K143713 · Ambicare Health, Ltd. · GEX · Jan 19, 2016 · General, Plastic Surgery
Device Facts
Record ID
K143713
Device Name
Lustre PRO Light System
Applicant
Ambicare Health, Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jan 19, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K143713 · Jan 19, 2016
Lustre PRO Light System
Ambicare Health, Ltd.
Post-market clinical experience
Post-market clinical experience data was used to support the safety and performance of the Lustre PRO Light System by demonstrating similarity to other commercial and clinically tested Blue LED light systems.
Post-market clinical experience; Safety and performance justification
Other commercial and clinically tested Blue LED light systems
Safety and performance
Indications for Use
The Lustre PRO light system is indicated for the treatment of mild to moderate acne.
Device Story
Lustre PRO Light System is a portable, non-invasive phototherapy device for home use. It consists of a rechargeable controller and therapy heads containing blue light-emitting diodes (LEDs) peaking at 415nm (FWHM 15nm). Therapy heads attach to acne-affected skin sites using disposable polyethylene film with medical-grade adhesive. The controller features an LCD interface, control buttons, and a programmable timer to manage treatment duration. Sensors monitor therapy heads for overheating or disconnection, providing audible and visible alarms. The device is operated by the patient. By delivering non-thermal blue light, the system targets acne lesions. Clinical benefit is derived from the phototherapeutic effect of blue light on acne-causing bacteria. The device is intended for over-the-counter use.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), EN 62471 (photobiological safety), IEC 62366 (usability), IEC 60601-1-11 (home healthcare environment), and ISO 10993 (biocompatibility of adhesive strips). A self-selection and usability study of 71 subjects demonstrated that labeling and user guides exceeded comprehension requirements, confirming safe and effective use by lay persons.
Technological Characteristics
Portable LED-based phototherapy device. Emits blue light at 415nm peak (15nm FWHM). Components: rechargeable controller, therapy heads, polyethylene adhesive film (ISO 10993 compliant). Connectivity: mini USB. Power: rechargeable battery. Standards: IEC 60601-1, IEC 60601-1-2, EN 62471, IEC 62366, IEC 60601-1-11.
Indications for Use
Indicated for treatment of mild to moderate acne in patients suitable for light-based therapy.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are black and have a flowing, abstract design. The profiles are enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 19, 2016
Ambicare Health Ltd. c/o Susan D'Arcy iSMART Marketing Services Ltd. 129 Green Lanes, Wylde Green, Sutton Coldfield, Birmingham, GB B735LT West Midlands
Re: K143713 Trade/Device Name: Lustre PRO Light System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: November 15, 2015 Received: November 23, 2015
Dear Ms. D'Arcy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K143713
Device Name Lustre PRO Light System
Indications for Use (Describe)
The Lustre PRO light system is indicated for the treatment of mild to moderate acne.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Ambicare Health. The word "ambicare" is written in gray, with a horizontal orange line above the "i" and "c". To the right of the word "ambicare" is a plus sign, also in orange. Below the word "ambicare" is the word "Health" in a smaller font, also in orange.
#### 510(k) Summary
Alba Innovation Centre Alba Campus Livingston EH54 7GA Scotland United Kingdom
This 510(k) Summary is submitted in accordance with 21 CFR Part 807, Section 807.92(c).
#### Submitter's Name:
Ambicare Health Ltd
#### Submitter's Address:
Ambicare Health Ltd 10 Lochside Place, Edinburgh, EH12 9RG Scotland, UK
SBDN: SBD158325
User Fee Organisation Number: 297701
#### Establishment Registration Number:
Still to be established
#### Contact Person:
Susan D'Arcy
Telephone +44 (0) 7880313315
## Date Prepared: December 16th 2014
Date Modified: May 30th 2015
November 13th 2015
Page 1 of 4 K143713-.510k SummaryFINAL-.doc
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Ambicare Health. The word "ambicare" is written in gray, with a horizontal orange line above the "care" portion of the word. To the right of the word is an orange plus sign. Below the word "ambicare" is the word "Health" in orange.
#### Device Trade Name:
Lustre PRO Light System
#### Device Common Name:
Lustre Pure Light system
#### Device Classification Information:
| Regulation<br>Number | Device | Device<br>Class | Product<br>Code | Classification<br>Panel |
|----------------------|----------------------------------------------------------------------------|-----------------|-----------------|---------------------------------|
| 21 CFR<br>878.4810 | Over the counter powered<br>light based laser for Acne,<br>21 CFR 878.4810 | Class 2 | OLP | General &<br>Plastic<br>Surgery |
#### Predicate devices
The Ambicare Health, Lustre PRO Light system is substantially equivalent to the Silk' n Blue (K121435) (Home skin innovations).
## Device Description:
The Lustre PRO light system is a small portable non-invasive device, comprising of a rechargeable controller and therapy heads containing light emitting diodes, emitting blue light at a peak wavelength of 415nm, and a Full Width Half Maximum (FWHM) of 15nm.
The therapy heads are designed to be attached to the site of the acne using a disposable polyethylene film impregnated with a medical-grade adhesive. The controller comprises of an LCD graphical user interface, 3 control buttons, a coaxial connector and 3 mini USB connectors. The controller switches the therapy heads on and off and includes a programmable timer, allowing modification of treatment time and indicators for battery level, therapy head status and treatment time remaining. The device contains sensors which monitor the therapy heads and emit audible and visible alarms relating to overheating and disconnection of the therapy heads.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Ambicare Health. The word "ambicare" is written in a gray sans-serif font. Above the word "Health" is a yellow line and a plus sign. The word "Health" is written in a smaller, lighter font than the word "ambicare".
#### Intended/ Indications Use:
The Lustre PRO light system is indicated for the treatment of mild to moderate acne.
#### Technological Characteristics:
A comparative review of the Lustre PRO Light System with the predicate devices found that the technologies, mode of operation, and general principles for treatment with this device were substantially equivalent to the predicate device.
#### Substantial Equivalence:
The Lustre PRO and Silk'n Blue are phototherapy devices that use non thermal blue light energy and are intended for the treatment of mild to moderate acne. They both use identical wavelengths and have identical light outputs. The technological characteristics of the Lustre PRO Light System are similar to the technological characteristics of the listed predicate device. Where there are differences between the Lustre PRO Light System and the predicate device these differences have been negated by non-clinical performance testing. The Lustre PRO light system raises no new questions of safety and efficacy and introduce no new risks.
## Non Clinical Performance & Safety Data:
The Lustre PRO Light system complies with IEC 60601-1: "Medical Electrical Equipment Part 1 - General Requirements for Basic Safety and essential performance" IEC/EN 60601-1-2: Collateral Standard: Electromagnetic Compatibility - Requirements and test.
The Lustre Pure Light System has been evaluated against EN62471:Photobiological Safety of Lamps and Lamp Systems" in consideration of maximum possible light exposure to users and the results demonstrate that it poses no risk of retinal injury due to either the blue-light phototoxic effect, or the thermal damage mechanism.
Lustre Pure Light System was assessed with regards to usability for compliance with IEC 62366: Medical devices - Application of usability engineering to medical devices and design verification for compliance with IEC 60601-1-11: General requirements for basic safety and essential performance - Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Ambicare Health. The word "ambicare" is written in gray, with a yellow line above the "i" and a yellow plus sign to the right of the word. Below the word "ambicare" is the word "Health" in a smaller, lighter font. The logo is simple and clean, and the colors are muted.
In addition, testing and analysis of the adhesive strips have demonstrated compliance to ISO 10993: Biological evaluation of medical devices.
A self-selection and usability study of 71 subjects demonstrated that the labelling and user guide for the Lustre PRO system exceeded comprehension levels determined a priori and demonstrated that the device could be used safely and effectively by lay persons.
Data generated from post market clinical experience does not generate any significant doubts to the similarity of the Lustre acne system to other commercial and clinically tested Blue LED light systems. The data presented is suitable to justify safety and performance of the device in relation to predicate devices with similar technology.
These non-clinical tests demonstrate that the Lustre PRO light system is as safe, as effective and performs at least as safely and effectively as the legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.