K140503 · Boston Scientific · GEX · Apr 15, 2014 · General, Plastic Surgery
Device Facts
Record ID
K140503
Device Name
FLEXIVA SF SIDE-FIRING SINGLE-USE LASER FIBER
Applicant
Boston Scientific
Product Code
GEX · General, Plastic Surgery
Decision Date
Apr 15, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Flexiva SF Side-Firing Single-Use Laser Fiber is designed for use with Ho:YAG lasers for indications that are cleared for these laser systems, including, but not limited to endoscopic, and open surgical procedures involving vaporization, ablation, coagulation, hemostasis, excision of soft and cartitaginous tissue, and fragmentation of urinary and biliary calculi (Ho:YAG wavelength only). The fiber is designed for use with a standard SMA-905 connector that has been cleared for surgical use.
Device Story
Flexiva SF is a single-use fiber optic laser energy delivery device; consists of coated silica core fiber with ETFE jacket, SMA-905 connector, metal cap, adjustable hand piece, and rigid tubing. Device connects to external Ho:YAG or Nd:YAG laser systems; delivers laser energy at approximately 70° angle to tissue. Used in endoscopic, laparoscopic, or open surgical procedures; operated by physicians. Metal cap features circumferential black line guides for endoscopic positioning; adjustable hand piece and rigid tubing facilitate tip manipulation and control at treatment site. Output is directed laser energy for tissue vaporization, ablation, coagulation, hemostasis, excision, or calculus fragmentation. Benefits include precise energy delivery to target tissue during minimally invasive or open surgery.
Clinical Evidence
Bench testing only. Testing included operating life, hand-piece retention strength, fiber power transmission, fiber control, protective cap retention, fiber length, distal fiber max OD, aiming beam, connector temperature/durability, connector tensile, and seal integrity. Samples tested at T=0 and T=6 months accelerated aging.
Technological Characteristics
Fiber optic delivery system; coated silica core fiber with ETFE jacket; SMA-905 connector; metal cap with 70° laser aperture; adjustable hand piece; rigid tubing. Single-use. Passive device; no internal energy source or software.
Indications for Use
Indicated for patients requiring endoscopic, laparoscopic, or open surgical procedures involving vaporization, ablation, coagulation, hemostasis, excision, or incision of soft and cartilaginous tissue, and fragmentation of urinary and biliary calculi using Ho:YAG or Nd:YAG laser systems.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Flexiva™ High Power Single-Use Laser Fiber (K100078)
Submission Summary (Full Text)
{0}------------------------------------------------
APR 1 5 2014
Side-Firing Laser Fiber Traditional 510(k)
K140503 Boston Scientific
#### 510K SUMMARY SECTION 5 . .
## 510(k) Summary for Flexiva™ SF Side-Firing Single-Use Laser Fiber
## A. Sponsor
Boston Scientific Corporation Urology and Women's Health Division 100 Boston Scientific Way Marlborough, MA 01756
## B. Contact
Jeanne O'Toole Senior Specialist, Regulatory Affairs' ·508-683-4271 jeanne.otoole@bsci.com
or
Nichole Riek Manager, Regulatory Affairs 508-683-4175 nichole.riek@bsci.com
## C. Device Name
Flexiva™ SF Side-Firing Single-Use Laser Fiber Trade name: Common usual/name: Laser Instrument, Surgical, Powered GEX - Laser surgical instrument for use in general and Classification: plastic surgery and in dermatology 21 CFR 878.4810. Class II
## D. Predicate Device(s)
| Trade name: | DuoTome™ SideLite™ Fiber (cleared under K011703 as<br>Lumenis delivery devices) |
|--------------------|---------------------------------------------------------------------------------------------------------------------------|
| Common usual/name: | Laser Instrument, Surgical, Powered |
| Classification: | GEX - Laser surgical instrument for use in general and<br>plastic surgery and in dermatology<br>21 CFR 878.4810, Class II |
Premarket Notification: Lumenis, K011703
and
| Trade name: | Flexiva™ High Power Single-Use Laser Fiber (cleared under K100078 as Modified Straight Fire Laser Fiber) |
|-------------------------|------------------------------------------------------------------------------------------------------------------------|
| Common usual/name: | Laser Instrument, Surgical, Powered |
| Classification: | GEX - Laser surgical instrument for use in general and plastic surgery and in dermatology<br>21 CFR 878.4810, Class II |
| Premarket Notification: | Boston Scientific K100078 |
Premarket Notification: Boston Scientific, K 100078
{1}------------------------------------------------
#### 510K SUMMARY SECTION 5
## E. Device Description
The Flexiva SF Side-Firing Single-Use Laser Fibers (Flexiva SF) are fiber optic laser energy delivery devices consisting of an SMA-905 connector and a coated silica core fiber jacketed with ethylene tetrafluoroethylene (ETFE). The Flexiva SF fibers are equipped with a metal cap with circumferential solid black line guides, an adjustable hand piece and rigid tubing. The laser aperture is located near the tip of the metal cap. Laser energy is delivered at an approximately 70° angle to tissue from the tip of the fiber. The line guides assist with correct positioning of the fiber tip within the endoscope. The adjustable hand piece enables manipulation of the laser fiber tip at the treatment site. The rigid tubing aids in control of the fiber tip. These fibers may be used in a variety of laserbased surgical cases as an integral part of laser systems.
## F. Intended Use
The Flexiva SF fiber is designed for use with Ho:YAG and Nd:YAG lasers for indications that are cleared for these laser systems, including, but not limited to endoscopic, laparoscopic, and open surgical procedures involving vaporization, ablation, coagulation, hemostasis, excision, resection, incision of soft and cartilaginous tissue, and fragmentation of urinary and biliary calculi (Ho:YAG wavelength only). The fiber is designed for use with a standard SMA-905 connector that has been cleared for surgical use.
## G. Technological Characteristics
The Flexiva SF fiber has the same technological characteristics and fundamental design as the predicate devices. The proposed Flexiva SF fiber is available in a single configuration.
## H. Substantial Equivalence
A direct comparison of key characteristics demonstrates that the Flexiva SF fiber is substantially equivalent to the predicate devices in terms of intended use, technological characteristics, and performance characteristics. The Flexiva SF fiber is as safe and effective as the predicate devices.
## I. Performance Testing (Bench Evaluation)
Boston Scientific has conducted performance testing with samples aged at T=0 and T=6 months accelerated aging in support of the proposed laser device. The following testing was completed to evaluate the effects of the design change:
- Operating Life .
- Hand-Piece Retention Strength ●
- Initial Fiber Power Transmission .
- Fiber Power Transmission .
- Fiber Control .
{2}------------------------------------------------
Side-Firing Laser Fiber Traditional 510(k)
#### SECTION 5 -510K SUMMARY 11
- Protective Cap Retention Strength .
- Fiber Length .
- Distal Fiber Max OD .
- . Aiming Beam
- . Connector Temperature/Durability
- Connector Tensile . .
- Seal Integrity .
- Seal Strength .
The results of the performance testing demonstrate that the Flexiva SF fiber is considered safe and effective for its intended use.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 15, 2014
Boston Scientific Corporation Ms. Jeanne O'Toole Senior Specialist, Regulatory Affairs 100 Boston Scientific Way Marlborough, Massachusetts 01756
Re: K140503
Trade/Device Name: Flexiva" SF Side-Firing Single-Use Laser Fiber Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: March 31, 2014 Received: April 1, 2014
Dear Ms. O'Toole:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH docs not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that I Dri has made a and regulations administered by other Federal agencies. You must or any Federal the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2 - Ms. Jeanne O'Toole
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# David Krause -S
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
510(k) Number (if known) K140503
## Device Name
Flexiva™ SF Side-Firing Single-Use Laser Fiber
## Indications for Use (Describe)
The Flexiva SF Side-Firing Single-Use Laser Fiber is designed for use with Ho:Y AG lasers for indications that are cleared for these laser systems, including, but not limited to endoscopic, and open surgical procedures involving vaporization, ablation, coagulation, hemostasis, excision of soft and cartitaginous tissue, and fragmentation of urinary and biliary calculi (Ho:Y AG wavelength only). The fiber is designed for use with a standard SMA-905 connector that has been cleared for surgical use.
Type of Use (Select one or both, as applicable)
[x] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Consistence of Center for Devices and Radiological Health (CDRH) (original)
# Neil R Oga 2014.04.14 1
FORM FDA 3881 (1/14) For BSA
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.