FAMILY OF SQUARE EPIL (Alex, Alex2, Nd:Yag, Alex+Nd:Yag)
K161632 · Bios S.R.L. · GEX · Dec 8, 2016 · General, Plastic Surgery
Device Facts
Record ID
K161632
Device Name
FAMILY OF SQUARE EPIL (Alex, Alex2, Nd:Yag, Alex+Nd:Yag)
Applicant
Bios S.R.L.
Product Code
GEX · General, Plastic Surgery
Decision Date
Dec 8, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Nd: Yag 1064 Laser is intended for: Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. The lasers are indicated on all skin types Fitzpatrick I-VI including tanned skin. Photocoagulation and hemostasis of benign pigmented and vascular lesions such as but not limited to port wine stains, hemangioma, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins. Coagulation and hemostasis of soft tissue. Treatment of benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, sebortheic keratosis, nevi, chloasma, verrucae, skin tags, keratosis, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques. The laser is indicated for benign pigmented lesions to reduce lesion size, for patients with benign lesions that would potentially benefit from aggressive treatment, and for patients with benign lesions that have not responded to other laser treatments. Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar. Treatment of wrinkles. The Nd:Yag 1064 Laser is intended for: Temporary increase of clear nail in patients with onychomycosis (e.g., derniatophyton rubrum and T. mentagrophytes, and for yeast Candida Albicans, etc.) The Alexandrite 755 laser is intended for: Temporary hair reduction. Stable long-term or permanent hair reduction through selective targeting of melanin in hair follicles. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6,9, and 12 months after the completion of a treatment regime. On all skin types (Fitzpatrick I - VI) including tanned skin. Treatment of benign pigmented lesions. Treatment of wrinkles. Photocoagulation of dermatological benign vascular lesions (such as port-wine stains, hemangiomas, telangiectasias).
Device Story
Family of Square Epil is a medical laser system utilizing solid-state Alexandrite (755nm) and/or Nd:YAG (1064nm) sources. Device configurations include single or dual-source setups; dual-source models operate one wavelength at a time. Operated by clinicians in medical settings, the system uses a touchscreen interface and control software to deliver laser energy for dermatological and surgical applications. Energy targets melanin in hair follicles, vascular structures, or pigmented lesions to achieve photocoagulation, hemostasis, or tissue reduction. Clinical benefits include permanent hair reduction, lesion clearance, and scar management. The device is intended for professional use only.
Clinical Evidence
No clinical data or animal studies were performed. Substantial equivalence is supported by bench testing, electrical safety/EMC compliance, and comparison of technical specifications to predicate devices.
Technological Characteristics
Solid-state laser system; Alexandrite (755nm) and Nd:YAG (1064nm) sources. Class II device. Features touchscreen interface and control software. Surface-contact accessories (≤ 24 hours). Complies with EN 60601-1 (safety) and EN 60601-1-2 (EMC).
Indications for Use
Indicated for patients of all Fitzpatrick skin types (I-VI), including tanned skin, requiring hair reduction, treatment of benign pigmented/vascular lesions, soft tissue coagulation, scar reduction, wrinkle treatment, or temporary increase of clear nail in onychomycosis.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol resembling a caduceus, with three figures in profile facing right, representing the department's focus on health and human well-being.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 8, 2016
Bios S.R.L. Manuela Brambilla Regulatory and Quality Assistant Via Guido Rossa, 10/12 Vimodrone, Milan I-20090 IT
Re: K161632
Trade/Device Name: Family Of Square Epil (Alex, Alex2, Nd: YAG, Alex+Nd:YAG) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: November 2, 2016 Received: November 14, 2016
Dear Manuela Brambilla:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{2}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Image /page/1/Picture/8 description: The image shows the name "Jennifer R. Stevenson - A" in a large, clear font. The text is arranged vertically, with "Jennifer R." on the top line and "Stevenson - A" on the bottom line. The background is plain and does not distract from the text.
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
### 510(k) Number (if known) K161632
### Device Name
FAMILY OF SQUARE EPIL (Square Epil Alex, Square Epil Nd:Yag, Square Epil Nd:Yag, Square Epil Alex+Nd: Yag)
Indications for Use (Describe)
The Nd: Yag 1064 Laser is intended for:
- Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
The lasers are indicated on all skin types Fitzpatrick I-VI including tanned skin.
- Photocoagulation and hemostasis of benign pigmented and vascular lesions such as but not limited to port wine stains, hemangioma, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins.
- Coagulation and hemostasis of soft tissue.
- Treatment of benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, sebortheic keratosis, nevi, chloasma, verrucae, skin tags, keratosis, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques.
The laser is indicated for benign pigmented lesions to reduce lesion size, for patients with benign lesions that would potentially benefit from aggressive treatment, and for patients with benign lesions that have not responded to other laser treatments.
- Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.
- Treatment of wrinkles.
The Nd:Yag 1064 Laser is intended for:
- Temporary increase of clear nail in patients with onychomycosis (e.g., derniatophyton rubrum and T. mentagrophytes, and for yeast Candida Albicans, etc.)
The Alexandrite 755 laser is intended for:
- Temporary hair reduction.
- Stable long-term or permanent hair reduction through selective targeting of melanin in hair follicles.
Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6,9, and 12 months after the completion of a treatment regime. On all skin types (Fitzpatrick I - VI) including tanned skin.
- Treatment of benign pigmented lesions.
- Treatment of wrinkles.
- Photocoagulation of dermatological benign vascular lesions (such as port-wine stains, hemangiomas, telangiectasias).
| Type of Use (Select one or both, as applicable) | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|
| <table border="0"><tr><td><div> <span> <span style="font-size: 10px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div></td></tr><tr><td><div> <span> <span style="font-size: 10px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div></td></tr></table> | <div> <span> <span style="font-size: 10px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 10px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
| <div> <span> <span style="font-size: 10px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | | |
| <div> <span> <span style="font-size: 10px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | | |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# Section 5 510(k) Summary
#### SUBMITTER I.
| Applicant Information | Bios s.r.l. |
|-----------------------|-------------------------------------------------------------------------------------------------------------|
| | Via Guido Rossa, 10/12 |
| | 20090 Vimodrone (MI) – Italy |
| Contact | Ing. Aldo Casalino |
| Date Prepared | 7 June 2016 |
| II. DEVICE | |
| Device Trade Name | FAMILY OF SQUARE EPIL (Square Epil Alex, Square Epil Alex2, Square<br>Epil Nd:Yag, Square Epil Alex+Nd:Yag) |
| Common Name | Medical Laser System |
| Regulatory class | II |
| Classification Name | Laser surgical instrument for use in general and plastic surgery and in<br>dermatology (21 CFR 878.4810) |
| Product Code | GEX |
#### PREDICATE DEVICE III.
| 510(k) Number | Pro Code | Trade Name | Applicant |
|---------------|---------------------|---------------------|------------------------------------------|
| K150516 | GEX 21 CFR 878.4810 | DEKA Synchro | El. En. Electronic<br>Engineering S.p.A. |
| K140122 | GEX 21 CFR 878.4810 | Candela Corporation | GentleMAX Family of<br>Laser Systems |
#### Device Description IV.
The FAMILY OF SQUARE EPIL LASER , is a laser equipped with one or two solid-state laser sources:
As shown in the previous table, the 4 configurations of the Family differ for the internal crystals of the sources, that can be Alexandrite and/or Nd:YAG. The emitted laser radiation has a wavelength of 755nm, respectively, and 1064nm:
-Alex configuration has only one source and one wavelength available (755 nm).
-Alex configuration has two sources of the same wavelength (755 nm), in order to increase the maximum power available.
{6}------------------------------------------------
-Nd:Yag configuration has only one source and one wavelength available (1064 nm).
-In the Alex+Nd: Yag configuration has both the two wavelengths available (755 nm and 1064 nm) but can act only individually, simultaneous operations are not allowed.
This device is intended for medical use only.
#### V. Indication for Use
The Nd:Yag 1064 Laser is intended for:
Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
The lasers are indicated on all skin types Fitzpatrick I-VI including tanned skin.
Photocoagulation and hemostasis of benign pigmented and vascular lesions such as but not limited to port wine stains, hemangioma, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins.
Coagulation and hemostasis of soft tissue.
Treatment of benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafè au lait macules, seborrheic keratosis, nevi, chloasma, verrucae, skin tags, keratosis, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques.
The laser is indicated for benign pigmented lesions to reduce lesion size, for patients with benign lesions that would potentially benefit from aggressive treatment, and for patients with benign lesions that have not responded to other laser treatments.
Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.
Treatment of wrinkles.
The Nd:Yag 1064 laser is intended for:
Temporary increase of clear nail in patients with onychomycosis (e.g., derniatophytes, Trichophyton rubrum and T. mentagrophytes, and for yeast Candida Albicans, etc.).
{7}------------------------------------------------
The Alexandrite 755 laser is intended for:
Temporary hair reduction. Stable long-term or permanent reduction through selective targeting of melanin in hair follicles. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing after a treatment regime. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. On all skin types (Fitzpatrick I- VI) including tanned skin.
Treatment of benign pigmented lesions.
Treatment of wrinkles.
The photocoagulation of dermatological benign vascular lesions (such as pant-wine stains, hemangiomas, telangiectasias)
#### VI. Comparison of Technological Characteristics with the Predicate Devices
The product specification, functionality, indication for use, and treatment parameters of the FAMILY OF SQUARE EPIL LASER are the same or very similar to the legally marketed, claimed predicate devices for the purpose of this 510(k) submission.
The general features of FAMILY OF SQUARE EPIL LASER are similar to those of the predicates in particular as regards the classification IV laser and the electrications. Also the predicate devices are equipped with a touch screen and with a control SW. The physical characteristics of FAMILY OF SQUARE EPIL LASER (weight and size) are also comparable with those of the predicates.
The intended use of each configuration of FAMILY OF SOUARE EPIL LASER are identical to those of the predicates and also the technical characteristics are the same.
In particular:
- Nd:Yag source of FAMILY OF SQUARE EPIL LASER with the wavelength of 1064 mm applies । a maximum energy in the same range of DEKA Synchro K150516 and the GentleMAX Family of Laser Systems K140122.
The emission frequency of the Nd: Yag source of FAMILY OF SQUARE EPIL LASER is the same of DEKA Synchro K150516 and the GentleMAX Family of Laser Systems K140122.
{8}------------------------------------------------
- Nd:Yag Source of FAMILY OF SQUARE EPIL LASER with the wavelength of 1064 nm, are shown operative parameters for a specific application (Temporary increase of clear nail in patients with onychomycosis). These parameters are identical to the predicate device GentleMAX Family of Laser Systems K140122.
- Alexandrite source of FAMILY OF SQUARE EPIL LASER with the wavelength of 755 nm applies a maximum energy in the same range of DEKA Synchro K150516 and the GentleMAX Family of Laser Systems K140122.
The emission frequency of the Alexandrite source of FAMILY OF SQUARE EPIL LASER are comparable to DEKA Synchro K150516 and the GentleMAX Family of Laser Systems K140122.
#### Performance Data VII.
# Biocompatibility testing
The FAMILY OF SQUARE EPIL LASER is classified as a surface device (≤ 24 hours contact duration) in contact with skin in accordance with ISO 10993.
The patient contacting materials for the device and accessories are all well known for their compatibility.
# Electrical safety and EMC
The FAMILY OF SQUARE EPIL LASER has been tested and is in compliance with:
- EN 60601-1, Medical electrical equipment. Part 1: General requirements for basic safety and essential performance 2006/A11:2010/A1:2013, EN 60601-2-22:2013 Medical electrical equipment;
- l EN60601-1-2 Medical electrical equipment. Part1: General requirements for safety - Collateral standard: Electromagnetic Compatibility - Requirements and tests.
# Software verification and validation testing
Software verification and validation was conducted according to FDA regulations, standards and guidance document requirements. The results of this testing conclude the software has met these requirements. Design verification and validation was also performed on the FAMILY OF SQUARE EPIL LASER in compliance with internal design control procedures.
{9}------------------------------------------------
# Usability testing
Usability of the FAMILY OF SQUARE EPIL LASER have been tested in accordance with the standard EN 60601-1-6:2010
### Pre series test
Pre-series tests have been performed to support the evidence that performances of FAMILY OF SQUARE EPIL LASER are aligned with the identified intended uses.
### Animal and clinical study
Animal and clinical studies were not necessary to demonstrate that the permormances are comparable to the predicate devices.
## VIII. Conclusions
BIOS s.r.l. has determined, by using comparisons and tests, that FAMILY OF SQUARE EPIL LASER is substantially equivalent to the listed predicate devices in terms of intended use, typical clinical use, operational characteristics, and fundamental technological characteristics. Any differences are considered minor and do not raise new issues of the safety and effectiveness of the FAMILY OF SQUARE EPIL LASER device when compared to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.