K161926 · Beijing Adss Development Co., Ltd. · GEX · Feb 22, 2017 · General, Plastic Surgery
Device Facts
Record ID
K161926
Device Name
ND YAG Q-switch Laser Therapy Machine
Applicant
Beijing Adss Development Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Feb 22, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ND YAG Q-switch Laser Therapy Machine is indicated for the treatment of: benign cutaneous lesions, such as Warts, Scars, Striae and Psoriasis; benign pigmented lesions, such as Lentigines, nevus, and birthmark; and the removal of black or blue tattoos.
Device Story
Nd:YAG Q-switch laser system; delivers 1064nm or 532nm laser energy; used for benign cutaneous/pigmented lesions and tattoo removal. System comprises main unit, articulated arm, treatment probe, and foot pedal. Operated by clinician in clinical setting. Laser energy targets lesions/pigment; Q-switching provides short pulse widths (5-8ns) for photomechanical effect. Clinician uses aiming beam to target treatment area; foot pedal controls pulse output. Output energy 100-1000mJ (1064nm) or 50-500mJ (532nm). Device benefits patients by providing non-invasive removal of tattoos and treatment of skin lesions.
Clinical Evidence
No clinical data. Bench testing only. Compliance verified with IEC 60601-1, IEC 60601-2-22, IEC 60825-1, IEC 60601-1-2, ISO 10993-5 (cytotoxicity), and ISO 10993-10 (irritation/sensitization). Performance testing included spot size accuracy and thermal energy distribution.
Technological Characteristics
Nd:YAG laser; 1064nm/532nm wavelengths; 5-8ns pulse width; 1-10Hz repetition rate. Materials: Stainless steel treatment probe. Energy source: Electrical. Form factor: Floor-standing unit with articulated arm. Safety: Complies with IEC 60601-1, 60601-2-22, 60825-1. Sterilization: Disinfection of handpiece via 75% medicinal alcohol.
Indications for Use
Indicated for treatment of benign cutaneous lesions (warts, scars, striae, psoriasis), benign pigmented lesions (lentigines, nevus, birthmarks), and removal of black or blue tattoos in patients requiring laser therapy.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 22, 2017
Beijing Adss Development Co., Ltd % Mr. Ray Wang Beijing Believe Technology Service Co., Ltd. 5-1206, Build 332, Dafangju No. 25 Banbidian Rd. LiYuan Town, Tongzhou District, Beijing, 101121 China
Re: K161926 Trade/Device Name: Nd Yag Q-switch Laser Therapy Machine Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: January 18, 2017 Received: January 23, 2017
Dear Mr. Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K161926
Device Name ND YAG O-switch Laser Therapy Machine
### Indications for Use (Describe)
The ND YAG Q-switch Laser Therapy Machine is indicated for the treatment of: benign cutaneous lesions, such as Warts, Scars, Striae and Psoriasis; benign pigmented lesions, such as Lentigines, nevus, and birthmark; and the removal of black or blue tattoos.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Tab #3 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: K161926
- 1. Date of Preparation
02/16/2017
- Sponsor 2.
## Beijing ADSS Development Co., Ltd
F6, Xiandao Bldg., Jinyuan Rd. 36, Daxing Economic Development zone, Beijing, China, 102628
Establishment Registration Number: Not yet registered or the Number Contact Person: Gao Yurong Position: sales manager Tel: 86-10-83625120 Fax:86-10-83625121 Email: gyr@adss.com.cn
- Submission Correspondent 3.
Mr. Ray Wang
Beijing Believe Technology Service Co., Ltd. 5-1206, Build 332, DaFangJu, No.25 BanBiDian Rd., LiYuan Town, TongZhou District, Beijing, 101121, China
Tel: +86-18910677558 Fax: +86-10-52214696 Email: ray.wang@believe-med.com
{4}------------------------------------------------
- 4. Identification of Proposed Device
Trade Name: ND YAG Q-switch Laser Therapy Machine Common Name: Powered Laser Surgical Instrument Model(s): FG 2010, FG 2010-B, FG 2010-C
Regulatory Information: Classification Name: Powered Laser Surgical Instrument Classification: II; Product Code: GEX; Regulation Number: 21 CFR 878.4810; Review Panel: General& Plastic Surgery;
Intended Use:
The ND YAG Q-switch Laser Therapy Machine is indicated for the treatment of: benign cutaneous lesions, such as Warts, Scars, Striae and Psoriasis; benign pigmented lesions, such as Lentigines, nevus, and birthmark; and the removal of black or blue tattoos.
- 5. Device Description
The ND YAG Q-switch Laser Therapy Machine is laser system which delivers laser at a wavelength 1064nm or 532nm.
There are 3 models included, FG 2010, FG 2010-B, FG 2010-C, the three models have same intended use, mechanism of action, principle and specification, only differences are the configurations. The detailed difference shown as following:
| Model | FG2010 | FG2010-B | FG2010-C |
|--------|---------------|---------------|---------------|
| Size | 860×310×830mm | 800×320×920mm | 830×320×830mm |
| Weight | 60kg | 65kg | 60kg |
Table 1 The Difference of Models
| Table 2 Main Components introduction | | |
|--------------------------------------|--|--|
|--------------------------------------|--|--|
| Component name | Function |
|-----------------|------------------------------------------------|
| Main unit | Main Interface |
| Articulated Arm | Articulated arm for holding of Treatment Probe |
| Treatment Probe | Laser Deliver |
| Foot Pedal | control pulse light output |
{5}------------------------------------------------
- 6. Identification of Predicate Device
Predicate Device: 510(k) Number K072564
Predicate Device Name Harmony XL Multiple Application Platform
### Manufacturer
Alma Laser, Ltd
- 7. Non-Clinical Test
Non clinical tests were conducted to verify that the proposed device met all design specifications and was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- > IEC 60601-1:2012 Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance;
- > IEC 60601-2-22:2007, Medical Electrical Equipment - Part 2-22: Particular Requirements For Basic Safety And Essential Performance Of Surgical, Cosmetic, Therapeutic And Diagnostic Laser Equipment;
- > IEC 60825-1: 2007, Safety of laser products - Part 1: Equipment classification and requirements.
- > IEC 60601-1-2:2007, Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility-Requirements and tests.
- > ISO 10993-5:2009, Biological Evaluation of Medical Device, Part 5-Tests for Vitro cytotoxicity
- > ISO 10993-10:2002/Amd. 1: 2006, Biological Evaluation of Medical Device, Part 10-Test for irritation and delay-type hypersensitivity AMENDMENT 1
- > Performance Testing for Spot Size Accuracy.
- > Thermal Energy Distribution Test: determine the difference for thermal energy distribution of proposed device with predicate device.
- 8. Clinical Test
No clinical study is included in this submission.
{6}------------------------------------------------
# 9. Substantially Equivalent (SE) Comparison
| ITEM | Proposed Device | Predicate Device | Remark |
|----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Product Code | GEX | GEX | SE |
| Regulation No. | 21 CFR 878.4810 | 21 CFR 878.4810 | SE |
| Class | 2 | 2 | SE |
| Intended Use | The ND YAG Q-switch Laser Therapy<br>Machine is indicated for the treatment of:<br>benign cutaneous lesions, such as Warts,<br>Scars, Striae and Psoriasis; benign<br>pigmented lesions, such as Lentigines,<br>nevus, and birthmark; and the removal of<br>black or blue tattoos. | The 1064 nm Long Pulsed and Q-switched<br>Nd:YAG Laser Module Handpieces<br>The 1064 nm Nd:YAG Laser Module<br>handpiece (Long Pulsed and Q-switched<br>with and without contact-cooling) are<br>indicated for treatment and clearance of<br>Benign vascular lesions such as, but<br>not limited to treatment of: Port wine stains Hemangiomas Warts Superficial and deep<br>telangiectasias (venulectasias) Reticular veins (0.1-4.0 mm<br>dia.) of the leg Rosacea Venus lake Leg veins Spider veins Angiomas Benign cutaneous lesions, such as,<br>but not limited to : Warts Scars Striae Psoriasis Benign pigmented lesions such as,<br>but not limited to: Lentigos (age spots) Solar lentigos (sun spots) Café-au- lait birthmarks Seborrheic keratoses Nevi and nevus of Ota Chloasma Verrucae | SE |
{7}------------------------------------------------
| | > Skin tage |
|---|-------------------------------------------|
| | > Keratoses |
| | > The removal of black, blue or |
| | green tattoos (significant |
| | reduction in the intensity of |
| | black and /or blue tattoos) |
| | > Plaques |
| • | Pigmented lesions to reduce lesion |
| | size, for patients with lesions that |
| | would potentially benefit from |
| | aggressive treatment, and for patients |
| | with lesions that have not responded |
| | to other laser treatments. |
| • | The non-ablative treatment of facial |
| | wrinkles, such as, but not limited to: |
| | > Periocular wrinkles |
| | > Perioral wrinkles |
| • | Laser skin resurfacing procedures for |
| | the treatment of: |
| | > Acne scars |
| | > Wrinkles |
| • | Reduction of red pigmentation in |
| | hypertrophic and keloid scars where |
| | vascularity is an integral part of the |
| | scar. |
| | Indicated for use on all skin types |
| | (Fitzpatrick I-VI), including tanned skin |
# Table 4 Performance Comparison
| ITEM | Proposed Device | Predicate Device 2<br>K072536 | Remark |
|---------------------------|---------------------------------------------|-------------------------------|----------|
| Laser Medium | Nd:YAG | Nd:YAG | SE |
| wavelength | 1064 nm<br>532 nm | 1064 nm<br>532 nm | SE |
| Output energy | 100-1000mJ for 1064nm<br>50-500mJ for 532nm | 400-1200mJ | SE |
| Max.<br>Energy<br>Density | 31.8J/cm²<br>15.9 J/cm² | 16.9 J/cm² | SE |
| Spot Size | 2-10mm | 3, 5 mm | Analysis |
{8}------------------------------------------------
| Pulse Width | 5ns-8ns | 20 ns | Analysis |
|-----------------------------------------------|-----------------------------------------------------------------------------------------|------------|----------|
| Repetition Rate | 1-10 Hz | 1, 2, 4 Hz | Analysis |
| Disinfection | Disinfect the handpiece before and<br>after every treatment by 75%<br>medicinal alcohol | --- | Analysis |
| Laser Class | Class 4 | Class 4 | SE |
| Cooling<br>method<br>for treated skin<br>area | N.A. | N.A. | SE |
| Aiming Beam | Red Laser, < 6mW | N.A | Analysis |
## Table 5 Safety Comparison
| Item | Proposed Device | Predicate Device | Remark |
|------------------------------------------------|--------------------------------------------|--------------------------------------------|--------|
| | | K072536 | |
| Patient Contact Materials and Biocompatibility | | | |
| Patient Contact<br>Materials | Treatment Probe (Stainless Steel) | Treatment Probe (Stainless Steel) | SE |
| Cytotoxicity | No Cytotoxicity | Comply with ISO 10993-1 | |
| Sensitization | No evidence of sensitization | | SE |
| Irritation | No evidence of irritation | | |
| EMC, Electrical and Laser Safety | | | |
| Electrical Safety | Comply with IEC 60601-1, IEC<br>60601-2-22 | Comply with IEC 60601-1, IEC<br>60601-2-22 | SE |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | SE |
| Laser Safety | Comply with IEC 60601-2-22, IEC<br>60825 | Comply with IEC 60601-2-22, IEC 60825 | SE |
## Analysis for difference
The proposed device is substantially equivalent to the predicate device. Based on the nonclinical tests performed, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device.
## 10. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.