dock", OmniDock"), and KTP ("KTPDock") laser systems when laser fibers are deployed.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 003 510k Summary
This 510(k)-summary information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| APPLICANT: | Patient Pocket LLC 405 Hialeah Dr Cherry Hill, NJ 08002 |
|------------------------|---------------------------------------------------------|
| DATE PREPARED: | April 18, 2017 |
| CONTACT PERSON: | Raymond Kelly |
| TRADE NAME: | LaserDock |
| | HolmiumDock |
| | CO2Dock |
| | OmniDock |
| | KTPDock |
| CLASSIFICATION NAME: | Powered Laser Surgical Instrument |
| DEVICE CLASSIFICATION: | Class II |
| REGULATION NUMBER | 21 CFR 878.4810 |
| PRODUCT CODE | GEX |
| PREDICATE DEVICES: | LaserDock (K152636) |
Substantially Equivalent To:
The LaserDock (HolmiumDock, CO2Dock, OmniDock, KTPDock) is substantially equivalent in intended use, principal of operation and technological characteristics to the previously cleared LaserDock (K152636).
#### Description of the Device Subject to Premarket Notification:
The LaserDock is a sterile holster accessory to be used with surgical lasers and fiber optic laser energy delivery devices (laser fibers). The FDA recognizes that laser fibers can pose a risk of surgical fires, patient burns, and injuries if they are not properly secured. As part of a broader surgical fire prevention initiative, the FDA recommends that ignition sources such as ESUs (laser fibers) be placed "in a holster, and not on the patient or drapes," when not in use. [1]
<sup>1</sup> Preventing Surgical Fires: FDA Safety Communication (Oct. 3, 2011).
{4}------------------------------------------------
The LaserDock can be marketed under the following brand names (HolmiumDock, CO2Dock, OmniDock, or KTPDock). During a clinical procedure, when the laser fiber is not deployed in the patient, insertion into the LaserDock provides a sterile and secure holster for temporarily protecting the laser fiber. The sterility of the laser fiber is maintained by the sterile LaserDock.
The LaserDock is a single use sterile device comprising of an inner medical grade silicone tube, a clamp for securing the laser fiber in place, and a clam shell housing to protect the inner silicone tubing. When not in use, the laser fiber is held securely within the sterile medical grade silicone tubing by a twist clamp. Upon insertion of the laser fiber into the Laserdock, accidental discharging of laser energy is prevented from escaping the LaserDock and injuring or burning users and patients or creating potential fire hazards within the sterile field or outside of the sterile field.
### Indications for Use
LaserDock is indicated to hold laser fibers and prevent any unwanted laser emissions from escaping LaserDock. The LaserDock is designed for use with Ho:Y AG ("HolmiumDock"), Nd:YAG, CO2 ("CO2Dock", OmniDock"), and KTP ("KTPDock") laser systems when laser fibers are deployed.
### Technological Characteristics:
The modified LaserDock has the same technological characteristics and is similar in design and performance compared to the previously cleared device.
Both the predicate and proposed LaserDock have a sterile chamber where the fiber wire is secured preventing it from becoming contaminated or exposed to non-sterile surfaces, it also protects it within a hard shell from being crimped or damaged or falling on the floor where it can be stepped on. The retention clamp secures the fiber wire firmly within the protective sterile chamber without damaging the fiber.
Both the predicate and proposed LaserDock are made from a medical grade ABS housing which contains an internal medical grade silicone tube. Although the device does not have patient contact the device is made from the highest medical grade materials and provided sterile.
Both the predicate and proposed LaserDock have a retention clamp to keep the fiber wire securely inside the protective housing and sterile tube.
#### Performance Data
Testing conducted on the LaserDock demonstrates the modified device is substantially equivalent to the previously cleared LaserDock.
{5}------------------------------------------------
### Basis for Determination of Substantial Equivalence:
Upon reviewing and comparing intended use, design, principle of operation and overall technological characteristics, the LaserDock is determined by Patient Pocket LLC., to be substantially equivalent to the existing legally marketed device LaserDock.
#### Sterile Protective Enclosure
Both the predicate LaserDock and the proposed LaserDock are single use devices which are terminally sterilized to achieve an SAL 10°. Both devices are packaged in a validated environmental barrier package which protects the sterility of the device until it is opened for use. Both the predicate and proposed LaserDock are intended to be opened and used within a sterile surgical field to prevent the device from becoming contaminated. Both devices have been validated for package integrity, shelf life validation, and sterility. The gamma sterilization validation followed the overkill method and consisted of a VDMax validation and the package was validated using peel tests and bubble emersion tests on the pouches.
### Biocompatible Material
Both the predicate and proposed LaserDock are made from a medical grade ABS housing and an internal medical grade silicone tube. Both the proposed LaserDock LaserDock have been validated to be biocompatible for cytotoxicity, irritation, sensitization, sensitization, and material mediated pyrogen.
#### Fiber Wire Securement
Both devices have been tested to validate the pull force to ensure the fiber does not become dislodged from the LaserDock if it is accidentally tugged by the users. Based on these treatment types the following laser fibers will be used 200-500um, and 1-2.1mm. We have based the largest force parameter (5 1b/f) based on the largest fiber intended to be used which is the KTP green light 180 moxy fiber and the smallest force parameter (1 1b/f) based on the smallest fiber intended to be used (200um) and on the maximum grip users can get with a wet glove which is 0.7 lb/f.
- o 5 lb/f for fiber diameters between 200 500 um
- o 2 5 lb/f for fiber diameters between 500 1000 um
- o 2 5 lb/f for fiber diameters between 1 2.1 mm
All pull testing passed for both the predicate and proposed LaserDock.
## Accidental Laser Fiber Suppression
The device holds the fiber securely in place to prevent practitioners from attaching the fiber to the patient's lower extremity drape or from placing instruments on top of the laser fiber potentially damaging the fiber ("ANSI Z136-3, Safe Use of Lasers in Hospital Facilities".) The Laser Dock has a containment area (tubing) within the device to absorb the wavelength if the laser is accidently fired during the procedure. The outer shell of the device is the housing and the inner containment tubing acts as a heat sink to dissipate any thermal buildup that mav occur from accidental activation of the laser. Testing of the Laser Dock consisted of live firing into the device for 10 seconds (worst case scenario) to test the Laser Dock ability to withstand direct laser energy. The decision to use
{6}------------------------------------------------
the 10 second time frame as "worst case" revolves around a combination of field experience and the maximum permissible exposure (MPE) for the referenced wavelengths as outlined in ANSI Z136.3, Appendix A and ISO 11810-1/11810-2. Validation testing was performed initially on Ho:YAG, Nd:YAG, and KTP Diode Lasers.
| Wavelength | Mode | Max Power | Exposure Time |
|------------|------------|-----------|---------------|
| 532nm | Continuous | 30 Watts | 10s |
| 980nm | Continuous | 180 Watts | 10s |
| 1064nm | Continuous | 100 Watts | 10s |
| 2100nm | Pulsed | 100 Watts | 10s |
Testing results all met acceptance criteria in that the containment material showed no visual signs of exposure stress after being exposed to each of the four wavelengths at maximum power and maximum exposure times. The outer shell does not have any heat transferred to it because the inner protective tubing blocks all heat, energy, and penetration of the laser energy.
### Primary and Secondary Ignition and Penetration
Fire safety and penetration safety testing was performed following ISO 11810-1:2005 and ISO 11810-2:2007 for both safety of primary ignition and secondary ignition and for penetration. The worst-case laser type intended to be used with the device is CO2, therefore all laser firing was performed at this stage on the CO2 laser fibers. CO2 has been selected as the most hazardous energy type for causing potential damage or fire risk to the Laser Dock. CO2 is a more aggressive energy type then the other energy types which Laser Dock is intended to safely block, including Ho:YAG, Nd:YAG, and KTP laser systems due to its higher output temperatures and the fact that CO2 has a narrower focused diameter at the output. The narrower focused diameter and higher output temperature creates a higher intensity energy point compared to other types, making it the worst case energy type for containment and highest risk for ignition source, and therefore it was chosen to represent the other intended energy types for this study (Ho:YAG, Nd:YAG, and KTP). Testing was performed using CO2 with a 20-watt intensity for firing periods of 10 seconds. Spot diameter on target was 2mm. Primary ignition testing was performed on the test articles per ISO to establish classification for ignition.
All samples were tested for primary ignition risk and were classified as Class I1 material does not ignite after 10 seconds of focused laser contact. Material is safe and does not present a risk as a primary ignition source. All samples were tested for secondary ignition risk and were classified as Class SI1 material, material does not ignite after 10 seconds of focused laser contact, and gauze does not ignite after 10s of energy contact. Material is safe and does not present a risk as a primary ignition source. All samples were tested for penetration risk and were classified as Class P1 material, no penetration observed after 10 seconds of focused laser contact. Material is safe and does not present a risk for penetration.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.