K161356 · Shanghai Wonderful Opto-Electrics Tech.Co.,Ltd · GEX · Sep 9, 2016 · General, Plastic Surgery
Device Facts
Record ID
K161356
Device Name
Matrix LS-40 CO2 Laser System
Applicant
Shanghai Wonderful Opto-Electrics Tech.Co.,Ltd
Product Code
GEX · General, Plastic Surgery
Decision Date
Sep 9, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
It is indicated for incision, excision, ablation, vaporization and coagulation of body soft tissues in medical specialties including aesthetic (dermatology and plastic surgery), podiatry, otolaryngology (ENT), gynecology, neurosurgery, orthopaedics, general and thorasic surgery (including open and endoscopic), dental and oral surgery and genitourinary surgery. The use with the scanning unit is indicated for ablative skin resurfacing. The CO2 System is intended for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in dermatology and plastic surgery, general surgery. Dermatology, Plastic Surgery and General Surgery procedures: - Laser skin resurfacing. - Treatment of furrows and wrinkles. •Removal of skin taqs, actinic keratosis, acne scars, keloids, tattoos, telangiectasia, squamous and basal cell carcinoma, warts and uneven pigmentation. - Treatment of cysts, abscesses, hemorrhoids and other soft tissue applications. - · Blepharoplasty. - · Site preparation for hair transplants. - The fractional scanner is for skin resurfacing.
Device Story
Matrix LS-40 CO2 Laser System; delivers 10,600nm wavelength laser energy via 7-joint articulated arm; activated by footswitch. Used in office/clinic settings by physicians for soft tissue surgery (incision, ablation, coagulation) and skin resurfacing. Device transforms electrical power into CO2 laser output; operates in Continuous Wave, Pulse, and Super Pulse modes. Healthcare providers use LCD touch screen interface to control parameters; output affects tissue via thermal ablation/vaporization. Benefits include precise tissue removal and resurfacing capabilities.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, and IEC 60825-1 standards for electrical, electromagnetic, and radiation safety.
Technological Characteristics
Sealed CO2 laser tube; 10,600nm wavelength; 0.5-40W power output; 7-joint articulated arm; closed-loop liquid cooling; LCD touch screen interface. Standards: IEC 60601-1:2005, IEC 60601-1-2:2007, IEC 60601-2-22:1995, IEC 60825-1:2007. Power input: 110-120 VAC, 50-60 Hz.
Indications for Use
Indicated for incision, excision, ablation, vaporization, and coagulation of soft tissues in patients requiring surgical intervention in dermatology, plastic surgery, general surgery, podiatry, ENT, gynecology, neurosurgery, orthopaedics, thoracic surgery, dental/oral surgery, and genitourinary surgery. Includes ablative skin resurfacing via scanning unit.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features the department's name in a circular arrangement around an emblem. The emblem consists of a stylized representation of three human profiles facing right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 9, 2016
Shanghai Wonderful Opto-Electrics Tech. Co., Ltd. Mr. Jonathan Hu Medwheat (Shanghai) Medical Technology Co., Ltd. Yangpu District Liaoyuan East Road Shuangyang First Suite No. 33 Room 303 Shanghai, Shanghai China 200093
Re: K161356
Trade/Device Name: Matrix LS-40 CO2 Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: July 25, 2016 Received: August 3, 2016
Dear Mr. Hu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{1}------------------------------------------------
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Christopher J. Ronk -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K161356
#### Device Name Powered Laser Surgical Instrument
#### Indications for Use (Describe)
It is indicated for incision, ablation, vaporization and coagulation of body soft tissues in medical specialties including aesthetic (dermatology and plastic surgery), podiatry, otolaryngology (ENT), gynecology, neurosurgery, orthopaedics, general and thorasic surgery (including open and endoscopic), dental and oral surgery and genitourinary surgery. The use with the scanning unit is indicated for ablative skin resurfacing.
The CO2 System is intended for use in surgical applications requiring the ablation, vaporization, and coagulation of soft tissue in dermatology and plastic surgery, general surgery.
Dermatology, Plastic Surgery and General Surgery procedures:
- · Laser skin resurfacing.
- Treatment of furrows and wrinkles.
· Removal of skin tags, actinic keratosis, acne scars, keloids, tattoos, telangiectasia, squamous and basal cell carcinoma, warts and uneven pigmentation.
- · Treatment of cysts, abscesses, hemorrhoids and other soft tissue applications.
- · Blepharoplasty.
- · Site preparation for hair transplants.
- · The fractional scanner is for skin resurfacing.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div> <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
|------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for SWOT. The logo is blue and features the word "SWOT" in a stylized font. There are two lines above and below the word "SWOT". The logo also includes the registered trademark symbol.
# 002_510(k) Summary
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for SWOT. The logo is blue and white, and it features the word "SWOT" in a stylized font. The letters are connected by a horizontal line. There is a registered trademark symbol in the upper right corner of the logo.
Date Prepared:September 9, 2016
# 510(k) Summary
[As required by 21 CFR 807.92]
# 1. Submitter's Information
| Name of Sponsor: | Shanghai Wonderful Opto-Electrics Tech. Co., Ltd. |
|------------------|--------------------------------------------------------------------------|
| Address: | 2F, Building 3, Lane 561, Nujiang Road (North),Shanghai<br>200333, China |
| Contact Name: | Amy Lee |
| Telephone No.: | 0086-021-62642623 |
| Fax No.: | 0086-021-52827988 |
| Email Address: | regulatory@wonderful-sh.com |
# 2. Correspondent's Information
| CompanyName: | Med-wheat Shanghai |
|---------------------|--------------------------|
| Correspondent Name: | Jonathan Hu |
| Telephone No.: | 0086-021-65181421 |
| Email Address: | Jonathan.hu@medwheat.com |
# 3. Trade Name, Common Name, Classification
| Trade Name: | Matrix LS-40 CO2 Laser System |
|---------------------------|-----------------------------------|
| Common Name: | Medical Laser System |
| Mode Name: | Matrix LS-40 |
| Regulation Classification | Powered Laser Surgical Instrument |
| Product Code: | GEX |
| Classification Panel: | General and Plastic Surgery |
| Device Class: | II |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for SWOT. The logo is blue and features the word "SWOT" in a stylized font. There are two lines above and below the word "SWOT". To the left of the word is a design that looks like two letter "S"s stacked on top of each other with a line going through them. There is a registered trademark symbol in the upper right corner of the image.
# 4. Identification of Predicate Device(s)
The identified predicatedevice within this submission is as follows:
TheCortex Laser System of Cynosure, Inc. dbaEllmanhas been cleared by FDA through 510(k) No.K150587 (Decision Date - July 21, 2015).
## 5. Description of the Device
The MATRIX LS-40 CO2 Surgical laser System incorporating a permanently articulated arm that delivers CO2 laser power in a wavelength of 10,600nm in its normal operations. The device is activated by means of a footswitch.
The Laser System is designed for use in an office or a clinic, which shall be a clean, dry and well-ventilated room with a door.
# 6. Indication for Use
It is indicated for incision, excision, ablation, vaporization and coagulation of body soft tissues in medical specialties including aesthetic (dermatology and plastic surgery), podiatry, otolaryngology (ENT), gynecology, neurosurgery, orthopaedics, general and thorasic surgery (including open and endoscopic), dental and oral surgery and genitourinary surgery. The use with the scanning unit is indicated for ablative skin resurfacing.
The CO2 System is intended for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in dermatology and plastic surgery, general surgery.
Dermatology, Plastic Surgery and General Surgery procedures:
- Laser skin resurfacing.
- Treatment of furrows and wrinkles.
•Removal of skin taqs, actinic keratosis, acne scars, keloids, tattoos, telangiectasia, squamous and basal cell carcinoma, warts and uneven pigmentation.
- Treatment of cysts, abscesses, hemorrhoids and other soft tissue applications.
- · Blepharoplasty.
- · Site preparation for hair transplants.
- The fractional scanner is for skin resurfacing.
## 510(k) Summary
{6}------------------------------------------------
# 7. Technological Characteristics
The subject device in this 510(k) uses the same technology that is utilized in the predicate device. A comparison of technological characteristics is provided in the following table:
| Items | Matrix LS-40<br>CO2Laser System | Cortex Laser System<br>in K150587 | |
|----------------------------|----------------------------------------------------------------|----------------------------------------------------------------|----------|
| Technology Characteristics | | | |
| Laser Type | Sealed CO2 Laser Tube | Sealed CO2 Laser Tube | |
| Wavelength | 10.6 micron (10,600nm) | 10.6 micron (10,600nm) | |
| Laser Operation<br>Mode | Continuous Wave (CW), Blend<br>(PW), Super Pulse (SP) | Continuous Wave, Blend (PW),<br>Super Pulse | |
| Power to<br>Tissue | CW | 0.5-40W | 0.5-40W |
| | PW | 0.5-30W | 0.5-30W |
| | SP | 0.5-15W | 0.5-15W |
| Frequency<br>Range | CW | N/A | N/A |
| | PW | 30Hz | 30Hz |
| | SP | 20-240Hz | 20-240Hz |
| Peak Power | 200W | 200W | |
| Repetition Rate | 1-8 Hz/240 Hz Super Pulse | 1-8 Hz/240 Hz Super Pulse | |
| SuperPulse Duration | 10-1000ms | 10-1000ms | |
| Aiming Beam | 3mW (650nm diode) adjustable | 3mW (650nm diode) adjustable | |
| Spot Size | 100µm for surgical handpiece<br>150µm for fractional handpiece | 100µm for surgical handpiece<br>150µm for fractional handpiece | |
| Articulated Arm | 7-joint articulated arm | 7-joint articulated arm | |
| Cooling Type | Closed Loop Liquid | Closed Loop Liquid | |
| User Interface | LCD Touch Screen | LCD Touch Screen | |
| Dimension | 34cm*46cm*96cm | 150cm*46cm* 31cm | |
| Weight | 46.5Kg | 60kg | |
| Power Input | 110 - 120 VAC, 10 A, 50-60 Hz | 110 - 120 VAC, 10 A, 50-60 Hz | |
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image is a logo for SWOT. The logo is blue and white. The word "SWOT" is written in blue letters, with a blue line underneath. There is a blue oval shape to the left of the word "SWOT" with the letters "S" written twice inside the oval. There is a registered trademark symbol in the upper right corner of the logo.
# 8. Discussion of Non-clinical Testing
The Matrix LS-40 CO2 Laser System has been conducted related non-clinical tests to identify the substantial equivalence from the predicate device. The tests include the concerning of electrical safety, EMC and the requirements of radiation safety, which contains the standards including IEC 60601-1:2005/COR1:2006/COR2:2007, IEC 60601-1-2:2007, IEC 60601-2-22:1995 and IEC 60825-1:2007.
## 9. Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, it is concluded, basing on performance testing, indication for use, and technology, the subject device is substantially equivalent to the predicate device K150587.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.